Cardiothoracic Transplantation and Mechanical Circulatory Support: Acquired
Callington et al
Computational fluid dynamic study of hemodynamic effects on aortic
root blood flow of systematically varied left ventricular assist device
graft anastomosis design
Andrew Callington, BS,a Quan Long, PhD,a Prashant Mohite, MD,b Andre Simon, MD,b and
Tarun Kumar Mittal, MDb
ABSTRACT
Objectives: To quantify the range of blood flow parameters in ascending aorta
that can result from various angulations of outflow graft anastomosis of a left
ventricular assist device (LVAD) to the aortic wall, as a means to understand
the mechanism of aortic valve insufficiency.
Methods: A realistic aorta model with LVAD anastomosis was generated from
computed tomographic images of a patient. Based on this model, the LVAD anastomosis geometry parameters, such as anastomosis locations, inclination angle,
and azimuthal angle (cross-sectional plane) of the graft, were varied, to create
21 models. With the assumption of no flow passing the aortic valve, and a constant
flow rate from the LVAD cannula, computational fluid dynamics simulations were
used to study the blood flow patterns in the ascending aorta. In addition, pulsatile
flows were assumed in the LVAD cannula, with the aortic valve opened during
peak systole, for 2 specific anastomosis configurations, to evaluate the influence
of the pulsatile flow profile and the transvalvular flow on the aortic flow patterns.
Results: Changes in the inclination angle, from 60 to 120 , or the azimuthal
angle, from 90 to 120 , or moving from a lower to a higher anastomosis position,
causes significant changes for all flow parameters. A lower anastomosis location,
an inclination angle 90 , and an azimuthal angle of 60 or 120 are all capable of
reducing blood flow stagnation in the aortic root and producing normal wall shear
stress and moderate pressure values in the region.
Conclusions: Carefully chosen anastomosis geometry is likely to be able to
generate a close-to-normal hemodynamic environment in the aortic root. Greater
knowledge of aortic valve remodeling may make possible the creation of favorable flow patterns in the aortic root, through optimization of surgical design to
reduce or delay the occurrence of aortic valve insufficiency. (J Thorac Cardiovasc
Surg 2015;150:696-704)
For advanced stages of heart failure, heart transplantation
remains the best treatment for suitable patients.1 However,
the increasing severity of heart failure as a public health
TX
From the aBrunel Institute for Bioengineering, Brunel University London, Uxbridge,
and bDepartments of Imaging and Surgery, Royal Brompton & Harefield National
Health Service Foundation Trust, Harefield Hospital, Middlesex, United Kingdom.
Received for publication Nov 20, 2014; revisions received April 10, 2015; accepted
for publication May 9, 2015; available ahead of print June 17, 2015.
Address for reprints: Quan Long, PhD, Brunel Institute for Bioengineering, Brunel
University, Kingston Lane, Uxbridge, UB8 3PH, United Kingdom (E-mail: quan.
long@brunel.ac.uk).
0022-5223/$36.00
Copyright Ó 2015 by The American Association for Thoracic Surgery
http://dx.doi.org/10.1016/j.jtcvs.2015.05.034
696
Reconstructed aorta with left ventricular assist device
anastomosis at 3 inclination angles.
Central Message
This study investigates the effects of varying
left ventricular assist device anastomosis geometry on local blood flow, using computational fluid dynamics.
Perspective
A systematic haemodynamic study demonstrated that the design of LVAD anastomosis
to the aortic wall will have significant influence
to the blood flow in ascending aorta.
See Editorial Commentary page 704.
problem and the major lack of donor organs has spurred
the evolution of mechanical circulatory support devices in
recent decades.2,3 These devices, such as ventricular assist
devices, are used as a ‘‘bridge-to-transplant,’’ but in
addition, increasingly as destination therapy in selected
patients. As of 2012, >4000 left ventricular assist devices
(LVADs) had been implanted in the United States alone,
and this number continues to rise.4,5
However, several adverse effects can result from this
treatment; these include high rates of thromboembolic
events, infection, bleeding, and device malfunction or failure. Another is aortic valve insufficiency (AVI), which is a
known complication after continuous flow LVAD support
The Journal of Thoracic and Cardiovascular Surgery c September 2015
Cardiothoracic Transplantation and Mechanical Circulatory Support: Acquired
Abbreviations and Acronyms
AVI ¼ aortic valve insufficiency
CFD ¼ computational fluid dynamics
CT
¼ computed tomography
LVAD ¼ left ventricular assist device
WSS ¼ wall shear stress
that may result in ineffective biomechanical output and endorgan malperfusion, with substantial effects on both quality
of life and survival.6 The mechanisms involved with the
initiation and progression of AVI are the main focus of
this study.
Clinical observation has shown that the start of AVI can
in some cases be delayed if blood is still flowing past the
aortic valve. Although the nonopening aortic valve is
believed to be the main culprit for this complication,
this observation indicates that maintaining a physiologic
hemodynamic flow in the aortic root after the introduction
of an LVAD may delay development of AVI. A few studies
published recently have evaluated the effects of both
LVAD–ascending aorta anastomosis configurations, along
with their long-term impact on the aortic valve,7-12
and cerebral embolization,13 using computational fluid
dynamics (CFD).
