Silicon Photomultipliers:
Characterization and Applications
1 Dipartimento
2 Istituto
Marco Ramilli1 , Alessia Allevi1 , Luca Nardo1 ,
Maria Bondani2 and Massimo Caccia1
di Fisica e Matematica - Università degli Studi dell’Insubria
di Fotonica e Nanotecnologie - Consiglio Nazionale delle Ricerche
Italy
1. Introduction
Silicon Photo-Multipliers (SiPMs henceover) are photo-detectors based on a technology
originally invented in Russia (Akindinov et al., 1997). They essentially consist of an array
of p-n junctions operated beyond the breakdown voltage (McKay, 1954), in a Geiger-Müller
(G-M) regime (Oldham et al., 1972), with typical gain of the order of 106 and on-cell integrated
quenching mechanisms. Silicon photo-detectors with internal multiplication are in use since
more than a decade (Lutz, 1995). Avalance Photo-Diodes (APDs) (Akindinov et al., 2005) are
operated in a proportional regime, with typical gains of 104 , Single-Photon Avalanche Diodes
(SPADs) (Cova et al., 1996) are endowed with single-photon sensitivity and are tailored for
high frequency counting with time resolutions down to 30 ps. However, being made of a
single cell operated in binary mode, they do not carry any information about the intensity of
the incoming light field.
SiPMs complement the family of existing sensors: with a cell density of ∼ 103 /mm2 , areas up
to 3 × 3 mm2 and a single output node, they offer the possibility of measuring the intensity of
the light field simply by counting the number of fired cells. The main features of SiPMs, due
to their structural and operational characteristics are:
• high gain, granted by G-M operating mode, comparable to the values achieved by standard
photomultipliers (PMTs);
• enhanced linearity, owing to cell structure, with deviations that become relevant when
the average number of detected photons approaches the same order of magnitude of the
number of cells of the device (Tarolli et al., 2010);
• large dynamic range, provided by the pixelated structure with a common output, spanning
from the single photon regime up to high intensities;
• operability in magnetic fields, compactness and relatively low cost, granted by
silicon-based technology.
On the other hand, since G-M avalanches that are triggered by electron-hole pairs extracted by
the impinging photons are obviously indistinguishable from the ones originated from other
processes, dealing with all possible sources of noise is far from being trivial:
78
2
Photodetectors
Will-be-set-by-IN-TECH
• thermally extracted electron-hole pairs cause high Dark Count Rates (DCR), with values
ranging from several hundreds of kHz up to the MHz level, depending on the total number
of cells, the operating temperature and the overbias;
• spurious signals are also due to optical cross-talk (Sciacca et al., 2008): photons emitted by
an avalanche can travel through silicon and reach the depleted region in a neighboring cell,
thus triggering another avalanche;
• carriers extracted during an avalanche process may be trapped in a false potential
minimum in the depleted region: escaping from that trap they can originate an afterpulse
avalanche, so called because it typically happens shortly after (or even during) the recovery
of the previous avalanche (Du et al., 2008; Eckert et al., 2010).
These apparently huge disadvantages can be overridden by means of an exhaustive
characterization of the SiPM performances. An example of this procedure will be given in
the following Section, where we present the characterization of the devices as a function of
temperature.
Another aspect of the characterization of SiPMs will be addressed by providing a description
of the G-M avalanches probability distribution functions, in order to reconstruct the statistics
of the impinging light (Ramilli et al., 2010).
Modeling the SiPM response is also preliminary to application of these sensors as detectors in
Fluorescence Fluctuation Spectroscopy biophysical experiments (Chen et al., 1999; Schwille,
2001), where the parameters describing the system under investigation are inferred by the
deviations of the fluorescence intensity of suitable probes around its mean value. A feasibility
study on this topic will be presented in the final part of the Chapter.
2. Characterization
Tests have been performed on existing devices with the main goal to define an exhaustive
characterization protocol and to produce a comparative study. Three kinds of detectors
from different manufacturers have been studied: SensL1 , Hamamatsu Photonics2 and
STMicroelectronics3. Detector characterization is a major task for all the applications of SiPMs
and in particular for the identification of critical parameters: for example, DCR is an important
parameter for low-rate-event applications, while thermal stability is essential for portable
devices. For this reason, the following characterization protocol has been developed assessing:
• Geometrical parameters (number of cells, size of detectors and occupancy factor);
• I–V measurements;
• Noise measurements: DCR, optical cross-talk, dependence on the environmental
parameters.
• Analysis of photon spectra: resolution power, gain, working point optimization (at low
and large flux), electronic noise measurement taking into account cell-to-cell variations,
dependence on the environmental parameters (temperature).
• Linearity and dynamic range.
• Spectral response measurement: photon detection efficiency (PDE).
1
2
3
see http://sensl.com/.
see http://www.hamamatsu.com/.
see http://www.st.com/stonline/.
Silicon
Photomultipliers:
Characterization and Applications
Characterization
and Applications
Silicon Photomultipliers:
793
In particular, studies of the main SiPM parameters (such as gain, PDE, DCR and optical
cross-talk) as functions of temperature will be described in details.
2.1 Experimental setup
In order to implement the characterization protocol, we have developed an experimental
setup in which a green-emitting LED (λ = 510 nm), coupled to a fast pulse generator
(PDL800-B PicoQuant), has been used as the light source. The SiPM output signal has been
elaborated in the following way:
• SiPM was directly connected to a first stage amplification board:
– in experimental situations where the SiPM was operated under a continuous light flux,
a transimpedance pre-amplifier provided by SensL has been used: this device converts
the raw current from the SiPM into a voltage, with an amplification of 470 V/A;
– when the SiPM was operated in the pulsed regime, a different pre-amplifier, called
Pulse Amplifier and also provided by SensL, has been used. It allowed the fast rise of
the detector to be exploited, providing an amplification factor of 20;
• a leading edge discriminator (Lecroy 821) with a user-defined voltage threshold that the
SiPM output has to exceed in order to provide a triggering signal has been then exploited
at the output of the amplifier;
• in case of frequency measurements (e.g. DCR measurements), the discriminator output
has been directly sent to a scaler;
• the output of the SiPM has been integrated by a CAEN QDC V792N board:
– the board provides a charge measurement performing the integration of the input
voltage signal, with a conversion time of 2.8 μs and a reset time of 4 μs;
– the integration gate is generated by a NIM timing unit and must precede the analog
input signal of at least 15 ns;
• data have been stored in a PC via a USB-VME Bridge (CAEN).
The system composed by the investigated SiPM and the first-stage amplification board has
been located in a metal box sealed with grease, in which air has been replaced with helium.
The cooling fluid has been pumped into the box through a copper pipe, allowing controlled
temperature variations. The temperature of the system has been measured by a thermistor
placed in contact with the external packaging of the sensor.
2.2 Studying temperature behavior
In order to quantitatively understand the effects of temperature changes on the SiPM main
figures of merit, a suitable set of measurements must be accomplished. The procedure has
been tested with SensL CSI 0747 015 A20 HD and the results regarding this sensor will be
presented in more detail in the following. Once the procedure was tested, it has been repeated
with different SiPM models: a SensL CSI 0740 001 A20 HD, a Hamamatsu S10362-11-100C and
a ST Microelectronics TO-8 prototype.
