Abstract
The ability to measure subtle changes in arterial pressure using devices mounted on the skin can be valuable for monitoring vital signs in emergency care, detecting the early onset of cardiovascular disease and continuously assessing health status. Conventional technologies are well suited for use in traditional clinical settings, but cannot be easily adapted for sustained use during daily activities. Here we introduce a conformal device that avoids these limitations. Ultrathin inorganic piezoelectric and semiconductor materials on elastomer substrates enable amplified, low hysteresis measurements of pressure on the skin, with high levels of sensitivity (~0.005 Pa) and fast response times (~0.1 ms). Experimental and theoretical studies reveal enhanced piezoelectric responses in lead zirconate titanate that follow from integration on soft supports as well as engineering behaviours of the associated devices. Calibrated measurements of pressure variations of blood flow in near-surface arteries demonstrate capabilities for measuring radial artery augmentation index and pulse pressure velocity.
Similar content being viewed by others
Introduction
Research over the last decade has established inorganic nanoparticles1,2,3, nanowires4,5,6 and nanomembranes7 as materials for high-quality electronic devices on unusual types of substrates, including plastic foils and rubber sheets8. A key opportunity and a realm ripe for further development and growth follows from an ability to integrate such technologies with the soft, curved contours of the human body9. Although semiconductor devices such as light-emitting diodes, solar cells and transistors have received much attention, application requirements often demand co-integration of other types of active materials and classes of components. Lead zirconate titanate (PbZr0.52Ti0.48O3, PZT), as one example, is a ferroelectric/piezoelectric material for ultrasonic transducers10, microelectromechanical devices and actuators11, as well as pressure12 and strain13 sensors, owing to its large piezoelectric and electromechanical coupling coefficients, high dielectric permittivity and significant remnant polarization14. Force sensing, mechanical energy harvesting and actuation represent promising possible roles in stretchable systems that interface with the body. Recent studies show the ability to build PZT-based piezoelectric systems for uses ranging from the measurement of deformations of neuronal cells15 to harvesting of electrical power from motions of the heart, lung and diaphragm16. Alternative options include barium titanate17, zinc oxide18,19 and other inorganics in nanostructured forms.
An important, unexplored area for application of such materials is in the continuous monitoring of pressure transients associated with arterial blood flow. The traditional cuff sphygmomanometer is one of the oldest and most established diagnostic tools for blood pressure, but it is not well suited for continuous monitoring and yields only two basic pressure values: systolic (greatest heart contraction) and diastolic (greatest heart dilation)20. As a result, capabilities for use in observation during hypertensive treatment, in evaluation for hemodynamic effects of atherosclerotic risk factors or in projections of life-threatening cardiovascular events21 are limited. A popular alternative, known as arterial tonometry, uses arrays of transducers that press against a radial artery to measure external blood pressure22. The transducers consist of dielectric capacitor structures formed in chip-scale geometries integrated with processing electronics. Pressure waves associated with blood flow appear as mechanically induced changes in capacitance. Enhanced embodiments exploit films of PZT in capacitor structures that connect to silicon metal oxide semiconductor field effect transistors (MOSFETs) for signal manipulation23. A drawback of such technologies for use on the skin follows from their planar, rigid formats. By consequence, monitoring schemes often require straps, tapes, pins or other mechanical fixtures that can cause irritation and discomfort during prolonged use.
Another device architecture that provides improved mechanics and potential for measurement capabilities in this context uses polypropylene ferroelectret films bonded by epoxy to the gate electrodes of amorphous silicon thin-film transistors on flexible polyimide substrates24. This structure represents one of a variety of devices based on organic materials for the semiconducting25,26,27,28 and/or piezoelectric29 components. Others exploit conductive elastomers30,31, triboelectric effects32 and air-gap capacitors33, with and without integrated organic transistors, to offer simple construction and natural compatibility with the skin. Systems that incorporate organic active materials, however, typically have slow response times, strongly temperature-dependent behaviours, uncertain fatigue and lifetime properties, sensitivity to parasitics and/or high operating voltages.
Here we present inorganic materials, heterogeneous designs and theoretical models for an ultrathin, compact device capable of softly laminating on the skin with capabilities and modes of use that lie outside of the scope of the above-mentioned approaches. These systems are small (~1 cm2), lightweight (2 mg), thin (25 μm) and capable of stretching (to ~30%, with system-level effective modulus of ~60 kPa) to conform to the skin, while providing high levels of pressure sensitivity (~0.005 Pa), fast response times (~0.1 ms), low hysteresis, superior operational stability and excellent fatigue properties. Ultrathin (400 nm) sheets of high-quality PZT serve as the active components of capacitor-type structures that connect to the gate electrodes of MOSFETs based on nanomembranes of silicon (SiNMs). Specifically, a SiNM n-channel MOSFET (that is, n-MOSFET) amplifies the piezoelectric voltage response of the PZT and converts it to a current output via capacitance coupling. Comprehensive electromechanical measurements and theoretical models provide complete descriptions of the principles of operation, including enhanced piezoelectric responses in PZT when mounted on soft substrates. Applications of the technology include skin-mounted sensors for measuring arterial pulse pressure waves and evaluating subtle motions of the throat associated with speech. Quantitative correlations of data from the former class of measurement to those of conventional devices suggest opportunities in continuous, non-invasive monitoring of pressure transients associated with arterial blood flow.
