PNNL-11221
UC-606
Instrument Validation Project
B. A. Reynolds
E. A. Daymo
J. G. H. Geeting
J. Zhang
June 1996
Prepued for
the U.S . Department of Energy
under Contract DE-AC06-76RLO 1830
Pacific Northwest National Laboratory
Richland, Washington 99352
DISCLAIMER
This =port was prepared as an account of work sponsored by an agency of the United
States Government. Neither the United States Government nor any agency thereof, nor
any of their employees, make any wamnty, express or implied, or a~sumes
any legal WiIity or respona'bilityfor the accuracy, compieteness,or usefulaess of any information,apparatus, product, or process disdose& or represents that its use wouid not infringe privately
owned rights. Ftefemm herein to any specific commenhl product, prucess, or service by
trade name, trademark, manuf-,
or otherwise does not neeessanly constitute or
imply its endomment, recommendatioa,or favoring by the United States Governmentor
any agency themof. The views and opinioos of aulhors expressed herein do not ~ e c e s s ~ c ily state or r e k t those of the United States Governmeator any agency thereof.
DISCLAIMER
Portions of this document may be illegible
electronic image products. Images are
produced from the best available original
document.
Summary
Westinghouse Hanford Company Project W-2 11 is responsible for providing the system capabilities to
remove radioactive waste from ten double-shell tanks used to store radioactive wastes on the Hanford Site
in Richland, Washington. The project is also responsible for measuring tank waste slurry properties prior
to injection into pipeline systems, including the Replacement of Cross-Site Transfer System (WHC Project
W-058).
This report summarizes studies of the appropriatenessof the instrumentation specified for use in
Project W-211. The instruments were evaluated in a test loop with simulated slurries that covered the
range of properties specified in the functional design criteria. The study also evaluated the design of the
test loop itself against a less compact design (an “Ideal”loop) and the specified instruments were compared
to several alternative pieces of instrumentation.
The results of the study indicate that the compact nature of the baseline Project W-211 loop does not
result in reduced instrumental accuracy resulting from poor flow profile development. Of the baseline
instrumentation, the Micromotion densimeter, the Moore Industries thermocouple, the Fischer & Porter
magnetic flow meter, and the Red Valve Pressure transducer meet the desired instrumental accuracy. An
alternate magnetic flow meter (Yokagawa) gave nearly identical results as the baseline Fischer & Porter.
The Micromotion flow meter did not meet the desired instrument accuracy but could potentially be
calibrated so that it would meet the criteria.
The Nametre on-line viscometer did not meet the desired instrumental accuracy and is not
recommended as a quantitative instrument although it does provide qualitative information. Measuring the
pressure drop over the test loop provides comparable data with comparable accuracies and directly
measures one of the most important parameters in the operation of the cross site transfer line: pressure
drop.
The recommended minimum set of instrumentation necessary to ensure the slurry meets the
Project W-058 acceptance criteria is the Micromotion mass flow meter (measuring density, flow rate, and
temperature) and delta pressure cells. The Moore Industries thermocouple is redundant. The
Micromotion flow meter can substitute for the Yokagawa and Fisher & Porter flow meters provided it is
properly calibrated.
...
111
Acronyms and Abbreviations
AC
alternating current
CP
centipoise
CRC
Handbook of Chemistry and Physics
DC
direct current
DP
delta pressure
DSTs
double-shell tanks
FBRM
focussed beam reflectance method
FDC
functional design criteria
HP
horsepower
Hz
hertz
ICFKH
ICF Kaiser Hanford
IVF
Instrument Validation Facility
P
pressure
Pa
Pascal
PCA
principle component analysis
PLC
programmable logic controller
PLS
partial least squares
PNNL
Pacific Northwest National Laboratory
PSD
particle size distribution
V
RPM
revolutions per minute
VIP
variable importance in the projection
WHC
Westinghouse Hanford Company
vi
Contents
...
Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
111
.............................................
V
Acronyms and Abbreviations
1.0 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1
1.1 Background . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1
...................................................
1.2
1.2 Objectives
1.2.1 Accuracy
...............................................
......................................
1.4
..................................
1.4
.................................
1.5
..........................................
1.5
.......................................
1.6
....................................
1.8
...............................
2. 1
.........................................
2.1
..........................................
2.1
..................................
2.4
........................................
2.5
1.2.2 Instrument Elimination
1.2.3 Instrument Loop Verification
1.2.4 Evaluate Alternate Instruments
1.3 Experimental Approach
1.3.1 Test Loop Description
1.3.2 Test Loop Instrumentation
2.0 Experimental Instrumentation and Procedures
2.1 Analytical Instrumentation
2.1.1 Haake Rheometer
2.1.2 Horiba Particle Size Analyzer
2.1.3 DYNAC Centrifuge
2.1.4 Pycnometer
..............................................
2.5
..........................................
2.5
......................................
2.8
...................................
2.9
2.2 Experimental Procedure
2.3 Slurry Preparation and Matrix
2.4 Experimental Text Matrix and Data
2.5 Statistical Analysis
1.3
..............................................
vii
2.9
..............................................
3.1
3.1 General Observations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.1
.............................................
3.1
3.1.2 Vibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.1
3.2 W-2 1 1 vs . “Ideal”Loop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2
3.3 Instrument Validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.3
3.3.1 Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.3
3.3.2 Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.4
................................................
3.5
3.0 Results and Discussion
3.1.1 Temperature
3.3.3 Density
3.3.4 Viscosity
................................................
3.9
..............................................
3.22
............................
3.27
...........................................
3.32
3.3.5 Flow Rates
3.3.6 Particle Size Distribution Measurement
3.4 Instrument Elimination
.............................................
3.33
................................
3.33
3.5.1 Viscosity Multivariate Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.33
.........................
3.35
3.5.3 Particle Size Multivariate Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.38
3.5.4 Flow Rate Error Multivariate Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.38
3.4.1 Temperature
3.5 Multivariate Statistical Analysis Results
3.5.2 Solids ConcentrationMultivariate Analysis
4.0 Conclusions and Recommendations:
.......................................
4.1
...................................
4.1
4.2 Instrument Elimination Conclusions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2
......................................
4.2
4.4 Evaluation of Alternate Instrumentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.3
4.1 Instrument Validation Conclusions
4.3 Test Loop Design Conclusions
...
Vlll
4.5 Other Conclusions and Recommendation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.3
5.0 References
. . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1
6.0 Appendices
. . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.1
Appendix A .Detailed Loop Designs
........................................
A .1
Appendix B .Simulant Development
.........................................
B .1
. . .........................................
c .1
Appendix D .Statistical Analysis Results . : . . . . . . . . : . . . . . . . . . . . . . . . . . . . . . . . . . .
D .1
Appendix E .Lasentec Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
E. 1
Appendix C .Experimental Data
ix
Figures
1.1 Designed Experiment Approach . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.6
1.2 Test Loop Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.7
1.3 Nametre Preferred Installation Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.10
1.4 Lasentec Particle Size Measurement Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.13
1.5 Lasentec Particle Size Distribution for Unknown Sample in Run 8C . . . . . . . . . . . . . . . . .
1.13
1.6 Lasentec Diagram
..................................................
1.15
2.1 Classes of Viscosities
...............................................
2.2
2.2 Classes of Viscosities
...............................................
2.2
...............................
2.6
......................................
3.3
...........................................
3.4
2.3 Typical Particle Size Distribution from Horiba
3.1 Rosemount vs . Red Valve Pressure
3.2 Thermocouple Comparison
3.3 Settling in Feed Tank
...............................................
3.4 Test Loop Density vs . Feed Tank Density
..................................
3.5
3.6
3.5 Micromotion Density vs . Average S-3& S-4 Density. Air Runs Included . . . . . . . . . . . . . .
3.7
3.6 Micromotion Density vs . Average S-3 & S-4 Density. Air Runs Excluded . . . . . . . . . . . . .
3.8
3.7 Micromotion Density vs . Calibrated Vessel Density
...........................
3.9
..........................................
3.11
...........................................
3.12
3.10 Nametre vs . Haake at 300 Sec-' . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.13
3.11 Nametre vs . Haake at lo00 Sec"'
3.14
3.8 'Nametre Vibration Tolerance
3.9 Nametre Viscosity Stability
.......................................
X
3.12 Nametre vs . Haake at 4084-' Seconds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.15
3.13 Pressure Drop vs . Nametre Measured Viscosity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.16
3.14 Haake Viscosity vs . Back Calculated Viscosity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.20
3.15 Back Calculated vs . Laboratory Measured Consistency Parameters . . . . . . . . . . . . . . . . .
3.21
3.16 Yokagawa Flow Rate vs . Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.24
................................
3.25
3.17 Yokagawa Instrument Volumetric Flow Rate
3.18 Fischer & Porter Volumetric Flow Rate
........................
. . . . . . . . . . 3.26
3.19 Particle Size distribution shift in Run 6C1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.28
..............................................
3.29
3.20 Lasentec Color Effects
...............................
3.30
..............................
3.31
..................................................
3.34
................................
3.34
.......................................
3.35
...........................................
3.36
3.21 Impact of Bubbles on Measured Particle Sizes
3.22 Lasentec vs . Horiba Particle Size Distributions
3.23 Model Overview
3.24 VIP Plot of Haake to Measured Parameters
3.25 Predicted vs . Observed Viscosity
3.26 Solids Concentration Model
3.27 VIP Plot for Solids Concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.36
3.28 Predicted Solids Loading vs . Observed Solids Loading
.........................
3.37
.................................................
3.38
3.29 Particle Size Model
xi
Tables
1.1 Project W-211 Design Criteria . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.2
1.2 Acceptable Accuracies for Instrumentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.3
.............................................
1.8
1.4 Instrument Cross-Check and Validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.9
1.3 Instruments to be Evaluated
3.1 Horiba PSD Instrument Repeatability with Run 4A “Ideal”Loop Sample . . . . . . . . . . . . . . . 3.30
1.0 Introduction
Westinghouse Hanford Company (WHC) Project W-2 11 is responsible for providing the system capabilities to remove radioactive waste from ten of the Hanford Site’s double-shell tanks (DSTs). The project
is also responsible for measuring tank waste slurry properties prior to injection into pipeline systems,
including the Replacement of Cross-Site Transfer System ( W C Project W-058).
WHC Project W-211 requested that Pacific Northwest National Laboratory (PNNL) construct and
operate an Instrument Validation Facility (IVF) test loop to determine if the selected instruments and
proposed instrument loop configuration function as intended to measure specific slurry characteristics and
to recommend the minimum suite of instruments to provide reasonable assurance that the waste can be
pumped.
1.l Background
Project W-211 is designed to remove radioactive waste from ten DSTs used to store radioactive wastes
on the Hanford Site and condition those wastes for transport via underground cross site transfer pipelines.
The project is responsible for ensuring that wastes are suitably conditioned and meet the W-058
specification for transport. Project personnel will also install an instrument loop with selected
instrumentation to measure the properties of importance for slurry transport and for use in process control
of the conditioning of the waste feed. ,
The pertinent slurry characteristicsfor which Project W-2 11 is being designed (Table 1.1) have been
set forth in the functional design criteria (FDC) for Project W-211, Initial Tank Retrieval Systems, (WHCSD-W-211-FDC-001, Rev. 2):
The Project W-211 Project specified the following instruments as part of the instrument loop:
Micromotion Mass Flow Meter, Densimeter
Nametre Viscometer
Moore Thermocouple
Red Valve Pressure
Fischer & Porter How Meter
Moore Industries Thermocouple
1.1
Table 1.1. Project W-211 Design Criteria
Process Parameter
Nominal
Range
Slurry Viscosity")
10 CP
1.0 - 30.0 CP
Solid Content@"'
20 vol%
0.0 to 30.0 volume%
Specific Gravity
1.25
1.0 - 1.5
Miller Number
I e 100
I
Particle Size(@
0.5 to 4OOO microns
Minimum pH:
11.0
Design Velocity:'")
I 6 ftlsec
I 4.5 - 6 ftlsec
Temperature (flushes)
35°F to 200°F (2°C to 93°C)
Friction Factor
Newtonian Flow
(a) Ranges were taken from Rev. 1 of the FDC.
(b) C, = ( l-M,/Vt p,) where, C, = solids concentration by volume, MI = mass of liquid,
V, = total sample volume p = liquid density.
(c) The 30 vol% solids for the purposes of this project is defined as true volume percent
solids. The volume percent settled solids and the volume percent centrifuged solids
were measured for comparison to the settled solids volume fraction, which is the basis
for the Project W-058 volume% solids slurry acceptance criterion.
(d) Guidance on the particle size distributions are not specified in the Project W-211 FDC.
However, since some guidance was needed to prepare the slurries, the information was
taken from the W-058FDC. Specifically, 4 0 micron particles will comprise
approximately 95 % of the total, 50-500 micron particles will comprise < 5 % of the
total and 500-4,OOOmicron particles will comprise < 1%of the total.
(e) 4.5 ft/sec is based on a slurry viscosity of 30 cP; 6.0 ft/sec is based on a slurry
viscosity of 10 cP.
1.2 Objectives
The four main objectives of the IVF project were to
determine the accuracy of the instruments
eliminate unnecessary instruments
1.2
verify adequacy of instrument design and ability for process control
evaluate alternate instruments.
Each of the four objectives will be discussed in the following subsections.
1.2.1 Accuracy
The first and principal goal of the IVF project was to determine whether the specific brands of
instruments specified for the WHC Project W-211 instrument loop adequately measured the desired slurry
properties over a broad range of physical properties. A corollary to this testing is that specific types of
technologies were also evaluated. Alternate brands of instruments that use these technologies could
potentially be substituted at a later date if needed.
Each instrument’s accuracy was determined by the degree to which the measurement matched that of a
“fundamentalmeasurement” (Le., a measurement that involves only length/volume, time, and mass) and/or
a laboratory measurement. ICF Kaiser Hanford (ICF KH) and WHC staff provided guidance on the level
of accuracy required for the instruments to be considered acceptable for use in the test loop (see
Table 1.2).
A secondary goal was to determine the instrument and instrument loop performance at off-normal
process conditions.
Table 1.2. Acceptable Accuracies for Instrumentation
I
Instrument
I
Reference Standard
Accuracy
Referencestandard
+o. 1%
Micromotion Density
+I- 5%
Laboratory Pycnometer
Micromotion Mass Flow Rate
+/- 5%
Volumetric Calibration
Tank
Pressure (Red Valve)
+/- 1%
Instrument Calibration
Not Applicable
Temperature (Moore Industries)
+/- 1%
Instrument Calibration
Not Applicable
Fischer & Porter Volumetric Flow
+/- 1%
Yokagawa Volumetric Flow
+I- 2%
~
Viscosity (Nametre)
1(a)
~~~~
+I- 10%
1
I
* 1.2%‘”)
* 1.2%
Volumetric Calibration
Tank
Volumetric Calibration
Tank
Haake Laboratory
Rheometer
~
~
+1.2%
+1.0%
I
See Section 3.3.5 for derivation of this value.
1.3
The final goal was to determine if the IVF operators and engineers could determine slurry characteristics on-line. This interpretive test used the knowledge generated in testing known, well quantified
slurries to interpret the readings of an unknown slurry.
There were three primary components to the experimental work to evaluate the accuracy goals:
1. Rheological simulants were prepared to span the range of slurry properties specified in the Project
W-211 FDC. The instruments were tested and the results compared to lab data and/or fundamental
measurements.
2. The flow rate was varied and air bubbles were introduced to test off-normal conditions. (Note that
bubbles are likely to always be present in tank wastes and may be normally present in the simulants as
well.)
3. Blind interpretive tests were run to find out whether operators could determine slurry characteristics
without prior knowledge of the feed with in-line instruments only. The blind test simulant, which may
not meet the W-058 transport criteria but was made from the solids tested, was injected into the pipe
loop. The test operator(s) and engineer(s) relied solely on the in-line instruments to measure the
physical properties of the slurry and determined whether the feed was within the allowable transport
specifications. After the blind tests were completed, the interpretive tests were treated like all of the
other simulants and were used in the statistical analyses.
1.2.2 Instrument Elimination
The second objective was to determine if the number of instruments specified by ICF KH and WHC
could be reduced. Instruments are expensive to service and replace in a radioactive environment and the
fewer instruments installed, the greater the cost savings to the operation of the system. Reducing the
number of instruments would reduce the space requirements in the tank farm valve pits, lower the operational cost of the facility, and lower the frequency of forced shutdowns due to instrument failure.
All the instruments were tested with a broad range of physical simulants and a multivariate analysis
was used to search for relationships between important process variables (e.g., viscosity, density, flow
rate, PSD, etc.) that might be difficult to find otherwise. Fewer instruments were recommended if strong
correlations can be made. Redundant instruments were recommended for removal where possible.
1.2.3 Instrument Loop Verification
The third objective was to verify that the proposed instruments function adequately in the valve pit as
designed. The instrument loop designed for WHC Project W-211 was believed to be adequate for measuring the desired slurry characteristics; however, instruments were placed in positions where it was
suspected that the flow would probably not be fully developed. (This loop is replicated in the IVF facility
and is called the Project W-211 loop. An alternate test loop was installed and called the “ideal”loop.) The
1.4
proximity of instruments to elbows and other instruments may have resulted in inaccurate readings if fullydeveloped flow was not achieved. If the instruments adequately measured the slurry properties to the
desired accuracy in the Project W-2 11 loop, the expense involved in designing and constructing a new
instrument pit could be avoided.
The design of the instrument loop was evaluated by comparing the performance of each instrument in
the “ideal”and Project W-211 loops to look for synergistic (or antagonistic) effects. This comparison
helped separate poor performance resulting from an inadequate test-loop design from an inadequate instrument. It also established a best-case baseline against which informed decisions can be made about whether
alternate instrument configurations in the Project W-211 test loop should be considered. Knowledge of this
baseline could prevent unnecessary trial and error runs if the instrument performance is not adequate in the
Project W-211 loop. The baseline was used to establish the best-case accuracy performance expected for
the various instruments.
The experimental approach used for this objective was to compare the instrument performance in the
Project W-211 loop to the performance in the best-case “ideal”loop. If poor performance was seen in the
Project W-211 loop, it could be determined if the poor performance was caused by flow conditions or by
instrument performance and if “adequate”performance (see Table 1.2) can be attained in the “ideal”test
loop.
1.2.4 Evaluate Alternate Instruments
The fourth objective was to evaluate selected alternate instruments. If the performance of any of the
specified instrumentation is not adequate, additional and/or different instruments would be evaluated to
measure the desired slurry characteristics.
