Embodiment
Below by embodiment to SF of the present invention
6Gas density relay checking instrument is further described.Referring to Fig. 1, tester of the present invention comprises that the source of the gas of an adjustable in pressure provides mechanism 15, at least one pressure transducer 16, temperature sensor 17, computer data processing system 18, operation keyboard 20 and a display screen 19, and computer data processing system is provided with the relay action signal input port.
Referring to Fig. 2, Fig. 2 is SF of the present invention
6Source of the gas in the gas density relay checking instrument provides the structural representation of mechanism 15.It mainly comprises piston 1, wireway 21,22,23, valve 3, casing 4, air reservoir 5, pressure transducer 6, four-way seat 7, reverse checkvalve 8, O RunddichtringO 91,92, adjuster bar 10, handwheel 11, installing plate 12, pressure regulating cylinder 13.Air reservoir 5 is connected with an end of wireway 22, communicates on gas circuit; The other end of wireway 22 is connected with an end of valve 3; The other end of valve 3 is connected with an end of wireway 21; The other end of wireway 21 is connected with an interface of four-way seat 7; Another interface of four-way seat 7 is connected with an end of wireway 23 again; Pressure transducer 6 is fixed on the 3rd interface of four-way seat 7; Reverse checkvalve 8 is fixed on the 4th interface of four-way seat 7.The other end of wireway 23 is connected with pressure regulating cylinder 13; Piston 1 is arranged in the pressure regulating cylinder, and O- ring seal 91,92 is set in the groove of piston 1, and major function is that piston 1 seals when pressure is regulated in pressure regulating cylinder 13 motions; Piston 1 is movably connected with adjuster bar 10 again and connects; Adjuster bar 10 is connected again on the end cap of pressure regulating cylinder 13, can free activity regulate pressure; Handwheel 11 is movably connected with adjuster bar 10 and connects, and handwheel 11 can be movable, during use and adjuster bar 10 combine, the time spent can be not for convenience detach.Installing plate 12 is fixed on air reservoir 5 and the pressure regulating cylinder 13, and valve 3 and four-way seat 7 are fixed on the installing plate 12; Above element is installed in again in the casing 4.By the air reservoir of favorable sealing property, pressure regulating cylinder, valve, piston (studding with O-ring seal), wireway, pressure transducer, reverse checkvalve, adjuster bar, handwheel etc. have just been set up totally enclosed SF
6Gas-circulating system can make things convenient for, and is trickle, slowly, repetition, light adjusting pressure makes and does not waste SF in the checking procedure
6Gas.
When using at the scene, SF
6Gas connects an end and the SF of tracheae
6SF on the electrical equipment
6Gas density relay verification mouth is connected SF
6The other end that gas connects tracheae provides the reverse checkvalve 8 in the mechanism 15 to be connected with source of the gas, after connecting, observes force value, if pressure is not enough, illustrates that gas is not enough, just opens valve, the gas of air reservoir is released, till gas is enough tested.Valve-off just can be regulated the SF that is tested by regulating piston 1 then
6The pressure of gas density relay, just can the verification density monitor.Its operation key step is as follows:
1. the aviation plug with test article (relay) 14 checkpoint contacts sampling lead docks with the density monitor contact socket on the tester, and verification point sampling lead other end redness (a pair of), black (a pair of) and blue (a pair of) fish folder are connected with warning (a pair of), locking (a pair of) and locking 2 or superpressure (a pair of) contact of density monitor respectively.
2. if use the AC220V power work, connect external power source, the power-on switch; If use self-contained battery work, direct power-on switch, preheating 5 minutes.
3. enter by menu operation and prepare the verification screen.
4. whether observe force value meets the requirements.If force value is on the low side, slowly open air reservoir valve F again, after the observation force value has arrived desired force value, just close air reservoir valve F.