Results
Fabrication and design
The sensor (Fig. 1a) consists of an array of square elements of PZT, collectively connected to the gate electrode of an adjacent SiNM n-MOSFET (inset; Fig. 1a), all mounted on a thin elastomer substrate of silicone (Smooth-on, Ecoflex 0030, Platinum Cure Silicone Rubber), as further illustrated in Supplementary Fig. 1a. The n-MOSFET consists of heavily phosphorus-doped regions of the SiNM for source and drain, and a lightly boron-doped region for the channel. A cross-sectional view and simplified equivalent circuit appear in Fig. 1b,c, respectively. Figure 1d shows a photograph of a typical integrated device in which serpentine conducing traces (optical micrograph shown in inset, top right of Fig. 1d) provide an ability to stretch the overall system. Anisotropic conductive film (ACF; Ellsworth Adhesives) serves as an interface to external power supplies and measurement systems, for application of voltage on the gate (VGS) and drain (VDS), and evaluation of current between drain and source (IDS). The conductive back side of the anisotropic conductive film cable is isolated with a thin layer of poly(dimethylsiloxane) (PDMS). A photograph of a device laminated on the wrist appears in Fig. 1e.
Device sensing properties
Because the overall device involves coupled operation of two component parts, their separate characterization represents a natural starting point for analysis of the behaviours and underlying physics. Application of calibrated weights (post structures made of PDMS) to the PZT elements after coating their top surfaces with an electrode enables measurement of their responses to controlled pressures between 1 and 10 Pa, as in Supplementary Fig. 2. Figure 2a shows the voltages (VPZT) recorded for PZT devices built on a silicon wafer and on a silicone slab, both formed after transfer printing PZT films from their growth substrate. Multiple (50) measurements of the dependence of VPZT on applied pressure for a PZT sensor on silicone establish the high level of repeatability in the response (Supplementary Fig. 3). It is immediately seen that the PZT devices on the silicone substrate offer responses that are more than two orders of magnitude greater than those of otherwise identical devices on silicon. Theoretical analysis yields a quantitative explanation. The dependence of the piezoelectric voltage on the contact force can be written (Supplementary Note 1)
where P is the contact pressure, Acontact is the contact area between the PZT element and the weight, and α is a parameter that depends on the materials constants for PZT, as well as the deformation mechanics of each substrate. For the case of the wafer, deformation in the PZT involves approximately uniaxial compression along the thickness direction, as illustrated in Fig. 2b obtained using the Finite Element Method (FEM). This circumstance corresponds to , where tPZT is the thickness of PZT, A is the total area of PZT, including regions in direct contact with the weight and those not in contact, and e33, c33 and k33 are the piezoelectric, elastic and dielectric constants, respectively (see Supplementary Note 1 for details). FEM results confirm this relationship34 (Supplementary Fig. 4). For devices on silicone, the deformation in the PZT is complex and three-dimensional, as shown in Fig. 2c. Subsequently, the coefficient α is computed using FEM, giving a value of 0.068 V/N for the device layout (Supplementary Fig. 5a,b). For A=6 × 2 mm2 as in experiments, Fig. 2a shows that equation (1) agrees well with measurements for both the silicon and silicone substrates, without any parameter fitting. This good agreement holds regardless of the contact location of the PDMS post on the array of PZT elements (Supplementary Note 1 and Supplementary Fig. 5c–g). The key finding is that the induced voltage associated with PZT on silicone is much larger than that on Si owing to the different modes of deformation. This is important as it yields approximately a 100-fold enhancement in sensitivity simply as a consequence of integrating devices of this type on compliant substrate, which is, in any case, required to allow soft lamination on the skin. We also note that the responses of the PZT elements associated with the pyroelectric effect (Supplementary Fig. 6) are nearly two orders of magnitude smaller than the piezovoltages reported in Fig. 2a, for ranges of both temperature (25–35 °C) and pressure (4–10 Pa) that are relevant for cutaneous monitoring. As a result, effects of temperature can be ignored in the following.
The piezoelectric voltage response from the PZT elements is further amplified by its integration with the flexible SiNM n-MOSFET. Figure 2d reveals IDS–VGS curves of representative transistors (channel width and length of 200 and 20 μm, respectively) that were measured in the common source configuration with a semiconductor parameter analyser (HP 4155C, Agilent) for values of VGS between 0 to 5 V in steps of 0.1 V at a VDS of 0.1 V. The full response corresponds to outputs measured at VDS from 0 to 5 V in steps of 0.1 V at different VGS (Supplementary Fig. 7). Similar measurements on devices with channel widths of 200, 400, 600 and 800 μm appear in Supplementary Fig. 7. Extracted electron mobilities lie in the range of 350–400 cm2 V−1 s−1 and the on/off ratios exceed 105. Unless otherwise noted, all of the results in the following use SiNM n-MOSFETs with 800 μm channel widths.