A Fischer & Porter DC-driven magnetic flowmeter was evaluated head to head against the Yokogawa
Dual Frequency magnetic flowmeter to determine if it enhanced the performance of the test loop. DCdriven magnetic flowmeters are known for having low drift, whereas AC-driven magnetic flowmeters
experience less noise when measuring slurry volumetric flowrate.
A Lasentec particle size distribution (PSD) analyzer was installed to determine if solids concentration
or PSD could be determined. The Replacement of Cross-Site Transfer System (Project W-058) has
delineated solids concentration and PSD limitations. Although no instruments specified for use in the
Project W-211 design, directly measure the solids concentration or PSD, it was postulated that viscosity,
density, mass flow rate, and perhaps other measured quantities could provide an adequate estimate for
these two unknown parameters.
1.3 Experimental Approach
The slurry properties varied and tested were the slurry specific gravity, viscosity, and solids content.
These properties were anticipated to be largely determined by the characteristics of the carrier
1.5
liquid, the solids selected, and the particle sizes of the solids tested. Each of these tkree parameters was
characterized by a low and a high specification limit. There were eight combinations of these three values
and they can be thought of as the corners of a cube (see Figure 1.1). Simulants were developed to
represent each combination of low and high values of these three parameters. The combination of high
specific gravity and low solids content was not feasible. The initial low solids content, low specific gravity
combination were modified during the course of a run by adding solids to test interior points in this cube.
The interior line in Figure 1.1 represents the relationship between solids loading and slurry density. See
Section 2.3 for a description of the experimental procedures.
Viscosity = f(Solids Loading,
Sluny Density) at constant
Temperature for a given solid
I1
1
1
1
Loading, Slurry Density
\I
Sokds Loadmg, vol%
\1
30
Figure -1.1. Designed Experiment Approach
1.3.1 Test Loop Description
The Instrument Validation Facility (IVF) is located on the mezzanine levels of the 336 Building.
Detailed drawings of the IVF are found in Appendix A. The pump, feed tank, and calibration tank are
located on the ground level and the instrumentation is located on the second and third mezzanine levels.
The piping is 3-inch schedule 40 stainless steel. The programmable logic controller and data acquisition
system are located on the ground floor near a display panel for many of the instruments used in the test
loop.
The IVF is a slurry test loop designed to evaluate various instruments at various slurry flow rates.
Figure 1.2 shows a schematic of the IVF. It consists of a 250 gallon feed tank and a 225 gpm centrifugal
pump which can pump simulant to either one of two test loops with the instruments installed. The
1.6
1
w-211 Loop
I
Red Valve
-
Micromotion
7
i
I
1
IdealLoop
i
To 1/4 Scale
Tank
F-3
r
f
Lasentec
I
Yokagawa Magmete
Calibration Vesse
Three Way Valv
'i.
I
Feed Tank
,-Sample
Variable Speed F'um
(S- 1)
7
I Not to scale or
exact position
i
Figure 1.2. Test Loop Schematic
two test loops are teed off common up-flow and down-flow pipes. The La-sentec Particle Size Analyzer
was mounted in the common down-flow leg of the two loops. The Yokagawa was mounted in the common
up-flow leg of the two loops.
1.7
1.3.1.1 Project W-211 Loop
The Project W-211 instrument loop is located on the third level, 42 feet above ground level. The pipe
size, instrument configuration, and spacing matches that proposed in the Title I design of the upgraded
valve pits for Project W-2 11.
1.3.1.2 “Ideal”Loop
The “ideal”flow instrument loop is located on the second level, approximately 33 feet above ground
level. Instruments in this loop are spaced so that a more fully developed flow profile exists at each measuring point. (Completely developed flow profiles would require approximately 10 feet between
instruments and this was not possible in any reasonable loop design.) Vibrational interference, between the
Nametre in-line viscometer and the Micromotion mass flow meter were expected to be virtually eliminated
because the instruments were adequately spaced.
1.3.2 Test Loop Instrumentation
The instrumentationevaluated is described in this section and shown in Table 1.3.
Table 1.3. Instruments to be Evaluated
Property
Mass Flow Rate
PNNL Supplemental
Project W-211 Specified
Micromotion mass flow meter
Yokogawa magnetic (AC) flowmeter
Fischer and Porter magnetic (DC)
flowmeter
Pressure
Red Valve pressure sensor
Rosemount Differential Pressure
Transmitter with Remote Seal
Assembly
Pressure Drop
Temperature
Moore Industries Thermocouple
I Micromotion Flow Meter
I
Viscosity
Nametre in-line vibrating ball viscometer
Density
Micromotion Flow Meter
Lasentec PSD analyzer
Particle Size
Distribution
~
Solids Loading
Lasentec PSD Analyzer
1.8
Table 1.4 lists the instruments evaluated and the planned method of cross checking and validating the
evaluated instrument measurement.
Table 1.4. Instrument Cross-Check and Validation
Instrument
Method of Measurement Cross-check and Validation
Micromotion Mass Flow Flowmeter
1. Divert system flow for set time to calibrated vessel and weigh.
m
2. Compare to other flowrate instruments (i.e., volumetric)
Density
1. Divert system flow to calibrated vessel, weigh.
2. Take grab sample, measure density in laboratory.
Moore Industries
1. System thermocouples.
Fischer & Porter Flowmeter
1. Divert system flow for set time to calibrated vessel and
weigh.
H
2. Compare to other flowrate instruments.
Yokogawa Flowmeter
1. Divert system flow for set time to calibrated vessel and
weigh.
2. Compare to other flowrate instruments.
I
I Red Valve Pressure Sensor
1 Nametre Viscometer
1. Compare to Pressure gauges in system.
1. Grab sample, measure viscosity in laboratory.
2. Measure AP across the Micromotion mass flowmeter.
Since the flowrate is known, the viscosity can be derived.
Lasentec PSD Analyzer
1. Knowledge of PSD in feed.
2. Grab sample, measure PSD in laboratory.
1.3.2.1 Nametre Viscoliner
The Nametre Viscoliner operates using the torsion-oscillation principle which moves the sensor at
resonance (-650 Hz)at constant amplitude (1 micrometer). The manufacturer states that this method of
operation gives the instrument excellent accuracy over a broad viscosity range. This, however, means that
the Nametre measurement is affected by high-frequency, high-energy, mechanical vibrations translated by
the process. When the viscosity changes, the Nametre Viscoliner varies the power input to maintain a
constant oscillation amplitude. The power requirement is proportional to the viscosity.
1.9
For mounting within a pipeline, Nametre recommends installing the viscometer in a piping tee with the
flow impinging on the sensor as shown in Figure 1.3. Flexible piping (3 feet minimum) or vibration
dampening baffles are recommended if mechanical noise dampening is required. These vibration
dampening devices were not used because they would not fit in the Project W-211 test loop and because
they would not stand up to the radiation environment.
A. This distance should be the minimum distance as dictated by the clearance needed to fasten the flange.
This eliminates stagnant flow between the flange face and the outer edge of incoming flow.
B. First hand tighten all nuts; then stagger tighten to 40 ft lbs. The viscometer flange must be flush
mounted to process flange--no uneven torque.
C. Flexible piping (3 ft minimum) or vibration dampening baffles.
D. Couple pipe section to a stable member such as a floor, wall or structural member of the building.
FLOW
FLOW
Figure 1.3. Nametre Preferred Installation Configuration
1.10
1.3.2.2 Fischer & Porter “Minimag”Flow Meter
Fischer & Porter’s Minimag flowmeter belongs to the general category of electromagneticflowmeters,
commonly referred to as magmeters. Magmeters measure the flowrate in electrically conductive liquids
by applying Faraday’s Law of Electromagnetic Induction. A voltage is induced as a conductive fluid
moves through a magnetic field. The magnitude of this induced voltage, E, is directly proportional to the
velocity of the fluid, V, the strength of the magnetic field, B, and the conductor width (Le, pipe diameter),
D. Since the direction of flow and the magnetic field are at right angles to each other, Faraday’s Law
reduces to
E
=
kBDV
where k is a proportionality constant.
Magmeters can be used to measure the flow rate of fluids with a minimum conductivity of 5 microSiemenslcm (5 micromhos/cm), making this instrument appropriate for most water-based applications.
Magmeters are a popular choice for flowmeter applications since they are non-intrusive and the measurement is not influenced by the slurry density and viscosity. Like most popular magmeters, the Fischer &
Porter Minimag uses a pulsed-direct current (pulsed-DC) system to excite the magnetic coil. The pulsedDC system is popular since the electronics can be re-zeroed between the 7-Hz pulses. This stable zero
feature makes the instrument less susceptible to the effects of coating of the electrode.
1.3.2.3 Yokogawa Flow Meter
As mentioned above, pulsed-DC magmeters can self-zero between each measurement. But for
slurries, the lower sampling frequency can cause problems.
While an AC magmeter reverses polarity at 60 cycles per second, a DC system pulses at 4-7 Hz. At
these low pulse frequencies, the instrument manufacturer claims that the instrument is sensitive to noise
from electrochemicalreactions, high viscosities, and/or low conductivity liquids. Traditionally, slurries
have caused noise in these pulsed-DC magnetic flow meters.
Yokogawa has introduced an instrument which is capable of dual “frequencyexcitation?”called the
Admag. The instrument has a carrier frequency of 7 Hz with a low pass filter, and a 72-Hz superimposed
high-frequency waveform. The two waveforms are then combined. The result is an instrument claimed
by the manufacturer to have all the benefits of zero correction of DC magmeters and the noise immunity of
AC magmeters.
1.3.2.4 Micromotion Coriolis Mass Flow Meter
A Coriolis mass flow meter measures mass flow rate by vibrating a sensing tube at its natural frequency. The flow of mass through the tube causes a twisting action due to the Coriolis effect. The twist
angle is measured by position sensors, which subsequently generate a voltage signal which is proportional
1.11
to mass flow rate. Since the density of the fluid affects the natural frequency of the tube, measuring the
driving frequency of the’tube yields the density of the fluid along with the mass flow rate.
1.3.2.5 Red Valve & Rasemount Pressure Meters
Two line pressure instruments were investigated: a Red Valve pressure transducer (Model 1151 Smart
series 48 with hypalon sleeve and silicone oil sensor fluid) and a Rosemount direct tap sensor (Model
3051CG). The Red Valve pressure instrument was chosen for Project W-211 since the diaphragm is
protected from the slurry process fluid. The Red Valve is certified to /- 1 % of full scale (1.O psig on a
1-100 psig scale). Unlike the Rosemount direct pressure tap, where slurry travels through a 1M-inch tube
to act against a diaphragm, a silicone oil acts as an intermediate (transmitting) fluid so that the slurry never
contacts the diaphragm. The principle behind the Red Valve pressure transducer suggests that plugging
and fouling issues should be eliminated when this instrument is used. Both instruments use the same
transmitter (Rosemount Model 3051CG), with a reported accuracy of 1.0% full scale (3.0 psig on a 1-300
psig scale).
+
1.3.2.6 Moore Thermocouple
The Moore Industries thermocouple assembly is a Type J thermocouple mounted in a 3/4-inch stainless
steel thermocouple well. The thermocouple well is inserted into the slurry stream by means of flange
mounting the assembly to the pipe loop. The signal from the thermocouple is sent to the TCX
thermocouple transmitter which converts the signal to a 4-20 mA which is then read by the PLC. The
transmitter has a zero and span adjustment to be used for calibration.
1.3.2.7 Lasentec Particle Size Analyzer
Only a few instrument manufacturers produce instruments that can measure PSD in-line. Lasentec’s
optical sensor is the only one which claims to measure in-line PSD in concentrated slurries (> 35 vol %)
that can change in composition (optically and chemically).
The Lasentec sensor uses a technique known as focused beam reflectance measurement (FBRM) to
measure the chord length distribution of particles which pass by the instrument. A focused laser is
reflected off of a rotating mirror resulting in a linear velocity of 6 ft/sec. This system reflects light off
particles as they pass. The time the laser light is reflected off a particular particle is recorded by the
sensors electronics. Since the traverse velocity of the laser light (v) is set at 6 ft/sec and the reflection time
is known (t), the size of the particle can be estimated by
s
=
v*t
1.12
Since both the orientation of the laser beam and particles are random, the chord length measured is also a
random chord length as illustrated in Figure 1.4. The instrument will tend to measure larger chord lengths
for larger particles, so there is a natural correlation between a chord length distribution and a PSD.
The Lasentec software prepares a chord length distribution (Le., a PSD) (Figure 1.5). The software
manipulates the chord lengths and prepares four different mean PSD values: average particle size
Figure 1.4. Lasentec Particle Size Measurement Method
Lasentec Particle Size htribution (Run SC)
m
Y
lo
$
I
Particle Size (Microns)
Figure 1.5. Lasentec Particle Size Distribution for Unknown Sample in Run SC
1.13
(number), length weighted average particle size, length squared weighted average particle size, and length
cubed weighted average particle size. These length weighted averages accentuate the presence of the
larger particles more than the simple number average particle size. Thus the presence of a few large
particles, which would not shift the number of the average size, would dramatically change the weighted
average. The length weighted average was used the most during the runs because it gave a good
compromise of not weighting large particles heavily while still emphasizing the effect of deposition and
resuspension of large particles on the average particle size.
The instrument was calibrated at the factory and again on-site by measuring the PSD of polyvinyl
chloride/water slurry. The particle size measurement graphs were virtually identical (see Appendix F).
The Lasentec is not advertised as a solids concentration instrument, but it can be used to determine
relative solids loading up to its saturation limit of 64,000 particles per scan. Since saturation was observed
only once, relative solids loading information should be available for analysis.
The Lasentec does not correct for the velocity of the particles as they move by the sensor window.
Increasing or decreasing the flow rate tends to shift the PSD slightly. Increasing the flow rate causes
particles to move by the sensor faster, thereby decreasing the average time for which the laser reflects off
of a particle. Thus faster flow rates result in slightly smaller measured particle sizes and vice versa for
slower flow rates
The Lasentec M300 Sensor is diagramed in Figure 1.6. It is normally installed in a vertical upflow
run with the sensor window pointed against the flow at a 45 degree angle. Since the laser has a linear
velocity of 6 ftisec (essentially the same as the flow rates tested), a pipe expansion spool piece was used to
locally lower the flow rate to about 4.5 ft/sec.
The probe is normally titanium, but other metals can be used for chemical compatibility. The window
at the end of the probe tip is sapphire, which is claimed by the manufacturer to be chemically compatible
with alkaline waste and radiation. (Recent Hanford experience with sapphire windows used in the velocity
density temperature trees for 101-SY is poor. Field experience initially indicated that the sapphire window
disintegrated; however, more detailed analysis showed that the bonding material between the sapphire
window and the metal mounting bracket failed and allowed tank waste to penetrate behind the sapphire
window.) The probe is positioned about 2 inches into the pipe and angled into the flow at a 45 degree
angle so that it is self-cleaning. (Even after the graphite run,which coated the inside of the Yokogawa and
Fischer & Porter magnetic flow meters, the Lasentec sapphire window remained clean.)
Vertical upflow mounting is recommended to prevent segregation of particles. The Lasentec was
actually installed in a downflow position in the common downflow leg based on process knowledge of
settling behavior at the processing conditions. (The original design phase assumption was that the solids
would be well mixed during both upflow and downflow. This assumption proved to be correct. See
Section 3.3.3 for details.)
1.14
i
!
I
I
I
i
I
I
I
!
1
!
j
I
I
Ii
I
1
i
1.15
The entire instrument is claimed to be safe for use in a radiation area. The M300 has only one moving
part: a bearing assembly which is used to rotate the laser light. This bearing assembly is rotated by 60 psi
instrument air. The electronics (including the diode laser) are located in a remote field unit connected by a
fiber optic cable to the sensor. The field unit is normally located no more than 10 meters away from the
probe, although this distance can be increased.
Lasentec recommends that the bearing assembly be replaced about once a year if the probe is continuously used. Normally the entire probe must be removed from the pipe in order to change the bearings,
however the probe can be engineered to avoid this requirement.
1.3.2.8 Pressure Drop
The pipe loop flow differential pressure was measured with a Rosemount 305 1CD transmitter with a
remote seal assembly with a 0.25 inches of water accuracy. The transmitter was calibrated such that the
4-20 mA output signal was proportional to 0 to 25 inches of water pressure. The remote seal assembly
consists of two flange mounted diaphragms connected to the transmitter via tubing which is filled with a
seal fluid, thus preventing the slurry solids from plugging the impulse tubing or the transmitter. During
operation, the isolating diaphragms and the fill fluid on the high and low sides transmit process pressure to
the oil fill fluid. The fluid in turn transmits the process pressure to a sensing diaphragm in the center of
the transmitter. The sensing diaphragm functions as a spring which deflects in response to differential
pressure across it. Capacitor plates on both sides detect the position of the sensing diaphragm. The
differential capacitance between the sensing diaphragm and the capacitor plates is converted to an output
signal to the PLC.
1.16
.
2.0 Experimental Instrumentation and Procedures
This section presents the instrumentation used in the analytical laboratories to compare the test loop
measurements. Also included is a description of how the experimental slurries were derived.
2.1 Analytical Instrumentation
The analytical equipment used in the PNNL Slurry Processing Laboratory to measure the slurry
simulants and the slurry samples is discussed in the following subsections.
2.1.1 Haake Rheometer
The Haake rheometer operates according to the Couette principle in which an outer cup rotates while
the inner cylinder, which measures the torque, remains static. The outer cylinder is driven by an electronically controlled motor while the resistance of the sample to flow causes a very small movement in a
torsion bar, mounted between the motor and the drive shaft. This movement is detected by an electronic
transducer. Signals proportional to the speed and torque are transmitted to the control unit for processing
and display.
The Haake rheometer has precise temperature control for the sensor system because the measured
viscosity is sensitive to temperature change. The sensor system used in this test was Mooney-Ewart
(ME) 45.
The Haake rheometer measures the shear stress as a function of shear rate from 0 to 300 sec-' and can
be used to analyze slurries that follow several types of rheological behaviors as illustrated in Figure 2.1.
The viscometer is not well suited for large particles with high settling rates. Simulants that contained the
large (500 micron) silica solids were difficult to measure and exhibited high levels of variability.
Figure 2.2 presents an illustrative rheology curve representing a typical rheogram measured with the
Haake rheometer. (The experimentally measured rheograms have a much faster initial slope than is
depicted in this figure.) The slurries evaluated in the instrument validation test loop (as well as actual tank
wastes) all are categorized as shear thinning slurries or Bingham plastics. The figure shows that shear
stress goes up as the shear rate is increased; however, the ratio of shear stresslshear rate goes down with
increasing shear rate.
