Abstract
A potential new laser-based air pollution measurement technique, capable of measuring ultralow concentrations of urban air toxins in the field and in real time, is examined. Cavity ringdown laser absorption spectroscopy (CRLAS) holds promise as an air pollution monitor because it is a highly sensitive species detection technique that uses either pulsed or continuous tunable laser sources. The sensitivity results from an extremely long absorption path length and the fact that the quantity measured, the cavity decay time, is unaffected by fluctuations in the laser source. In laboratory experiments, we reach detection limits for mercury of the order of 0.50 parts per trillion. We developed a CRLAS system in our laboratory and measured Hg with the system, investigating issues such as background interference. We report experimental results for mercury detection limits, the dynamic range of the sensor, detection of Hg in an absorbing background of ozone and SO2, and detection of a mercury-containing compound (HgCl2 in this case).
© 2000 Optical Society of America
Full Article | PDF ArticleMore Like This
Andreas Karpf, Yuhao Qiao, and Gottipaty N. Rao
Appl. Opt. 55(16) 4497-4504 (2016)
Takuya Nayuki, Kohji Marumoto, Takashi Fujii, Tetsuo Fukuchi, Koshichi Nemoto, Akira Shirakawa, and Ken-ichi Ueda
Appl. Opt. 43(35) 6487-6491 (2004)
Hans Edner, Gregory W. Faris, Anders Sunesson, and Sune Svanberg
Appl. Opt. 28(5) 921-930 (1989)