The nature of the coupling between the PZT elements and the n-MOSFET can be understood by reference to the diagram in Fig. 1b and the simplified equivalent circuit in Fig. 1c. During use of the sensor, the n-MOSFET operates in an active mode, with VGS set to 3.5 V through a 100-MΩ resistor and VDS set to 0.1 V. This condition yields a transconductance (gm) of ~11 μA V−1. Applying pressure to the PZT array generates a voltage owing to a decrease in the spontaneous polarization of PZT (δP) and a concomitant change in the compensating surface charge (δq), as indicated in Fig. 1b. Since the bottom electrode is shared with the gate of the n-MOSFET, redistribution of surface charge appears as a change in voltage on the gate (ΔVGS), which is amplified by the action of the n-MOSFET. The equivalent circuit (Fig. 1c) involves a PZT capacitor in series with the gate oxide capacitor of the n-MOSFET. For such a series connection, modulation of the gate bias (ΔVGS) with applied pressure is therefore related to VPZT as
where CPZT and CGox are the capacitances associated with the PZT and the gate oxide, respectively. For the devices examined here, CPZT/CGox=0.15 × 106 (see Supplementary Note 2 for details). The change in current flow from drain to source is therefore
Equations (1)–(3), , yield ΔIDS due to applied pressure as ΔIDS,pressure according to
Measured values of ΔIDS,pressure under various applied pressures for a n-MOSFET with 200 μm channel width (gm=11 μA V−1; Supplementary Note 3) appear in Fig. 2e. Equation (4) gives a linear relationship between ΔIDS,pressure and pressure, with a slope of ΔIDS,pressure/P=1.36 μA Pa−1. Operation at various VGS (1, 2 and 3 V) yields consistent results, as a consequence of the nearly constant value of gm over this range (Fig. 2d). The device also exhibits low hysteresis behaviour, as seen in Supplementary Fig. 8a. Repeated application and release of pressure using the same setup (Supplementary Fig. 2) show stable behaviour over 1,000 cycles (Supplementary Fig. 8b). Scaling studies of responses of separate sensors using SiNM n-MOSFETs with different channel widths (200, 400, 600 and 800 μm), and a fixed channel length (20 μm) indicate expected behaviours (Supplementary Fig. 9).
An audio speaker system (iHM79; Supplementary Fig. 10) enables examination of the sensor response under low pressures and high frequencies. Sound decibel levels (Radio Shack, 33–2055) recorded at the position (~1 cm from the speaker) of the sensor define the pressures (Supplementary Note 4). Figure 2e shows a linear relationship between the sensor response and pressure, with a slope of ~2 μA Pa−1. The IDS recorded during exposure to a middle C note (1 kHz, 80 dB) appears in Fig. 2f. The frequency response is the same over the audio range, as shown in Fig. 2g. The region highlighted by the red dashed box of Fig. 2h (IDS measured using the sensor as a function of time during exposure of 1 kHz, 80 dB sound by a stereo speaker) is given in Fig 2i. The response time is ~0.1 ms (Fig. 2i), which exceeds requirements for pressure monitoring of blood vessels of the body. As illustrated in Supplementary Fig. 8c, the response remains the same under 1 kHz, 80 dB sound for 24 h.
Mechanical stretchability and flexibility
Serpentine traces for gate, drain and source and a silicone substrate allow stretching at the system level, without affecting the performance of the PZT arrays or the SiNM n-MOSFET. Supplementary Fig. 11 illustrates the ability of the device to accommodate 30% uniaxial strain, without any degradation in operating characteristics. Optical images and mechanics modelling of the serpentine traces illustrate reversible buckling during deformation, in a way that minimizes strains in the materials. For stretching up to 25%, the computed maximum principal strains in the Au are less than the failure strains (~5% for Au; ref. 35). At 30%, both experiment and simulation indicate fracture of certain parts of the serpentine (Supplementary Fig. 11f).
Sensors with this construction can wrap easily around a finger or a pen, as in Fig. 3a,b, respectively. Bending-induced strains can, of course, influence the response of the sensor; such effects are important to understand, as they can influence the operation when mounted on the skin. Studies that involve mounting a sensor on cylindrical supports with different radii reveal the change in current flow from drain to source due to bending, ΔIDS,bending, as a function of tube radius (Fig. 3c). Theory suggests that it is possible to write (see Supplementary Notes 5 and 6 for details)
where ē=e31−(c13/c33)e33 and are effective piezoelectric and dielectric constants, tPZT is the thickness of PZT element, R is the bend radius, h is the distance from the neutral mechanical plane (NMP) to the mid-plane of the PZT element (Fig. 3d and Supplementary Fig. 12), where η depends on the bending stiffness ratio in the sections with and without the PZT elements and is given in the Supplementary Note 5. The coefficient in equation (5) is for gm=33.2 μA Pa−1 and CGox=3.72 pF; CPZT=0.14 μF; tPZT=400 nm; and h=1.19 nm, ē=−34 C m−2, , and η=327 obtained from Supplementary Note 5. As shown in Fig. 3c, equation (5) agrees reasonably well with experimental results, without any parameter fitting, and both show that ΔIDS,bending is inversely proportional to the bend radius R (Fig. 3c). Equation (5) implies a means to eliminate the effect of bending on pressure measurement from the ΔIDS,pressure–P relation in equation (4). In particular, with an appropriate value of tPI, h can be zero (that is, the NMP coincides with the mid-plane of the PZT layer), and therefore ΔIDS,bending=0. For the material properties and thicknesses used in experiments, a plot of h versus tPI (Fig. 3d) suggests that h=0 occurs when tPI=3.59 μm, which is close to the value used in experiments (tPI=3.6 μm). For this tPI, Fig. 3e shows ΔIDS,pressure as a function of P for various outward (Out) and inward (In) bending radii. The data establish a linear relationship between ΔIDS,presssure and P with a coefficient of proportionality consistent with that in Fig. 2f. This behaviour is the same for all bend radii, ranging from 14.8 to 50.5 mm in Fig. 3c, that are larger than the in-plane dimensions of each PZT element (557 × 625 μm2, spacing of 170 μm). The conclusion is that the devices can measure pressure accurately on any body part with a radius of curvature larger than a few millimetres (for example the wrist). For sensors that do not use an NMP design, the relationship between ΔIDS,pressure and P remains linear, but with different coefficients of proportionality (Supplementary Fig. 12b).
