Yan et al., 2014 - Google Patents

A 680 nA ECG acquisition IC for leadless pacemaker applications

Yan et al., 2014

View PDF
Document ID
5806422943372814231
Author
Yan L
Harpe P
Pamula V
Osawa M
Harada Y
Tamiya K
Van Hoof C
Yazicioglu R
Publication year
Publication venue
IEEE transactions on biomedical circuits and systems

External Links

Snippet

A sub-μW ECG acquisition IC is presented for a single-chamber leadless pacemaker applications. It integrates a low-power, wide dynamic-range ECG readout front end together with an analog QRS-complex extractor. To save ASIC power, a current-multiplexed channel …
Continue reading at research.tue.nl (PDF) (other versions)

Classifications

    • AHUMAN NECESSITIES
    • A61MEDICAL OR VETERINARY SCIENCE; HYGIENE
    • A61BDIAGNOSIS; SURGERY; IDENTIFICATION
    • A61B5/00Detecting, measuring or recording for diagnostic purposes; Identification of persons
    • A61B5/04Detecting, measuring or recording bioelectric signals of the body of parts thereof
    • A61B5/0402Electrocardiography, i.e. ECG
    • A61B5/0452Detecting specific parameters of the electrocardiograph cycle
    • AHUMAN NECESSITIES
    • A61MEDICAL OR VETERINARY SCIENCE; HYGIENE
    • A61BDIAGNOSIS; SURGERY; IDENTIFICATION
    • A61B5/00Detecting, measuring or recording for diagnostic purposes; Identification of persons
    • A61B5/72Signal processing specially adapted for physiological signals or for diagnostic purposes
    • A61B5/7235Details of waveform analysis
    • AHUMAN NECESSITIES
    • A61MEDICAL OR VETERINARY SCIENCE; HYGIENE
    • A61BDIAGNOSIS; SURGERY; IDENTIFICATION
    • A61B5/00Detecting, measuring or recording for diagnostic purposes; Identification of persons
    • A61B5/72Signal processing specially adapted for physiological signals or for diagnostic purposes
    • A61B5/7225Details of analog processing, e.g. isolation amplifier, gain or sensitivity adjustment, filtering, baseline or drift compensation
    • AHUMAN NECESSITIES
    • A61MEDICAL OR VETERINARY SCIENCE; HYGIENE
    • A61BDIAGNOSIS; SURGERY; IDENTIFICATION
    • A61B5/00Detecting, measuring or recording for diagnostic purposes; Identification of persons
    • A61B5/72Signal processing specially adapted for physiological signals or for diagnostic purposes
    • A61B5/7203Signal processing specially adapted for physiological signals or for diagnostic purposes for noise prevention, reduction or removal
    • HELECTRICITY
    • H03BASIC ELECTRONIC CIRCUITRY
    • H03FAMPLIFIERS
    • H03F3/00Amplifiers with only discharge tubes or only semiconductor devices as amplifying elements
    • H03F3/45Differential amplifiers
    • H03F3/45071Differential amplifiers with semiconductor devices only
    • H03F3/45076Differential amplifiers with semiconductor devices only characterised by the way of implementation of the active amplifying circuit in the differential amplifier
    • AHUMAN NECESSITIES
    • A61MEDICAL OR VETERINARY SCIENCE; HYGIENE
    • A61BDIAGNOSIS; SURGERY; IDENTIFICATION
    • A61B5/00Detecting, measuring or recording for diagnostic purposes; Identification of persons
    • A61B5/72Signal processing specially adapted for physiological signals or for diagnostic purposes
    • A61B5/7232Signal processing specially adapted for physiological signals or for diagnostic purposes involving compression of the physiological signal, e.g. to extend the signal recording period
    • AHUMAN NECESSITIES
    • A61MEDICAL OR VETERINARY SCIENCE; HYGIENE
    • A61BDIAGNOSIS; SURGERY; IDENTIFICATION
    • A61B5/00Detecting, measuring or recording for diagnostic purposes; Identification of persons
    • A61B5/04Detecting, measuring or recording bioelectric signals of the body of parts thereof
    • A61B5/0476Electroencephalography
    • AHUMAN NECESSITIES
    • A61MEDICAL OR VETERINARY SCIENCE; HYGIENE
    • A61BDIAGNOSIS; SURGERY; IDENTIFICATION
    • A61B5/00Detecting, measuring or recording for diagnostic purposes; Identification of persons
    • A61B5/04Detecting, measuring or recording bioelectric signals of the body of parts thereof
    • A61B5/04012Analysis of electro-cardiograms, electro-encephalograms, electro-myograms
    • AHUMAN NECESSITIES
    • A61MEDICAL OR VETERINARY SCIENCE; HYGIENE
    • A61BDIAGNOSIS; SURGERY; IDENTIFICATION
    • A61B5/00Detecting, measuring or recording for diagnostic purposes; Identification of persons
    • A61B5/145Measuring characteristics of blood in vivo, e.g. gas concentration, pH value; Measuring characteristics of body fluids or tissues, e.g. interstitial fluid, cerebral tissue
    • A61B5/1455Measuring characteristics of blood in vivo, e.g. gas concentration, pH value; Measuring characteristics of body fluids or tissues, e.g. interstitial fluid, cerebral tissue using optical sensors, e.g. spectral photometrical oximeters
    • A61B5/14551Measuring characteristics of blood in vivo, e.g. gas concentration, pH value; Measuring characteristics of body fluids or tissues, e.g. interstitial fluid, cerebral tissue using optical sensors, e.g. spectral photometrical oximeters for measuring blood gases
    • HELECTRICITY
    • H03BASIC ELECTRONIC CIRCUITRY
    • H03FAMPLIFIERS
    • H03F3/00Amplifiers with only discharge tubes or only semiconductor devices as amplifying elements
    • H03F3/45Differential amplifiers
    • H03F3/45071Differential amplifiers with semiconductor devices only
    • H03F3/45479Differential amplifiers with semiconductor devices only characterised by the way of common mode signal rejection
    • H03F3/45928Differential amplifiers with semiconductor devices only characterised by the way of common mode signal rejection using IC blocks as the active amplifying circuit
    • AHUMAN NECESSITIES
    • A61MEDICAL OR VETERINARY SCIENCE; HYGIENE
    • A61BDIAGNOSIS; SURGERY; IDENTIFICATION
    • A61B5/00Detecting, measuring or recording for diagnostic purposes; Identification of persons
    • A61B5/02Detecting, measuring or recording pulse, heart rate, blood pressure or blood flow; Combined pulse/heart-rate/blood pressure determination; Evaluating a cardiovascular condition not otherwise provided for, e.g. using combinations of techniques provided for in this group with electrocardiography or electroauscultation; Heart catheters for measuring blood pressure
    • AHUMAN NECESSITIES
    • A61MEDICAL OR VETERINARY SCIENCE; HYGIENE
    • A61BDIAGNOSIS; SURGERY; IDENTIFICATION
    • A61B5/00Detecting, measuring or recording for diagnostic purposes; Identification of persons
    • A61B5/05Detecting, measuring or recording for diagnosis by means of electric currents or magnetic fields; Measuring using microwaves or radiowaves
    • A61B5/053Measuring electrical impedance or conductance of a portion of the body
    • AHUMAN NECESSITIES
    • A61MEDICAL OR VETERINARY SCIENCE; HYGIENE
    • A61BDIAGNOSIS; SURGERY; IDENTIFICATION
    • A61B5/00Detecting, measuring or recording for diagnostic purposes; Identification of persons
    • A61B5/68Arrangements of detecting, measuring or recording means, e.g. sensors, in relation to patient
    • AHUMAN NECESSITIES
    • A61MEDICAL OR VETERINARY SCIENCE; HYGIENE
    • A61BDIAGNOSIS; SURGERY; IDENTIFICATION
    • A61B5/00Detecting, measuring or recording for diagnostic purposes; Identification of persons
    • A61B5/0002Remote monitoring of patients using telemetry, e.g. transmission of vital signals via a communication network