Sensors have been placed in the cooling box described in Section 2.1 and several spectra have
been acquired at different bias voltages and different temperatures in order to obtain the
values of gain and PDE. Moreover, DCR measurements at different thresholds were used to
evaluate both DCR and cross-talk contributions. To analyze the procedure in detail, a typical
80
Photodetectors
Will-be-set-by-IN-TECH
4
1600
1400
1200
1000
800
600
400
200
0
100
150
200
250
300
350
QDC counts
Fig. 1. Example of a low flux spectrum of a SiPM; each histogram bin represents a single
QDC channel corresponding to 0.11 pC.
low-flux spectrum is presented in Fig. 1 that shows the charge measurements performed by
the CAEN QDC: a waveform generator triggered both the light source and the gate needed to
perform the charge measurement. The structure of the histogram reflects the characteristics of
the SiPM, since the n-th peak position represents the most probable output value (in released
charge) of n cells firing simultaneously and its Gaussian broadening is due to stochastic noise
sources. In this experimental situation, the very first peak represents the zero-photon peak, i.e.
the output of the QDC board with no SiPM signal. The broadening of the n-th peak σn can be
described with good approximation as:
σn2 = σ02 + nσ12 ,
(1)
where σ0 is the variance of the zero-photon peak (giving the noise contribution due to the
electronic chain), and σ1 is the variance of the first photon peak, providing an estimation of
the SiPM noise.
2.2.1 Gain
The gain has been evaluated by illuminating the SiPM with a low photon flux, thus obtaining
a spectrum whose peaks are clearly recognizable (see Fig. 1). As the n-th peak corresponds to
the mean charge released by n G-M avalanches, the gain G can be computed in the following
way:
QDC
(2)
G = − cal Δ PP ,
e K amp
where QDCcal = 0.11 pC/channel is the charge corresponding to one QDC unit, e− is the
elementary charge, K amp is the global amplification factor of the electronic setup, and Δ PP is
the distance in QDC units between two adjacent peaks of the collected spectra.
The gain behavior as a function of the bias voltage, at fixed temperatures, is shown in Fig. 2: a
linear dependence in the range of interest is clearly observable with a slope independent of the
temperature within the experimental errors. In Fig. 3 the same gain values are presented as
functions of temperature, for fixed bias voltages: a linear behavior is still clearly recognizable;
in this case as well the slope is not affected by the change of the applied bias.
This analysis suggests that the gain can be expressed as a linear function of a variable which
can be re-scaled with temperature. Since it is well known that the breakdown voltage has, in
815
Silicon
Photomultipliers:
Characterization and Applications
Characterization
and Applications
Silicon Photomultipliers:
SiPM Model
mBD (mV/degree)
23.2 ± 1.4
23.6 ± 0.9
61.9 ± 0.7
31.5 ± 0.1
SensL CSI 0747 015 A20 HD
SensL CSI 0740 001 A20 HD
Hamamatsu S10362-11-100C
STM TO-8 prototype
Table 1. Rate of change of the breakdown voltage with temperature.
our range of interest, a linear dependence on temperature (Goetzberger et al., 1963), the over
voltage, defined as the difference between the applied bias voltage and the breakdown one, is
thus a suitable candidate. The conditions can be summarized as follows
G (V, T ) = mV (V − VBD ( T )),
(3)
G (V, T ) = mT T + G ( T0 , V ),
(4)
where mV and mT are the slopes, VBD ( T ) is the breakdown voltage and T0 is a reference
temperature. Solving these equations for VBD ( T ), the rate of change of the breakdown voltage
with temperature can be expressed as:
mBD = −
mT
.
mV
(5)
The procedure has been applied to the different SiPMs. From the results in Table 1, we can see
that the values of mBD are technology dependent.
By using the breakdown voltage value measured at room temperature (RT) as the reference
χ2 / ndf
Gain
×103
−16° C
−10° C
+1 ° C
+12 ° C
+21 ° C
3000
2500
0.772 / 8
p0
−1.766e+07± 4.107e+05
p1
6.478e+05 ± 1.357e+04
2000
1500
χ2 / ndf
1000
0.06658 / 5
p0
−1.796e+07± 6.189e+05
p1
6.383e+05 ± 2.085e+04
500
29
29.5
30
30.5
31
31.5
32
bias voltage [V]
Fig. 2. Gain as a function of bias voltage, for different temperatures; a linear fit has been
performed for each temperature set: the obtained fit slopes are in agreement.
value, the breakdown values for each temperature can be written as:
VBD ( T ) = mBD ( T − TRT ) + VBD ( TRT ).
(6)
From Equation 6 it is possible to express the gain and all the other measured parameters
as a function of the over voltage: in Fig. 4 the gain is plotted as a function of the over
voltage, showing a global linear behavior independent of the temperature. Equation 6 can
be used to fix the operational parameters of a SiPM regardless of the varying environmental
conditions: if mBD is known with enough precision and frequent measurements of the SiPM
temperature are performed, the over voltage across each cell can be maintained fixed by
simply continuously adjusting the applied bias according to the temperature variations. This
82
Photodetectors
Will-be-set-by-IN-TECH
6
×103
χ2 / ndf
Gain
2200
2000
0.3038 / 4
bias 29 V
p0
1.845e+06 ± 1.757e+04
bias 29.5 V
p1
−1.467e+04 ± 998.5
bias 29.7 V
bias 30 V
bias 30.5 V
1800
1600
1400
1200
1000
800
600
χ2 / ndf
0.5858 / 4
p0
8.721e+05 ± 1.651e+04
p1
−1.407e+04± 931.2
−15
−10
−5
0
5
10
15
20
temperature [° C]
Fig. 3. Gain as a function of temperature (expressed in Celsius degrees), for fixed bias
voltages; each set has been fitted using a linear law: the obtained slopes are in agreement.
×103
−16° C
−10° C
1° C
12 ° C
21 ° C
global fit
Gain
3000
2500
2000
1500
χ2/ndf =
5.8 /
43
p0 −6.253e+04± 1.642e+04
1000
p1 6.505e+05 ± 6.613e+03
500
1
1.5
2
2.5
3
3.5
4
4.5
over voltage [V]
Fig. 4. Gain as a function of the over voltage: values are obtained by calculating the
breakdown voltage for each temperature; all the data acquired at different temperatures have
been fitted with the same linear law.
idea has led to a collaboration between Università dell’Insubria and CAEN for the realization
of CAEN SP5600 General Purpose Power Supply and Amplification Unit module with an
integrated threshold discriminator.
2.2.2 Dark count rate
As explained in Section 1, DCR is the frequency of the G-M avalanches triggered by thermally
extracted carriers. A scan of the DCR at different thresholds can be done, and the resulting
plot (an example is shown in Fig. 5) is usually referred to as staircase function.
DCR has been measured at different voltages and different temperatures, also setting different
discrimination thresholds; in Fig. 6, for example, the results for a “half photon threshold” are
shown. A clear and expected dependence on temperature is recognizable.
An exhaustive knowledge of DCR behavior is of utmost importance for a complete
characterization of all the noise sources of the detector. It can be a fundamental figure of
merit for low flux applications.
2.2.3 Optical cross-talk
An electron avalanche is modeled as a microplasma (Oldham et al., 1972): photons emitted by
the accelerated carriers during this event have a certain probability to reach the neighboring
837
Silicon
Photomultipliers:
Characterization and Applications
Characterization
and Applications
DCR [kHz]
Silicon Photomultipliers:
102
10
1
50
100
150
200
250
300
Threshold [mV]
DCR [MHz]
Fig. 5. Example of staircase curve, obtained after a three-stage amplification.
5
−16° C
−10° C
+1 ° C
+12 ° C
+21 ° C
4
3
2
1
0
1
1.5
2
2.5
3
3.5
4
4.5
over voltage [V]
Xtalk [%]
Fig. 6. DCR at different temperatures as a function of the over voltage, for a SensL SiPM.