5. counter-clockwise direction is shaken handwheel then, reduces pressure, to during the ratings of the density monitor tested, and according to the prompting on the LCD screen, first verification ratings.And then according to the prompting on the LCD screen, counter-clockwise direction is shaken handwheel again, reduces pressure, and to during the locking value of the density monitor tested, clockwise direction is shaken handwheel again, and pressure boost is up to the warning rreturn value of the density monitor of being tested.
6. print or stores it according to screen prompt.
7. verification is opened air reservoir valve F after hours, shakes handwheel clockwise, till active on earth, closes air reservoir valve F then.This moment the SF in the pressure regulating cylinder
6Gas pushes back SF again
6Among the gas air reservoir.Throw off and connect tracheae.
8. turn off power switch, restore.
Below among each embodiment, the connection of all having omitted instrument stresses the correct acquisition of corresponding 20 ℃ of force value of measured relay, promptly how tester accurately measures absolute pressure relay and relative pressure density monitor.
Embodiment 1
At the SF that is tested
6Density monitor is absolute pressure relay or relative pressure relay, utilize man-machine communication, (man-machine conversation or artificial selection form can be versatile and flexible, and be unrestricted, for example can adopt external selector button, keyboard or a switch promptly to utilize man-machine interface dialogue or artificial selection; Also can adopt the keyboard on the liquid crystal, also can adopt and put question to the mode of answering, also can adopt the touch-screen on the touch LCD to operate or the like), confirm to instrument.
Adopt absolute pressure transducer to measure to the absolute pressure relay, obtain absolute pressure value, temperature value (obtaining) and SF during then according to test by temperature sensor
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value automatically, finishes this SF
6The performance verification of gas density relay.
This absolute pressure relay is if represent that with absolute pressure value its test result also is corresponding 20 ℃ of absolute pressure value; If represent that with the relative pressure value its test result also can be converted into corresponding 20 ℃ of relative pressure values, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Adopt relative pressure sensor to measure to the relative pressure relay, obtain the relative pressure value, temperature value and SF during then according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure values automatically, finishes this SF
6The performance verification of gas density relay.
This relative pressure relay is if represent that with the relative pressure value its test result also is corresponding 20 ℃ of relative pressure values; And if represent that with absolute pressure value its test result also can be converted into corresponding 20 ℃ of absolute pressure value, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relatively Pressure+ P
Standard atmospheric pressure
In the present embodiment, instrument can be according to the SF that is tested
6Gas density relay is absolute pressure relay or relative pressure relay, utilize man-machine communication, promptly utilize man-machine conversation or artificial selection, finish accurate test with corresponding, suitable pressure transducer respectively separately the density monitor performance of institute's verification.Can not influence measuring accuracy because of atmospheric influence.
Embodiment 2
Utilize the sensor of two kinds of measured pressure value methods, promptly relative pressure sensor and absolute pressure transducer apply same pressure (relative pressure can be 0) and measure, and obtain local atmospheric pressure P with this
Local air pressure=P
Absolute sensor pressure-P
Relative cell pressureAnd this local atmospheric pressure P
Local air pressureBe stored in the instrument.
At the SF that is tested
6Density monitor is absolute pressure relay or relative pressure relay, utilize man-machine communication, (man-machine conversation or artificial selection form can be versatile and flexible, and be unrestricted, for example can adopt external selector button, keyboard or a switch promptly to utilize man-machine conversation or artificial selection; Also can adopt the keyboard on the liquid crystal, also can adopt and put question to the mode of answering, also can adopt the touch-screen on the touch LCD to operate or the like), confirm to instrument.
All utilize the sensor of measuring absolute pressure to test to two types density monitors.
Adopt absolute pressure transducer to measure to the absolute pressure relay, obtain absolute pressure value, temperature value and SF during then according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value automatically, finishes this SF
6The performance verification of gas density relay.
This absolute pressure relay is if represent that with absolute pressure value its test result also is corresponding 20 ℃ of absolute pressure value; If represent that with the relative pressure value its test result also can be converted into corresponding 20 ℃ of relative pressure values, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Adopt absolute pressure transducer to measure to the relative pressure relay, obtain absolute pressure value, utilize the P that is stored in advance in the instrument then
Local air pressureRevise, promptly by relational expression P
The relative pressure of test=P
The absolute pressure of test-P
Local air pressureObtain the relative pressure value, temperature value and SF during again according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure values automatically, makes test result accurate, finishes this SF
6The performance verification of gas density relay.