Similar Documents

Publication Publication Date Title
Yan et al. A 680 nA ECG acquisition IC for leadless pacemaker applications
Song et al. A 769 μW battery-powered single-chip SoC with BLE for multi-modal vital sign monitoring health patches
Kim et al. Sub-$\mu $ V rms-Noise Sub-$\mu $ W/Channel ADC-Direct Neural Recording With 200-mV/ms Transient Recovery Through Predictive Digital Autoranging
Xu et al. A 36 μW 1.1 mm 2 reconfigurable analog front-end for cardiovascular and respiratory signals recording
Van Helleputte et al. A 345 µW multi-sensor biomedical SoC with bio-impedance, 3-channel ECG, motion artifact reduction, and integrated DSP
Lee et al. An ultra-high input impedance analog front end using self-calibrated positive feedback
Da He et al. A 58 nW ECG ASIC with motion-tolerant heartbeat timing extraction for wearable cardiovascular monitoring
Yan et al. 24.4 A 680nA fully integrated implantable ECG-acquisition IC with analog feature extraction
Mondal et al. A dynamically reconfigurable ECG analog front-end with a 2.5× data-dependent power reduction
Agarwala et al. A 0.6 V 785-nW multimodal sensor interface IC for ozone pollutant sensing and correlated cardiovascular disease monitoring
Bai et al. A $64.8~\mu $ w> 2.2 g $\omega $ DC–AC configurable CMOS front-end IC for wearable ECG monitoring
Yan et al. Challenges of physiological signal measurements using electrodes: Fudamentals to understand the instrumentation
Saadeh et al. A 0.5 V PPG-based heart rate and variability detection system
Kim et al. Multi-channel biopotential acquisition system using frequency-division multiplexing with cable motion artifact suppression
Tang et al. Group-Chopping: An 8-Channel, 0.04% Gain Mismatch, 2.1 µW 0.017 mm 2 Instrumentation Amplifier for Bio-Potential Recording
Tian et al. A low-noise and low-power multi-channel ECG AFE based on orthogonal current-reuse amplifier
Yang et al. A 4-μW analog front end achieving 2.4 NEF for long-term ECG monitoring
Kim et al. Real time digitally assisted analog motion artifact reduction in ambulatory ECG monitoring system
CN117770826A (en) Wearable electrocardiosignal reading circuit
Rezaeiyan et al. A 4.5$\mu $ W Miniaturized 3-Channel Wireless Intra-Cardiac Acquisition System
Mahajan et al. A 64-channel ultra-low power bioelectric signal acquisition system for brain-computer interface
Pan et al. A 97.3-dB SNR Bioimpedance AFE With− 84-dB THD Segmented-ΔΣ M Sinusoidal Current Generator and Passing-Through Instrumentation Amplifier
Buaban et al. A low-power high-input-impedance ECG readout system employing a very high-gain amplification and a signal-folding technique for dry-electrode recording
Koo Design of low power and low noise instrumentation amplifier for biopotential acquisition: a review
Chebli et al. Low noise and high CMRR front-end amplifier dedicated to portable EEG acquisition system