−16° C
−10° C
+1 ° C
+12 ° C
+21 ° C
35
30
25
20
15
10
5
1
1.5
2
2.5
3
3.5
4
4.5
over voltage [V]
Fig. 7. Cross-talk as a function of the over voltage, at different temperatures, for a SensL
SiPM.
cell diodes thus triggering a second avalanche. The quantity named optical cross-talk (in short
cross-talk, XT ) is simply the percentage of avalanches triggered by such a mechanism.
Cross-talk has been calculated starting from DCR measurements by taking the ratio between
the DCR frequencies with the discriminator threshold respectively at “one-and-half photon”
and at “half photon”. This method is based on the assumption that the probability that
two uncorrelated thermally-triggered avalanches are generated within the same rise time is
negligible, so that all the second photon events are due to cross-talk.
84
Photodetectors
Will-be-set-by-IN-TECH
8
In Fig. 7 the results of the cross-talk evaluation are presented: it is worth noting that, within
the experimental errors, cross-talk does not seem to suffer a strong temperature dependence.
2.2.4 Photon detection efficiency
Photon detection efficiency (PDE) is a key parameter for every light detector and is defined as
the product of three terms:
- quantum efficiency, that is the probability that the impinging photon produces a charge
carrier, which is a function of the incoming light wavelength;
- the G-M probability that the extracted carrier generates an avalanche that depends on the
over voltage;
- the so-called Fill Factor (FF), a geometrical factor that expresses the portion of the active area
of the sensor with respect to the total area.
The PDE value thus represents the fraction of impinging photons that are actually detected.
To measure this quantity, SiPMs have been illuminated with a light intensity Φ which has been
previously measured with a calibrated PMT HAMAMATSU H5783 and the resulting spectra
have been acquired. PDE has been evaluated by estimating the mean number of detected
photons < n > meas with respect to the incoming light intensity:
PDE =
< n > meas
Φ
(7)
In the case of spectra in which each photon peak could be resolved, < n > meas has been
obtained by fitting the peak positions with a Poissonian curve and by evaluating the mean
value of the fit distribution. In the case in which the peaks could not be resolved, < n > meas
has been estimated as
QDCcal
< n > meas = −
ΔQDC
(8)
e G K amp
where G is the proper gain value, e− is the elementary charge, QDCcal = 0.11 pC is the
charge per QDC unit, K amp is the amplification factor of the electronics chain, and ΔQDC
is the difference between the mean value of the obtained spectrum and the pedestal position.
However, this method provides a zero-order approximation of the PDE because it does not
take into account cross-talk effects that can be as large as 40% (see Fig. 7 as a reference). To
properly estimate the “true” number < n >0 of impinging photons triggering an avalanche,
the following relation has been used:
< n > meas =
< n >o
< n > o (1 + X T ).
1 − XT
(9)
The approximation is valid for small values of the cross-talk; the previously obtained PDE
values have thus been corrected using < n >0 as the mean number of detected photons as
shown in Fig. 8. Again, PDE does not seem to have any remarkable temperature dependance,
thus confirming the agreement of our results with the definition of PDE.
3. Photon-number statistics
Due to their good linear response, SiPMs can be considered as the ideal candidates for the
reconstruction of the statistics of any light state (Afek et al., 2009). Obviously, the presence
of DCR and cross-talk effects must be taken into account. A proper model of the detector
859
Silicon
Photomultipliers:
Characterization and Applications
Characterization
and Applications
PDE0 [%]
Silicon Photomultipliers:
−16° C
−10° C
+1 ° C
+12 ° C
+21 ° c
12
10
8
6
4
1
1.5
2
2.5
3
3.5
4
4.5
over voltage [V]
Fig. 8. PDE values corrected taking into account cross-talk effects, for a SensL SiPM. The
wavelength of the impinging light is 510 nm.
Number of Diodes:
Area:
Diode dimension:
Breakdown Voltage:
Dark Count Rate:
Optical Cross-Talk:
Gain:
PDE (green):
Hamamatsu MPPC S10362-11-100C
100
1 mm × 1 mm
100 μm × 100 μm
69.23 V
540 kHz at 70 V
25 % at 70 V
3.3 · 106 at 70 V
15 % at 70 V
Table 2. Main features of the SiPM MPPC S10362-11-100C (Hamamatsu). The data refer to
room temperature.
response is thus required in order to properly assess the statistical properties of the light field
under investigation (Ramilli et al., 2010). In the following Section we present two different
analysis methods, both validated by suitable measurements performed by using a SiPM
produced by Hamamatsu Photonics. In particular, to keep DCR and cross-talk effects at
reasonable levels and to optimize the PDE, a detector endowed with 100 cells was chosen
(MPPC S10362-11-100C, see Table 2). To test the model, the sensor was illuminated by the
pseudo-thermal light obtained by passing the second-harmonics (@ 523 nm, 5.4-ps pulse
duration) of a mode-locked Nd:YLF laser amplified at 500 Hz (High Q Laser Production)
through a rotating ground-glass diffuser (D in Fig.8). The light to be measured was delivered
to the sensor by a multimode optical fiber (1 mm core diameter). The detector response was
integrated by the charge digitizer V792 by Caen. The signal was typically integrated over a
200-ns-long time window synchronized with the laser pulse.
3.1 Detector response modeling
The response of an ideal detector to a light field can be described as a Bernoullian process
(Mandel & Wolf, 1995):
n
η m (1 − η ) n − m
(10)
Bm,n (η ) =
m
being n the number of impinging photons over the integration time, m the number of detected
photons and η < 1 the PDE. Actually, η is a single parameter quantifying detector effects and
86
Photodetectors
Will-be-set-by-IN-TECH
10
losses (intentional or accidental) due to the optical system. As a consequence, the distribution
Pm,det of the number of detected photons has to be linked to the distribution Pn,ph of the
number of photons in the light under measurement by (Agliati et al., 2005; Mandel & Wolf,
1995; Zambra et al., 2004):
Pm,det =
∞
∑
n=m
Bm,n (η ) Pn,ph
(11)
It can be demonstrated (Bondani et al., 2009a) that for a combination of classical light states
the statistics is preserved by the primary detection process. This simple description has to be
further developed to link Pm,det to the probability density distribution of the G-M avalanches
of any origin. First we must take into account spurious hits and cross-talk effects, not
negligible in the detectors under study. The DCR results in a Poissonian process that can
be described as:
m
/m! exp(− mdc )
(12)
Pm,dc = mdc
( 2)
( 3)
where mdc is the mean number of dark counts during the gate window and σm,dc = σm,dc =
mdc . As a consequence, the statistics of the recorded pulses may be described as:
m
∑ Pi,dc Pm−i,det
Pm,det+dc =
(13)
i =0
obviously shifting the mean value and increasing variance and third-order central moment in
( 2)
( 2)
( 3)
( 3)
the following way: mdet+dc = mdet + mdc , σm,det+dc = σm,det + mdc and σm,det+dc = σm,det +
mdc . As a further step, cross-talk effects must be taken into account. Cross-talk is a genuine
cascade phenomenon that can be described at first order as (Afek et al., 2009)
l
xtk−l (1 − xt )2l −k .
(14)
Ck,l ( xt ) =
k−l
being xt the (constant) probability that the G-M avalanche of a cell triggers a second cell
(which becomes equivalent to the cross-talk probability XT in the limit of XT → 0), l
the number of dark counts and photo-triggered avalanches and k the actual light signal
amplitude. Within this first-order approximation, the real sensor response is described by
Pk,cross =
k
∑
m =0
Ck,m ( xt ) Pm,det+dc
( 2)
(15)
( 2)
characterized by kcross = (1 + xt )mdet+dc, σk,cross = (1 + xt )2 σm,det+dc + xt (1 − xt )mm,det+dc
( 3)
( 3)
( 2)
and σk,cross = (1 + xt )3 σm,det+dc + 3xt (1 − x2t )σm,det+dc + xt (1 − 3xt + 2x2t )m m,det+dc.