Blood pressure wave measurements and vibration detection
Photographs of a sensor on the wrist appear in Fig. 4a,b. Demonstrations involved a healthy female in her late twenties, under approval by the Institutional Review Board of the University of Illinois (IRB no. 13561). No allergic reactions, redness or damage to the skin were observed in any of our studies. At a gate voltage of 3.5 V, without variation in pressure, the value of IDS is constant over time, as shown in Supplementary Fig. 13. Changes in IDS result from variations in pressure associated with blood flow, as in Fig. 4c. The blood vessel examined here covers an area of ~0.6 × 0.2 cm2 that affects ~20 PZT elements in the array. The region indicated by the dashed box of Fig. 4c appears in Fig. 4d. The peak labelled P1 is the sum of the incident (ejected) wave and reflected wave (from the hand); P2 is the peak of the reflected wave from the lower body minus end-diastolic pressure; P2/P1 is the radial artery augmentation index (AIr); and ΔTDVP is the time difference between these two peaks and represents a measure of arterial stiffness36. The measured values of AIr ~0.45 and ΔTDVP ~0.2 s are compatible with a person in their late twenties, as reported by Nichols via tonometry37. The device of Fig. 4a,b can also determine small variations in the diastolic tail of the pulse pressure wave; such features provide additional detailed information and are undetectable with conventional arterial tonometry36. The ΔIDS,pressure for P1 corresponds ~4 Pa, with similar responses from sensors that use SiNM n-MOSFETs with different channel widths (400, 600 and 800 μm; Supplementary Fig. 14). As a further check of behaviour, poling the PZT in the opposite direction (Supplementary Fig. 15) yields waveforms that have the same magnitudes but with forms inverted compared with those observed in Fig. 4c and Supplementary Fig. 14.
The fast response time allows sensing of blood pressure during transient events, such as those associated with cuffing of the arm. The data of Fig. 4e correspond to measurements on the wrist between 0 and 2 s, cuffing the arm (Manual Inflate Blood Pressure Kit, Walgreens, model WC010) between 2 and 8 s and then releasing the cuff. The region indicated by the dashed box of Fig. 4e appears in Fig. 4f. P2/P1 and ΔTDVP are ~0.45 and 0.2 s, respectively, consistent with Fig. 4d. Furthermore, the sensor can be placed on nearly any location on the body, including the neck (Fig. 4g) and throat (Fig. 4h), to allow measurement not only of blood pressure (Fig. 4i) but also vibrations associated with speech (Fig. 4j). The former shows that the peak shape of Fig. 4i is more rounded and the magnitude of P1 is higher by ~75% than that in Fig. 4c due to carotid artery blood pressure37.
Long-term pressure wave monitoring
The complete pulse pressure waveform is known to provide valuable information for diagnostics and therapy of cardiovascular diseases such as arteriosclerosis, hypertension and left ventricular systolic dysfunction37,38,39,40. Many hemodynamic parameters such as arterial index, upstroke time, stroke volume variation and cardiac output can be directly calculated or estimated in real time from the pressure waveforms41,42. The type of skin-mounted sensor reported here provides a promising path towards long-term diagnostics based on this measurement. Figure 5a shows measured waveforms, normalized to the first measured response, for weekly measurements over the course of a month. Corresponding blood monitoring results (an average systolic pressure of 110 mm Hg and diastolic pressure of 65 mm Hg; Blood Pressure Dock BP3, iHealth Lab Inc.) are shown in Supplementary Fig. 16. Systolic, diastolic and pulse pressure conversion factors are ~0.81, ~0.50 and ~9 mm Hg μA−1, as shown in Fig. 5b. These conversion factors correspond to the observed IDS values as systolic (~135 μA), diastolic (~130.5 μA) and pulse pressure (~4.5 μA). The collective results in Supplementary Fig. 17 suggest that the sensor is robust and suitable for long-term monitoring. Data on three human subjects (each in their early twenties) were collected with a 2-min interval in between four consecutive measurements, while the subjects were in a relaxed state and the blood pressure remained relatively constant. A analysis summary appears in Supplementary Fig. 17d,e. Supplementary Figure 18a–c shows measured waveforms, normalized to the first measured response, for weekly measurements on these three subjects. The repeatability in the pulse wave velocity (PWV) and the AIr (P2/P1:(AIr)) is consistent with insignificant variations in blood pressure over this same period (measured with a cuff device). Simultaneous use of multiple sensors at different locations enables determination of the PWV. PWV is accepted as the most simple, robust and reproducible method to determine the regional arterial stiffness43. PWV is important because it can be quantitatively related to BP when calibrated using a conventional blood pressure cuff measurement device44. Calculating the PWV involves measurements at two sites with a known distance between them. This distance divided by the time separation associated with a given feature (for example, P1) in the waveform defines the PWV. Figure 5c–e presents pictures of three cases: Case I: sensors on the carotid artery and the lateral epicondyle vessel; Case II: on the lateral epicondyle vessel and the radial artery of the wrist; and Case III: on the near epicondyle vessel and the radial artery of the wrist; respectively. PWV profiles of Cases I, II and III appear in Fig. 5f–h. As seen in Fig. 5i–k, PWVCase I is ~5.4 ms−1 where the distance between sensors is ~43 cm and time difference is ~0.08 s; PWVCase II is ~5.8 ms−1 where the distance between sensors is ~23 cm and time difference is ~0.04 s; and PWVCase III is ~6.5 ms−1 where the distance between sensors is ~6.5 cm and time difference is ~0.01 s. These PWV values are comparable with the typical aortic PWV (~4.5 ms−1) measured by tonometry in a healthy 20-year-old person45.