To the absolute pressure value that the relative pressure relay adopts absolute pressure transducer to measure, can also handle according to the following steps: temperature value and SF during promptly earlier according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value P with the absolute pressure value that measures
The absolute pressure of 20 tests, utilize the P that is stored in advance in the instrument again
Local air pressureRevise, obtain corresponding 20 ℃ of relative pressure value: P
The relative pressure of 20 tests=P
The absolute pressure of 20 tests-P
Local air pressureFinish this SF
6The performance verification of gas density relay.
This relative pressure relay is if represent that with the relative pressure value its test result also is corresponding 20 ℃ of relative pressure values; If represent that with absolute pressure value its test result also can be converted into corresponding 20 ℃ of absolute pressure value, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Embodiment 3
Utilize the sensor of two kinds of measured pressure value methods, promptly relative pressure sensor and absolute pressure transducer apply same pressure (relative pressure can be 0) and measure, and obtain local atmospheric pressure P with this
Local air pressure=P
Absolute sensor pressure-P
Relative cell pressureAnd this local atmospheric pressure P
Local air pressureBe stored in the instrument.
At the SF that is tested
6Density monitor is absolute pressure relay or relative pressure relay, utilize man-machine communication, (man-machine conversation or artificial selection form can be versatile and flexible, and be unrestricted, for example can adopt external selector button, keyboard or a switch promptly to utilize man-machine conversation or artificial selection; Also can adopt the keyboard on the liquid crystal, also can adopt and put question to the mode of answering, also can adopt the touch-screen on the touch LCD to operate or the like), confirm to instrument.
All utilize the sensor of measuring relative pressure to test to two types density monitors.
Adopt relative pressure sensor to measure to the relative pressure relay, obtain the relative pressure value, temperature value and SF during then according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure values, finishes this SF
6The performance verification of gas density relay.
This relative pressure relay is if represent that with the relative pressure value its test result also is corresponding 20 ℃ of relative pressure values; If represent that with absolute pressure value its test result also can be converted into corresponding 20 ℃ of absolute pressure value, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Adopt relative pressure sensor to measure to the absolute pressure relay, obtain the relative pressure value, then with the P that is stored in advance in the instrument
Local air pressureRevise, promptly pass through relational expression: P
The absolute pressure of test=P
The relative pressure of test+ P
Local air pressure, obtain absolute pressure value, temperature value and SF during again according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value, makes test result accurate, finishes this SF
6The performance verification of gas density relay.
To the relative pressure value that the absolute pressure relay adopts relative pressure sensor to measure, can also handle according to the following steps: temperature value and SF during earlier according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure value: P
The relative pressure of 20 tests, utilize the P that is stored in advance in the instrument again
Local air pressureRevise, obtain corresponding 20 ℃ of absolute pressure value: P
The absolute pressure of 20 tests=P
The relative pressure of 20 tests+ P
Local air pressureThereby finish this SF
6The performance verification of gas density relay.
This absolute pressure relay is if represent that with absolute pressure value its test result also is corresponding 20 ℃ of absolute pressure value; And if represent that with the relative pressure value its test result also can be converted into corresponding 20 ℃ of relative pressure values, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Press relatively Power+ P
Standard atmospheric pressure
Embodiment 4
Is 0 o'clock with absolute pressure transducer at applied pressure, measures local atmospheric pressure: P
Local air pressure=P
Definitely pass this moment Sensor pressureAnd this local atmospheric pressure P
Local air pressureBe stored in the instrument.