In the following we refer to this analytical model as Method I and compare it with a better
refined model (Method II). This second method offers, in principle, an extended range of
application even if it is limited to a numerical rather than an analytical solution. Irrespective
of the model, the amplification and digitization processes that produce the output x can be
simply described as a multiplicative parameter G:
Px,out = GPGk,cross
( 2)
( 2)
(16)
( 3)
( 3)
whose moments are xout = Gkcross , σx,out = G2 σk,cross and σx,out = G3 σk,cross .
Silicon
Photomultipliers:
Characterization and Applications
Characterization
and Applications
Silicon Photomultipliers:
87
11
3.1.1 Method I: an analytical evaluation of the second and third order momenta
In Ref. (Bondani et al., 2009b) we presented a self-consistent method aimed at reconstructing
the statistics of detected photons. According to this method, the analysis of the output of the
detector is based on the assumptions that the detection process is described by a Bernoullian
convolution and that the overall amplification-conversion process is given by a very precise
constant factor G, which allows the shot-by-shot detector output to be converted into the
number of detected photons. Experimentally, the value of G can be obtained by detecting
a light field at different optical losses and keeping the detector parameters fixed. Once the
amplification-conversion factor has been evaluated, the detected-photon distribution can be
achieved by dividing the output values by G and re-binning the data in unitary bins. Here
we present the extension of the method to detectors with a significant DCR and first order
cross-talk effects. The second-order momentum of the recorded pulse distribution Px,out can
be used to evaluate the Fano factor:
( 2)
Fx,out =
σx,out
Q
1 + 3xt
= det+dc x out + G
,
xout
mdet+dc
1 + xt
(17)
( 2)
where Qdet+dc = σm,det+dc /mdet+dc − 1 is the Mandel factor of the primary charges. Note
that, due to dark counts, the coefficient of xout in Equation 17 cannot be written as Qph /n
(Andreoni & Bondani, 2009; Bondani et al., 2009b), that is, the coefficient Qdet+dc /mdet+dc
does not only depend on the light field to be measured. Similarly we can calculate a sort
of symmetry parameter
( 3)
S x,out =
σx,out
Q
− 3Qdet+dc 2
1 + 3xt Qdet+dc
1 + 7xt
= s,det+dc2
x out + G
xout + G2
,
xout
1 + xt mdet+dc
1 + xt
mdet+dc
where Qs,det+dc
=
( 3)
σm,det+dc /mdet+dc − 1.
(18)
In the presence of dark counts both
coefficients ( Qs,det+dc − 3Qdet+dc )/m2det+dc and Qdet+dc /mdet+dc depend on the light being
measured (Bondani et al., 2009a).
3.1.2 Method II: a numerical evaluation based on the photon-number resolving properties of
SiPMs
The above mentioned self-consistent method is very powerful, but requires the acquisition
of several histograms at varying η, which might not always be easily performed in practical
applications: from this point of view, the possibility to analyze each spectrum independently
looks complementary to the self-consistent approach. This analysis has been performed with
a two-step procedure:
• the areas of the spectrum peaks have been computed in order to obtain an estimation of
the number of counts per peak;
• the obtained data points have been fitted with a theoretical function, which takes into
account the statistics of the light, the detection and all the deviations of the detectors from
ideality, such as DCR and cross-talk effects.
To evaluate the area of each peak, we performed a multi-peak fit of the spectrum histogram
modeling each peak with a Gauss-Hermite function (Van Der Marel et al., 1993):
2
(19)
GH = Ne−w /2 1 + 3 hH3 (w) + 4 hH4 (w)
88
Photodetectors
Will-be-set-by-IN-TECH
12
where w = ( x − x̄ )/σ, N is a normalization factor, x̄ the peak position and σ the variance of
the Gaussian function. H3 (w) and H4 (w) are the third and the fourth normalized Hermite
polynomials and their contribution gives the asymmetry of the peak shape, whose entity is
regulated by the pre-factors 3 h and 4 h, with values in the range [−1, 1]. The global fit function
of the spectrum is a sum of as many Gauss-Hermite functions as the number of resolved peaks.
By using Equation 19 it is possible to calculate the area An of the n-th peak as:
√
An = Nn σn ( 2π + 4 hn ).
(20)
The error σAn has been calculated by propagating the errors on the fit parameters.
This analysis is also useful to estimate the system gain G: in fact, from the fitted values of the
peak positions x̄n , the peak-to-peak distance Δ for all the resolved peaks is:
Δ n,n+1 = x̄n+1 − x̄n .
(21)
The associated error σΔn,n+1 can be once again obtained by propagating the fit errors;
furthermore, the value of G can be derived as a weighted average on all the peak-to-peak
values obtained from the analysis of the histograms.
The effect of detection, DCR and amplification have been modeled as described in the
previous Sections (see Equations 10-13 and Equation 16).
The effect of cross-talk has been described by using a Bernoullian process, in a way analogue
to what has been done in Equation 14. However, as the cross-talk process is intrinsically a
cascade phenomenon, its contribution has been calculated by adding higher order effects:
Pk,cross =
k
m
n
∑ ∑ ∑
m =0 n =0 j =0
Pk−m−n− j,det+dc Bm,k−m−n− j( xt ) Bn,m ( xt ) B j,n ( xt );
(22)
where the terms like B j,n ( xt ) stand for the Bernoullian distribution
B j,n ( xt ) =
n
j
x t (1 − x t ) n − j .
j
(23)
Such a higher order expansion is not trivial to be achieved by the self-consistent approach of
Method I, in which an explicit analytic expression of Px,out is needed in order to calculate its
moments. Here, as all the elements of interest (mel , mdc , xt , the number of modes μ) will be
obtained as parameters of a fit, this is not necessary and therefore Px,out can be just numerically
computed as the fitting function.
The major limit of this approach is obvious: as all the information on the statistics of the
system is obtained from the peak areas, this method can only be applied to peak-resolving
histograms with a number of peaks larger than the number of free parameters of the fitting
function, which, in the present analysis, can rise up to five.
3.2 Experimental results
In Ref. (Ramilli et al., 2010), we presented the experimental validation of the two
complementary methods by using both coherent and pseudo-thermal light. In particular, we
have demonstrated that it is not possible to use coherent light to derive the DCR contribution,
89
13
Silicon
Photomultipliers:
Characterization and Applications
Characterization
and Applications
Silicon Photomultipliers:
Fig. 9. Experimental setup. Nd:YLF: laser source, ND: variable neutral density filter, SiPM:
detector.
which is described by a Poissonian distribution and is thus indistinguishable from the light
under study. So we think that, within this context, it is more exhaustive and intriguing to
show the results obtained with thermal light. The photon-number distribution of a field made
of μ independent thermal modes is given by:
Pn,ph =
( 2)
with n = Nth , σn
( n + μ − 1) !
,
n! (μ − 1)! ( Nth /μ + 1) μ ( μ/Nth + 1) n
(24)
( 3)
= Nth ( Nth /μ + 1) and σn = Nth ( Nth /μ + 1) (2Nth /μ + 1).