The test data can be regressed to determine structure coefficients, yield points, or viscosity temperature coefficients. Ten different models and an automatic evaluation of best fit are available.
The Nametre Viscoliner operates at a single shear rate of 4084 sed'. It is believed to report the slope
of the line between 0 and 4084 sec-'. Therefore, a strategy of extrapolating the Haake viscosity data based
on the best fit to 4084 sec-' and plotting the two against each other was selected. This strategy was selected
2.1
b
Extrapolation from best Fit
~
1 4084 sec I
-I
Figure 2.1. Classes of Viscosities
Extrapolation from best fit to 4084 sec
Typical IVF Sluny or Tank Waste
Sluny (Le., Shear Thinning)
( i.e., Shear StressiShear Rate)
Viscosity at 1000 sec .';
(Le., Shear StredShear Rate)
I
I
I
Gammaldot, Le., shear Rate' in sec
-' '
Figure 2.2. Classes of Viscosities
2.2
b
despite the dangers of extrapolating an order of magnitude even for very good fits because any errors are
magnified. This was the only known method to compare the laboratory based viscosities to the Nametre
viscosity.
The WHC Project W-211 design criteria requires that the viscosity of the slurry be less than 30 CPat a
slurry.velocity of 4.5 ft/sec, or 10 CPat a slurry viscosity of 6 ft/sec. These viscosity limits were used to
calculate a Reynolds number (and hence estimate the friction factor) when determining the pressure drop
across the line. Viscosity is defined here as the shear stresdshear rate. Since the pressure drop measured
in the pipe is due to the friction of the fluid at the wall, the shear stress at the wall, t, is directly
proportional to the pressure drop:
D
T~=~APL
The shear rate at which the 10 or 30 CPviscosity limit is measured at must therefore be the shear rate
at the wall. Unfortunately, it is difficult to precisely measure the shear rate at the wall since the flow is
turbulent and non-Newtonian. It is possible, however, to obtain an order-of-magnitude estimate of the
shear rate of the wall by considering Newtonian turbulent flow. The purpose of estimating the shear rate
at the wall is to demonstrate that the viscosity of a non-Newtonian (i.e., shear rate dependent) fluid measured by the Haake at 300 sec-' or by the Nametre at 4084 sed' will probably not correlate to the shear rate
of the fluid at the wall (i.e., the shear rate at which the Cross-Site Transfer System viscosity limit is
imposed).
At the fluid velocities tested in the test loop (6 ftkecond), the expected shear rate is on the order of
lo00 sec'' and a corresponding shear stress can vary between 1 and 100 Pa depending on the fluid.
The loo0 sed' shear rate was determined from a rough approximation of the fluid velocity near the
wall of the pipe. Bird et al. (1960) give the velocity profile of a turbulent Newtonian fluid to be
and vZ,- equal to
and vBthe bulk velocity.
e
2.3
It is further assumed that the turbulent core gives way to a boundary layer at a radius of 0.05 R (Bird
et al. 1960, Figure 5.1-1). This boundary layer is responsible for most of the fluid resistance, and hence
the pressure drop. Thus, the fluid viscosity associated with the pressure drop is directly related to the
shear rate at the wall. The shear rate at the wall is then approximated as the fluid velocity near the wall
divided by the distance from the wall:
Again, this estimate makes several assumptions which may not be rigorously correct for the slurries
tested within the IVF. In particular, some rather large assumptions are made about the boundary layer
resistance for a non-Newtonian fluid. However, for purposes of this report, this shear rate estimate will
serve to illustrate why the 10 and 30 CP limits may not be measured directly by either the Haake or
Nametre viscometers.
2.1.2 Horiba Particle Size Analyzer
The Horiba CAPA 700 was used to measure the PSD of the slurry samples in the laboratory. It uses a
non-contact measuring method based on liquid phase sedimentation and light transmission techniques. The
slurry samples were extremely diluted to avoid saturation of the detector. In this method, Stokes' sedimentation equation is combined with the proportional relationship between the absorbency and particle
concentration. A particle having diameter (D) and density (r) in a solvent of density (ro) and viscosity
coefficient (&) will settle at a constant velocity according to the Stokes sedimentation law by the effect of
gravity. Particles with a larger diameter settle faster than particles with a smaller diameter if all particles
have the same density.
For centrifugal sedimentation (CS), the relation between particle size (D) and particle density (p),
solvent density (po) , solvent viscosity(q o), centrifugal rotational angular velocity (o(t)), distance between
center of rotation and sedimentation plane (X,), and distance between center of rotation and measuring
plane (Xd is given as follows:
where w(t), xl, and x, are constant instrument factors during the entire centrifugal sedimentation measurement. The sample characteristicsare described by both solid and solvent physical properties. Water was
used as the dispersion medium for all samples used in the data analysis. For a single component slurry
system, particle size is a key factor affecting sedimentation velocity, which in turn influences the measured
optical properties. However, for a multi-component slurry system, both density and particle size can control the particle sedimentation velocity. The equation for calculating the average density in a mixture of
solid particles is
2.4
The optical transmission method measures the degree of particle sedimentation by measuring the
amount of light transmitted. The intensity of the transmitted light is proportional to particle concentration
and size. The relationship between measured optical property and particle size is given below:
where I,, is the incident light intensity, 4 is the transmitted light intensity through particle Di, K is the
optical coefficient of the cell and particle form, k(D> is the absorptivity of particle Di, Ni is the number of
particle Di, and Di is the particle diameter.
Figure 2.3 shows a typical PSD taken during testing. The graph shows both an incremental and
cumulative PSD, respectively, as a function of particle size in microns.
2.1.3 DYNAC Centrifuge
A D m A C Centrifuge (60 Hz,from Becton Dickinson 8z Company, Model # 420101) was used in the
measurement of the slurry samples solid volume percentage. Each sample was centrifuged at 1350 RPM
rotation (centrifugal radius of 16.3 cm) for 5 minutes and resulted in a very good phase separation. The
separated solids were dried in an oven at 60°C for 3 to 4 days. To ensure that no residual water remained
in the solid, the samples were put back into the oven, and re-weighed the following day to see if there was
any additional drying. The solids volume concentration was obtained by dividing the measured solids
weight, its density, and the total volume and weight of the sample.
2.1.4 Pycnometer
The pycnometer is used to take fast measurements of the density of a slurry. It works by filling a
container of a known volume, weighing the mass of the slurry, and dividing the mass by the known
volume. These density data are taken within a few minutes of taking the samples and are measured
essentially at the temperature of the slurry in the test loop.
2.2 Experimental Procedure
The location of the instruments can be seen in the test loop schematic shown in Figure 1.2 for the
“ideal”and Project W-211 loops. A throttle valve controls the flowrate and a pneumatically operated
3-way valve directs the simulant either to the feed tank or to a 100 gallon (16-inch diameter schedule
2.5
Figure 2.3. Typical Particle Size Distribution fiorn Horiba (Run 530)
2.6
30 pipe 10 feet high flow calibration tank). Control of the 3-way valve is through the computer which
measures the duration the flow is diverted to the flow calibration tank. The computer also integrates the
signal from all the flowmeters during the diversion and displays the total flow for each flowmeter during
the flow diversion. The level in the flow calibration tank corresponds to a volume and can easily be
measured within 1/4 inch or 0.75 liters. The flow calibration tank is mounted on load cells so that the
mass of the slurry can be measured and the slurry density can be determined by dividing the mass by the
volume. After the flow and density calibrations, the simulant is drained back to the feed tank. Bubbles
can be introduced into the IVF just downstream of the pump through a 1/4-inch tubing connection.
Tests were generally carried out as follows. The simulant was prepared in the feed tank which has a
1.5 HP agitator to keep the solids suspended. The desired solids were weighed and added 9 a known
quantity of water in the feed tank. After the solids were added to the feed tank, the pump was started
directing flow from the feed tank to the desired loop ("ideal" or Project W-211) and back to the feed tank.
The throttle valve was adjusted to achieve the desired flowrate. System parameters were then monitored
until steady-state is reached, as determined by a stable density and PSD. Typically the system required
from five to ten minutes to reach steady-state.
After steady state was achieved, three flow calibrations were conducted by diverting flow from the
feed tank to the flow calibration tank for 20 to 30 seconds. Since the estimated error in reading the
volume is 0.75 liters in the nominal 20 seconds diversion at 9 liters per second, the estimated percentage
error in flow rate is 0.4%. The error at the 6 liter per second conditions is also 0.4%. The volume in the
flow calibration tank was compared with the volume obtained from the computer (which integrated the
instantaneous flow signals from the flowmeters). The mass and volume of the slurry in the flow calibration tank was used to determine the slurry density. This measured density is compared with the density
reported by the Micromotion densimeter. The flow rate data from the Yokagawa, Micromotion, and
Fischer & Porter flow meters is expressed as a percent error from the calibrated vessel flow rate and not
as volumetric flow per time. The reason for this is that the flow rate in the loop was not constant when the
flow is diverted to the calibration tank.
After the flow calibrations, individual density measurements are taken from each of the four sample
locations (S-1 through S-4) and recorded with the time and temperature. This information is used as a
check and to ensure that the density at the top (sample S-1) and bottom (sample S-2) of the feed tank are
similar (a large difference would indicate significant solids settling in the feed tank). Samples were also
taken from the upflow sample locations (S-3) and downflow sample iocation (S-4) in the test loop for
analytical laboratory testing (Haake viscosity, PSD, vol% centrifuged solids).
The flowrate, density, Nametre viscosity, Moore Instrument temperature, and the Micromotion temperature are time averaged for ten seconds via the computer and stored electronically. The differential
pressure measured over a ten-foot section of straight pipe was also recorded in the Project W-211 loop.
These values were recorded manually at least three times during the run.
2.7
2.3 Slurry Preparation and Matrix
Physical properties of approximately 35 solids were obtained through literature searches and product
brochures and by personal communication with the manufacturer or manufacturer representatives. A
subset of nine solids was selected for further analysis. Samples were obtained and tests were performed in
the laboratory to verify the manufacturers claims of physical properties. Slurries of some of the solids that
were interesting were prepared and viscosity measurements were performed in the laboratory. (Detailed
information on the testing of these solids and slurries is available in the Project files but will not be
presented in detail here.)
Based upon solids and slurry characterization work and on the hazards/toxicity of the solids, slurries were
selected for use in the IVF. The process of selecting the exact'simulant and combinations of simulants
was intuitive but was guided by the designed experiment objective and the known properties of the simulants previously tested. Combinations of types of solids, concentrations, size distributions, and slurry
viscosities were evaluated until a set of simulants were found that represented each corner of the design
cube. Interior points were evaluated by adding solids to a low solids loading simulant or alternatively by
adding water to a high solids loading simulant.
Simulants were specifically selected for the following:
I. Some simulants were selected to include a variety of solid densities which when combined with water
in solids loading concentration of 0 to 30 vol% solids were constrained by the Project W-211 density
specification of 1.0 and 1.5 gm/cm3. These simulants were made up from either plastic or silica
powders to form the low (1.0 gm/cm3)and high (1.5 gm/cm3)density simulants. The tests with the
neutrally buoyant plastic powders were designed to create a simulant where settling of the solids did
not occur in the upflow and downflow portions of the test loop.
2. Some simulants were selected to include a variety 6f solids which when combined with water were
constrained by the Project W-211 solids loading constraint of 0 to 30 vol% solids and not the density.
These simulants were primarily made up of Gibbsite.
3. Some simulants were selected to include a variety of solids which when combined with water were
constrained by the Project W-211 viscosity constraint and not by either the solids loading or slurry
density. These simulants were primarily made up of combination of solids with small particle sizes
(i.e., Bentonite, Graphite, and Mica).
4. Some simulants were selected to give a variety of slurry PSDs, including some bi-modal distributions
by having multiple solids with different PSDs. These simulants were primarily made up of small
particle size silica and solids with a very small particle size (Le., Bentonite or Graphite). The simulant used in Run 5C1 also contained the W-058 specification limit of 1 vol% 500 micron particles (Le.>
large silica powder).
2.8
5 . The simulants described above were also used to test effects of the shape of the solid particle. No
simulants were prepared with the express explicit purpose of testing particle shapes. The original test
plan called for the evaluation of particle shapes (Le., spherical, irregular, rod-like, plate-like). All of
these shapes are represented in the simulants with the exception of the rod-like particles. Suitable
particles for this geometry could not be located.
Processing conditions were chosen to cover the range of operational parameters that were used to
design the Project W-211 test loop.
The slurry simulant and processing conditions test matrix is shown in a table in Appendix B.
2.4 Experimental Text Matrix and Data
All the runs listed in the test matrix in Appendix B were executed with one exception. The exception
was one of the unknown slurries to determine if the operator could tell the appropriate slurry characteristics based purely on the instrumentation. This condition (Run 8D) was designed to be a very viscous
slurry that would challenge and potentially plug the test loop. During the switch over from Run 8C to 8D,
the operators observed tremendous increases in the reported viscosity (Le., from a nominal 40 CPto over
100 cP). The operators felt that this would plug the test loop and shut the run down. The experimental
condition was not deleted on the test matrix, however. Appendix C shows the experimentally measured
and laboratory based analytical data collected.
2.5 Statistical Analysis
Two types of statistical analyses are used in the analysis of the data. The first statistical test is a paired
t-test that compared two sets of values. It calculates the difference between the ordered sets of values and
determines if the average difference is statistically different from zero at the 95% confidence level. These
tests were used to determine if settling occurred in the feed vessel and if the laboratory measured analytical
values are different from the values measured at the test loop.
Multivariate statistical analysis is the second type of analysis. This analysis was used to determine if
the performance of any specified measured response is a function of any other or any other grouping of
predictor variables. These analyses are used to help pinpoint where the sources of error in the flow rate
measurements, A P, and viscosity measurements derived. A software package' called SimcaP was chosen
to analyze the project data.
"
Multivariate statistical methods are used since the amount of data is large and the process variables are
interrelated. Software packages like SimcaP essentially reduce the number of independent variables by
creating new variables which are nothing more than linear combinations of the original, measured variables and measured slurry properties. For example, the 13 measured variables from the IVF (Yokogawa
flow rate; Moore Industries Temperature; Lasentec mean particle size; laboratory mean particle size;
2.9
Nametre viscosity; Haake viscosities at 300, 1O00, and 4084 sed'; line pressure; Lasentec small particle
count; differential pressure, vol% solids; and Micromotion density) were reduced to five statistically relevant independent variables by the SimcaP software.
SimcaP performs two types of multivariate analyses: principle component analysis (PCA) and partial
least squares analysis (PLS). PCA is used to find outliers, recognize groups of data, and determine the
general relationships between process variables. PLS, on the other hand, performs regression to relate
input parameters to designated output parameters. For example, PLS can be used to find a quantitative
relationship between viscosity and the rest of the measured variables.
2.10
3.0 Results and Discussion
The results of the test loop testing are discussed in this section. The discussion is split into general
observations that were either associated with the startup of the test loop or cut across all runs, Specific
discussion of the performance of the individual test loop instrumentation is presented in Sections 3.2
through 3.5.
3.1 General Observations
General observations associated with the start-up and testing of the test loop and observations that
occurred for all runs are discussed in Section 3.1.
3.1.1 Temperature
In every run,the temperature of the slurry increased from a nominal starting temperature of 10°C to
20°C and leveled out at about 40°C to 45°C. While a temperature rise was considered in the preliminary
design of the test loop, no temperature control system was installed. The magnitude of the temperature
increase was not anticipated and led to a large increase in laboratory analytical work and in data analysis/
modeling work to determine the viscosity vs. temperature dependence of the slurry samples. If the slurries
were shown to be temperature sensitive in laboratory tests, this correlation was then used to correct the
observed viscosity data to the temperature at which the Nametre viscosities were recorded. The temperature corrections in the viscosity values are sources of small errors in the analysis.
Similarly, the Micromotion densities were recorded at a different temperature than the pycnometer
samples. The Micromotion densities were either interpolated or extrapolated to the temperatures that the
pycnometer samples were taken at. These interpolations or extrapolations were generally within a range of
5°C or less. The temperature corrections in the density values are also sources of small errors in the
analysis.
3.1.2 Vibration
Viscosity standards were tested with the Nametre viscometer prior to installation to verify the viscometer worked properly. A static test stand site next to the Nametre was installed in the test pipe loop so
that the Nametre could be merely unbolted and placed in the stand while conducting static tests. The
Nametre provided steady viscosity values that were 1 to 2 CPhigher than the accepted standard. Because
these tests were conducted in the 336 Building High Bay, which is unheated during December, the value
was expected to be a little high as the viscosity standards cooled slightly during testing.
After the static tests, the viscometer was installed in the pipe loop in accordance with factory instructions. However, during shakedown testing the Nametre viscometer did not work properly with flow in the
pipe loop regardless of flow rate, as the Nametre was subject to severe vibrations. For example, it rapidly
3.1
fluctuated between 10 and 27 CPwhile running a simulant which in the lab had a viscosity of ‘5 cP. The
Nametre was again tested statically with viscosity standards and satisfactory results similar to the previous
static tests were obtained. The Nametre was re-installed and various equipment was turned off that may be
causing interference with no success (e.g., Micromotion mass flowmeter). The Nametre was sent back to
the factory to check the system out. They reported the unit was operating properly, but the symptoms
persisted.
The cause of the poor readings was determined to be vibrational interference. The vibration in the
Project W-211 loop was measured and the vibrations present at various frequencies neared or exceeded the
acceptable level defined by Nametre (see Figure 3.8). The pump was isolated with rubber expansion
joints, causing a substantial decrease in the measured vibration in the Project W-211 loop, and the Nametre
began operating properly.
Unfortunately, the Nametre could not be made to work in the “ideal”loop even with the pump
isolation. Vibration measurement in the “ideal”loop showed it to be within the acceptable level as defined
by Nametre. Nevertheless, despite all efforts, the unit did not function properly in the “ideal”loop and
was not used.
3.2 Project W-211 vs. “Ideal”Loop
One of the lesser objectives of this project was to determine if the design of the Project W-211 test
loop was satisfactory as compared to the “ideal” test loop where the instruments were spaced out and flow
was allowed to more fully develop. To test this, the pycnometer densities, Micromotion densimeter, and
the Micromotion flow meter in the two loops were compared against the laboratory data. This comparison
used a paired t-test previously described with one added feature. All data collected in both loops were
used in the analysis and data was segregated by “Groups”(Le., Project W-211 or “ideal”)and then the
paired t-tests were run. The results showed the following:
1. The average of the upflow density (S-3) and downflow density (S-4) measured with the pycnometer
(labeled S-31s-4 in subsequent graphs) compared to the lower feed tank density (S-2) (measured with a
pycnometer) were not statistically different. (See Grouped t-Tests in Appendix D-3).