CFD is a technique used to predict fluid flow patterns and
distributions of flow parameters, such as velocity, pressure,
and fraction force of blood to the arterial wall, which is also
called wall shear stress (WSS), by using a computer to solve
fluid flow governing equations. A geometry model (ie, patient aorta) is required in CFD simulation. Together with
predefined flow information at the model inflow and outflow
planes (also called dynamic boundary conditions), CFD
simulation can provide flow parameter distributions in
the model; these can be changed with time if a timedependent boundary condition is used.
Studies to date have considered only a limited number of
LVAD anastomosis designs. Furthermore, in cases in which
aortic WSS and pressure distributions were resolved, localized distributions were not analyzed. The CFD analysis
performed in these studies used imaging data from
computed tomography (CT) or magnetic resonance imaging to generate the geometry model. Thin volumetric CT
data of the thoracic aorta, with electrocardiogram-gating,
provides high-resolution 3-dimensional reconstructions
and improved model geometry accuracy.
The present study explores the effect of a wide range of
LVAD outflow graft anastomotic angles to the ascending
aorta, and its location distal to the aortic root, to understand
the mechanism of AVI using CFD simulation on patientspecific CT data. Aortic pressure and WSS distributions
are resolved at the LVAD jet impact region and in the aortic
root. The research presented is used to study the hypothesis
that long-term alteration of the local flow environment is
associated with initiation of AVI in LVAD patients. In
addition, the current work can help in future long-term
studies to determine if an optimal anastomosis to reduce
the incidence of AVI is obtainable.
METHODS
The use of the data for this study was approved by the local ethical
committee, and a consent waiver was granted because anonymized image
data were used. The 3-dimensional geometry of the aorta was extracted
from CT scan data in a patient with an LVAD. The scan was performed
for clinical reasons, with electrocardiogram-gating, using 85 mL of
iodinated contrast (given intravenously), and included the thoracic aorta
and proximal branches of the arch, using an Aquilion 64-slice CT scanner
(Toshiba Medical Systems Corporation, Otawara, Tochigi, Japan). The
images were reconstructed at 0.5-mm slice thickness, with a 0.3-mm
increment, resulting in an isotropic spatial resolution of 0.35 mm. These
data were used as a template for subsequent geometric reconstructions.
Branching arteries, ie, left/right coronary, brachiocephalic, subclavian,
and carotid arteries, were included in the flow domain, which consisted of
the aortic root, arch, and ascending and descending aorta. A section (approximately 75 mm) of the LVAD outlet conduit, before the aortic anastomosis,
was removed from the original CT geometry, and new LVAD anastomoses
were reconstructed in its place. These were described as: lower ¼ just above
the aortic sinuses; middle ¼ at the original graft site; and upper ¼ just before
the origin of the brachiocephalic trunk (Figure 1). The reconstructed
anastomotic angles are described as ‘‘inclination’’ and ‘‘azimuthal,’’ on
the coronal and transversal planes, respectively, as shown in Figure 2.
A total of 21 anastomosis designs were studied, exploring the inclination (f) and azimuthal (q) angles of 60 , 90 , and 120 . An inclination
angle (f) of 60 would represent the jet flow pointing away from the aortic
root, whereas 120 would mean that the flow direction is toward the aortic
root. As a result of unrealistic LVAD graft profiles, f ¼ 60 and f ¼ 120
were omitted from the lower and upper anastomosis designs, respectively.
All geometric reconstructions and mesh generations were carried out using
ANSYS ICEM CFD (ANSYS, Inc, Canonsburg, Pa).
Quantification of arterial WSS and pressure distributions in the
ascending aorta, aortic root, and LVAD jet impact region are the primary
focus of this study. To accurately determine these, circumferential
segmentation was carried out for 6 locations on the ascending aorta, with
the first band located 6 mm from the aortic valve. Subsequent bands
were set at a spacing of 10 mm. Each circumferential band was further
segmented 16 times, based on the angle from the lumen center (shown in
Figure 1, B, colored bands). WSS and pressure at the aortic root
(WSS_root, pressure_root) values, presented in the Results section, are
the area-weighted average value at the lowest band.
A mesh sensitivity study was carried out to assess the extent to which the
computational solution was affected by mesh parameters. To ensure
mesh-independent data, numerical calculations were run on 3 separate
mesh grids, consisting of, respectively: 700,000; 1 million; and 1.5 million
elements. A few points on the descending aorta wall and in ascending aorta
were used to compare the variations in WSS and velocity for the three mesh
models. According to the comparison results, the models used in the present
study have a mesh of about 1 million elements. To ensure the validity of the
study, solver settings, such as the discretization schemes and convergence
criteria, remained the same.