We reproduced a single mode of this pseudo-thermal light field by passing the coherent light
of our laser source through an inhomogeneous diffusing medium and then selecting a single
speckle with an aperture (∼ 150 μm diameter), much smaller than the coherence area of the
speckle pattern produced (Arecchi, 1965). By delivering this pseudo-thermal light to the SiPM
sensor, we measured the values of the output, x, at 50000 subsequent laser shots and at 10
different mean values, obtained by means of a variable neutral-density filter (ND in Fig. 9). In
Fig. 10 the plot of the experimental values of Fx,out and S x,out is shown as a function of xout ,
along with the fitting curves evaluated from Equations 17 and 18:
Fx,out =
x
1 − dc
xout
2
xout + B
3
2
x
x
x2out + 3B 1 − dc
x out + C ,
S x,out = A 1 − dc
xout
xout
(25)
where μ, the number of independent thermal modes, has been set equal to one and, for
simplicity of notation, the parameters A = 2, B = G (1 + 3xt )/(1 + xt ) and C = G2 (1 +
7xt )/(1 + xt ) have been introduced. We fitted the data to Fx,out thus obtaining the values of
x dc = (5.82028 ± 1.34015) and B = (87.805 ± 2.09009) ch.
Then the data for S x,out have been fitted substituting the above values of x dc ch and B in
order to obtain A = (2.34754 ± 0.091576) and C = (8531.48 ± 419.571) ch2 . These values
were then used to evaluate G and xt , obtaining G = (74.2785 ± 1.76982) ch and xt =
(0.100174 ± 0.050529). The x-values were then divided by G and re-binned in unitary bins
(Bondani et al., 2009b) to obtain the Pk,cross distribution of the actual light signal amplitude
measured in the presence of dark counts and cross-talk. In Fig. 11 six different Pk,cross
distributions are plotted as bars at different mean values. Superimposed to the experimental
values, two theoretical distributions are shown: the first one (white circles) is evaluated
including the dark-count contribution that modifies the statistics of a single-mode thermal
90
Photodetectors
Will-be-set-by-IN-TECH
14
Fig. 10. Plot of Fx,out and S x,out as functions of xout for pseudo-thermal light.
distribution (see Equation 24) into
Pm,det+dc =
=
m
∑ Pk,dc Pm−k,det =
k =0
e−mdc
−μ
−m
μ
m
1 + det
( μ − 1) !
mdet
μ
μ
,
U − m, 1 − m − μ, mdc 1 +
mdet
1+
(26)
where U ( a, b, z) is the confluent hypergeometric function. The parameters are evaluated as
mdc = xdc /( G (1 + xt )) and mdet = ( xout − xdc )/( G (1 + xt )). The second curve (full circles) is
evaluated from Equation 15 to take into account the cross-talk. Unfortunately, the calculation
does not yield an easy analytical result, and hence it has been evaluated numerically.
The comparison between the data and
the theoretical functions can be estimated through the
m
evaluation of the fidelity f = ∑k=0 Pk,exp Pk,theo (see Fig. 11).
Turning now to the other approach, it is worth noting that the number of fit parameters is
large: the probability distribution is described by the expectation value mdet of the avalanches
generated by detection, the expectation value mdc of DCR contribution, the number of modes
μ, the probability xt of triggering a cross-talk event (up to three “iterations”) and again a global
normalization factor, for a total of 5 fit parameters: obviously, this puts a severe limit on the
applicability of this method, needing at least 6 resolved peaks.
As it can be noted from the fit results in Fig. 12, the results obtained by using Method II are
compatible within errors with what we found by applying Method I. However, even if the
global fits present a very low χ2 value for degree of freedom, the obtained fit parameters
present large uncertainties, probably indicating the presence of very large off-diagonal
elements in the minimization matrix and suggesting a strong correlation between the various
parameters. This problem can be avoided by fixing some of the fit parameters (such as DCR
or cross-talk), once their values have been retrieved from an accurate direct measurement (see
Section 2).
Silicon
Photomultipliers:
Characterization and Applications
Characterization
and Applications
Silicon Photomultipliers:
91
15
Fig. 11. Experimental Pk,cross distributions at different mean values (bars) and theoretical
distributions evaluated according to Method I: thermal modified by dark count distribution
(white circles), thermal modified by dark counts and cross-talk effect (full circles). The
corresponding fidelity values of the reconstruction are also shown.
Fig. 12. Experimental results for Method II applied on two of the histograms acquired with
thermal light. Upper panels: result of the multi-peak fit procedure; lower panels: fitted
theoretical function. The corresponding fidelity values of the reconstruction are also shown.
92
Photodetectors
Will-be-set-by-IN-TECH
16
3.3 Discussion
It is worth comparing the results of the two analysis methods on the same data sets (see
Fig. 13). It is possible to notice that the agreement is better for the mean values of detected
photons (panel (a)) and for the DCR (panel (b)), while the estimated values of xt with the two
Methods definitely disagree. This can be due to the different approximations adopted by the
two Methods (first order vs third order).
In Table 3 we summarize the results of the two Methods. Both Methods work in a self-consistent
way, even if they have two definitely different approaches. Method I does not need peak
resolving capability, but requires the acquisition of several histograms at varying η. Once
determined the parameters xt and DCR, all the datasets in a series can be analyzed,
independent of the number of distinguishable peaks in the pulse-height spectrum. Method II
works analyzing each histogram independently, but, as the G-M-avalanches distribution is
obtained with a fit of the data, it requires at least a number of resolved peaks larger than the
number of free parameters.
The fact that the two Methods give very similar results for mean photon numbers is particularly
important as in most applications this is the only important parameter. Merging the two
Methods we can develop an optimal strategy based on a self-consistent calibration performed
by measuring a known light and analyzing the data with Method I: once xt and DCR are
known, the determination of the mean photon number is independent of the specific light
statistics. Hence the information on the mean photon number can be obtained from each
single measurement, even when the fitting procedure of Method II cannot be applied.
Fig. 13. Left panel: Cross- talk values of xt obtained by applying Method II to thermal light
(full circles) and their weighted average (full line). Dashed line: value of xt obtained by
Method I. Central panel: Values of DCR evaluated for the same data as in the left panel by
applying Method II (full circles) and their weighted average (full line). Dashed line: value of
DCR obtained by Method I. Right panel: Values of mean photon numbers evaluated by
applying Method II (full circles) and by applying Method I (white circles).
Pseudo-thermal
Method I
Method II
DCR
0.071 ± 0.017 0.094 ± 0.035
Cross-talk 0.091 ± 0.042 0.035 ± 0.004
Table 3. Comparison between the global DCR and cross-talk values obtained with Method I
and weighted average of the values obtained with Method II.
Silicon
Photomultipliers:
Characterization and Applications
Characterization
and Applications
Silicon Photomultipliers:
93
17
4. Photon counting histogram with SiPM
Another instance in which the capability of resolving pulse-by-pulse the number of photons
in high-repetition-rate periodic light signals is the analysis of fluorescence fluctuations.