2. The Micromotion densimeter performance compared to the average of the upflow (S-3) and downflow
(S-4) density (measured with a pycnometer) were not statisticaily significantly different. (See Grouped
t-Tests in Appendix D-4).
3. The Micromotion density performance compared to the density from the calibrated feed vessel was not
statistically significantly different. (See Grouped t-Tests in Appendix D-5).
4. The Yokagawa and Micromotion flow rate meters are statistically different from each other and are
also statistically significantly different in each loop. In the Project W-211 loop, the Micromotion flow
meter measured 5.9% less flow than the Yokagawa flow meter. In the “ideal”loop, the Micromotion
flow meter measured 7.9%less flow than the Yokagawa flow meter. These regressions were poor
3.2
Appendix D-7.) Since the Yokagawa is in a common location for the testing in both loops, the test
loop must have affected the Micromotion flow rate. The relevance of this observed difference is discounted because the grouped t-Tests on the Micromotion density did not differ between the two loops
and because the temperature reported by the Mcromotion thermocuple consistently agreed with the
Moore Industries thermocouple.
The only grouped t-Test analysis that showed that there was a difference between the two test loops is
item 4 above and it had relatively poor correlations with wider confidence limits. Since the other five had
very good regressions with tight confidence limits and showed no difference between the test loops, it is
not felt that there is a difference between the two flow loops. Therefore, the data analyses in all of the
following sections will be on the full data set comprised of both loops.
3.3 Instrument Validation
The following sub-sections describe the results of each instrument being evaluated separately.
3.3.1 Pressure
The Rosemount and Red Valve instruments functioned consistently well and gave measurements that
were consistently within about one psi of each other (Figure 3.1). Typical operating line pressures were
60-90 psig.
100
90
80
70
60
50
40
40
50
60
70
80
RedValve Pressure, psi
Figure 3.1. Rosemount vs. Red Valve Pressure
3.3
90
100
3.3.2 Temperature
Two thermocouples were installed in the slurry pipe test loop: the Moore Industries thermocouple (as
specified in the Project W-211 FDC) and the Micromotion thermocouple. The Moore Industries thermocouple was recorded by the data acquisition system. The Micromotion transmitter only allows two
4-20 mA signals to be sent to the PLC, consequently the thermocouple reading is only available as a
readout from the rack mounted display panel and is not logged to the PLC data acquisition system. Micromotion temperatures were manually recorded whenever samples were taken for density measurements and
just before flow calibration tests.
The Micromotion thermocouples measured temperatures consistently between 0.2 and 0.9 degrees
higher than the Moore Industries thermocouple (Figure 3.2). The Micromotion thermocouple has a
reported accuracy of & 1"C (CMF 300 Product Brochure), and the Moore Industries thermocouple has a
reported accuracy of 0.1% of full scale (full scale = 200"C), or L- 2°C. Thus, the nominal error for the
two thermocouples overlaps the reported difference between the temperature measurements. No other
independent measurement of temperature was made.
50
45
G
Y
40
e 35
10
5
5
10
15
20
25
30
35
40
Moore Industries Thermocouple Temperature (Deg C)
Figure 3.2. Thermocouple Comparison
3.4
45
50
3.3.3 Density
The results of the density analyses showed that settling did not occur in the feed tank or in the test loop
itself. The Micromotion densities agree very well with the reference densities and to the densities derived
from the feed tank volumetric calibration tests.
3.3.3.1 Settling in Feed Tank
Two sample ports were installed into the feed tank to determine if solids settled during the course of a
run. Samples from the upper (S-1) and lower (S-2) sample ports were taken during each run. The samples were analyzed in the field with a pycnometer and in the laboratory. If the solids settled, the solids
concentration or density of sample S-2 should be greater than in S - 1.
Analysis of the laboratory based density as measured with a pycnometer is shown in Figure 3.3. The
largest percentage increase observed in the density of the S-2 sample was less than 1% and the average
was calculated to be 0.03%with a standard deviation of 0.24%. A paired t-Test showed that the difference between the s-1density and s-2 density was not statistically significantly different from 0 (see
-0 6
08
09
I
11
12
S2 Density, @cm3
Figure 3.3. Settling in Feed Tank
3.5
13
14
15
Appendix D-1). Also, a paired t-Test of the volume% solids measured in the laboratory showed that the
difference between the S-1 solids volume% and S-2 solids volume% was not statistically different from 0
(see Appendix D-2). Therefore, no settling is observed in the top layer of the feed tank.
The average density of the upflow (S-3) and downflow (S-4) samples taken in the test loop legs was
plotted against the average density of the feed tank samples (S-1 and S-2) (labeled S-US-2 in subsequent
graphs) to test whether settling may have occurred in the bottom of the feed tank (see Figure 3.3). The
results show a near perfect correlation with no evidence of settling in the tank. Also, the results of a
paired t-test showed that the difference between the average of the S-3 and S-4 samples and the average of
the S-1 and S-2 samples was not statistically significantly different from 0 (see Appendix D-3).
3.3.3.2 Micromotion vs. Analytical Values
Figure 3.4 shows a regression analysis of the Micromotion density plotted against the average laboratory density measured by the pycnometer from Samples S-3 and S-4. The line should have a slope of one.
The fit was not expected to be perfect because the temperature of slurry in the pycnometer samples may
have cooled slightly during the measurement process, because of random errors in the weiglung/filling
process and because the S-3 and S-4 samples were not taken at the same temperatures as the Micromotion
08
09
1
12
11
13
SllS2 Awrage Density, gdcm3
Figure 3.4. Test Loop Density vs. Feed Tank Density
3.6
14
15
densities. The Micromotion densities were interpolated (or extrapolated) based on the slurry temperature
taken simultaneously with the Micromotion density measurements to correct them to the S-3 and S-4
temperatures.
The observed regression line has a slope of 1.122. The paired t-Test indicates that the difference
between Micromotion density and the average of the S-3 and S-4 density is statistically significantly different from 0 (see Appendix D-4). This result is consistent with the regression line for the Micromotion data
in Figure 3.5 because the Micromotion appears to read slightly high at all densities but the offset appears to
be a consistent percentage high. All Micromotion densities were observed to be within the 5% allowable
error tolerance specified with the exception of Run 2. Run 2 with the plastic powder produced a thick
foam in the feed vessel with large amounts of air trapped in the feed and resulted in the S - 3 / S - 4 density of
0.88 gm/cm3. The S-3 and S-4 measured densities are about 0.1 gm/cm3lower than measured by the
Micromotion because the feed to the test loop is taken from the bottom of the tank and does not pick up
any of this entrained air foam.
1.6
1.5
1.4
{
1.3
d
'B&
1.2
5
z
z2
1.1
1
0.9
0.8
0.8
0.9
1
1.1
1.2
1.3
1.4
S3/S4 Analytical Densities, g m / c d , Air Runs Included
Figure 3.5. Micromotion Density vs. Average S-3 & S-4 Density, Air Runs Included
3.7
1.5
t
Figure 3.6 is the same data as in Figure 3.5 except that runs where air bubbles were deliberately
introduced were excluded. The observed regression line has a slope of 1.1654 and is a slightly better
correlation than in Figure 3.5.
Figure 3.7 shows a regression analysis of the Micromotion measured densities compared to the slurries
densities calculated from the calibrated vessel tests. In theory, the line should have a slope of one. In this
case, the Micromotion densities and calibrated vessel derived densities were measured at the same temperatures. No temperature interpolation or extrapolation was required.
The observed regression line has a slope of 1.1044. The observed fit is not as good as the previous
regression. The paired t- test indicated that the difference between the Micromotion density and the measured density is statistically significantly different from 0 (see Appendix D-5). This is not surprising
because there is more random error in measuring the calibrated feed vessel volume than the pycnometer
and because the Micromotion densimeter has already been shown to read a little high. Since the calibration derived densities are derived from the measurements of the flow rates, the detailed discussion of
sources of errors is presented in Section 3.3.5 for flow rates. The Micromotion appears to read slightly
high at all densities but the offset appears to be a consistent percentage high. All Micromotion densities
1.6
1.5
1.4
.-d
1.3
E
1.2
nB
8
E
1.1
1
0.9
0.9
1
1.I
1.2
1.3
1.4
1.5
1.6
S 3 6 4 Analytical Densities, gm/cm3, Air Runs Excluded
Figure 3.6. Micromotion Density vs. Average S-3 & S-4 Density, Air Runs Excluded
3.8
1.6
1.5
1.4
5
a0
.
I
1.2
E
0
*
.
I
sb
0
1.1
s
1
0.9
0.8
0 .a
0.9
1
1.1
1.2
1.3
1.4
1.5
CalitxatedVessel Density, gm/cm3, Air Runs Included
___________~
~
Figure 3.7. Micromotion Density vs. Calibrated Vessel Density
were observed to be within the 5% allowable error tolerance specified with the exception of Run 2. Run 2
with the plastic powder produced a thick foam in the feed vessel with large amounts of air trapped in the
feed. The S-3 and S-4 measured densities for Run 2 are about 0.1 gm/cm3lower than measured by the
Micromotion.
3.3.4 Viscosity
The Nametre Viscoliner was difficult to install and make operational due to the vibrations in the test
loop. Once operational, large spikes in viscosity were seen. The cause of these spikes was not determined
but may have been random vibrational effects. The Nametre was found to give a qualitative agreement
with the laboratory measurements but did not meet the desired acceptability criteria.
3.3.4.1 Temperature Change Impacts
Most tests started with a loop temperature of 10°C to 20°C, but ended at around 40°C -to 45°C
because of the heat added to the fluid by the pump. Between 10°C and 40"C, the viscosity of water drops
3.9
from 1.307 to 0.6529 CP(CRC Handbook of Physics). Since most empirical relationships show a direct
relationship between the slurry viscosity and the carrier liquid viscosity, it was assumed that the viscosity
would decrease appreciably at higher temperatures. The simplest such relationship is the Einstein equation, which describes slurry viscosity as a function of the carrier liquid viscosity and the solids volume
fraction:
where p is the slurry viscosity, p, is the carrier liquid viscosity (i.e., water), and @ is the solids volume
fraction.
The Einstein equation applies only to dilute suspensions of spherical particles up to a void volume of
approximately 0.10. Other correlations, which are functions of only the carrier liquid viscosity and the
void fraction, have been developed. The Thomas (1965) correlation, for example, was developed with
particles ranging in size from 0.1 to 435 microns, and solids concentrations up to 80 volume%:
p = pc[l+2.5@ + 10.05@2 + 0.00273exp(16.6@)]
Since the slurry viscosity was expected to decrease as the temperature increased, and temperature
increases as the day progresses, the run order was adjusted so that the more concentrated slurries (Le.,
high viscosity) were run first (e.g., Runs 5AO and 5A). For other runs,the more dilute slurries were run
before the concentrated slurries (e.g., Run 4A - 4A3). By alternating runs in this fashion, the effects of
temperature and solids concentration can be separated in the statistical analyses (Le., decoupled) but the
effects are still seen in the test data.
3.3.4.2 Observed
This section presents viscosity results from the available techniques to measure the on-line viscosity.
Nametre Vibrational Interference. Mechanical noise in the system caused problems with the
Nametre readings in both the Project W-211 and the “ideal”loops. Nametre provided guidelines on the
vibrational interferences that are acceptable for acceptable performance. The guideline is illustrated in
Figure 3.8. The figure also plots IVF vibration (Project W-211 loop) measurements before and after
mechanical isolation of the pump. Before pump isolation, the amplitude of the deflections were near or
above the unacceptable region over a range of frequencies. Before pump isolation the Nametre viscosity
measurements were extremely erratic, fluctuating from “1 to -27 CPwhile running water in the system,
regardless of system flowrate. After isolation, the amplitude of the deflections were reduced into the
acceptable region and the Nametre viscosity measurements were generally stable within 2 cP.
3.10
3.11
Figure 3.9 shows a plot of the Nametre viscosity versus time throughout Run 3. The solids loading
was changed twice during this run. The Nametre reported viscosity jumps at each change. Within these
three time periods, the Nametre fluctuated around a mean value. There were numerous spikes in the
reading and then it decayed back to the nominal reading. These spikes may have been due to random
vibrations entering the system from outside of the test loop but have never been satisfactorily explained. If
the Nametre is to be used, it is important that an averaging routine be used so that a control action is not
taken based on one of these instantaneous spikes.
0 35
I
03
35
U
2 25
3,
P
I
2 20
e
-p
.-
j
Ct:
5
0.15
I
u
15
10
0.05
5
0
0
Figure 3.9. Nametre Viscosity Stability
Nametre Viscosity Correlation. The Nametre torsion-oscillation principle applies a shear to the fluid
which is significantly different than the shear experienced by the slurry at the pipe wall. Slurries expected
in t h cross-site
~
transfer system are generally shear thinning, so the measured viscosity is likely to be shear
dependent. The Nametre viscosity is measured at a shear rate of 4084 sed', while the estimated shear at
the wall in the ISF at 6.0 ft/sec is on the order of lo00 sed'. Since most of the slurries run in the IVF
were shear thinning, it is not surprising that the Nametre viscosity readout was consistently less than
expected from the laboratory Haake instrument with a maximum shear rate of 300 sed'. Figure 3.10
shows the raw data obtained from the Haake in the lab against the measured Nametre viscosity in the test
loop.
3.12
75
b
.I
25
.Y
>
5
-5
0
20
40
60
80
100
120
140
160
Haake Visccsityat300 Reciprocal Seconds
Figure 3.10. Nametre vs. Haake at 300 Sec-'
Figure 3.11 presents the Nametre viscosity against the Haake viscosities extrapolated to loo0 sec-',
which is the estimated shear rate at the wall in a 3-inch pipe at 6 ftlsec. The pressure drop measured in
the test loop should correlate to this approximate shear rate.
Figure 3.12 presents the Narnetre viscosity against the Haake viscosity when extrapolated to 4084-'
reciprocal seconds. Note that a data point at a Haake viscosity of 65 CPvs. a Nametre viscosity of 59 CP
is nearly obscured by the best-fit line and the lower adequacy limit line. The fit is surprisingly good considering ,the uncertainties in the extrapolation process and the operational difficulties with the Nametre. The
Nametre does not, however, meet the +I- 10% accuracy criteria as specified by Project W-211 and as
illustrated with the Upper and Lower adequacy limit lines. The Nametre appears to be useful as a general
guide for determining if the viscosity of the fluid changes during transport and may play an important role
in process debugging and control.
3.13
0
10
20
30
W
40
50
60
70
EO
90
1M3
e Viscaritylkhpolatedto 1000 Reciprocal SeconL
Figure 3.11. Nametre vs. Haake at lo00 Sed’
Nametre Discussion. Figure 3.12 showed that the Nametre provides a good qualitative agreement
with laboraton measured viscosities and could therefore be suitable as an on-line indicator. The Nametre
was used for this purpose in the testing program. However, the difficulty in installing and getting the viscometer to work in the Project W-211 loop and inability to make it work at all in the “ideal”loop must be
considered equallyh the decision on whether to install or not install the Nametre in a high-radiation field
environment where there is limited ability to debug the unit.
If the decision is that the qualitative indicator of the viscosity is needed, it is critical that the mechanical vibration in the Project W-211 test loop be quantified and reduced to well below the manufacturers recommendations before installation. The vibrational interference was registered in the Nametre like a
switch, causing either extremely erratic reading when a threshold was exceeded or no problems when the
mechanical noise was dampened.
The Nametre may be more useful for a single reasonably homogeneous real tank wastes slurry than the
slurries tested because there would be a longer baseline and a more uniform feed. Any change in the
Nametre value would provide information about changes in the slurry being transported.
3.14
10
0
0
10
20
40
30
50
60
70
Haake Viscosity Ektrapdated to 4084 reciprocal seconds
Figure 3.12. Nametre vs. Haake at 4084-1Seconds
Viscosities from Pressure Drop. Figure 3.13 shows the measured pressure drop as a function of the
Nametre viscosity. A positive correlation exists although there is some scatter in the data. Based on this
data, the possibility exists for using a pressure drop measurement as a replacement for the Nametre
viscometer. To further explore this possibility, the literature was searched for correlations that would
enable back-calculation of the slurry viscosity from pressure drop measurements. The prime driver for
this is the fact that since the viscosity limits of 10 CPat 6 ft/sec and 30 CPat 4.5 Wsec were derived from
the Cross-Site Transfer Line pressure drop calculations. Back-calculation of viscosity from line pressure
drop is the most direct method to ensure that slurry properties do not exceed the line design criteria.
For ali fluids, the pressure drop can be calculated from
AP =
2pv zfL
3.15
D
25
20
0
10
20
30
40
50
60
Nametre Viscosity, CP
Figure 3.13. Pressure Drop vs. Nametre Measured Viscosity
where p is the slurry density, L is the length of pipe over which d p is measured, D is the pipe diameter, v
is the fluid’s average velocity, andfis the Fanning friction factor. The friction factor, in turn, is a function of the Reynolds number (Re), slurry rheological properties (e.g., the consistency parameter, yield
stress, and the flow behavior index). The Colebrook formula is a popular correlation for the friction factor of a Newtonian (Denn 1980), turbulent fluid in a rough pipe (4.E3 < Re < 2.E7):
where k is the surface roughness parameter for pipe (taken as 0.05 mm for rough steel). The Reynolds
number, Re, is commonly defined as
3.16
where p is the slurry viscosity. Although the Colebrook formula is a p o p u r correlation, it is also a
transcendental equation that requires a root finding algorithm to obtain each new value of friction factor.
An explicit equation for the friction factor of a Newtonian fluid in a rough pipe that agrees well with
the Colebrook formula is given by Serghides (Serghides 1984):
k
120
k
2.51A
1
Re
A= -2.Olog,,( -+i
)
3.7 Re
B
=
-2.010g,,(-+3.7
4
c
=
-2.Olog,,(-
k
3.7
2.51B
Re
+ -)
Even though non-Newtonian models best describe the slurries that were run through the slurry pipe test
loop, using a non-Newtonian friction factor calculation is likely to prove unwieldy for field use because the
non-Newtonian rheological parameters may be difficult to obtain (these parameters are normally obtained
from laboratory- generated rheograms). However, a good non-Newtonian model for the friction factor
should yield similar results to a Newtonian model when Newtonian parameters are used with the nonNewtonian model.