The CFD solver ANSYS Fluent (2014 release; ANSYS, Inc) was used
for the simulations. Blood was modeled as a Newtonian fluid, with a
constant density of 1060 kg/m3, and a constant viscosity of 0.004 Pa$s.
Arterial walls were treated as rigid, with no-slip boundary conditions.
The first part was steady flow simulation, in which flow rate from the
LVAD cannula was constant throughout the cardiac cycle, and the aortic
valve was closed. A zero gauge pressure mass flow-rate boundary condition
The Journal of Thoracic and Cardiovascular Surgery c Volume 150, Number 3
697
TX
Callington et al
Cardiothoracic Transplantation and Mechanical Circulatory Support: Acquired
Callington et al
FIGURE 1. A, Original geometry; B, C, and D, Geometric reconstructions; lower, middle, and upper anastomoses. Six 2-mm wide bands, with 10-mm
spacing, are shown near the aortic root, for WSS and pressure data area weighted average. WSS and pressure values at the aortic root are the results
from the lowest band.
TX
of 0.08833 kg/s was set only at the cannula inlet (providing a volumetric
flow rate of 5l per minute to the ascending aorta), normal to the boundary.
At this flow rate, the Reynolds’s number was 1840, which is smaller than
the critical value of 2300 so that the flow could be treated as a laminar flow.
Flow-rate ratios were prescribed for the descending aorta and all other
branching arteries as outflow boundary conditions, according to previously
published data.13 Specifically, the flow rate ratios for the left coronary
artery, right coronary artery, brachiocephalic artery, left common carotid
artery, left subclavian artery, and descending aorta outlet, respectively,
are 0.591%, 0.589%, 18.40%, 8.49%, 9.92%, and 62%. To determine
the total pressure drop of the main flow along the ascending aorta, total
pressure (the sum of static and dynamic pressure, representing the energy
level) values were taken at the lumen center of the aortic root, and
approximately three-quarters of the way down along the descending aorta,
for each case. The pressure difference of the former minus the latter value
was calculated to represent flow resistance.
Owing to the variation of left ventricular pressure during a cardiac cycle,
the pressure loading condition on the continuous flow LVAD pump changes
with time, causing a pulsatile flow output in the LVAD graft. Simulations
with pulsatile flow rate from an LVAD graft were performed. Two conditions were mimicked: (1) Pul_B, a weak pulsatile wave in LVAD graft
flow with the aortic valve closed to mimic the high–pump-rate setting
(12,000 rpm); (2) Pul_A, a medium strong wave in the LVAD jet, with
698
the aortic valve opened at around peak systole for a duration of 150 ms
(with a heartbeat of 90 bpm),14 with 5%, 10%, and 20% of total flow
rate through the aortic valve, to mimic the medium–pump-rate setting
(9000 rpm). The detailed flow rate curve was derived from Figure 6 of
Estep and colleagues15 and is shown in Figure 3. The aortic valve opening
is circular at the center of the aortic valve plane, with an area in which the
diameter is one-half of the aortic root diameter.
Two specific anastomosis configurations were chosen for pulsatile flow
simulation: f/q/anastomosis position ¼ 90 /60 /low; and 60 /60 /mid to
represent the 2 extreme cases (higher aortic root flow vs low jet resistance
flow) obtained from steady flow study. The flow rate curve was derived
from the pulse Doppler measurement result of Estep and colleagues,15
which has a higher mean flow rate 9.35 L/min. A turbulence model named
‘‘Realisable k-ε turbulence model’’ was chosen in the CFD solver. The
outflow boundary conditions are the same as those in steady flow of fixed
outflow rate ratio.
RESULTS
Qualitative Flow Pattern
Transversal plane secondary velocity vectors showed that
the LVAD flow resembled a jet as it entered the aorta, hitting
The Journal of Thoracic and Cardiovascular Surgery c September 2015
Callington et al
Cardiothoracic Transplantation and Mechanical Circulatory Support: Acquired
FIGURE 2. Definitions of inclination angle and azimuthal angle used in the study. A, Diagram for the definition of inclination angle 4. B, Diagram for
azimuthal angle q. LVAD, Left ventricular assist device.
impact on particle distribution and the level of exposure to
shear forces to which the arterial walls were subjected. In
addition, the cross-sectional velocity vectors showed
visibly higher near-wall velocity gradients as the flow
swirled around the aorta (Figure 4, A-C).