Traditionally, fluorescence techniques are applied to the study of systems constituted by
huge ensembles of particles. For this reason, the probability distribution of the photon
emission rate is peaked around its average value and fluctuations in the fluorescence signal
are negligibly small. Nevertheless, if a microscopic system in which only few fluorophores
are excited at a time is considered, the fluctuations in the fluorescence intensity become
comparable to the average fluorescence intensity value, and can thus be measured. As these
fluctuations are due to either the diffusion of the fluorophores in and out from the part
of the specimen from which the fluorescence is collected and conveyed to the fluorescence
detector, the so-called observation value, or to physical/chemical reactions inducing changes
in the emission rate, their analysis can be applied both to characterize the diffusion kinetics
and to study the physical/chemical reactions affecting emission. Such techniques are
collectively called fluorescence fluctuation spectroscopy (FFS) techniques. This sort of analysis
is particularly useful in probing biological systems both in vitro and in vivo. Indeed, many
biomolecules (e.g. enzymes) and substances (such as minerals or ions) that are present in
traces in cells have a determinant role in the cellular metabolism, and techniques allowing
to study their action at physiological concentrations are most desirable. The most popular
FFS technique is fluorescence correlation spectroscopy (FCS), which was developed in the
early 1970 s (Gosch et al., 2005; Schwille, 2001), and consists in the analysis of the temporal
correlations of fluorescence fluctuations in small molecular ensembles, combining maximum
sensitivity with high statistical confidence. Among a multitude of physical parameters that
are in principle accessible by FCS, it most conveniently allows to determine characteristic
rate constants of fast reversible and slow irreversible reactions of fluorescently labelled
biomolecules at very low (nanomolar and sub-nanomolar) concentrations. Moreover, the
technique is a non-invasive one, which means that the above parameters can be assessed
without perturbing the system equilibrium and, possibly, under physiological conditions. The
technique also allows the determination of the mobility coefficients and local concentrations
of fluorescent (or fluorescently labelled) molecules in their natural environments, provided
that the exact value of the observation volume is known. More recently, at the end of the
1990 s, another FFS technique, called photon counting histogram (PCH), was developed by
Chen et al. (Chen et al., 1999; 2005). This technique is based on the analysis of the statistical
distribution of the fluctuations in amplitude of the fluorescence rate and is thus suited to
yield steady state parameters of the system to be investigated, such as the number of different
fluorescent species present in a solution, the average number of molecules of each species
in the observation volume, and their molecular brightness, rather than dynamic parameters.
This technique is thus to a certain extent complementary to FCS. Until now, both FCS and
PCH measurements have been made by using APDs or other single-photon counters as the
light detectors. Information on either the time-correlations or the statistical distribution of
the fluorescence intensity fluctuations have been derived by counting the detection events
in consecutive and equally long time intervals. As single-photon counters must be used
only in the single photon regime (that is keeping the overall count rate very low in order
to be sure that impingement on the detector sensitive area of photons during their dead
time is a negligibly rare event), this sampling procedure is extremely time consuming and
the sampling interval width must not be too small. Namely, in the most frequent case of
pulsed excitation, it has to correspond to tens or even hundreds of pulse periods, which
94
Photodetectors
Will-be-set-by-IN-TECH
18
means tens of microseconds or more. Obviously, this puts an inferior limit on the sensibility to
fluorescence fluctuations occurring on time scales comparable, or shorter than, the sampling
time. Conversely, many biological reactions occur on time scales of hundreds of nanoseconds
to few microseconds (Nettels et al., 2007), that is slightly beyond the detection limit with
the current procedures. We believe that the usage, for FFS techniques, of detectors able to
count how many photons are contained in each fluorescence pulse could allow avoiding
sampling. This will significantly and inherently enhance the temporal resolution of both
techniques, virtually pulling it on the scale of the excitation source pulse period. Another
advantage connected to the release of the single-photon regime operation condition is the
possibility of performing statistically reliable measurements in much shorter times, which is
notably beneficial for in-vivo analyses, as the samples quickly deteriorate. Recently, we have
addressed the task of performing PCH analysis with SiPM detectors. Firstly, the standard PCH
model developed for fitting data acquired with APDs (Chen et al., 1999; 2005) is not suitable
to properly analyze the SiPM output; thus, the phenomenological model for describing the
SiPM response presented above has been used to modify the PCH equations so as to take
into account the deviations for ideal detection caused by the presence of optical cros-talk and
sizeable DCR, leading to the following fitting model:
Π ( k; N̄, , xt , n dcr ) =
=
∞
∑
N =0
p# ( N; N̄ )
k
∑
m =0
Ck,m ( xt )
∞
∑
n =0
Pn ( n dcr ) p( N ) (m − n; )
(27)
where xt is the parameter representing the probability for a cell of triggering a neighbour
cell by cross-talk and Ck,m ( xt ) is the Bernoullian-like distribution defined in Section 3.1
describing the probability of measuring k G-M avalanches out of n previously triggered,
with k > m. Secondly, a two-photon excitation setup has been commissioned and
tested by performing standard PCH measurements on Rodamine B water solutions with
a single-photon avalanche diode (PDM50, MPD) and a PC board allowing for real-time
PCH reconstruction (SPC150, Becker & Hickl GmbH). The setup relied on fluorescence
excitation by means of a continuous-wave SESAM mode-locked Ti:sapphire laser (Tiger-ps
SHG, Time Bandwidth Product) delivering 3.9 ps pulses of 840 nm wavelength at a repetition
rate of 48 Mhz. A 1.25 NA, 100X oil-immersion microscope objective (Nikon) was used
both to focus the excitation beam onto the sample and to collect the emitted light in
epifluorescence configuration. The fluorescence signal was separated from the back-scattered
excitation light by using a dichroic mirror. Residual stray light was removed by using
a short-wavelength-pass filter with cut at 700 nm. The fluorescence was delivered to
the detector, through a 1 mm diameter multimode fiber. With this setup, appreciably
superPoissonian PCH distributions were obtained for <300 nM dye solutions. Finally, a series
of feasibility studies on the same dye were performed by using as the detector a Hamamatsu
SiPM (MPPC S10362-11-100C). Given the 48 MHz repetition rate of the laser light source,
the SiPM is expected to detect a fluorescence light pulse every 20 ns. In such experimental
conditions, the usual detector signal processing seemed unfeasible with the instrumentation
at hand by both the integration method and the free-time signal acquisition method. Indeed,
in the first case the CAEN V792N QDC used for the quantum-light statistics measurements
presented above requires an integration gate starting at least 15 ns before the rise of the
signal, meaning that the board would not be able to process every pulse; moreover, with
a detector recovery time of ∼ 60 ns, the output signal may exhibit pile-up features, which
95
19
Silicon
Photomultipliers:
Characterization and Applications
Characterization
and Applications
Silicon Photomultipliers:
in turn would lead to difficulties in reconstructing the photon statistics. In the second case,
the DT5720A Desktop Digitizer at our disposal samples the output signal every 4 ns, which
means that the signal integration would be performed using only 5 points. For these reasons,
a completely different approach to process the detector signal was followed. Namely, the
coincidence frequency ν( x ) between a threshold scan of the SiPM output and a trigger signal
νT synchronous with the laser pulses (but with a threefold reduced frequency) has been
acquired. The measured frequency can be described by the following relation:
ν( x ) = νT 1 − e−tC νS
+∞
x
dyP ( y )
+ νT tC νS
+∞
x
dyP (y);
(28)
where P (y) is the probability of measuring a value y of the detector output signal, ν( x ) is the
threshold value, νS is the laser repetition rate and tC is the coincidence time window. Equation
28 is based on the fact that a coincidence occurs both if after a trigger signal at least a SiPM
signal over the threshold x is triggered within the coincidence time window or if the trigger
signal rises within the coincidence time after the SiPM signal crosses the threshold. For small
expectation values, Equation 28 can be approximated to
ν( x ) ≈ 2νT νS tC
+∞
x
dy P (y).