Discussion
Results presented here demonstrate the ability to exploit enhanced piezoelectric effects in PZT elements on soft supports for precision, skin-mounted sensors for pressure. When coupled to SiNM-based field effect transistors, these devices offer unique capabilities in accurate measurements of subtle effects of motion on the surface of the skin, ranging from blood pressure pulse waves in near-surface arteries to vibrations on the throat associated with speech. The thin, conformal nature of the devices, together with their predictable responses, consistent with theoretical models, offer high performance operation, without measurable side effects during use in daily activities. The collective results suggest that these materials and the resulting sensor capabilities could be valuable for a range of applications in continuous health/wellness monitoring and clinical medicine. Current work focuses on long-term device characterization and wireless communication.
Methods
Fabrication of SiNM MOSFETs
The fabrication began with a silicon-on-insulator (top silicon thickness ~300 nm, p-type, SOITEC, France) wafer, consisting of lightly p-type doped (1 × 1015 per cm3) top silicon and a 1-μm-thick buried silicon dioxide (SiO2) layer, as shown in Supplementary Fig. 1a. Cleaning the silicon-on-insulator wafers involved removal of organic and ionic contaminants by sequential immersion in: (1) 1:1:5 solution of NH4OH (ammonium hydroxide):H2O2 (hydrogen peroxide):H2O (de-ionized (DI) water) at 80 °C for 10 min; (2) 1:50 solution of HF (hydrofluoric acid, 49%):(DI water) for 1 min; and (3) 1:1:6 solution of HCl (hydrochloric acid):H2O2:H2O (DI water) at 80 °C for 10 min. Thermal oxidation produced a layer of SiO2 on the top silicon, as a mask for phosphorous doping. This thermal oxide was patterned by reactive ion etching (RIE; Plasma-Therm) with CF4/O2 through a photoresist (PR) mask. The wafer was then cleaned with acetone and isopropyl alcohol to remove the remaining PR. Another RCA cleaning preceded solid state diffusion doping of phosphorous (2 × 1020 per cm3) in a 6-inch tube furnace (Lindberg/Tempress Model 8500 Dual-Stack Diffusion/Oxidation Furnaces) at 1,000 °C. A series of wet-etching steps using HF (49%) and piranha solution (3:1=H2SO4 (sulphuric acid):H2O (DI water)) eliminated surface contamination. The top silicon was then patterned by RIE to form an array of 3 μm diameter circular openings. Immersion in HF (49%) removed the buried oxide to allow transfer printing of the resulting SiNM via a PDMS stamp onto a carrier wafer with a spin-cast layer of polyimide as an adhesive and an underlying layer of poly(methylmethacrylate) (495 A2). The SiNM was then patterned by RIE to define active regions for transistors. The gate dielectric consisted of a sandwich structure of SiO2 (140 nm thick) with aluminium oxide (Al2O3; 10 nm thick) above and below; the Al2O3 is in contact with the Si. The SiO2 and Al2O3 were deposited by plasma-enhanced chemical vapour deposition (Plasma-Therm/Unaxis SLR730) and by atomic-layer deposition (Cambridge Nanotech), respectively. The gate dielectric was patterned by etching in buffered oxide etchant (BOE) (6:1). Deposition of 5 nm of Cr and 150 nm of Au followed by wet etching of the Au (Transene Company Inc., USA) and the Cr (OM Group, USA) defined electrodes aligned to the silicon and gate dielectric. The entire device was then encapsulated with a 1.2-μm-thick layer of PI to position the SiNM near the NMP. Etching the underlying PI with an O2 plasma (Asher (LFE)) exposed the poly(methylmethacrylate) layer to enable its removal in hot acetone (110 °C) and to partially release the device, with PI above and below, from the wafer. The entire structure was then retrieved onto a piece of PDMS and pressed against a 20-μm-thick silicone (Smooth-on, Ecoflex 0030, Platinum Cure Silicone Rubber) substrate. A bilayer of 4 nm of Ti and 40 nm of SiO2 deposited onto the back side of the device while on the PDMS enabled formation of strong bonds to the silicone. As a final step, the device, with bonded silicone, was transfer printed to a 200-μm-thick layer of poly(vinylalcohol) (A-30031, Best Triumph Industrial Ltd). Integration onto the skin involved placing the sensor on a transparent plastic frame and physically laminating the sensor on the skin. Rinsing with water removed the poly(vinylalcohol), leaving an ultrathin device in intimate, conformal contact with the skin. Devices mounted in this way exhibited excellent compliance and ability to follow skin motion without constraint or delamination. The plastic frame facilitated removal and re-application for repeated measurements. Separate adhesives were not needed to bond the devices to the skin. The adhesion strength was measured using a digital force metre (Mark-10, USA), using the following procedures46: the wrist was secured and the device substrate was connected to a force metre. The device was peeled away at a maximum speed of 1,000 mm min−1 in an upward direction, parallel to the width of the device, at room temperature. The maximum value of the force during this process defined the adhesion force. A typical device with a 20-μm-thick silicone substrate and lateral dimensions of ~1.5 cm (length) × 1 cm (width) showed an average adhesion force of 0.032 N. Photographs collected during mounting and peeling appear in Supplementary Fig. 19.