The influence of inclination angle f on the flow can be
seen qualitatively in Figure 4, D: f ¼ 60 ; E: f ¼ 90 ;
and F: f ¼ 120 (while q ¼ 90 , middle anastomosis). At
f ¼ 60 , the jet flow from LVAD cannula followed the
curvature of the aorta arch and did not hit the aortic wall
until it reaches the aortic arch. At f ¼ 90 , the jet flow
was divided into 2 parts: most of the flow was still going
TX
the opposing inner wall (Figure 4, A; q ¼ 60 ; (B) q ¼ 90 ;
and (C) q ¼ 120 , while f ¼ 90 , middle anastomosis). For
azimuthal angle q ¼ 90 , the jet flowed across the center of
the aorta, creating recirculation regions at either side. Varying q away from 90 changed the radial jet impact region, ie,
the azimuthal angle changed from 90 , ie, to 60 and 120 ;
consequently, one recirculation region was suppressed
while the other grew in size and intensity. The flow was
directed around the circumference of the aorta, which produced higher WSS values than in the q ¼ 90 cases, likely as
a result of the swirling effect of particles. The addition of
centrifugal forces of the jet flow here would have a notable
FIGURE 3. Velocity temporal wave curve in LVAD cannula used as boundary condition in pulsatile flow simulation. For Pul_A, the aortic valve opens
within the shadow period. The aortic valve remains closed for the whole cycle for Pul_B. LVAD, Left ventricular assist device; Pul_A, medium strong pulsatile wave curve; Pul_B, weak pulsatile wave curve.
The Journal of Thoracic and Cardiovascular Surgery c Volume 150, Number 3
699
Cardiothoracic Transplantation and Mechanical Circulatory Support: Acquired
Callington et al
FIGURE 4. A, B, and C, Example of velocity vectors at transversal plane of anastomosis, when 4 ¼ 90 , middle anastomosis; 3-dimensional streamlines
for the middle anastomosis case with (D) 4 ¼ 60 ; (E) 4 ¼ 90 ; and (F) 4 ¼ 120 , with q ¼ 90 . Color code: velocity.
into the descending aorta; a small portion of flow turned
toward the direction of the aortic root after the jet hit the
aortic wall, which provided some ‘‘washing out’’ force in
the aortic root. At f ¼ 120 , the flow was directed toward
the valve, creating large vortices near the center of the aortic
root.
The anastomosis location had a substantial impact on
root dynamics. A lower anastomosis provided greater root
WSS, which was reduced as the anastomosis location was
moved up the ascending aorta. A similar relationship was
found with aortic root static pressures. Furthermore, bulk
fluid velocity below the graft increased, as the angle of
inclination increased, which substantially enhanced the
blood particle washout in the aortic root (Figure 4, D-F).
TX
Quantitative Results for Steady Flow at the LVAD
Cannula, With the Aortic Valve Closed
Quantitative results of the effects of LVAD anastomosis
design on the flow parameters, such as WSS and pressure
at the aortic root and the anastomosis jet impact region,
and the pressure difference between the aortic root and
the descending aorta, are shown in Table 1. In general, a
lower anastomosis will produce a higher WSS at the root
700
region and greater root pressure. A higher inclination angle
f will normally produce higher values of WSS and pressure
in the aortic root region, for all anastomosis locations,
because of the partial backward flow from the anastomosis
to the aortic root. For the same anastomosis location and
inclination angle, the change in azimuthal angle q has
additional noticeable influences on the aortic root WSS
value. For cases of a lower inclination angle (f ¼ 60 , or
90 ), a change of q away from 90 (ie, to 60 or 120 ),
can almost double the WSS value in the aortic root region,
owing to the strong swirling flow effect.
With various anastomosis configurations, the range of
the flow parameter change is: (0.082 Pa, 3 Pa) for WSS;
and ( 11.7 Pa, 130 Pa) for pressure. In the jet impact
region, ranges for WSS and pressure are (6 Pa, 20 Pa)
and (47 Pa, 280 Pa), respectively. The range of pressure
drop from the aortic root to the descending aorta is
( 55.7 Pa, 108 Pa). The model generally created high
values of all flow parameters at a lower anastomosis site,
with f ¼ 120 , q ¼ 120 . We obtained the lowest values
at a middle anastomosis site, with f ¼ 60 , q ¼ 90 , which
happens to be similar to the anastomosis configuration used
for the patient.
The Journal of Thoracic and Cardiovascular Surgery c September 2015
Callington et al
Cardiothoracic Transplantation and Mechanical Circulatory Support: Acquired
TABLE 1. Quantitative results of aorta hemodynamic parameters for all testing cases
Test case (4 /q )
Parameter
Lower
WSS_root
Pressure_root
WSS_jet
Pressure_jet
Pressure_drop
Middle
WSS_root
Pressure_root
WSS_jet
Pressure_jet
Pressure_dropy
Upper
WSS_root
Pressure_root
WSS_jet
Pressure_jet
Pressure_drop
60/60
60/90
60/120
90/60
90/90
90/120
120/60
120/90
120/120
0.809
60
12
190
23
0.784
72
12
190
42
1.703
90
17.5
250
50
1.845
66.5
20
280
22.7
1.201
103.7
17
270
60
3.027
130
20
233
108
1.26
72.2
12.5
215
28
0.603
76.5
11
198
33
1.292
83.1
12
198
41
0.228
14.6
13
110
28.1
0.113
11.7*
6
47
55.7
0.31
27
11
137
16.3
0.45
40.4
11.7
162
7.6
0.2
18.4
9.1
118
22.3
0.702
48.4
12
180
0.6
0.133
15
9.4
117
34
0.082
5.8
7.5
70
49
0.183
33
11
147
9.53
0.119
66.5
12
162
21
0.111
64.3
11
148
8.9
0.356
71.8
12.8
195
17
Values are given in Pa units. Except for pressure_drop, pressure values are static pressure. WSS, Wall shear stress. *Zero pressure was assigned at the joint plane of the left
ventricular assist device graft and the ascending aorta. Therefore, if a lower pressure occurs at the aortic root, it is negative. yThe pressure drop was comparing the total pressure
difference between the aortic root and the descending aorta. If the pressure at the aortic root is lower than the pressure at the descending aorta, a negative pressure drop was
expected.