(29)
Because of the Fundamental Theorem of Integration the detector output signal probability
distribution can be written as
−1 dν( x )
P(x) =
(30)
2νT νS tC dx
which can be processed in the same way as the experimental detector output signal histograms
experimentally obtained in the case of quantum-light statistics measurements to yield the
probability distribution of the number of detected photons (see Section 3). In this set of
measurements, the detector was controlled by the CAEN SP5600 General Purpose Power
Supply and Amplification Unit module with an integrated threshold discriminator. The
digital NIM output of the discriminator was put in coincidence with a synchronism signal
triggered by the laser by using an EG&G CO4010 module. The coincidence frequency was
sampled by a CAEN VME V1718. The fluorescence emission of two solutions of rhodamine
B in water, with concentrations of 2 μM and 200 nM, respectively, have been analyzed with
the threshold scan method described above. For the 2 μM concentrated solution a 1 optical
density neutral filter was placed in front of the fiber, to reduce the mean detected fluorescence
intensity to the value of the 200 nM concentrated solution. This makes the comparison
between the two resulting distributions easier. Indeed, for equal expectation value, the 200 nM
solution PCH should feature larger deviations from Poisson distribution, therefore showing
a higher tail. Hereafter, the pertaining results are briefly summarized and their significance
on the way towards implementation of PCH with SiPMs is outlined. Using Equation 30 the
multi-peak spectra reported in Fig. 14 were obtained from the staircases for the 200 nM and
the 2 μM solution, respectively. The PCH were derived from the spectra by applying the
same procedure implemented in the quantum-light statistics data analysis. The zero-photons
frequency was determined by imposing the normalization condition:
Π (0) = 1 −
L
∑ Π (k)
k =1
(31)
96
Photodetectors
Will-be-set-by-IN-TECH
20
histo
Entries
38
Mean
26.62
RMS
5.97
200 nM
histo
10-3
10-4
10-5
10-6
20
30
40
50
60
70
80
Threshold (mV)
Fig. 14. Multipeak spectra of the fluorescence obtained from the threshold scans using
Equation 30; in the upper panel spectrum from the 200 nM solution is shown, while in the
lower panel the one obtained from the 2 μM solution is presented. Since the staircases has
been acquired with a 2 mV step, histograms have 2 mV large bins. Spectra have been fitted
with a multi-Gaussian function and the areas of each peak have been used as the PCH
entries.
where L is the last detectable peak. The PCH are displayed in Figs. 15, for the 200 nM and
the 2 μM solution, respectively. Fitting to Equation 27, with the average number of dark
counts fixed to the experimental value preliminarily determined by performing a coincidence
measurement of the dark counts, yields the fitting curves reported as dashed lines in Figs. 15.
The fitting parameters are summarized in Table 4.
Parameters
N̄
xt
n dcr
Concentrations (nM)
200
2000
5.24 ± 0.03
71.70 ± 0.03
3.99 ×10−3 ± 2.4 ×10−5 1.536 ×10−4 ± 7 ×10−7
0.08085 ± 5×10−5
0.0637 ± 1 ×10−4
−
3
3.24 ×10
3.24 ×10−3
Table 4. Results for the fit of the PCH distributions. The errors listed are obtained from the fit
procedure. The results for the mean number of particles and the molecular brightness are
consistent with what expected, even though there is not exactly a factor 10 between the
parameters of the two data sets.
97
21
Silicon
Photomultipliers:
Characterization and Applications
Characterization
and Applications
Silicon Photomultipliers:
data
Entries
4
Mean
0.02617
RMS
0.1748
data
1
10-1
10-2
10-3
10-4
-0.5
0
0.5
1
1.5
2
2.5
3
3.5
data
4
4.5
Photons
data
Mean 0.01519
1
RMS
0.1313
10-1
10-2
10-3
10-4
-0.5
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
Photons
Fig. 15. Fit (red dashed line) of the experimental PCH data (blue continuous line) for 200 nM
concentration (upper panel) and 2 μM concentration (lower panel).
The brightness and average particle number values are fairly consistent with the experimental
conditions. Namely, the mean numbers of particles in the excitation volume differ by
approximately one order of magnitude, reflecting the 10:1 concentration ratio between the
two samples. To this purpose it should be noted that, while the concentration of the 2
μM solution could have been determined spectrophotometrically, 200 nM represents only a
nominal value of the concentration of the less concentrated sample, obtained by dilution of
the first, as the absorbance of it is below the sensitivity of our spectrophotometer (≈ 0.001) in
this case. Also the difference in molecular brightness is roughly consistent with the insertion
of the 1-optical-density filter. Indeed, the fact that the molecular brightness measured for the
2 μM concentrated sample is found to be more than twenty times lower could be ascribed to a
slight misalignment of either the laser or the microscope objective during the measurement
session. The latter lasted for many hours due to the necessity of determining with high
statistical reliability the very low frequencies of coincidences occurring at a high threshold
value. Even a slight degeneration in the quality of the optical setup alignment would
determine a notable decrease in the two-photon absorption cross section, which scales as the
square power of the excitation intensity. This would be reflected in the brightness value.
A broader focus is also consistent with the mean number of molecules in the excitation
volume detected for the 2 μM solution, which is slightly larger than expected. Finally, the
difference between the cross-talk values is consistent with a lower over-biasing condition for
the 2 μM set, suggesting that a worsening of the photon detection efficiency could also have
contributed to the lower molecular brightness. It must be admitted that the above presented
procedure is very far from ideal pulse-by-pulse PCH acquisition. However, these preliminary
data demonstrate that despite the relatively high dark count rates with respect to APDs
98
22
Photodetectors
Will-be-set-by-IN-TECH
and the additional artifice constituted by optical cross-talk, SiPMs are detectors capable of
discriminating the tiny deviations from Poisson distribution typically displayed by the PCH
of diffusing fluorophores in solution at concentrations relevant for biophysical applications.
Moreover, we notice that our fitting model yields consistent values of brightness and average
fluorophore number. The above conclusions suggest that pulse-by-pulse reconstruction
of PCH should be straightforwardly obtained with our detector and electronics apparatus
for excitation rates <10 MHz, according to both the recovery time of the SiPM detector
and the requirements of the CAEN integrator. An optical pulse-picker may be used to
customize the Ti:sapphire repetition rate. Even working at such rates would allow to speed
up PCH reconstruction by more than two orders of magnitude with respect to state-of-art
APD-based methods. Alternatively, high-repetition rate measurements might be performed
by reconstruction methods based on the analysis of the detector free-run output signal, as
recorded by fast-sampling-rate digitizers. In such a case, the issues connected with detector
pile-up effects would be overcome, thus confining the limits on the acquisition rate to the
excitation rate and to the intrinsic decay time of the fluorophores under analysis.
5. Conclusions
We have presented the characterization of a class of photodetectors, namely the Silicon
Photomultipliers, that has acquired great importance in the last decades because of its good
photon-counting capability. Besides this feature, the compactness, robustness, low cost,
low operating voltage and power consumption are added value that would make these
detectors suitable for many applications. To this aim, we have implemented an exhaustive
characterization protocol providing a quantitative evaluation of the main figures of merit
these detectors are endowed with. In particular, we have studied the dependence of the
photon detection efficiency, gain, dark count rate and optical cross-talk on temperature. The
protocol has been applied to three SiPM detectors produced by different manufactures and the
results have been then employed in a partnership with CAEN to develop the SP5600 General
Purpose Power Supply and Amplification Unit module.
The versatility of SiPMs has been investigated in the reconstruction of the statistics of light
states: two complementary models that take into account the DCR and the cross-talk effects
have been developed in order to describe the output signal of the detectors. The correctness
of these methods has been experimentally validated by illuminating the SiPM sensor with
a single-mode pseudo-thermal light. However, further improvements could be achieved by
including in the models the afterpulsing effect, which would become important in high-rate
events.
The applicability of SiPMs in Fluorescence Fluctuation Spectroscopy studies has been
preliminary evaluated by analyzing Rhodamine solutions by means of a Two-Photon
Excitation setup: the capability of the sensors to reconstruct Photon Counting Histogram and
in particular to provide the estimation of the mean number of fluorophores in the excitation
volume and their molecular brightness has been proved.