The non-Newtonian friction factor correlation recommended in a previous PNNL paper"' (Erian 1994)
is the Torrance equation:
(a) Erian, F. F., R. L. McKay and E. A. Daymo. 1994. TWRS Retrieval Technology Project: NonNewtonian Hanford Waste Slurry Transpon Calculations Using Rheological Data of Tanks 101-SY,
101-AZ,and 102-AZ. Letter Report dated August 1994.
3.17
E
(-) = (A-l.SnI3)
+
B2.3031og(l -x)
+
B 2 . 3 0 3 1 0 g ( R e ( , , ~ ~+~ 0.347(5n-8)B
)
where n is the flow behavior index. A , B, and x are given by
where ,t is the wall shear stress and
zy is the slurry yield stress.
-
tw-
t, is defined as
fpv
2
The modified Reynolds number (RepLc)for the Torrance equation is given by
where K is the consistency factor (Le., Bingham viscosity).
Note that when the slurry is Newtonian, K = p, Re = Re,,,,
equation becomes (Denn 1980)
3.18
n = 1, zy = 0, and the Torrance
1
- = 4.53log(Re[Q-2.3
$f
This equation is similar in form to the von Urmdn-Nikuradse equation (KN equation), which is an
empirical correlation for Newtonian fluids over the entire turbulent region (Re > 4000) :
1 = 4.01og(Re[Q-0.4
Jf
To back-calculate the viscosity from the pressure drop, a golden section search root finding technique
was used to find the viscosity, p (or K),which predicts the measured pressure drop. It is assumed that the
temperature, pressure drop, fluid velocity, and density are known (constant) parameters: temperature
(Moore Industries thermocouple), pressure drop (DP cells), flow rate (Yokogawa), and density (Micromotion) were averaged over ten seconds to smooth out normal fluctuations in the instrument readings.
The roughness parameter for the Serghides friction factor equation was determined empirically from a
water-only test. Specifically, the pressure drop was measured during Run 1A when the temperature was
16.5"C. The viscosity of water at 16.5"C is approximately 1.09 CP(CRC Handbook of Physics and
Chemistry), so the empirical roughness factor can be estimated such that the back-calculated viscosity is
calibrated to a water standard at this temperature, flow rate, and pressure drop. The empirical roughness
factor was determined to be 0.036 mm (compared to a literature value of 0.05 mm for commercial steel).
The Torrance relationship does not have an empirical factor to "normalize" the equation to the IVF
system. Consequently, the length over which the pressure drop was measured was increased (from 3.03 m
to 3.55 m). After adjusting the roughness parameter for the Newtonian friction factor correlation, and the
effective pipe length for the non-Newtonian friction factor correlation, the viscosity of water in Run 1A
matched that of the CRC Handbook.
Figure 3.14 is a plot of viscosity back-calculated from the pressure drop using a Newtonian friction
factor correlation vs the viscosity calculated from an extrapolation of the Haake rheogram to lo00 sed'.
As discussed previously, loo0 sec'' is the assumed shear rate at the wall, and the principal viscous resistance that causes the observed pressure drop. Notice that there is little correlation between the two viscosity measurements. The lack of a correlation between the two viscosity measurements could either be
because the shear rate at the wall is different from loo0 sec'' and/ci because the Newtonian friction factor
relationship is insufficient to describe the pressure drop of a non-Newtonian slurry.
There is, however, a correlation between the consistency factor of a yield-pseudoplastic material and
the consistency factor back-calculated from the pressure drop calculation using the Torrance friction factor
relationship (Figure 3.15).
3.19
ALL RUNS: (Nemtonian f vs. Re relationship)
0
50
100
150
200
250
300
350
400
Viscosity from DP, cP
Figure 3.14. Haake Viscosity vs. Back Calculated Viscosity
The relationship is not perfect, though, for several reasons. First, the Torrance equation assumes that
the rheogram is of the form of a yield-pseudoplastic (which includes Bingham plastics, where n = 1, and
power law fluids, where z, = 0):
z = zo + K j "
However, not all of the slurries were described by this rheological equation. Graphite, gibbsite-graphite,
bentonite, and bentonite-mica slurries were better described by the Casson or logarithmic shear stress
(shear rate relationships). The Casson and logarithmic equations are of drastically different form than the
rheogram for a yield-pseudoplastic, so the fitted parameters from the Haake instrument for these rheological equations are not applicable for the Torrance equation:
Casson Equation:
3.20
Consistency Factor from NowNewtonianf-Re (Torrance) Comlation vs.
Laboratory Measured Corsistency Factor
4000
3500
3000
E
2500
e
2
2
3
2000
$
m
1500
500
0
0
500
1000
2000
1500
2500
3000
3500
LabParam
Figure 3.15. Back Calculated vs Laboratory Measured Consistency Parameters
Logarithmic Rheogram:
z = K + bl n(y)
As a result, rheological parameters (z,,,n) for the Torrance equation were obtained from the Haake fit
of either the Bingham, power law, or yield-pseudoplastic equations, even though the R2for these fitswas
often between 0.7 to 0.8.
Another reason why it was difficult to compare the back-calculated “viscosity”from the Torrance
equation to laboratory measurements is because the Torrance equation is highly sensitive to the value of the
exponent n. A 0.1 change in the (unitless) value of n could shift the back-calculated “viscosity”by as much
as 100%. For some runs, the Haake fitted value of n varied by as much as 0.1.
Considering the errors which were faced in determining the back-calculated parameters, Figure 3.15
shows a surprisingly good correlation between the back-calculated “viscosity”and the labmatory measured
parameter, K.
3.21
Even though there s e e m to be a good qualitative correlation between the pressure drop and the lab
tory measured viscosity, the usefulness of this correlation must be questioned. It is expected that the ta
waste slurries will exhibit non-Newtonian behavior, but the Newtonian friction factor correlation did nc
correlate well with the laboratory measured viscosity extrapolated to lo00 sed'. However, the nonNewtonian friction-factor Reynolds number correlation requires information about the yield stress and
rheological exponent, n. These two parameters may not be available since they can only be determine(
from laboratory measurements. In addition, the non-Newtonian back-calculation method obtains the vi
of the constant K in the yield-pseudoplastic rheological equation. Depending on the value of the expon(
n, K may not have units of viscosity (e.g., cP, Pa-sec). Since the back-calculated parameter will have
same units of viscosity only when the fluid is a Bingham plastic (n = l), it is difficult to compare the c
culated constant K to the 10 or 30 CPlimit of the Cross-Site Transfer System.
In summary, extracting useful quantitative information by the back-calculation methods presented 1
is unlikely to be of real use to Project W-211 operators. In fact, perhaps the entire approach of measu
the viscosity should be abandoned in favor of measuring the allowable AP per length of pipe. After all
W-058 viscosity specification is really to make sure that the AP per length of pipe is not exceeded. Th
viscosity is not particularly relevant provided that the AP per length is not exceeded.
3.3.5 Flow Rates
The volumetric flow rates of the Fischer & Porter and Yokagawa flow meters were compared to a
flow rate measured in a calibrated volumetric vessel by measuring how much volume was delivered in
specific period of time.
The flow calibration tank volume was established mathematically from the geometry of the vessel.
The column tank was constructed from a ten foot length of 16-inch schedule 30 pipe. This geometry 'R
chose to minimize air entrainment during the slurry discharge into the tank. The gas bubbles are confi
to a thin layer near the surface but since the area is small, the error in determining the volume is also
small.
The average wall thickness was 8.376 inch which corresponds well to the published wall thickness
16-inch schedule 30 pipe) of 0.375 inch. The cross sectional area of such a pipe, if it is a perfect cylin
is 182.65 in'. This cross sectional area represents the maximum area barring any severe stresses to thc
pipe which might cause a ballooning distortion. Because the wall thickness of the pipe matched the pul
lished values, the possibility of a ballooning stress was not considered further.
The inside diameter of the pipe was measured with micrometers. The smallest and largest diamek
measured were 14.950 inches and 15.406 inches, respectively. These two diameters were perpendicul
one another. Assuming the flow calibration tank to be an ellipse, the cross sectional area would be 18(
in2. The difference in cross sectional areas between the perfect cylinder (maximum cross sectional are
and the elliptical cylinder was assumed to be our volumetric uncertainty. Comparing the elliptical croe
sectional area with that of the perfect cylinder one finds a 0.97% difference. The error in reading
3.22
the height measurement is thought to be less than 1/4 inch or +/- 0.2% of full scale. The addition of these
two uncertainties (1.2 % of full scale) was assumed to be the error associated with the flow calibration
tank.
The column tank was used to calibrate the load cells. The column tank was filled with water to various heights and the calculated mass was compared to the load cell displayed mass. All displayed masses
from the load cells matched the calculated mass within 0.8 lbs between 0 and 757 lbs.
During the experiment, a three-way valve was thrown to divert the flow of slurry from returning to the
feed tank to the calibrated vessel. The three-way valve was diverted for an operator selected period of
time to yield a reasonable height of slurry in the vessel. The height of this interface was recorded. Simultaneously, the original weight of the vessel before diversion and after diversion was recorded. Based on
the volume of material added to the vessel and the time of diversion, the volumetric flow rate could be
determined. The volumes from the Yokagawa and Micromotion flow meters were recorded simultaneously. The percentage error from the Yokagawa or Micromotion was calculated and used as the primary
data for flow rate evaluation.
These experimental flow rate error determinations are subject to random experimental error. Specifically, for some slurries it was very difficult to measure the interface height in the sight glass because solids
collected on the walls and obscured the interface. In this case, the height of the interface was measured
from the inside top of the calibrated vessel. Also, as the higher solids loading slurries were very viscous,
there was a significant time lag between the end of the diversion time and when the height of the interface
in the sight glass finally equilibrated.
3.3.5.1 Yokagawa How Meter
Figure 3.16 shows a run chart of the Yokagawa flow rate over a 30-minute time period for Run 4Alb
(25 vol% gibbsite). The steady downward drift is from the throttle valve shifting slightly. The data was fit
and the residuals were examined to determine the noise inherent in the Yokagawa flow rate measurement.
The average residual had a value of 0.0 and a normal distribution indicating a very good fit. The standard
deviation of the residual was 0.0155 (literkec). Since events more that three standard deviation away from
the mean occur only 0.03% of the time, any change from an established Yokagawa flow rate that exceeds
0.0465 (liter/sec) is a good indicator that a change has occurred in the process and should be watched very
closely.
Figure 3.17 plots the average error in the Micromotion volumetric flow rate error against the average
Yokagawa volumetric flow rate error. Two individual data points that were rerun were excluded because
they were not consistent with the three other observations. Data from Run 5B1 (gibbsite/graphite) in the
Project W-211 loop test and “ideal”loop test were excluded because the observations were extreme outliers
with values of -22.9%, -30.7%, -36.4% for the Project W-211 loop test and -36.4%, -45.3%, -51.1%,
-36.5% for the “ideal”loop test. Both of these runs were graphite runs.
3.23
6.4
6.35
6.3
6.15
6.1
Figure 3.16. Yokagawa Flow Rate vs. Time
Of all the solids run in the IW, only graphite caused operational problems. Graphite was observed to
form a slimy coating on the inside of the slurry pipe, interfering with the electrodes and producing artificially low readings. Twice after graphite runs, the Yokagawa was removed from the pipe loop and
cleaned even though the performance was recovered by back-flushing the instrument in the loop. After
cleaning, the observed flow rate returned to normal. Any material that could coat the electrodes would
interfere with the performance of the flow meters. Since it is not possible to know whether Hanford Site
tank wastes would form these slimy coatings, it is recommended that any exposed electrodes be coated to
prevent erroneous operation.
If the on-line flow meters report the same flow rates as was measured in the flow calibration tank, then
the flow rate error for the flow meters should be equal to 0. This was not observed with either flow
meter. Specifically, the Yokagawa was observed to be on average 2.1% low. Since the error in the flow
calibration tank was determined to be about f 1.2%,the Yokagawa error could be as low as 0.9%low or
as high as 3.3% low compared to the reference. This range overlaps the acceptability limit of &2%(see
Appendix D-8 for the distributional statistics). The above analysis includes pure water runs.
3.24
'6
4
2
0
si
E
&
I
.c
-2
f
-4
2
-6
g
.-
!
': -8
-10
-12
-14
-10
-a
-6
-4
-2
0
2
4
6
Average Yokagawa Ehor,
Figure 3.17. Yokagawa Instrument Volumetric Flow Rate Errors
For the water runs, the Yokagawa flow error ranged from -0.43 to -1.47%low and averaged -0.99%
low. Therefore, the performance with slurries is about 1.1% low compared to testing with water. This is
within the desired accuracy limit of +2%.
The Micromotion was observed to be on average 8.5 % low. Since the error in the flow calibration
tank was determined to be about f 1.2%,the Micromotion error could be as low as 7.3%low or as high
as 9.7% low compared to the reference. This is not within the 5% acceptability limit and it cannot be recommended as is as a suitable instrument.
For the water runs, the Micromotion flow error ranged from -4.88 to -8.56% low and'averaged -6.2%
low. Therefore, the performance with slurries is about 2.3%low compared to testing with water. The
implication is that if the flow meters are calibrated to pure water, rather than relying on the factory calibrations, the performance with slurries should be within 5 % acceptability limit. The manufacturer states
that the unit can be user-calibrated and therefore, it should fit the desired accuracy criteria.
If the discrepancy between the measured flow rate errors for either instrument is caused by the method
for determining the flow rate in the volumetric tank, then the two instruments should give the same error
3.25
and fall along the “Y = X” line in Figure 3.17. (This figure was prepared from the average value
observed in a run and not the individuals.) However, the Micromotion tends to have a higher error fairly
consistently across all of the ru~lsbut with some scatter.
3.3.5.2 Fischer and Porter Flow Meter
Figure 3.18 plots the average error in the Fischer & Porter volumetric flow rate against the average
Yokagawa volumetric flow rate error Yokagawa. One data point in the Run 5a “ideal” loop that was rerun
was excluded because it was not consistent with the three other observations. Another data point was
excluded because it was not consistent with the other two data points and was also exactly equal to four
significant digits to the error from the Micromotion flow meter.
Data from Run 5B1 was excluded because the three you list four values observations were extreme
outliers with values of -36.4%, -45.0%,-49.3%, and -52.8 %. Data from Run 5B0 was also excluded
(values of -26.9%, -27.0%, - 28.0%,and -28.6%)because it was run several hours after Run 5B1. Both
of these runs were graphite runs.
-10
-8
-6
-4
-2
0
2
4
Anrage Yokagawa Dror, %
Figure 3.18. Fischer and Porter Volumetric Flow Rate Errors
3.26
6
0
After the normal back-flushing, the observed flow rate did not return to normal and the flow meter had
to be removed to clean it for further use. It is recommended that if either magnetic flowmeters are
selected for use in the cross-site transfer system, that unit should have electrodes that are covered. The
manufacturers state that such a selection should prevent interference with the measurements if a coating
should form.
If the online flow meters report the same flow rates as was measured in the flow calibration tank, then
the flow rate error for the flow meters should be equal to 0. This was not observed with either flow
meter. The Fischer & Porter was observed to be on average 2.0 vol% low. Since the error in the flow
calibration tank was determined to be about & 1.2%,the Yokagawa error could be as low as 0.8%low or
as high as 3.2% low compared to the reference. This range overlaps the acceptability limit of f I % , but
just barely.
The flow errors for pure water ranged from -1.38% to -2.01% with an average of -1.6% low.
Therefore, the performance with slurries is about 0.4% low compared to testing with water. This is within
the desired accuracy limit of 1% .
If the discrepancy between the measured flow rate errors for either instrument is caused by the method
€or determining the flow rate in the volumetric tank, then the two instruments should give the same error
and fall along a “Y = X line in Figure 3.18 (Note that Figure 3.18 was prepared from the average value
observed in a run and not all of the individuals.) The agreement with the “Y = X line is reasonable.
The paired t-Test revealed that the Fischer & Porter and Yokagawa flow meters gave the same reading at
the 95% confidence level (see Appendix D-6). This indicates that the Yokagawa performs as well as the
Fischer & Porter and would be a suitable replacement flow meter. It is not clear why the acceptability
limit for the Fischer & Porter was set at 1% when the Yokagawa was set at 2%.
A principal component analysis for the Fischer & Porter flow rate errors was performed and reported
in Section 3.5.4. This analysis indicates that the errors observed in the flow calibration tests are random
and do not depend on the simulant material or properties.
3.3.6 Particle Size Distribution Measurement
This section will discusses the PSD test performed with the Lasentec and Horiba particle size analyzers. As has been previously discussed, the two techniques employ different methods to measure solids
concentrations. The Horiba essentially measures equivalent spherical diameters based on sedimentation in
very dilute slurries and the Lasentec measures chord lengths by laser scanning in highly concentrated slurries. Since the two techniques operate in extremely different solids concentration regimes, it is conceivable that both results are correct and are accurately measuring the PSD that they see.
3.27
3.3.6.1 Bi-Modal Particle Size Distributions
During Runs 5C and 6C1,two or more solids were mixed in an attempt to create a bimodal PSD. By
varying the relative proportions of the solids, the shape of the distribution could be made to change. The
PSDs from Run 5C containing graphite and Gibbsite from the Horiba and Lasentec did not clearly show
that a bi-modal PSD was created or that the mean particle size changes with the varying proportions of
solids. The probable cause is that the mean particle size of the graphite (2.9 microns) and gibbsite
(4 microns) were too close and with too broad of a PSD to show up as separated peaks in the PSD.
However, the Lasentec PSDs from Run 6C1 containing bentonite (mean particle size of 0.8 microns)
and Mica (mean particle size of 6.3 microns) shows a substantial shift in the mean particle size. It is not
clear that the resulting distribution is bimodal (see Figure 3.19).
I*'"
I
Figure 3.19. Particle Size Distribution Shift in Run 6C1
3.3.6.2 Color Changes
During Run 7C1 (4vol% bentonite), a total of 25 red and orange food coloring bottles (0.2gal) were
added to the feed tank (containing 210 gallons of slurry) to change the color of the carrier liquid from clear
to orangish. The simulant changed color from olive-green to peach. This test was to see if the change in
color contrasts resulted in changes in the measured PSDs. The Lasentec showed no change in the average
size distribution compared to the data scatter as a result of this color change, as indicated in Figure 3.20.
3.28
17.5
17
e
e
'I
v
16.5
15
14.5
Figure 3.20. Lasentec Color Effects
3.3.6.3 Bubbles
The presence of bubbles can potentially affect the PSD measurement (since air bubbles also reflect
light). In Figure 3.21, the average particle size of a 2 vol% graphite slurry is compared when the bubbler
is both off and on. The measurement of average particle size did not change.