Fabrication of arrays of PZT capacitor structures
The PZT was formed by sol–gel techniques on an oxidized silicon wafer coated with Pt/Ti (300 nm/20 nm). The PZT (INOSTEK; thickness 400 nm) was formed by a sol–gel process on an oxidized silicon wafer coated with Pt/Ti (300 nm/20 nm), following the procedures of Li et al.47 A 0.4-M precursor solution consisted of zirconium n-propoxide, titanium isopropoxide and lead acetate trihydrate with acetic acid and propyl alcohol as the solvents. The prepared solution was spin-cast onto the Pt/Ti/SiO2/Si (100) substrate at 3,000 r.p.m. for 40 s. Pyrolysis and pre-crystallization occurred at 400 °C for 20 min. After deposition of five consecutive layers in this fashion, the sample was annealed at 680 °C for 10 min. The total thickness of the PZT was ~400 nm. The array of PZT elements consisted of square regions 557 × 625 μm2 with a spacing of 170 μm formed by wet chemical etching in HNO3:BHF:H2O (nitric acid: buffered HF:DI water=4.51:4.55:90.95 through a hard-baked mask of PR (AZ4620, Clariant) as shown in Supplementary Figs 1b and 20. The spacing facilitates access of concentrated HF for undercut etching of the PZT elements. The array structure provides a large sensing area. The hard baking process involved 80 °C for 5 min, 110 °C for 30 min and 80 °C for 5 min. The bottom Pt/Ti electrode was patterned by etching in HCl:HNO3:H2O(DI water)=3:1:4 at 95 °C also through a hard-baked mask of AZ4620. A layer of PI was spin cast on the sample at a thickness (3.6 μm) to position the PZT at the NMP. The PI was removed by O2 plasma (Asher (LFE)) everywhere except for regions with sizes of 650 × 550 μm2 centred on the PZT. These structures were protected by a PR mask during partial removal of the underlying SiO2 with diluted HF (H2O:HF (49%)=1:3). The hard-baked PR was then removed in acetone (3 h). The fabrication yield was ~80%, limited mainly by cracking that sometimes occurred during the retrieval stage of transfer printing the PZT.
Transfer printing PZT structures on a stretchable substrate
As illustrated in Supplementary Fig. 1b, a PDMS stamp retrieved the partially undercut PZT/PI/Pt structures and delivered them to a layer of silver epoxy (Ted Pella Fast Drying Silver Paint (160040-30)) coated on an extended gate electrode of a SiNM n-MOSFET. The PZT elements bond at their base regions to a uniform conductive coating Ted Pella Fast Drying Silver Paint (160040-30) applied to an extended gate electrode of a SiNM n-MOSFET. This arrangement electrically connects the bottom Pt electrodes of the PZT elements. A 3.6-μm-thick layer of PI applied to the entire array locates the NMP at the mid-plane of the PZT (Fig. 1a,b). The choice of thickness involved balanced considerations in bendability, where thinner is better, and in voltage output, where thicker is better. A thickness of 400 nm represents a good choice.
Investigation of properties of PZT and the device responses
PZT capacitor structures with Ti/Pt (20 nm/300 nm) bottom electrodes and Cr/Pt (10 nm/200 nm) top electrodes were poled at 100 kV cm−1 for 2 h at 150 °C. Afterward, the top electrode was removed by etching the chromium and platinum. X-ray diffraction (Philips X'pert) was used to confirm the crystal structure of PZT, corresponding to two main peaks, belonging to (100) and (111) as seen in Supplementary Fig. 21a. A scanning electron microscope (Hitachi S4800) was used to examine cross-sections of the device stacks on a silicon wafer with the SiO2 sacrificial layer and the underlying silicon substrate, as in Supplementary Fig. 21b. The hysteresis loop of the PZT in Supplementary Fig. 21c was obtained by using a ferroelectric tester (Radiant Technologies, Inc.). A semiconductor parameter analyser (4155C, Agilent) was used to characterize the SiNM n-MOSFETs and to perform open-loop voltage measurements for the response of the pressure sensor. The signal-to-noise ratios for typical measurements are ~50 (Supplementary Fig. 22). PDMS posts served as calibrated weights in the milligram range, for applying well-defined levels of pressures to the PZT elements. A computed map of contact pressure between a representative post (1 mm thick) and an array of PZT elements on silicone appears in Supplementary Fig. 23. Each PDMS post was placed on the PZT region with a vacuum tweezers with bent metal probe, 1-inch long, 3/32-inch cup diameter (Ted Pella, Inc., Vacuum Pickup System, 115 V) from a fixed distance. The voltage response to changes in temperature was investigated using the continuous heating method, with a heating rate of 1 °C min−1. A small stage (All Electronics Corp.; 40 × 44 mm2) was used to measure effects of temperature on the PZT active region of a device. Human subject demonstrations were performed under approval by the Institutional Review Board of the University of Illinois (IRB no. 13561). The informed consent was obtained from all subjects in human trials.
Additional information
How to cite this article: Dagdeviren, C. et al. Conformable-amplified lead zirconate titanate sensors with enhanced piezoelectric response for cutaneous pressure monitoring. Nat. Commun. 5:4496 doi: 10.1038/ncomms5496 (2014).