6. References
Afek, I.; Natan, A.; Ambar, O.; & Silberberg, Y. (2009). Quantum state measurements using
multipixel photon detectors. Phys. Rev. A., Vol. 79, No. 4, April 2009, 043830, ISSN
1050-2947.
Silicon
Photomultipliers:
Characterization and Applications
Characterization
and Applications
Silicon Photomultipliers:
99
23
Agliati, A.; Bondani, M.; Andreoni, A.; De Cillis, G.; & Paris, M. G. A. (2005). Quantum and
classical correlations of intense beams of light investigated via joint photodetection.
J. Opt. B Quantum Semiclass. Opt., Vol. 7, No. 12, November 2005, S652–S663, ISSN
1464-4266.
Akindinov, A. V.; Martemianov, A. N.; Polozov, P. A.; Golovin, V. M.; & Grigoriev, E. A. (1997).
New results on MRS APDs. Nuclear Instruments and Methods in Physics Research Section
A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 387, No. 1-2,
March 1997, 231–234, ISSN 0168-9002.
Akindinov, A; Bondarenko, G.; Golovin, V.; Grigoriev, E.; Grishuk, Yu.; Malkevich, D.;
Martemiyanov, A.; Ryabinin, M.; Smirnitskiy, A.; & Voloshin, K. (2005). Scintillation
counter with MRS APD light readout. Nuclear Instruments and Methods in Physics
Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol.
539, No. 1-2, February 2005, 172–176, ISSN 0168-9002.
Andreoni, A. & Bondani, M. (2009). Photon statistics in the macroscopic realm measured
without photon counters. Phys. Rev. A, Vol. 80, No. 1, July 2009, 013819, ISSN
1050-2947.
Arecchi, F. T. (1965). Measurement of the Statistical Distribution of Gaussian and Laser
Sources. Phys. Rev. Lett., Vol. 15, No. 24, December 1965, 912–916, ISSN 0031-9007.
Bondani, M.; Allevi, A.; & Andreoni, A. (2009). Light Statistics by Non-Calibrated Linear
Photodetectors. Adv. Sci. Lett., Vol. 2, No. 4, December 2009, 463–468, ISSN 1936-6612.
Bondani, M.; Allevi, A.; Agliati, A.; & Andreoni, A. (2009). Self-consistent characterization
of light statistics. J. Mod. Opt., Vol. 56, No. 2, January 2009, 226–231, ISSN 0950-0340
print/ 1362-3044 online.
Chen, Y.; Müller, J. D.; So, P. T. C.; & Gratton, E. (1999). The Photon Counting Histogram in
Fluorescence Fluctuation Spectroscopy. Biophys. J., Vol. 77, No. 1, July 1999, 553–567,
ISSN 0006-3495.
Chen, Y.; Tekmen, M.; Hillesheim, L.; Skinner, J.; Wu, B., & Muller, J. D. (2005). Dual-color
photon counting histogram. Biophys. J., Vol. 88, No. 3, March 2005, 2177–2192, ISSN
0006-3495.
Cova, S; Ghioni, M.; Lacaita, A.; Samori, C.; & Zappa, F. (1996). Avalanche photodiodes and
quenching circuits for single-photon detection. Appl. Opt., Vol. 35, No. 12, April 1996,
1956–1976, ISSN 1559-128X print/ 2155-3165 online.
Du Y. & Reti‘ere, F. (2008). After-pulsing and cross-talk in multi-pixel photon counters. Nuclear
Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers,
Detectors and Associated Equipment, Vol. 596, No. 3, November 2008, 396–401, ISSN
0168-9002.
Eckert, P.; Schultz-Coulon, H.-C.; Shen, W.; Stamen, R.; & Tadday, A. (2010). Characterisation
studies of silicon photomultipliers. Nuclear Instruments and Methods in Physics
Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol.
620, No. 2-3, August 2010, 217–226, ISSN 0168-9002.
Goetzberger, A.; McDonald, B.; Haitz, R. H.; & Scarlett, R. M. (1963). Avalanche Effects in
Silicon p-n Junctions. II. Structurally Perfect Junctions. Journal of Applied Physics, Vol.
34, No. 6, June 1963, 1591–1600, ISSN 0021-8979.
Gösch, M. & Rigler, R. (2005). Fluorescence correlation spectroscopy of molecular motions and
kinetics. Adv. Drug. Del. Rev., Vol. 57, No. 1, January 2005, 169–190, ISSN 0169-409X.
100
24
Photodetectors
Will-be-set-by-IN-TECH
Lutz, G. (1995). Silicon radiation detectors. Nuclear Instruments and Methods in Physics Research
Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 367, No.
1-3, December 1995, 21–33, ISSN 0168-9002.
Mandel, L. & Wolf, E. (1995). Optical Coherence and Quantum Optics, Cambrige University Press,
ISBN 0521417112, New York.
McKay, K. G. (1954). Avalanche Breakdown in Silicon. Phys. Rev., Vol. 94, No. 4, May 1954,
877-884.
Nettels, D.; Gopich, I. V.; Hoffmann, A.; & Schuler, B. (2007). Ultrafast dynamics of protein
collapse from single-molecule photon statistics. Proc. Nat. Acad. Sci. USA, Vol. 104,
No. 8, February 2007, 2655–2660, ISSN 0027-8424.
Oldham, W. G.; Samuelson, R. R.; & Antognetti, P. (1972). Triggering phenomena in avalanche
diodes. IEEE Trans. Electron. Dev., Vol. 19, No. 9, September 1972, 1056-1060, ISSN
0018-9383.
Ramilli, M.; Allevi, A.; Chmill, V.; Bondani, M.; Caccia, M.; & Andreoni, A. (2010).
Photon-number statistics with Silicon photomultipliers. J. Opt. Soc. Am. B, Vol. 27,
No. 5, May 2010, 852–862, ISSN 0740-3224.
Schwille, P. (2001). Fluorescence correlation spectroscopy and its potential for intracellular
applications. Cell Biochem. Biophys., Vol. 34, No. 3, June 2001, 383–408, ISSN 1085-9195.
Sciacca, G.; Condorelli, E.; Aurite, S.; Lombardo, S.; Mazzillo, M.; Sanfilippo, D.; Fallica, G.; &
Rimini, E. (2008). Crosstalk characterization in Geiger-mode avalanche photodiode
arrays. IEEE Electron. Dev. Lett., Vol. 29, No. 3, March 2008, 218-220, ISSN 0741-3106.
Tarolli, A.; Dalla Betta, G.-F.; Melchiorri, M.; Piazza, A.; Pancheri, L.; Piemonte, C.; & Zorzi,
N. (2010). Characterization of FBK SiPMs under illumination with very fast light
pulses. Nuclear Instruments and Methods in Physics Research Section A: Accelerators,
Spectrometers, Detectors and Associated Equipment, Vol. 617, No. 1-3, May 2010, 430–431,
ISSN 0168-9002.
Van Der Marel, R. P. & Franx, M. (1993). A new method for the identification of non-gaussian
line profiles in elliptical galaxies. Astrophysical Journal, Vol. 407, No. 2, April 1993,
525–539, ISSN 0004-637X.
Zambra, G.; Bondani, M.; Spinelli, A. S.; Paleari, F.; & Andreoni, A. (2004). Counting
photoelectrons in the response of a photomultiplier tube to single picosecond light
pulses. Rev. Sci. Instrum., Vol. 75, No. 8, August 2004, 2762–2765, ISSN 0034-6748