A partial least square (multivariate statistical) analysis for the mean particle size was performed and
reported in Section 3.5.3. This analysis shows that none of the instruments tested in the IVF measure
properties which adequately correlate to the laboratory measured mean particle size. Interestingly, the
multivariate analysis alsow showed that the Lasentec and Horiba lab instruments did not correlate well with
each other either, probably because different particle sizing techniques are used by each manufacturer.
3.3.6.4 Comparison with Laboratory Data
Comparison of the median particle size between the Lasentec and Horiba analyzers must be tempered
with a knowledge of the measurement techniques. The Horiba analyzer assumes a constant density for all
solids, hence it can confuse a smaller but denser particle with a larger but less dense solid. The Horiba
also measures an equivalent spherical diameter based on the principle of sedimentation rates. Tests were
run on a selected slurry simulant to determine the repeatability of the testing (see Table 3.1).
3.29
Figure 3.21. Impact of Bubbles on Measured Particle Sizes
Table 3.1. Horiba PSD Instrument Repeatability with Run 4A “Ideal”Loop Sample
‘;“
Mean (Microns)
1
4
k
Standard Deviation
(Microns)
7.43
4.86
5.58
4.2
6.5
4.41
8.16
4.33
8.34
4.95
7.73
4.48
The mean particle size is about 7.29microns with a standard deviation of the means of 1.06 microns.
Therefore, the true mean is 7.29 +/- 1.11 microns at the 95% confidence level (see Appendix D-8). This
variability must be considered in the interpretation of any results based on the laboratory PSDs. A comparable analysis should have been performed on the Lasentec but this was not possible to perform in the
field.
3.30
If settling is occurring at the lower flow rates, both the Horiba and the Lasentec should be able to
detect these changes. Only for Gibbsite did the Horiba consistently report a lower average particle size
1 micron) at the lower flow rate (6.25 litershec). The Lasentec, however, did not report a lower
average particle size at the lower flow rates. Since the mean particle size change of 1 micron is within the
95 % confidence limits of 1.11 microns, it is very likely that the shift is an artifact of the laboratory measurement technique and does not reflect settling in the pipe loop.
(I
A comparison between the median particle sizes measured by the Lasentec and Horiba is presented in
Figure 3.22. The Lasentec did not measure the expected mean particle size as the reference Horiba particle size analyzer. Particle size measurement is a difficult analytical measurement to perform and even
more difficult to perform on-line. It is tempting to say that the Lasentec does not measure the correct
mean particle size but that cannot really be stated with high confidence because of the difference in analytical techniques and slurry environments.
10
5
4
0
2
4
6
8
10
12
14
16
IlarihWanPartide Size, dcrm
Figure 3.22. Lasentec vs. Horiba Particle Size Distributions
3.31
18
20
Both the laboratory and Lasentec measurements showed that the average particle size was fairly
consistent for each material, regardless of flow rate, solids concentration, and the presence of bubbles.
Since the laboratory and Lasentec data showed no evidence of settling in-line, the Lasentec may not have
been tested to its fullest capacity. That is, if the Lasentec is to be further evaluated in the future, much
lower flow rate tests are recommended so that solid particles are purposely settled out when the flow rate
is below that of the critical velocity for solids suspension.
In all of the PSDs, both analytical techniques detected small particles less than 1 micron in diameter.
However, there was no indication that either particle size analyzer was able to distinguish the low concentration of 500-micron particles specifically added in run 5C 1. No change in the flow behavior was
observed with the inclusion of this low concentration of very large solids.
3.3.6.5 Recommendation
The mean measured particle size from the Lasentec did not correspond to the mean measured particle
size from the Horiba particle size analyzer. The reasons for this have been previously discussed.
Bubbles barely influenced the Lasentec, so their presence did not make the instrument unreliable.
Alternatively, the instrument was not effective in determining if bubbles were present in the system. The
color of the carrier fluid in the simulant also had little bearing on the ability of the instrument to consistently measure the same average particle size.
The Lasentec particle size measurement instrument was useful in determining when steady state was
reached. Slight shifts in the PSD were noticed as solids were added to the system and as the flow rate
increased or decreased drastically (e.g., 6.25 litedsec to 9 litershec). Typically, the Lasentec reported
steady-state values within minutes of system startup throughout the entire operation, across all solids.
However, this finding is of relatively little value to Project W-211.
What would be of high value to Project W-211 is the ability to detect crystallization or precipitation of
solids. Since the simulants tested did not have soluble salts that could precipitate or form gels in the test
loop, the ability of the instrument to detect crystallization or precipitation of solids in solution was not
tested. (One of Lasentec’s primary markets is the measurement of PSD in-line systems that precipitate.)
The precipitation of material would both increase the particle count and shift the PSD. These deviations
from steady state simultaneously with an increase in viscosity and line pressure could indicate the onset of
adverse precipitation. The instrument is well suited for remote installation in the valve pit if properly
engineered .
3.4 Instrument Elimination
Since the Micromotion thermocouple corresponded well with the Moore Industries thermocouple and
since the Micromotion also provides good measurements of the density as well as potentially adequate
measures of the mass flow rate, the Moore Industries thermocouple can be eliminated from the actual
3.32
instrument tree. Unfortunately, Micromotion does not produce a transmitter capable of sending three
4-20 mA signals to the PLC. On the IVF system, density and flow rate were sent over the only two
4-20 mA outputs.
3.4.1 Temperature
There are three methods to send temperature, density, and flow rate signals to the PLC:
1. Move the flow rate output to the pulse output on the transmitter. The PLC will require a high speed
counter card to convert the signal to the appropriate flow rate. For example, a 8000-Hz signal from
the pulse output on the Micromotion transmitter can correspond to a 14-literlsecond flow rate.
2. Move the flow rate output to the pulse output on the transmitter and then use an external (third party)
signal converter to change the frequency output to a 4-20 mA signal. The converted flow rate signal,
along with the normal 4-20 mA signals for temperature and density from the Micromotion transmitter,
can all be interfaced with a PLC using a normal 4-20 mA input board.
3. Purchase a MODBUS board (or some board which uses the HART protocol) to allow a full digital
connection between the transmitter and the PLC. A digital connection will allow all information
shown on the rack mounted display (mass flow rate, volumetric flow rate, density, temperature, etc) to
be processed by the PLC. The 505-5184 MODBUS card for the Siemens SimaticTI-545 PLC has a
list price of $1825 (as of April 1996).
3.5 Multivariate Statistical Analysis Results
The following sections presents the results of the multivariate statistical analyses for viscosity, solids
concentration, PSDs, and flow rate errors. Good models were prepared for the viscosity and solids concentration. Very poor models were prepared for PSDs and flow rate errors.
3.5.1 Viscosity Multivariate Analysis
Figure 3.23 is a model overview showing the percent of the Haake viscosity data at 300 sed' which is
explained by the cumulative sum of the principle components. The first component alone explains 80% of
the variation in the data, whereas all three components explain nearly 90% of the variation. A general rule
of thumb is that a good model can explain at least 90% of the variation in the data. A poor model, on the
other hand, would explain less than 60-70% of the variation of the data.
Figure 3.24 is a VIP (Variable Importance in the Projection) plot, showing the relative importance of
the measured parameters to the prediction of the Haake (laboratory measured) viscosity at 300 sec-'. As
expected, the direct tap DP cells and the Nametre in-line viscometer correlate best to the measured
viscosity at 300 sed'.
3.33
Viscosity Multivariate Model Cumulative Overview
Component 1
Component 2
Component 3
Figure 3.23. Model Overview
Viscosity Multivariate Model: VIP Plot
2
1.8
1.6
1.4
1.2
1
0.8
0.6
0.4
0.2
0
Figure 3.24. VIP Plot of Haake to Measured Parameters
3.34
SIMDAT2.M4 (PLS),UnTitled, Workset
_ _ - 300s-1. Comp 3(Cum)
200
, ,
3
6
160
5b
160
26c
Predicted
Figure 3.25. Predicted vs. Observed Viscosity
Figure 3.25 shows that the observed viscosity matches well with the predicted viscosity. Here the
predicted viscosity is a linear combination of process variables:
pHaake,300sec. =0.85602 +O.
57276AP +O. 41 032pNmetre+f (other variables)
Note that this function should not necessarily be used to predict the laboratory measured viscosity of
actual tank waste: this is a correlation that explains the laboratory data only from the IVF experiments.
Another model was also created without the Nametre viscosity as an input parameter; it too showed that
the Haake viscosity at 300 sed' correlated extremely well with the pressure drop. Thus, even if a direct
quantitative relationship is not well established between the pressure drop and the slurry viscosity (i.e.,
through a friction factor/Reynolds number correlation), a multivariate statistical package (such as Simca-P)
could develop an empirical correlation between pressure drop and lab measured viscosity if enough data is
available.
I
3.5.2 Solids Concentration Multivariate Analysis
The overall fit of another model to determine the solids concentration is shown in Figure 3.26. Note
that this model is not as good as the previous one: only around 75% of the variation in the data is
explained by all four components (reduced variables).
From the VIP plot (Figure 3.27), the Micromotion density is the biggest contributor to the correlation
between solids concentration and the measured process variables. The laboratory measured mean particle
i
3.35
Solids Concentration Multivariate Model Cumulative Overview
1
0.9
0.8
0.7
0.6
rd 0.5
0.4
0.3
h
0.2
0.1
0
Component 1
Component 2
Component 3
Figure 3.26. Solids Concentration Model
Solids Concentration Multivariate Model: VIP Plot
1.4
12
1
’
n 08
06
04
02
0
E
0
‘5
Figure 3.27. VIP Plot for Solids Concentration
3.36
size, the Red Valve line pressure, and the Lasentec small particle count also are shown to have a significant correlation (VIP > 1.0) to the solids concentration. While it makes sense that the number of small
particles counted by the Lasentec would increase with solids loading, the laboratory mean particle size and
the line pressure are practically poor indicators of solids concentration. The positive correlation with laboratory mean PSD may be an artifact of the measurement technique: the Horiba lab instrument determines
the number of particles in each size classification through an optical absorbance technique. As the solids
concentration increases, the absorbance of the solution changes. Since the absorbance changes, the
reported number of particles in each size channel is also different, as is then the PSD. As the distribution
changes, the mean also shifts. The Red Valve line pressure, on the other hand, should increase as the
solids concentration increases.
However, it is doubtful whether a quantitative correlation between solids loading and density would be
successful, unless the feed is fairly uniform with respect to solids density. A combination of the slurry
density, line pressure, and the small particle size count may do a good enough job to allow qualitative
trends to be recognized. Figure 3.28 is a plot of the predicted solids loading vs. observed solids loading.
The correlation is not “ideal”,but at least there is hope that general solids concentration trends could be
seen with these measured process variables.
SIMDAT2.M3 (PLS), UnTitled, Workset
3
urn
0
40
30
2
0
20
%
c1
’
10
587
0
40
2Io
$0
Predicted
Figure 3.28. Predicted Solids Loading vs. Observed Solids Loading
3.37
4
k
3.5.3 Particle Size Multivariate Analysis
Figure 3.29 is the model overview for the prediction of the laboratory measured mean particle size
from the measured parameters, including the Lasentec measured mean particle size. As can be readily
seen from the bar graph, the model is very poor. This suggests that there is no correlation between the
laboratory measured mean particle size and other process variables. Interestingly, the Lasentec measured
mean particle size does not necessarily compare well with the laboratory (Horiba) measured mean particle
size. This discrepancy may be due to the fact that the lab data has a 95% confidence error of
1.11 microns when the same sample is repeatedly analyzed. The Lasentec mean particle size is more
constant over the duration of the run.
Particle Size Multivariate Model Overview
1
0.9
E
5
U
a
5
n
0.8
0.7
0.6
0.5
0.4
2 0.3
>
2
0.2
0.1
0
Corrponent 1
Component 2
Component 3
~
~~
Figure 3.29. Particle Size Model
3.5.4 Flow Rate Error Multivaraite Analysis
Partial least squares analyses of the Yokagawa, Micromotion, and Fischer & Porter flow rate errors
were conducted to determine if the error was associated with any of the measured slurry or processing
parameters. The analysis showed that the error in flow rate could not be modeled well by the other measured parameters. Specifically, the Yokagawa flow rate error model explained 20% of the variation in
flowrate error (relative to the calibrated standard), the Micromotion flow rate error model explained 35%
of the variation in the flowrate error (relative to the calibrated standard), and the Fischer & Porter flow
rate error model explained 40% of the variation in the flow rate error (relative to the calibrated standard).
The presence of bubbles and viscosity were the largest contributors to the Yokagawa and Micromotion
models. Temperature and viscosity were the largest contributors to the Fischer & Porter model. In all
three cases, the model was so poor that no conclusions will be drawn from the analysis.
3.38
4.0 Conclusions and Recommendations
Four test objectives were outlined in Section 1.2. The conclusions from the testing of each objective
will be discussed in the following four sub-sections.
4.1 Instrument Validation Conclusions
1. The Micromotion Density analyzer is recommended as meeting the desired accuracy for density
measurements.
2. The baseline Fischer & Porter magnetic flow meter provides good measurements of the flow rate.
The Yokagawa flow meter is as accurate as the baseline Fischer & Porter. Both instruments were
observed to form a slimy solids deposit with one solids slurry that catastrophically interfered with the
proper measurement of flow rates. While this may be unique to graphite, the instrument to be
installed in the Project W-211 test loop should have the electrodes coated to prevent this from
happening with the solids in tank wastes.
3. The Micromotion mass flow meter does not meet the desired accuracy limit; however, had it been user
calibrated for any of the slurries or water (rather than simply relying on the factory calibration), it
probably would have met the desired accuracies.
4. The Moore Industries Thermocouple and Micromotion densimeter thermocouple give equivalent and
adequate temperature results.
5. The Red Valve pressure analyzer is recommended for use in the Project W-211 test loop.
6. The Lasentec particle size analyzer is not recommended to measure the absolute particle size of a
given slurry or the solids concentration. It can be used qualitatively to monitor gross process changes
in the slurry being injected into or transported through the cross-site transfer line.
7. The Nametre on-line viscometer, while providing a qualitative indication of the viscosity of a slurry,
does not meet the desired accuracy limits. It is very sensitive to vibration and easily damaged. The
Nametre is not recommended for use in the Project W-211 loop unless there is a critical need to have
an on-line viscosity measurement.
8. An alternative method of indirectly monitoring the viscosity is to measure the pressure drop across any
straight run of transport loop or, although not tested, across the instrument loop or the cross-site
transfer line itself. The correlation with pressure drop is nearly as good, the instrumentation is far
simpler than the Nametre, easier to maintain, and is a direct measurement of the key process parameter of interest (Le., the allowed pressure difference across the cross-site transfer line).
4.1
4.2 Instrument Elimination Conclusions
Project W-211 desires to reduce the number of instruments to the bare minimum while ensuring that
the cross-site transfer line does not plug. The original intent was to measure or infer the value of the three
primary process variables (Le,, density, viscosity, and solids concentration) specified in the FDC. This
work suggests that the instrumentation to measure two of these primary process variables (i.e., solids
concentration and viscosity) does not meet the desired accuracy limits. However, the original intent may
not be the only control method to use to assure that the cross-site transfer line does not plug.
For example, an alternate scenario to radically reduce the required instrumentation is to simply measure the flow rate and monitor the pressure drop across the instrument loop and at various points in the
cross-site transfer line. If the flow rate is measured in the test loop and is maintained constant and the
pressure drop across each section of pipe remains constant, no plugging occurs. This scenario is more
credible with multiple pressure drop measurements installed at various points in the cross-site transfer line
or as the length of pipe over which the pressure drop is measured is increased. This scenario also provides
information as to where the plug is developing.
Another alternate scenario is to measure the flow rate and pressure entering the cross-site transfer line
and measure them again in an identical test loop at the exit of the cross-site transfer line. Any increase in
the pressure drop and a simultaneous decrease in flow rate exiting the test loop are indicative of plugging
occurring.
Both of these alternate scenarios can be used to monitor the development of plugs and with a minimum
of required instrumentation. Both, however, require a rethinking of the control scheme and cross-site line
design.
If the original objective of measuring all slurry properties at a single point for ensuring that a plug does
not form is maintained, it is still possible to somewhat simplify the instrument loop design and eliminate
some of the original instrumentation. The Micromotion mass flow meter can be re-calibrated to correct
the offset observed in this testing, therefore this single instrument can provide density, flow rate, and
temperature information. (This could eliminate the need for the Yokagawa flow meter in the pump pit.)
When the Micromotion results are combined with a pressure drop measurement (which was shown to be
strong function of the viscosity and solids loading in the multivariate statistical analyses) across the test
loop and with the process model, it may be possible to detect pluggage of the cross-site transfer line with
only these two instruments. The ability of the model to explain the variation improves as the ability to
independently measure the viscosity or solids loading is improved.
4.3 Test Loop Design Conclusions
The design of the current instrument test loop (Le., the Project W-211 loop) is adequate for measuring
the specified slurry properties.
4.2
4.4 Evaluation of Alternate Instrumentation
The Yokagawa is recommended as an alternate flow meter and performs as well as the Fischer &
Porter.
The Moore Industries thermocouple is not recommended as the Micromotion thermocouple performs
equally well and is part of the density measurement.
4.5 Other Conclusions and Recommendation
If further testing is to be performed on the test loop, a temperature control system should be installed
to prevent un-intentional temperature changes and the resulting complex data analysis.
4.3
5.0 References
Bird, B., W. Stewart, and E. Lightfoot. 1960. Transport Phenomena. John Wiley and Sons, New York.
Denn, M. 1980. Process Fluid Mechanics. Prentice-Hall, Inc., Englewood Cliffs, New Jersey.
Liu, K.,U. Choi, and K. Kasza. 1988. Measurements of Pressure Droppnd Heat Transfer in Turbulent
Pipe FZows of Particulate Slurries. ANL-88- 15, Argonne National Laboratory, Argonne, Illinois.
Serghides, T. 1984. “Estimate friction factor accurately.” Chemical Engineering March 5 , 1984.
Thomas, D.G. 1965. “Transport Characteristics of Suspension: VIII. A Note on the Viscosity of
Newtonian Suspensions of Uniform Spherical Particles. J. Colloid Sci. 20:267-277.