References
Koh, W., Saudari, S. R., Fafarman, A. T., Kagan, C. R. & Murray, C. B. Thiocyanate-capped PbS nanocubes: Ambipolar transport enables quantum dot based circuits on a flexible substrate. Nano Lett. 11, 4764–4767 (2011).
Son, J. S., Lee, J. S., Shevchenko, E. V. & Talapin, D. V. Magnet-in-the-semiconductor nanomaterials: High electron mobility in all-inorganic arrays of FePt/CdSe and FePt/CdS core–shell heterostructures. J. Phys. Chem. Lett. 4, 1918–1923 (2013).
Kango, S. et al. Surface modification of inorganic nanoparticles for development of organic-inorganic nanocomposites-A review. Prog. Polym. Sci. 38, 1232–1261 (2013).
Cohen-Karni, T. & Lieber, C. M. Nanowire nanoelectronics: building interfaces with tissue and cells at the natural scale of biology. Pure Appl. Chem. 85, 883–901 (2013).
Duan, X., Huang, Y., Agarwal, R. & Lieber, C. M. Single-nanowire electrically drive lasers. Nature 421, 241–245 (2001).
Hyun, J. K., Zhang, S. & Lauhon, L. J. Nanowire heterostructures. Annu. Rev. Mater. Res. 43, 451–479 (2013).
Rogers, J. A., Lagally, M. G. & Nuzzo, R. G. Synthesis, assembly and applications of semiconductor nanomembranes. Nature 477, 45–53 (2011).
Rogers, J. A., Someya, T. & Huang, Y. Materials and mechanics for stretchable electronics. Science 327, 1603–1607 (2010).
Kim, D.–H., Ghaffari, R., Lu, N. & Rogers, J. A. Flexible and stretchable electronics for biointegrated devices. Annu. Rev. Biomed. Eng. 14, 113–128 (2012).
Li, J., Wang, C., Ma, J. & Liu, M. Micromachined ultrasonic transducers based on lead zirconate titanate (PZT) films. Microsyst. Technol. 19, 211–218 (2013).
Maeda, R., Tsaur, J. J., Lee, S. H. & Ichiki, M. Piezoelectric microactuator devices. J. Electroceram. 12, 89–100 (2004).
Zarnik, M. S., Belavič, D., Maček, S. & Holc, J. Feasibility study of a thick-film PZT resonant pressure sensor made on a prefired 3D LTCC structure. Int. J. Appl. Ceram. Technol. 6, 9–17 (2009).
Soman, J. & O’Neal, C. B. Fabrication and testing of a PZT strain sensor for soil applications. IEEE Sens. J. 11, 78–85 (2011).
Damjanovic, D. Ferroelectric, dielectric and piezoelectric properties of ferroelectric thin films and ceramics. Rep. Prog. Phys. 61, 1267–1324 (1998).
Nguyen, T. D. et al. Piezoelectric nanoribbons for monitoring cellular deformations. Nat. Nanotech. 7, 587–593 (2012).
Dagdeviren, C. et al. Conformal, multilayer piezoelectric energy harvesting and storage from motions of the heart, lung and diaphragm. Proc. Natl Acad. Sci. USA 111, 1927–1932 (2014).
Park, K. I. et al. Flexible nanocomposite generator made of BaTiO3 nanoparticles and graphitic carbons. Adv. Mater. 24, 2999–3004 (2012).
Dagdeviren, C. et al. Transient, biocompatible electronics and energy harvesters based on ZnO. Small 9, 3398–3404 (2013).
Xu, S. & Wang, Z. L. One-dimensional ZnO nanostructures: solution growth and functional properties. Nano Res. 4, 1013–1098 (2011).
Avolio, A. P., Butlin, M. & Walsh, A. Arterial blood pressure measurement and pulse wave analysis-their role in enhancing cardiovascular assessment. Physiol. Meas. 31, R1–R47 (2010).
Nelson, M. R. et al. Noninvasive measurement of central vascular pressures with arterial tonometry: clinical revival of the pulse pressure waveform? Mayo Clin. Proc. 85, 460–472 (2010).
Terry, S., Eckerle, J. S., Kornbluh, R. S., Low, T. & Ablow, C. M. Silicon pressure transducer arrays for blood-pressure measurement. Sens. Actuat. A 23, 1070–1079 (1990).
Ravariu, C. et al. A designing roule for a pressure sensor with PZT layer. Proc. Int. Semicond. Conf., CAS 2, 379–382 (2001).
Graz, I. et al. Flexible ferroelectret field-effect transistor for large-area sensor skins and microphones. Appl. Phys. Lett. 89, 073501 (2006).
Someya, T. et al. A large-area, flexible pressure sensor matrix with organic field-effect transistors for artificial skin applications. Proc. Natl Acad. Sci. USA 101, 9966–9970 (2004).
Someya, T. et al. Conformable, flexible, large-area networks of pressure and thermal sensors with organic transistor active matrixes. Proc. Natl Acad. Sci. USA 102, 12321–12325 (2005).
Kim, J.-H., Sun, Q. & Seo, S. Organic Electronics Vol.11, 964–968Elsevier (2010).
Schwartz, G. et al. Flexible polymer transistors with high pressure sensitivity for application in electronic skin and health monitoring. Nat. Commun. 4, 1859 (2013).
Persano, L. et al. High performance piezoelectric devices based on aligned arrays of nanofibers of poly(vinyldenefluoride-co-trifluoroethylene). Nat. Commun. 4, 1633 (2013).
Takei, K. et al. Nanowire active-matrix circuitry for low-voltage macroscale artificial skin. Nat. Mater. 9, 821–826 (2010).