At the SF that is tested
6Density monitor is absolute pressure relay or relative pressure relay, utilize man-machine communication, (man-machine conversation or artificial selection form can be versatile and flexible, and be unrestricted, for example can adopt external selector button, keyboard or a switch promptly to utilize man-machine conversation or artificial selection; Also can adopt the keyboard on the liquid crystal, also can adopt and put question to the mode of answering, also can adopt the touch-screen on the touch LCD to operate or the like), confirm to instrument.
All utilize the sensor of measuring absolute pressure to test to two types density monitors.
Adopt absolute pressure transducer to measure to the absolute pressure relay, obtain absolute pressure value, temperature value and SF during then according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value, finishes this SF
6The performance verification of gas density relay.
This absolute pressure relay is if represent that with absolute pressure value its test result also is corresponding 20 ℃ of absolute pressure value; If represent that with the relative pressure value its test result also can be converted into corresponding 20 ℃ of relative pressure values, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Adopt absolute pressure transducer to measure to the relative pressure relay, obtain absolute pressure value, utilize the P that is stored in advance in the instrument then
Local air pressureRevise, promptly pass through relational expression: P
The relative pressure of test=P
The absolute pressure of test-P
Local air pressure, obtain the relative pressure value, temperature value and SF during again according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure values, makes test result accurate, finishes this SF
6The performance verification of gas density relay.
To the absolute pressure value that the relative pressure relay adopts absolute pressure transducer to measure, can also handle according to the following steps: temperature value and SF during promptly earlier according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value P with the absolute pressure value that measures
The absolute pressure of 20 tests, utilize the P that is stored in advance in the instrument again
Local air pressureRevise, obtain corresponding 20 ℃ of relative pressure value: P
The relative pressure of 20 tests=P
The absolute pressure of 20 tests-P
Local air pressureFinish this SF
6The performance verification of gas density relay.
This relative pressure relay is if represent that with the relative pressure value its test result also is corresponding 20 ℃ of relative pressure values; If represent that with absolute pressure value its test result also can be converted into corresponding 20 ℃ of absolute pressure value, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Embodiment 5
By man-machine communication, promptly utilize man-machine conversation, earlier the local atmospheric pressure P of input
Local air pressureAnd this local atmospheric pressure P
Local air pressureBe stored in the instrument.
At the SF that is tested
6Density monitor is absolute pressure relay or relative pressure relay, utilize man-machine communication, (man-machine conversation or artificial selection form can be versatile and flexible, and be unrestricted, for example can adopt external selector button, keyboard or a switch promptly to utilize man-machine conversation or artificial selection; Also can adopt the keyboard on the liquid crystal, also can adopt and put question to the mode of answering, also can adopt the touch-screen on the touch LCD to operate or the like), confirm to instrument.
All utilize the sensor of measuring absolute pressure to test to two types density monitors.
Adopt absolute pressure transducer to measure to the absolute pressure relay, obtain absolute pressure value, temperature value and SF during then according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value, finishes this SF
6The performance verification of gas density relay.
This absolute pressure relay is if represent that with absolute pressure value its test result also is corresponding 20 ℃ of absolute pressure value; If represent that with the relative pressure value its test result also can be converted into corresponding 20 ℃ of relative pressure values, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Adopt absolute pressure transducer to measure to the relative pressure relay, obtain absolute pressure value, utilize the P that is stored in advance in the instrument then
Local air pressureRevise, promptly pass through relational expression: P
The relative pressure of test=P
The absolute pressure of test-P
Local air pressure, obtain the relative pressure value, temperature value and SF during again according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure values, makes test result accurate, finishes this SF
6The performance verification of gas density relay.
To the absolute pressure value that the relative pressure relay adopts absolute pressure transducer to measure, can also handle according to the following steps: temperature value and SF during promptly earlier according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value P with the absolute pressure value that measures
The absolute pressure of 20 tests, utilize the P that is stored in advance in the instrument again
Local air pressureRevise, obtain corresponding 20 ℃ of relative pressure value: P
The relative pressure of 20 tests=P
The absolute pressure of 20 tests-P
Local air pressureFinish this SF
6The performance verification of gas density relay.