Pulsatile Flow Results
Figure 5 shows the average value of WSS at the aortic
root at various points in a cardiac cycle, for all pulsatile
cases. First, Model_1 (f ¼ 90 ; q ¼ 120 ; low anastomosis
site) produced a significantly higher WSS, compared with
model_2 (f ¼ 60 ; q ¼ 60 ; middle anastomosis site); the
same was true for the steady flow simulation. The WSS
values for the pulsatile condition are generally higher than
those for the steady flow condition, possibly because
of the almost-doubled LVAD temporal mean flow rate
(9.4 litre per minute vs 5 litre per minute), but are still
low compared to the physiological WSS range, especially
for the model_2.
Second, no noticeable difference was found in the aortic
root WSS value when the flow rate passing though the aortic
valve was increased. The aortic valve opening area may
have played a role in this result. In addition, the differences
in WSS between the 2 pulsatile index cases, as described
earlier, for each model are minor. Third, when the LVAD
flow rate curve (Figure 3) is compared with the WSS
variation (Figure 5), WSS can be seen to have more complex time variations during a cycle, compared with the input
flow rate variation. They do not peak at the same time.
DISCUSSION
The current work presents a systematic study of the
hemodynamic properties that have been hypothesized to
be associated with initiation of AVI after LVAD
implantation. The general flow features of WSS
distributions and flow patterns in the ascending aorta and
aortic arch are in agreement with the result of Inci and
colleagues.11 In the present work, more emphasis was put
on the aortic root region, especially the root WSS and
pressure given various anastomosis configurations. The
study demonstrated the significant influence of the type of
anastomosis design on the aortic root flow.
The Journal of Thoracic and Cardiovascular Surgery c Volume 150, Number 3
701
TX
The results indicate that more blood flow can be induced
in the aortic root when f 90 . A drawback of this
arrangement, however, is the increased flow resistance of
the main flow stream through the aortic arch, or the
increasing pressure drop from the aortic root to the
descending aorta. To quantify flow resistance, the difference
between the total pressure at the aortic root and the
descending aorta was calculated (Table 1). A higher
pressure drop means a higher flow resistance from the aortic
root to the descending aorta. In general, the flow resistance
increased with increasing inclination angle f.
The influence of the anastomosis location on the pressure
drop was smaller, compared with that of the inclination
angle change. Further analysis showed that anastomoses
displaying less-complex rotational flow structures produced
lower total pressure drops through the aorta, as
demonstrated in a case with a middle anastomosis location
of f ¼ 60 ; q ¼ 90 . The LVAD flow in this case was
directed almost entirely around the arch, resulting in
negative pressure in the root region (ie, negative
compared with the zero pressure at the LVAD inlet plane
set as a boundary condition), where the likelihood of
near-stagnant particles was high.
Cardiothoracic Transplantation and Mechanical Circulatory Support: Acquired
Callington et al
FIGURE 5. Average WSS variations in a cardiac cycle for pulsatile flow condition. Model_1: 4 ¼ 90 ; q ¼ 60 ; low anastomosis configuration. Model_2:
4 ¼ 60 ; q ¼ 60 ; middle anastomosis configuration. Pul_A: lower temporal mean flow rate but higher pulsatile index at LVAD cannula with aortic valve
open; Pul_B: higher temporal mean flow rate, low pulsatile index at LVAD cannula with a closed AV. WSS, Wall shear stress; AV, aortic valve; Pul_A,
medium strong pulsatile wave curve; Pul_B, weak pulsatile wave curve.
TX
A well-documented potential obstacle to the success of
long-term LVAD support is the native heart’s inability to
withstand the hemodynamic changes that result from
sustained mechanical assistance. A serious complication is
the AVI. Studies have indicated that one cause of AVI is a
closed aortic valve. LVADs, on the other hand, introduce
complex nonphysiologic hemodynamics into the aortic
root, which are directly or indirectly involved in the
instigation and development of AVI and several cardiovascular diseases.16-18 Flow patterns that exhibit disturbed
or turbulent characteristics, stagnation, recirculation, as
well as altered aortic WSS and pressure distributions, are
the main areas of investigation. WSS is an important
hemodynamic measure that is difficult to accurately
quantify using clinical techniques, such as magnetic
resonance imaging, alone.19,20 Low and/or oscillating WSS
distributions have been associated with the pathogenesis of
atherosclerosis.21 Studies have additionally shown that
nonphysiologic WSS could lead to arterial remodeling.22
Given its ability to resolve quantitative flow information,
the CFD technique has been used, in recent years, to
investigate the influence of an LVAD on aortic flow.7 The
most active group working in this area recently is Karmonik
and colleagues.8-10 In 2012, they published8 a CFD analysis
of blood flow in ascending aorta on data from 2 patients.