Lipomi, D. J. et al. Skin-like pressure and strain sensors based on transparent elastic films of carbon nanotubes. Nat. Nanotech. 6, 788–792 (2011).
Fan, F. R. et al. Transparent triboelectric nanogenerators and self-powered pressure sensors based on micropatterned plastic films. Nano. Lett. 12, 3109–3114 (2012).
Mannsfeld, S. C. B. et al. Highly sensitive flexible pressure sensors with microstructured rubber dielectric layers. Nat. Mater. 9, 859–864 (2010).
Dassault Systèmes. Abaqus Analysis User's Manual V.6.10 Dassault Systèmes Simulia Corp. (2010).
Timpano, K. Mechanical Characterization of Gold Thin Films for RF-MEMS Proc. Virginia Space Grant Consortium Student Research Conf. (2005).
Chen, C. H. et al. Validation of carotid artery tonometry as a means of estimating augmentation index of ascending aortic pressure. Hypertension 27, 168–175 (1996).
Nichols, W. Clinical measurement of arterial stiffness obtained from noninvasive pressure waveforms. Am. J. Hypertens 18, 3S–10S (2005).
Millasseau, S. C., Patel, S. J., Redwood, S. R., Ritter, J. M. & Chowienczyk, P. J. Pressure wave reflection assessed from the peripheral pulse: is a transfer function necessary? Hypertension 41, 1016–1020 (2003).
Pauca, A. L., O'Rourke, M. F. & Kon, N. D. Prospective evaluation of a method for estimating ascending aortic pressure from the radial artery pressure waveform. Hypertension 38, 932–937 (2001).
Denardo, S. J., Nandyala, R., Freeman, G. L., Pierce, G. L. & Nichols, W. Pulse wave analysis of the aortic pressure waveform in severe left ventricular systolic dysfunction. Circulation 3, 149–156 (2010).
Miki, T. et al. Radial Augmentation Index: a useful and easily obtainable parameter for vascular aging. Am. J. Hypertens. 18, 11–14 (2005).
Klintland, N. et al. Increased arterial stiffness and indication of endothelial dysfunction in long-standing rheumatoid arthritis. Scand. J. Rheumatol. 37, 1–5 (2008).
Antonini-Canterin, F. et al. Arterial stiffness and ventricular stiffness: a couple of diseases or a coupling disease? A review from the cardiologist’s point of view. Eur. J. Echocardiogr. 10, 36–43 (2009).
Moon, J., Mccombie, D., Dhillon, M. & Banet, M. Vital sign monitoring system. US Patent WO2010135516 A2 (issued 25 November 2010).
DeLoach, S. S. & Townsend, R. R. Vascular stiffness: its measurement and significance for epidemiologic and outcome studies. Clin. J. Am. Soc. Nephrol. 3, 184–192 (2008).
Yeo, W.-H. et al. Multifunctional epidermal electronics printed directly onto the skin. Adv. Mater. 25, 2773–2778 (2013).
Li, J., Wang, C., Liu, M. & Ma, J. Effect of annealing process on properties of Pb (Zr0.52 Ti0.48) O3 thin films prepared by sol-gel method. Ferroelectrics 445, 32–38 (2013).
Acknowledgements
This work was supported by the U.S. Department of Energy, Office of Science, Basic Energy Sciences, under Award no. DE-FG02-07ER46471, through the Frederick Seitz Materials Research Laboratory at the University of Illinois at Urbana-Champaign. A.R.D. and L.W.M. acknowledge support for their inputs on device modeling from the National Science Foundation under grant DMR-1149062. We thank Tony Banks for assisting to set up electrical connections of the semiconductor parameter analyser (HP 4155C, Agilent). C.D. thanks H.E. Dagdeviren for useful medical discussions.
Author information
Authors and Affiliations
Contributions
C.D. and J.A.R. designed the research; C.D., P.J., R.Y., Y.L., Y.-S. K., A.R.D, L.W.M. and J.A.R. performed the research; Y.S., Y.A.H., J.X. and Y.H. performed mechanical modelling; C.D., Y.S., P.J., Y.H. and J.A.R. analysed the data; and C.D., Y.S., Y.H. and J.A.R. wrote the paper.
Corresponding authors
Ethics declarations
Competing interests
The authors declare no competing financial interests.
Supplementary information
Supplementary Information
Supplementary Figures 1-23, Supplementary Tables 1-2, Supplementary Notes 1-6 and Supplementary References (PDF 1930 kb)
Rights and permissions
About this article
Cite this article
Dagdeviren, C., Su, Y., Joe, P. et al. Conformable amplified lead zirconate titanate sensors with enhanced piezoelectric response for cutaneous pressure monitoring. Nat Commun 5, 4496 (2014). https://doi.org/10.1038/ncomms5496
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/ncomms5496
This article is cited by
-
3D-assembled microneedle ion sensor-based wearable system for the transdermal monitoring of physiological ion fluctuations
Microsystems & Nanoengineering (2023)
-
Fast and versatile electrostatic disc microprinting for piezoelectric elements
Nature Communications (2023)
-
Interface engineering for substantial performance enhancement in epitaxial all-perovskite oxide capacitors
NPG Asia Materials (2023)
-
A conformable phased-array ultrasound patch for bladder volume monitoring
Nature Electronics (2023)
-
Hierarchically resistive skins as specific and multimetric on-throat wearable biosensors
Nature Nanotechnology (2023)