‘I
5.1
I
:
c
Appendix A
Detailed Loop Designs
FROU VALVE TO V N M OF
LOOP SHAU BE W
L
m
4 ''
- --
7
1'
6'
Ii1
\
.L
I I
u
i-
PLAN
VIEW
(c
Figure A.1. Ideal Lc
As NEEDCD
RED
VALVE PRESSURE
ROTATED 9G' FOR C m l T Y
1
LA.jER/DIOGE:
SOU05
CONCENTFATION C R * D I € M
OEECnON. INSTRUUEHT
NOT SHOWN
rusS R O W UEiER CORIOUS S
BE ROTATED AT 1G' *BoM
HORIZONWL W E
i
S-1
W
r--
.-.-I
J.6.33t.06
1/27
-
I ENGINEERWG
WORKSHEET
I
A. 1
Design
-4
1-
5'4
P
R
2'-11
1/22.4-
PLAN
VIEW (n
Figure A.2. Project W2-11 I
+
B'
p--1,-11-
/-
*s
I
5'-3
I
op Design
A. 3
1/1r
Appendix B
Simulant Development
Figure B.l. Instrument Validation Si
Test 'Sub 'Test
;Test 'Loop Liquid
,
!Composition 'Color
'Solid (1)
IDenstty: Solids
:Measured Loading
jviscosity Solid ( I )
-
___
Liquid Characteristics
iParticie Size
iDis1.lMean:
IMeasured
Solid Characteristics
iSolid(2) i
/Density: !Solids
'Solid (2) :Measured ;Loading
!
!
:Shape
/Particle Size j
Dtsl./Mean:
IMeasured 'Shape Solid (3) :Del
!
i
I
i
ia
jW
;a1
!W -Water
a2
:W
y,
Water
I
-~
iOO#.Ciear
, I cps
1WX'Clear
, I cps
100% Clear
:1 cps
Silica
- 100"h:Clear
'Icpe
. Silica
.
!
I
I
-
I
Water
i
2.88 5 ~01%
130 microns
2.885voiSb
'30microns
,Spherical :
I
I
a3
:"
Water
I
.
,a4
Water
.W
lW%:Ctear
~
1 cps
2.8820 vol%
Silica
I
l W % Clear
:I
cps
Silica
Waler
I
100XlCkar
'1 cps
Silica
:
2.8835~01% i3Omicrons
Silica
.
2.8835volX
!W
,
,a6
IW
i
i
ia7
: W_ Water
.
8
;
j
jw
-
Water
IW
Water
IW
I,
IW
Water
i
la3
3'
1
1
i
i
I
ia2
,
I
I
I
lOO%.Clear
[I
cps
IOOXlClear
[Icps
!
t i cps
10Oo,GClear
Water
-
!
I
,IW
:
I
ja5
.
,
4,
I1 cps
Graphite.
/lBmlcrons
:TED
l18rnicrons
/TED
!
I
I
0.99 35 volX
118 microns
I
0.9935~01% il8rnicrons
i
Water
I
100X;Clear
!I
cps
Graphile
I
I
lW%lClear
I
I
i
2.2:2 vol%
i
2.2'2~01%
100X'Clear
.ala _
.E
~
IDO%!Clear
WO?h!Ckar
I
__
a2
.a3
12.9 miuons
i l cps
Graphite
'
2.2 7.5~01% i2.9microns-~lnegular
!I
cps-
Graphite
!
2 2 7 . 5 v o l % j2.9microns
I
jlrregular
Graphite
i
2.2 7.5~01% i2.9microns
llrregular
[I
cps
i
i l cps
I1 cps
10056,Clear
1 cps
a0
.a
Waler
_ _ _ _'
Water,-___
;E.--
~
100% Clear
Water
Water
'E
~
bl
co
B
100% Ckar
__
100% Clear
--.__I_
.C1
~
~..Water
_______--
'
' 1 cps
__
I~%Clear
100%
Clear
__
__
~
~
-_
-I
_
._
.-
Graphite
~~
4 7 VOlX
2.2 6 VOl%
~--__
1cps - _____
3 7 VOl%
1 cps
100% Ckar
' 1 cps
Graphite
2.2 5 vol%
IGibbsite
1
I
!
I
i
I
I
I
___
~
___
--
._.-..
__
..-.-
IGibbsile
__
1
!
I
!
____
I
:
I
I
{
I
1
A
L
I
i
!
!
I
i
I
l
,
1
I
,
i
1
_
I
_
' -
_____
r
---__
._
I
j
i
I
I
,
1
i
!
8-
~
Gibbsite !
__
2.5 5 VOW
i
2 . 5 5 vol%
j
2.5125vol%
I
I
I4 microns
I
14 microns
'
IlRR
I
I
/ 4 miuons
2.5 25 volX
I
1
I
1
R
R
Y
-__
IRR
,
,
14 microns
IIRR
:4 microns
ilRR
I
2.532 volX
?, ---
2.5 32~01% ;4 microns
~_
__
:IRR
,
I
-.L
L
-
1
6.3 vol%
-______
2.5 8 vol%
14 microns IRR
~
___
:133vOl%:
Gibbsite
Irregular
Glbbsite i
:4 microns
_2.5 18_volX - 4 microns
- ~
IRR
~
IRR -..
-.
-.
1
100% Clear
100% Clear
.'cps1 cps
Graphite
Waler
:
100% Clear
' 1 cps
- Graphae
067 vot%
Note 1) All QuanldledValues May change based on Slmutanl Development Tesling
!
I
2.2 0.75 volX 2.9 Microns
3' VOi%
~
'
irregular
'Gibbsite
2.5 26 vol%
i4 microns
-IRR
Irregular
Gibbslle :
2 5 26 vOl%
4 microns
~
2.2 0 75 volX 2 9 Miuons
T
d
-
---
j
2.5 18 vol%
'Irregular
2.9 microns
I
I
I
~
'I
i
I
i
'Irregular
i
!
I
__
2.9 Microns
,
I
I
I
i
'Gibbsite
:
1
i
!
IGibbsile
t
i
,
I
,
!
:
i
i
i
I
I
!Gibbsile
i_.-_
Graphite
.
.
.
.
.
.
5_.-v0lX
.
_
_
.
_
_
_
_
I
_
~ 2.9 miuons
___ -.---..2.2
--.-_c-_--.__
I
I
!
1
I
I
I
I
IGibbsile !
__
--__-__
--_..
I
!
~..
Water
e
6
~
-__
'1 cps
__
-
_-__--~
__
l
I
!
j
I
-~
_
i
1-cps
I
I
~
I
I
-
-_
____
~
bO
._I_-___
I
I
_ _ _ _ _ _ ~
1 cps
___
100% Clear
._____
b
-
__
_ _ ^ _
__I____
!
I
,
I
!
I
I
I
I
I
~
!
i
I
i
I
I
/Irregular
i2.9microns
I
-
,
i
1
2.25volSb
,
v
I
I
j
I
I
I
,
1
I
Graphite
I
i
,
!
ilrregular
I
!
I
!
1
I
I
/Z.Qmicrons
i
-
1
[Irregular
:I
cps
100%
Clear
' 1 cps
________
:E
Water
,
.
___
___._
'B
Water
,
,
I
~
alb
-.
~ _
_ E
_ . Waler
,
!
I
jlrregular
1005b'Ckar
~~
~
__
/Irregular
2 . 2 5 ~ 0 1 % /2.9microns
I
i
_ Water
1
I
i
100SbiCtear
-Water
__
~
jTBD
Graphile
11 cps
I
1.
Water
/
!
i
!
IOOXiClear
I
I
12.9 Microns
I
I
1
IB
1
lTBD
I
. .
!*
I
I
i
. ,
I r
I
I
i
i
0.995volX
2.2 2 vol%
I
___L_
-
~
Graphite
Water
______
ial
iB
5
I
!1cps
_
I
_
_
_
I
i
i
I
0.9920vol%
_
i
~
i
~
#a
-
Plaslic
~
I
1[M%lClear
i
i
Plastic
i
I
Water
!
I
!
_ 'a6_ IW_ Water
- 1
~
iW
_i
1
Water
:W
'
Plastic
Water
Water
I
!a7
_
I
IW
.ja4
Plastic
~
100X:Orange llcps
I
la3
I
I
I
IW
i
i
iTBD
I
i
i
!Spherical
I
I
I
I
100Xklear
;W
I
-I
I
i
-
/TED
j
I
TBD/Clear(?) i30cps
~
I
la1
\TED
i
I
I
;a2
'TED
TBD~Clear(7)!15cps TBD
i
i
-
i
ISphericaI !
i30miaons
I
G
Water
jal
v
-
_
/
I
!
iW
i
i
I
I
Icps
IWsb'Clear
Glyceroll,
Water
ibl
i
2:
ia
.
~
:Spherical
i
I
IW
I
!
2.88 20 vol%
!Spherical
I
j
lb
i30 microns
i
!30 microns
Water
;a5
,
j
I
!
-
~
.
p
-
~
'j
!Spherical
,
IRR
.-
___
..~
.Silica
2.55(7)
/
I
Process
~
Slurry
!
i
;Solids !Particle Size
!Loading /Dist.(Mean Shape
Velocity
~
I
_
6.2 fps !off
6.2 fps
--
15 v o l X
_____
!N/A
6.2fps
____
_____-__.Off
~,
. . .
.
/N/A
43fps
IOff'
;N/A
6.2fps
103
lNlA
4.3fps
.
u
I
!N/A
6.2fps
INlA
1
53fps
:N/A
6.2fps
!
j
4
I
!
!Low (1.00)
Low (2cP)
[Low (5%).
lM. Low (1,094)
Evaluate ability to detect air bubbles
'Want Bubble flow, Not Slug Flow
m u a t e Solids Loading Effects
Low (ZcP)
!Low (5%).
iM Low (1,094)
z u a t e Lower FlowRate
-
_-_____
S y t e m Checkout and Baseline testing with no solids
__
,
.
.
..
!
Med. Low (4cP) IMed (20%)
IM. High (1.376)
Evaruate Solids Laadin Effects
IM. High (1.376)
Evaluate Lower Flow Rate
I
!Off
Med. Low (4cP) ;Med (20%)
I
I
'Off
Med (-2OcP)
'High (35%)
jV. High (1.658)
!
1
!Off
Mad (-2OcP)
!High (35%)
/V. High (1.658)
I
/Off
Medium
Evaluate Solids Laadin Effects
-
Evaluale Lower Flow Rate
-
I
!
!Low
:Low
System
Checkout and viscosity Affects with no solids
-
ILOW
Evaluate intermediate viscosity
i
iNlA
6.2fps
!
High
JOff
I
!
low (ZcP)
ILow (5%)
I
6.2fps
loff
Med (6cP)
[Med (20%)
I
I
I
I
!
L
6.2fps IOff
High (-14cP)
i
High (-14cP)
I
!
j
i
1
:
I
I
~
_ _ _6.2fps
_ _ _iMf
I
_
_
1
_
_
!High (35%)
.!Low (-1.0)
E t r u m e n t checkouUtes1solids loading effects
/High (35%)
iLow (-1.0)
Tesl Carrier Liquid Color Effects on Lasentec Performance
!
Slip particles at same conditions as 2a
_
1
_
6 . 2 1 ~ 15vol
~
%
Low(2-3cP)
h
!
I
42fPS ;off
Low (2-3cP)
1
1
I
-_1___--
-_ 6.2fps
_-_
___
loff
!V Low (2%)
/Low(l.O24)
!V. Low (2%)
/Low (1.024)
,
i
Med (12-13cP) ILOW(5%)
ILow ( 1 . E )
1Supplement: Air Bubblsr on at 5~01%
lEvaluate Ability lo discriminate between air and 50%
!Test Low Flow Rates
I
Test Slip panicles at same conditions as 2a
1
I
_-
__________
Med(12-13cP) ~Low(5%)
- 4.3fps
1
__
!-A
I
/Low (-1.0)
I
__.__-
I
I
I
I
I
Instrument Checkout with NonSlip Solids particles
:Plastic i s S-Nauba 5021 from Shamrock Industries
Instrument Checkout nest solids loading effects
!Low (-1.0)
I
+ 6
1
-!
I
!Off
I
J
1
IVery Low (0%)
6.2fps
1i
,
1
Low (1cP)
I
I
;
'Low (1.00)
i
1
!
_I
I
i.
:Very Low (0%)
I
1
I
1
i Sp. Grav.
!
j
i
!
Low (1 cp)
I
I
.-
j
kolids Loading
I
j
I
__
-:Comments
Viscosity
_
I
I
purpose^^
Expected Properties
Control Variables __-___
-
;Low(l.ffi)
ITesI Low Flow Rates
I
I
I
_____
6.2fps ,off
High
_ ~ _(-35cP)
_ _ 'M.Low (7.5%) IM. Low (1.09) Test Slip particles atsame conditions as Zb
--___
!Also determines solid loading effeds
-L_.--2---!.-___-6.2 Ips ' 5 vol %
High (-35cP)
!M Low(7.556) IM Low (1 09)
.__!Evaluate Ability to discriminate betwee air and solids
I .-__
-i
~
---
B. 1
Figure B
___-
Liquid Characteristics
Test Loop Liquid
.
'Composrtion Color
100% Clear
,
'W
:.-1--b
iW
.__c.__._-
.
.__
Water
-
IW
c
said
Density
Solids
Measured Loading
2 2 0 75 vol%
_-__11~cps _ Bentonite
_
-L- -&. - -
100% Clear
~
2.4.1 volY
________
I
.
-
Water
100X.Ciear
-A---
--
viscosity Solid (11
1 cps
Graphite
Water
Bentonite
____
100% Clear
,I
cpi_
I
_-/-
-
F
IW
Water
100% Orange ' l c p s
I
Particle Size
Dist / Mean
Measured
2 9 microns
10.8 micron
Shape
Irregular
2.4,2.5 vol% i0.8 miaon
v-
!Disclike
I
2.4 4 v o l X
0.8 micron
Bentonite
2 4 4~01%
O8mlaW
I
-___
IW
Enknown
7
4
.
___
b
IC
i~
Id
/W
/W
iw
TED
-___-TBD
Unknown:
I
Unknown
Unknown
;
I
DisGLlke
-
_
_
_
.
I
_
___--~____
__
_
_
I
!
1
I
I
I
TED
I
TED
ITBD
~
1
TED
;TED
'TBD
Note. 1) Ail Quantified Values May change based on Simulant Development Testing
I
I
I
I
I
i
I
I
1
I
~ '
!
~
__...
--
_
___
_
_
I
_
_
-
___
I
-_
I_
_
-__
I
I
I
I
I
I
I
I
_
!Particle Size
Dist Mean
Measured
Shape Solld (3) 'Density
4 microns
IRR
Sti~ca
2 55(7)
I
1
a
I
c
--__
Disc-Like
!
i
ISolids
/Solid (2) IMeasumd Loadng
Gibbsite I
2 5 26 vol%
-____.i__-
Bentonite
! Solld(2)
Density
:Disc-Like I
,
~
c1
_
Solid Characteristics
cry
I
c2 B
Water
-
7.
ia
,
__
Test Sub Test
,
I
I
I
I
1
I
II
I
--
(contd)
-
___
)lids Parlicle Size
lading Disl /Mean Shape
vot%
_- 4 0 0 - 5 2 0 s p h e r i d
~
-_
__
__-
___
~-
____
_-
Velocity Bubbler
62fps
-- Off
62fps
-- -_
6.2fps
6.2 fps
--
__
____-_
__
-
_ _ Expected
- ~ Properties
_ _ _ _ ~ Purpose
_
_
~
'Comments
Slurry ___________
____
-
I
- 162fps
__
process- ~
Control VarfablE
Off
--_
1
Solids Loading
,High (27 75)
Sp Grav
H!gh (7 418)
_____
LOW
_____
__-_ __Low
LOW
_ _ ______
_ _ LO_\"__---__
Off
~.
Off
6-2 f p s--_Off
-
Viscasily
5*/ High (-35cP)
--- 6 2 fps
6 2 fps
62fps
62fps
-- -
Low (-3
Low
Low
T e a abdity to disutminate a r bubbks from sotids n campicaled mixtures
I-
____
Test ability 10 discriminatebetween calor of solids and lquid al low solids loads
-~
-~
Low (1.07)
Test ability of Lasentec to measuzhlghly irregular particle
Low ( l . l 3 ) - I ~ ~ - m o d a Idistributionconslg of Mat particles
_ _ ~ _ _ _ ~
cP)
Low (4%)
__ _____
Low (2 5%)
_______
______
(54%-
Med (-15cP)
Low (1 035)
H
Low (1 056)
High(47c.P)
Low
Test Viscosity Effects
______
Test Viscosity Effects
Low (1.056)
_________
__
-
~
C'P%Oft--
7fPS
S
7
I
I
__-__
--__-
___
-
-_.__
__
P TestabilQ lo discruninate between color of solids and lquid at low soltds loads
I
_______
Test a unknown slurry simulant lo test operator abtlity
IEvaluatessolids loading effetts
1
.--
- .____ -
-
___
I
_____
____
Testability of Lasentec to m e G r e highly irregular particles
-
-___-
I
-€valuates
-
solids loading effects
I
B.3
__
Appendix C
Experimental Data
Appendix C
Experimental Data
The table is composed of four sections.
Section 1: The first section contains the run numbers, test numbers, and what loop was tested (“W”
stands
for W-211 and “I”stands for ideal). The run number is identified by a number (e.g., 3). The
test number is identified by an alphanumeric code (e.g., 3a, 3a1, 3a2). The alpha part of the
test number is typically changed when the composition of the slurry is modified. (Since this
table has been modified from the original test plan submitted on October 1, 1995, based on
additional simulant characterization tests, the combination of run number and test number is
confusing. It is also generally not used within this report.)
Section 2: The second section contains the liquid and solid characteristics that constitute the slurry.
Section 3: The third section contains the processing conditions to be evaluated during the run.
Section 4: The fourth section contains the expected slurry properties based on laboratory testing and the
goal of each experiment.
1
Table C.1. Instrument Valid
3n Experimental Data
c.3
:ontd)
c.5
Table C.1. (contd
c.7
Table C . l . (cont
c.9
-
Table C.l. (contd)
t
C . 13
Appendix D
Statistical Analysis Results
Appendix D-1, t-Test for S-1 vs. S-2 Density
D.l
(fS2 Density By S 1 Density)
."0
z
n
(Y
v1
.8
.9
-Paired t-Test
1.0
1.1
1.2
S1 Density
1.3
1.4
1.5
I
(Paired t-Test]
-
S1 Density S2 Density
0.000253
Mean Difference
0.000341
Std Error
0.74061
t-Ratio
DF
48
Prob > It\
Prob > t
Prob < t
0.4625
0.2313
0.7687
Figure D-1: t-Test for S1 vs S2 Density
Since the Y = X line falls within the best fit upper and lower confidence limits, the difference between
the S1 and S2 densities are not statistically significant. Therefore, no settling is observed.