More recently, they increased the number of patients to 5
and analyzed the correlation between the flow pattern and
AVI9; in addition, they have studied the influence of
pulsatile and continuous flow LVAD on the flow patterns
702
in the ascending aorta.10 Their work demonstrated the
influence of anastomosis on blood flow in the ascending
aorta. However, with the very small number of patients,
drawing conclusions from this work is difficult. With regard
to development of AVI after LVAD implantation, some
researchers believe that the primary factor may be high
aortic root pressures caused by the jet flow.23-25 The
current study demonstrates that configurations with a high
inclination angle f will indeed produce higher aortic root
pressures. However, the highest root pressure value from
all simulation cases was 130 Pa, or 1 mm Hg, compared
with the anastomosis plane, where the pressure was
assigned to be zero in the CFD simulations. Thus, if the
diastolic blood pressure is 80 mm Hg at the anastomosis
location for a patient, the aortic root pressure will be
81 mm Hg, a minor aortic root pressure increase.
A clinically desirable setting of the LVAD device is at a
level that allows the aortic valve to open in systole
intermittently, and maintains the valve leaflet motion. In
practice, ensuring that this occurs is difficult, because of
the change in the patient’s condition. Kirmonik and
colleagues9 conducted a CFD simulation in which the aortic
valve was open during systole and had approximately 5%
of normal flow passing the valve. They found that this
transvalvular flow has only a minor impact on model
WSS distribution. Based on 5 case comparisons, they
suggested that an anastomosis design similar to that in our
model_2 (f ¼ 60 ; q ¼ 60 ; middle anastomosis site; as
in Figure 5) may be better because it produces the least
The Journal of Thoracic and Cardiovascular Surgery c September 2015
Cardiothoracic Transplantation and Mechanical Circulatory Support: Acquired
disturbance to the LVAD graft jet flow that flows smoothly
toward the aortic arch. However, the present study shows
that with a closed aortic valve and no transvalvular flow,
anastomosis setting as in model_2 would represent the
worst case for blood stagnation in the aortic root, according
to Table 1. Even with flow passing through the aortic valve
for 15% of a cycle duration, the WSS value at the aortic root
is still significantly smaller than the value with a low anastomosis, as in Figure 5. A long blood residential time in the
aortic root region can be expected in this graft
configuration, increasing the potential for thrombosis
formation and aortic wall remodeling, a potential risk factor
for development of AVI.
If LVAD-induced nonphysiologic hemodynamic
conditions in the aorta are the factors to be blamed for
AVI development, then maintaining a physiologic level of
WSS and pressure in the aortic root may delay aortic wall
remodeling and prevent thrombosis formation. The present
study indicated that few designs can be used to produce a
physiologic WSS (in the range of 0.35 to 1 Pa)26,27 in the
aortic root, even with a closed aortic valve. A low
anastomosis position can generally produce a WSS in that
range, in the root. If a high inclination angle (ie, 120 ) of
the guiding jet that is heading directly toward the valve
creates too violent a motion, then the low anastomosis
position f ¼ 90; may be a good choice. This position
produces a WSS_root of 0.78 to 1.7 Pa, and a moderate
aortic root pressure of 60 to 90 Pa. Higher anastomosis
locations will have difficulty creating an aortic root WSS
value that is in the physiologic range, unless a high
inclination angle is chosen.
This study has limitations, owing to several assumptions
that were made. First, the low anastomosis position may be
a bit too low in terms of actual surgical practice; the position
in practice may be a few millimeters higher than the
position used here. Second, we have applied relatively
simple treatment to the model boundary conditions, such
as a fixed flow rate ratio at all outflow vessels. Third, blood
viscosity value of normal blood was used in the CFD
simulation; LVAD patients are often subject to strict
anticoagulant regimes that would lower blood viscosity.
CONCLUSIONS
This study, which uses CFD, demonstrates that the
hemodynamic environment can be changed significantly
by varying the geometry of the outflow graft of LVADs.
Aortic root WSS can be altered from almost zero (blood
stagnation) to the normal range of blood flow in aorta,
with only a small rise in pressure in the root region. Among
all test scenarios, a low anastomosis position, an inclination
angle f ¼ 90 , and an azimuthal angel q ¼ 60 result in an
aortic root WSS in the normal range, and a moderate aortic
root pressure. More knowledge of aortic valve remodeling
may make possible the creation of a favorable flow pattern
in the aortic root, through optimization of surgical design to
prevent or reduce the occurrence of AVI.
Conflict of Interest Statement
Authors have nothing to disclose with regard to commercial
support.
References
1. McMurray JJ, Adamopoulos S, Anker SD, Auricchio A, B€ohm M, Dickstein K,
et al; ESC Committee for Practice Guidelines. ESC guidelines for the diagnosis
and treatment of acute and chronic heart failure 2012: The Task Force for the
Diagnosis and Treatment of Acute and Chronic Heart Failure 2012 of the European Society of Cardiology. Developed in collaboration with the Heart Failure
Association (HFA) of the ESC. Eur Heart J. 2012;33:1787-847.
2. George TJ, Arnaoutakis GJ, Shah AS. Surgical treatment of advanced heart
failure: alternatives to heart transplantation and mechanical circulatory assist
devices. Prog Cardiovasc Dis. 2011;54:115-31.
3. Komoda T, Drews T, Hetzer R, Lehmkuhl HB. New prioritization of heart transplant candidates on mechanical circulatory support in an era of severe donor
shortage. J Heart Lung Transplant. 2010;29:989-96.
4. Kirklin JK, Naftel DC, Kormos RL, Stevenson LW, Pagani FD, Miller MA, et al.
The Fourth INTERMACS Annual Report: 4,000 implants and counting. J Heart
Lung Transplant. 2012;31:117-26.
5. Rogers JG, Aaronson KD, Boyle AJ, Russell SD, Milano CA, Pagani FD, et al.
Continuous flow left ventricular assist device improves functional capacity and
quality of life of advanced heart failure patients. J Am Coll Cardiol. 2010;55:
1826-34.
6. Patil NP, Sabashnikov A, Mohite PN, Garcia D, Weymann A, Zych B, et al. De
novo aortic regurgitation after continuous-flow left ventricular assist device implantation. Ann Thorac Surg. 2014;98:850-7.
7. May-Newman K, Hillen B, Dembitsky W. Effect of left ventricular assist device
outflow conduit anastomosis location on flow patterns in the native aorta. ASAIO
J. 2006;52:132-9.
8. Karmonik C, Partovi S, Loebe M, Schmack B, Ghodsizad A, Robbin M, et al.
Influence of LVAD cannula outflow tract location on hemodynamics in the
ascending aorta: a patient-specific computational fluid dynamics approach.
ASAIO J. 2012;58:562-7.
9. Karmonik C, Partovi S, Loebe M, Schmack B, Weymann A, Lumsden AB, et al.
Computational fluid dynamics in patients with continuous-flow left ventricular
assist device support show hemodynamic alterations in the ascending aorta. J
Thorac Cardiovasc Surg. 2014;147:1326-33.e1.
10. Karmonik C, Partovi S, Schmack B, Weymann A, Loebe M, Noon GP, et al. Comparison of hemodynamics in the ascending aorta between pulsatile and continuous flow left ventricular assist devices using computational fluid dynamics
based on computed tomography images. Artif Organs. 2014;38:142-8.
11. Inci G, Sorg€uven E. Effect of LVAD outlet graft anastomosis angle on the aortic
valve, wall, and flow. ASAIO J. 2012;58:373-81.
12. Benk C, Mauch A, Beyersdorf F, Klemm R, Russe M, Blanke P, et al. Effect of
cannula position in the thoracic aorta with continuous left ventricular support:
four-dimensional flow-sensitive magnetic resonance imaging in an in vitro
model. Eur J Cardio-Thorac Surg. 2013;44:551-8.
13. Osorio AF, Osorio R, Ceballos A, Tran R, Clark W, Divo EA, et al. Computational fluid dynamics analysis of surgical adjustment of left ventricular assist device implantation to minimise stroke risk. Comput Methods Biomech Biomed
Engin. 2013;16:622-38.
14. Stainback RF, Croitoru M, Hernandez A, Myers TJ, Wadia Y, Frazier OH. Echocardiographic evaluation of the Jarvik 2000 Axial-Flow LVAD. Tex Heart Inst J.
2005;32:263-70.
15. Estep JD, Chang SM, Bhimaraj A, Torre-Amione G, Zoghbi WA, Nagueh SF. Imaging for ventricular function and myocardial recovery on nonpulsatile ventricular assist devices. Circulation. 2012;125:2265-77.
16. Davies PF, Remuzzitt A, Gordon EJ, Dewey CF, Gimbrone MA. Turbulent fluid
shear stress induces vascular endothelial cell turnover in vitro. Proc Natl Acad Sci
U S A. 1986;83:2114-7.
17. Moore JE, Xu C, Glagov S, Zarins CK, Ku DN. Fluid wall shear stress measurements in a model of the human abdominal aorta: oscillatory behavior and relationship to atherosclerosis. Atherosclerosis. 1994;110:225-40.
18. Stein PD, Sabbah HN. Measured turbulence and its effect on thrombus formation.
Circ Res. 1974;35:608-14.
The Journal of Thoracic and Cardiovascular Surgery c Volume 150, Number 3
703
TX
Callington et al
Cardiothoracic Transplantation and Mechanical Circulatory Support: Acquired
19. Boussel L, Rayz V, Martin A, Acevedo-Bolton G, Lawton MT, Higashida R, et al.
Phase-contrast magnetic resonance imaging measurements in intracranial aneurysms in vivo of flow patterns, velocity fields, and wall shear stress: comparison
with computational fluid dynamics. Magn Reson Med. 2009;61:409-17.