D.3
r
1
[S2 Density By S 1 Density]
I
1.082
1.081
1.080
'E1.079
a
r-4
rn
1.078
1.077
1.076
1.075
1.075
1.076
1.077
1.078
1.079
1.080
S1 Density
I
1.081
1.082
1.083
I
p
1-l
-
S1 Density S2 Density
Mean Difference
0.000253
Std Error
0.000341
t-Ratio
0.74061
DF
48
Prob > It1
Prob > t
Prob .c t
0.4625
0.2313
0.7687
J
\
Figure D-l(Blow Up): t-Test for S1 vs S2 Density
Since the Y = X line falls within the best fit upper and lower confidence limits, the difference between
the S1 and S2 densities are not statistically significant. Therefore, no settling is observed.
D.4
Appendix D-2; t-Test for S1 vs. S2 Volume % Solids
D.5
32 Solids, vol% By S1 Solids, ~ 0 1 % )
0
=IPaired t-Test
Paired t-Test ]
15
20
S1 Solids, vol%
10
5
25
30
35
I
.
-
3 1 Solids, vol% S2 Solids, vol%
Mean Difference
Std Error
t-Ratio
DF
0.194
0.148299
1.308167
Prob > It[
Prob > t
Prob < t
0.1996
0.0998
0.9002
34
Figure D-2;t-Test for S1 Volume % Solids vs S2 Volume YOSolids
Since the Y = X line falls within the best fit upper and lower confidence limits, the difference between
the $1 and S2 analytically measured solids loadings are not statistically significant. Therefore, no
settling is observed.
I
I
D.7
52 Solids, vol% By S1 Solids, ~ 0 1 % )
16.0
17.0
19,O
18.0
20.0
SI Solids, ~ 0 1 %
1
_
1
>
S1 Solids, vol% - S2 Solids,vol%
Mean Difference
0.194
Prob It1
Std Error
0.148299
Prob r t
Prob < t
t-Ratio
1.308167
DF
34
0.1996
0.0998
0.9002
1
Figure D-2(Blow Up); t-Test for S 1 Volume YOSolids vs S2 Volume YOSolids
Since the Y = X line falls within the best fit upper and lower confidence limits, the difference between
the S1 and S2 analytically measured solids loadings are not statistically signifcant. Therefore, no
settling is observed.
1
I
D.8
Appendix D-3; t-Test for S1/S2 vs. S3/S4 Average Densities
D.9
S3/S4 Pycnometer Density By S2 Density)
U.Y
I
I
I
i
I
I
I
.9
1.o
1.1
1.2
1.3
1.4
1.5
S2 Density
881 G P a i r e d
(Paired t-Test 1
S2 Density - S3/S4 Pycnometer Density
BFitting
Mean Difference
Std Error
t-Ratio
DF
0.000828
0.000751
1.102681
T
Prob > It1
Prob > t
Prob < t
0.2750
0.1375
0.8625
55
Figure D-3: t-Test for S3/M Density vs S2 Density
Since the Y = X line falls within the best fit upper and lower confidence limits, the difference
between the S2 and S3/S4 densities are not statistically signficant. Therefore, no settling is
observed.
D.11
33/S4 Pycnometer Density By S2 Density
I
1.07
1.08
S2 Density
E Paired t-Test
I
‘FzmGi-~.
S2 Density - S3/S4 Pycnometer Density
Mean Difference
Std Error
t-Ratio
DF
0.000828
0.000751
1.102681
Prob > It1
Prob > t
Prob < t
0.2750
0.1375
0.8625
55
Figure D-3 (Blow-Up): t-Test for S3/S4 Density vs S2 Density
Since the Y = X line falls within the best fit upper and lower confidence limits, the difference
between the S2 and S3/S4 densities are not statistically signficant. Therefore, no settling is
observed.
I
D. 12
r [s3/s4 density By s2 density)
Paired t-Test Loop=Ideal
Paired t-Test Loop=W-211
a
92 density - s31s4 density
Mean Difference
0.000321
Std Error
0.000904
t-Ratio
0.354723
\
DF
Prob > It1
Prob > t
Prob < t
0.7274
0.3637
0.6363
Prob > It1
Prob > t
Prob < t
0.1171
16
((Paired t-Test Loop=W-2 11
s2 density - s3fs4 density
Mean Difference
Std Error
t-Ratio
DF
0.001109
0.000919
1.206761
42
1
0.2343
0.8829
Figure D-3 (Blow-Up); Grouped t-Test for S31S4 Density vs S2 Density
Since the best fit line of the two groups of data fall within the best upper and lower confidence limits
of the two groups, the difference between the two groups is not statistically significant. Therefore, the
S3fS4 average density is not affected by the design of the loop.
I
I
D. 13
;3/s4 density By s2 density)
1.o
.9
1.2
1.1
1.3
1.4
1.5
s2 density
[Paired t-Test Loop=Ideal
-
1
s2 density s3is4 density
Mean Difference
Std Error
t-Ratio
DF
0.000321
0.000904
0.354723
16
Prob > It1
Prob > t
Prob < t
0.7274
0.3637
0.6363
Prob > It1
Prob > t
Prob < t
0.2343
0.1171
[Paired t-Test Loop=W-2 11 )
-
92 density s3is4 density
Mean Difference
Std Error
t-Ratio
DF
0.001109
0.000919
1.206761
0.8829
42
Figure D-3; Grouped t-Test for S31S4 Density vs S2 Density
Since the best fit line of the two groups of data fall within the best upper and lower confidence limits
of the two groups, the difference between the two groups i s not statistically significant. Therefore, the
S3lS4 average density is not affected by the design of the loop.
D. 14
Appendix D-4; t-Test for Micromotion vs. S3/S4 Average Densities
D. 15
MM Density By s3fs4 density]
1.o
.9
1.1
1.2
1.3
1.4
1.5
s31s4 density
] -P[
-
s31s4 density MM Density
Mean Difference
Std Error
t-Ratio
DF
-0.03526
0.003809
-9.25687
58
Prob > (tl
Prob > t
Prob < t
e.0001
1.0000
<.0001
Figure D-4; t-Test for Micromotion Density vs S31S4 Density
Since the Y = X line does not fall within the best fit upper and lower confidence limits, the difference
between the Micromotion density and the S3IS4 average density is statistically significant.
I
I
D.17
dh4 Density By s3/s4 density]
-
1.2
1.1
1.1
1.2
s3ls4 density
[Paired.)
s3Is4 density - MM Density
Mean Difference
Std Error
t-Ratio
DF
-0.03526
0.003809
-9.25687
Prob > It1
Prob > t
Prob < t
<.0001
1.0000
e.0001
58
Figure D-4 (Blow-up); t-Test for Micromotion Density vs S3/S4 Density
Since the Y = X line does not fall within the best fit upper and lower confidence limits, the difference
between the Micromotion density and the S3/S4 average density is statisticallysignificant.
D.18
I
(MM Density By s3ls4 density)
1
1.5
1.4
1.3
13
1.1
1.0
1.0
.9
Mean Difference
-0.04293
0.006631
1.1
1.2
s3ts4 density
Prob > It1
Prob > t
Prob < t
<.0001
1.0000
<.0001
Prob > It1
Prob > t
Prob .e t
<0001
1.3
1.4
1.5
r
[Paired t-Test Loop=W-211]
-
s3ts4 density MM Density
Mean Difference
Std Error
t-Ratio
DF
i\
-0.03215
0.00459
-7.00512
41
1.0000
<.0001
J
Figure D-4; Grouped t-Test for Micromotion Density vs 53154 Density
ince the best fit line of the two groups of data fall within the best fit upper and lower confidence limits of
the two groups, the difference between the two groups is not statistically significant. Therefore, the
Micromotion Densimeter works equally well in either loop.
D. 19
MM Density By s3Is4 density)
1.14
1.13
1.12
.-0rn
1.11
I 1.10
n
3
1.09
1.08
1.07
1.06
1.03
1.04
1.05
1.06
1.07
1.08
1.09
1.10
1.11
1.12
1.13
s3Is4 density
aFitting
Paired t-Test Loop=Ideal
%$i EPaired t-Test Loop=W-211
\
[Paired t-Test Loop=Ideal
-
s31s4 density MM Density
Mean Difference
Std Error
t-Ratio
DF
-0.04293
0.006631
-6.47384
16
Prob > It1
Prob > t
Prob < t
1.0001
1.0000
<.0001
Prob > It1
Prob > t
Prob < t
<.0001
1.0000
Paired t-Test Loop=W-211
-
s3Is4 density MM Density
Mean Difference
-0.03215
Std Error
0.00459
t-Ratio
-7.00512
DF
41
<.0001
Figure D-4(Blow-Up); Grouped t-Test for Micromotion Density vs S3lS4 Density
ince the best fit line of the two groups of data fall within the best fit upper and lower confrdence limits of
the two groups, the difference between the two groups is not statistically significant. Therefore, the
Micromotion Densimeter works equally well in either loop.
D.20
Appendix D-5; t-Test for Micromotion vs. Calibration Tank Densities
D.2 1
dM Density By Meas. Density)
U.Y
I
I
I
.9
1.0
1.1
Mean Difference
Std Error
t-Ratio
DF
'
-0.0275
0.0023
-11.958
189
Prob > It1
Prob>t
Prob < t
I
i.2
Meas, Defisity
<.0001
1.0000
<.0001
1
I
I
1.3
1.4
1.5
I
J
Figure D-5:&Test for Micromotion Density vs Calibration Tank Measured Density
Since the Y = X line does not fa11 within the best fit upper and lower confidence limits, the
difference between the Micromotion density and the Denisty measured in the calibration
feed tank is statistically significant.
I
D.23
VllM Density By Meas. Density)
1.14
1.13
1.12
1.11
0 1.10
.
I
v1
8
1.09
pi
1.08
1.07
1.06
1.05
1.04
1.04
1.05
1.06
1.07
1.08
1.09
1.10
1.11
1.12
Meas. Density
Paired t-Test
(Pairedt-Test]
Meas. Density - MM Density
Mean Difference
-0.0275
Std Error
0.0023
t-Ratio
-11.958
DF
189
1
Prob > It1
Prob > t
Prob < t
I
<.0001
1.0000
<.0001
Figure D-5 (Blow-up): t-Test for Micromotion Density vs Calibration Tank Measured
Density
Since the Y = X line does not fall within the best fit upper and lower confidence limits, the
difference between the Micromotion density and the Denisty measured in the calibration
feed tank is statistically significant.
D.24
1.13
MM Density By Meas. Density)
1.5
1.4
0
2
B
E
E
1.3
1.2
1.1
1.o
.9
aFitting a
1.0
1.1
1.2
Meas. Density’
1.3
1.4
1.5
Paired t-Test Loop=Ideal
=Paired t-Test Loop=W-211
I
(Paired t-Test Loop=Ideal
1
Meas. Density - MM Density
Mean Difference
-0.03148
Std Error
0.004379
-7.1882
t-Ratio
DF
57
Prob > It1
Prob > t
Prob < t
<.0001
Prob > It1
Prob > t
Prob < t
<A001
1.0000
<.0001
fPaired t-Test Loo~=W-211 I
Meas. Density - MM Density
Mean Difference
-0.02575
Std Error
0.002691
t-Ratio
-9.57045
DF
131
1.0000
cOOOl
Figure D-5;Grouped t-Test for Micromotion Density vs Measured Density:
Since the best fit line of the two groups of data fall within the best fit upper and lower confidence limits
of the two groups, the difference between the two groups is not statistically significant. Therefore, the
Micromotion Densimeter works equally well in either loop.
D.25
1.060
aFitting
1.080
Meas. Density-
[Paired t-Test Loop=Ideal
Meas. Density - MM Density
Mean Difference
Std Error
t-Ratio
DF
3
-0.03148
0.004379
-7.1882
57
7
Prob > It]
Prob > t
Prob < t
<.0001
1.0000
<.0001
.
3
(Paired t-Test Loop=W-2 11 )
Meas. Density - MM Density
Mean Difference
-0.02575
Std Error
0.002691
t-Ratio
-9.57045
DF
131
c
1.090
fPaired t-Test Loop=Ideal
E Paired t-Test Loop=W-211
7
c
1.070
Prob > It1
Prob > t
Prob < t
<.0001
1.0000
e.0001
I
J
Figure D-5(Blow-Up) ;Grouped t-Test for Micromotion Density vs Measured Density:
Since the best fit line of the two groups of data fall within the best fit upper and lower confidence limits
of the two groups, the difference between the two groups is not statistically significant. Therefore, the
Micromotion Densimeter works equally well in either loop.
D.26
Appendix D-6; t-Test for Yokagawa vs. Fisher & Porter Flow Rates
D.27
;&P By Yokagawa)
I
I
I
I
I
I
II
I
I.
I
I
-5
I
I
-5
-15
I
I
5
Yokagawa
MFitting
Mean Difference
Std Error
t-Ratio
DF
Paired t-Test
0.389167
'0,316167
1.230891
47
Prob > It1
Prob > t
Prob < t
0.2245
0.1122
0.8878
I
Figure D-6;t-Test for Yokagawa vs. Fischer & Porter Flow Meter
Since the Y = X line falls inside the best fit upper and lower cofidence limits, the difference between the
two flow meters is not statistically signifcant. This analysis is with the data from runs 5BO and 5B1
excluded due to the fouling of the electrodes and due to the exclusions of an outlier in run SA0 and 4A3
D.29
Appendix D-7; t-Test for Micromotion vs. Yokagawa Flow Rates
D.31
.
'FMicromotion By Yokagawa)
.-*B
0
E
eu
E
-10
-15
mFitting
111
-5
0
Yokagawa
5
10
15
Paired t-Test
I
@Gimz]
-
Yokagawa Micromotion
Mean Difference
Std Error
t-Ratio
DF
6.426215
0.213954
30.03543
Prob > It1
Prob > t
Prob < t
<.0001
<.0001
1.0000
176
Figure D-7; t-Test for Micromotion Flow rate vs Yokagawa Flow Rate
Since the Y = X line does not fit within the best tit upper and lower confidence limits, the difference
between the two flow meters is statistically significant. This analysis was done with Run 5Bl from the
W-211loop testing, Runs SB and 5B1 from the "ldeal" Loop testing and one outlier from SA "Ideal"
excluded.
D.33
[Micromotion By Yokagawa)
e
.-0
0
+I
f"
e
z
i
V
-10
0
-5
-10
5
10
15
Yokagawa
WFitting
EPaired t-Test Loop=Ideal
fl E Paired t-Test Loop=W-211
\
(Paired t-Test Loop=Ideal ]
Yokagawa - Micromotion
Mean Difference
Std Error
t-Ratio
DF
,.-
7.871633
0.294831
26.69883
48
Prob > It1
Prob > t
Prob t
<.0001
<0001
1.0000
5
[Paired t-Test L O O P = ~1]- ~ ~
-
Yokagawa Micromotion
Mean Difference
Std Error
t-Ratio
DF
5.872891
0.257756
22.78467
127
Prob > It1
Prob > t
Prob < t
<.0001
<.0001
1.0000
i
Figure D-7; Grouped t-Test for Micromotion vs Yokagawa Flow Rates
Since the best fit line from the two groups do not fit within the upper and lower confidence limits of the
ther group, the difference between thm is statistically significantly. This analysis was done with Run 5B1
from the W-211 loop testing, Runs 5B and 5B1 from the "Ideal" Loop testing and one outlier from 5A
"Ideal" excluded.
1
D.34
Appendix D-8; Yokagawa, Micromotion,
Fischer & Porter Data Distributions
D.35
(Micromotion)
5
1
10
I
0
-5
1
-10
-I
-
-15
I
I
!
-15
(Quantiles 1
maximum
100.0%
99.5%
97.5%
90.0%
quartile
median
quartile
minimum
75.0%
50.0%
25.0%
10.0%
2SoA
0.5%
0.0%
13.010
13.010
3.945
2.682
-0.515
-2.085
-3.895
-6.885
-9.081
-14.670
-14.670
maximum
quartile
median
quartile
100.0%
99.5%
97.5%
90.0%
75.0%
50.0%
25.0%
10.0%
2.5%
0.5%
minimum
0.0%
6.090
6.090
-1.799
-5.298
-6.900
-8.700
-10.713
-11.656
-13.596
-14.350
-14.350
J
Mean
Std Dev
Std Error Mean
Upper 95% Mean
Lower 95% M e a n
N
Sum Weights
-2.0998
3.6067
0.2750
-1.5570
-2.6427
172.0000
172.0000
Mean
Std Dev
Std Error Mean
Upper 95% Mean
Lower 95% Mean
N
Sum Weights
D.37
-8.5544
2.8315
0.2159
-8.1282
-8.9806
172.0000
172.0000
L
[F&p)I
i1
i
!
iI
I
1!
I
I
I
I
I
j
I
i
-
l
O
i
.-___----
.
.
(-Om.
5.5
Quantiles 1
maximum
quartile
median
quartile
minimum
100.0%
99.5%
97.5%
90.0%
75.0%
50.0%
25.0%
10.0%
25.?0/
0.5%
0.0%
. 7.200
maximum
7.200
6.676
97.5%
90.0%
1.270
-1385
-1.915
-3.280
-4.470
quartile
median
quartile
N
Sum Weights
\
-2.01864
2.67470
0.40323
-1.20546
-2.83182
44.00000
44.00000
50.0%
25.0%
2.5%
minimum
\
~M.pll
75.0%
10.0%
-10.018
-10.420
-10.420
\
Mean
Std Dev
Std Error Mean
Upper 95% Mean
Lower 95% Mean
100.0%
99.5%
0.5%
0.0%
83400
83400
83400
83400
8.2050
7.5800
6.2700
5.5800
5.5800
5.5800
5.5800
Moments)a(
Mean
Std Dev
Std Error Mean
Upper 95% Mean
Lower 95% Mean
N
Sum Weights
D.38
7.290000
1.060264
0.432851
8.402663
6.177337
6.000000
6.000000
Appendix E
Lasentec Calibration
Appendix E
Lasentec Calibration
12%
!
I
I
j
!
i
i
!
Percent 6%
I
I
I
!
Ij
3x
,
I
I
i
0%
0.8
/
4
8
16
n
.
.
I
I
63
I
*
,
,
,
I
,
125
Std Mean
Figure E.l. LASENTIZC@Measurement Analysis
E. 1
*
250
,
c
.
,
500
,
,
,
II
1000
c
0)
0
a
696 _ _ _ _ _ _ _ - _ _ _ _ _
0.8
4
a
16
31
63
125
Std Mean
Figure E.2. LASENTC" Compare Analysis
E.2
250
500
1000