20. Petersson S, Dyverfeldt P, Ebbers T. Assessment of the accuracy of MRI wall
shear stress estimation using numerical simulations. J Magn Reson Imaging.
2012;36:128-38.
21. Davignon J, Ganz P. Role of endothelial dysfunction in atherosclerosis. Circulation. 2004;109(23 Suppl 1):III27-32.
22. Dolan JM, Sim FJ, Meng H, Kolega J. Endothelial cells express a unique transcriptional profile under very high wall shear stress known to induce expansive
arterial remodeling. Am J Physiol Cell Physiol. 2012;302:C1109-18.
23. Bryant AS, Holman WL, Nanda NC, Vengala S, Blood MS, Pamboukian SV,
et al. Native aortic valve insufficiency in patients with left ventricular assist devices. Ann Thorac Surg. 2006;81:e6-8.
24. Cowger J, Pagani F, Haft J, Romano MA, Aaronson K, Kolias T. The development of aortic insufficiency in left ventricular assist device-supported patients.
Circ Heart Fail. 2010;3:668-74.
Callington et al
25. Samuels LE, Thomas MP, Holmes EC, Narula J, Fitzpatrick J, Wood D,
et al. Insufficiency of the native aortic valve and left ventricular assist system inflow valve after support with an implantable left ventricular assist system: signs, symptoms, and concerns. J Thorac Cardiovasc Surg. 2001;122:
380-1.
26. Meierhofer C, Schneider EP, Lyko C, Hutter A, Martinoff S, Markl M, et al. Wall
shear stress and flow patterns in the ascending aorta in patients with bicuspid
aortic valves differ significantly from tricuspid aortic valves: a prospective study.
Eur Heart J Cardiovasc Imaging. 2013;14:797-804.
27. Ooij PV, Potters WV, Nederveen AJ, Collins JD, Carr JC, Malaisrie SC, et al.
Thoracic aortic wall shear stress atlases in patients with bicuspid aortic valves.
J Cardiovasc Magn Reson. 2014;16(Suppl 1):P161.
Key Words: left ventricle assistant device, LVAD anastomosis geometry, aortic insufficiency, computational fluid
dynamics, wall shear stress at aortic root
EDITORIAL COMMENTARY
Computational fluid dynamics: Solidifying fluid concepts in left
ventricular assist device therapy
Hari R. Mallidi, MD
See related article on pages 696-704.
The opinions that are held with passion are always
those for which no good ground exists; indeed the
passion is the measure of the holders lack of rational
conviction.
—Bertrand Russell, Sceptical Essays, 1928
The study of viscous fluids, predicting their behavior under
the influence of outside forces and constraints, is of everincreasing importance for the everyday cardiac surgeon,
and is especially important to the heart failure surgeon. To
understand the short- and long-term consequences of
TX
From the Department of Surgery, Division of Transplant and Assist Devices, Baylor
College of Medicine, Houston, Tex.
Disclosures: Author has nothing to disclose with regard to commercial support.
Received for publication May 28, 2015; accepted for publication June 3, 2015;
available ahead of print July 3, 2015.
Address for reprints: Hari R. Mallidi, MD, Department of Surgery, Division of
Transplant and Assist Devices, Baylor College of Medicine, One Baylor Plaza,
MS: BCM390, Houston, TX 77479 (E-mail: mallidi@bcm.edu).
J Thorac Cardiovasc Surg 2015;150:704-6
0022-5223/$36.00
Copyright Ó 2015 by The American Association for Thoracic Surgery
http://dx.doi.org/10.1016/j.jtcvs.2015.06.007
704
differing device designs and anastomotic factors (angle,
length, location in relation to structural components), and
the impact that conduit selection, design, size-matching,
and tailoring have on outcomes after cardiac surgery, requires an appreciation for the parameters that influence
blood flow.
The first description of the physiology of the circulatory
system, where the blood is described as a fluid pumped by
the heart, was made by William Harvey in 1628 in the Exercitatio Anatomica de Motu Cordis et Sanguinis in Animalibus. Theoretical advances in our understanding of blood
flow, and the ability to predict the impact of various
anatomic defects or structural lesions on blood flow, would
have to wait for developments in the fundamental understanding of fluid dynamics as ultimately described by the
Navier-Stokes equations.
The Navier-Stokes equations are mathematical relationships describing the behavior of viscous fluids, applying
the general principles of the conservation of mass, energy,
and momentum within the context of Newtonian dynamics. This set of nonlinear partial differential equations
describes a velocity field in a given volume of space. The
velocities at various points can be utilized to calculate
other useful characteristics of the fluid, such as pressure,
stress, or shear. Given that they are nonlinear equations,
the solutions are cumbersome to derive, and whether
The Journal of Thoracic and Cardiovascular Surgery c September 2015