This relative pressure relay is if represent that with the relative pressure value its test result also is corresponding 20 ℃ of relative pressure values; If represent that with absolute pressure value its test result also can be converted into corresponding 20 ℃ of absolute pressure value, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Embodiment 6
Utilize absolute pressure transducer and external relative pressure table meter, apply same pressure survey and go out local atmospheric pressure: P
Local air pressure=P
Absolute sensor pressure-P
The relative pressure gauge pressureBy man-machine conversation, import local atmospheric pressure P then
Local air pressureAnd this local atmospheric pressure P
Local air pressureBe stored in the instrument; Perhaps utilize man-machine conversation directly to import P
The relative pressure gauge pressure, calculate local atmospheric pressure P automatically by instrument
Local air pressure=P
Absolute sensor pressure-P
Relative pressure Gauge pressure, and this local atmospheric pressure P
Local air pressureBe stored in the instrument.
At the SF that is tested
6Density monitor is absolute pressure relay or relative pressure relay, utilize man-machine communication, (man-machine conversation or artificial selection form can be versatile and flexible, and be unrestricted, for example can adopt external selector button, keyboard or a switch promptly to utilize man-machine conversation or artificial selection; Also can adopt the keyboard on the liquid crystal, also can adopt and put question to the mode of answering, also can adopt the touch-screen on the touch LCD to operate or the like), confirm to instrument.
All utilize the sensor of measuring absolute pressure to test to two types density monitors.
Adopt absolute pressure transducer to measure to the absolute pressure relay, obtain absolute pressure value, temperature value and SF during then according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value, finishes this SF
6The performance verification of gas density relay.
This absolute pressure relay is if represent that with absolute pressure value its test result also is corresponding 20 ℃ of absolute pressure value; If represent that with the relative pressure value its test result also can be converted into corresponding 20 ℃ of relative pressure values, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Adopt absolute pressure transducer to measure to the relative pressure relay, obtain absolute pressure value, utilize the P that is stored in advance in the instrument then
Local air pressureRevise, promptly pass through relational expression: P
The relative pressure of test=P
The absolute pressure of test-P
Local air pressure, obtain the relative pressure value, temperature value and SF during again according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure values, makes test result accurate, finishes this SF
6The performance verification of gas density relay.
To the absolute pressure value that the relative pressure relay adopts absolute pressure transducer to measure, can also handle according to the following steps: temperature value and SF during promptly earlier according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value P with the absolute pressure value that measures
The absolute pressure of 20 tests, utilize the P that is stored in advance in the instrument again
Local air pressureRevise, obtain corresponding 20 ℃ of relative pressure value: P
The relative pressure of 20 tests=P
The absolute pressure of 20 tests-P
Local air pressureFinish this SF
6The performance verification of gas density relay.
This relative pressure relay is if represent that with the relative pressure value its test result also is corresponding 20 ℃ of relative pressure values; If represent that with absolute pressure value its test result also can be converted into corresponding 20 ℃ of absolute pressure value, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Embodiment method 7
According to different areas, instrument is in advance the atmospheric pressure P of locality
Local air pressureBe stored in the instrument.
At the SF that is tested
6Density monitor is absolute pressure relay or relative pressure relay, utilize man-machine communication, (man-machine conversation or artificial selection form can be versatile and flexible, and be unrestricted, for example can adopt external selector button, keyboard or a switch promptly to utilize man-machine conversation or artificial selection; Also can adopt the keyboard on the liquid crystal, also can adopt and put question to the mode of answering, also can adopt the touch-screen on the touch LCD to operate or the like), confirm to instrument.
All utilize the sensor of measuring absolute pressure to test to two types density monitors.
Adopt absolute pressure transducer to measure to the absolute pressure relay, obtain absolute pressure value, temperature value and SF during then according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value, finishes this SF
6The performance verification of gas density relay.
This absolute pressure relay is if represent that with absolute pressure value its test result also is corresponding 20 ℃ of absolute pressure value; If represent that with the relative pressure value its test result also can be converted into corresponding 20 ℃ of relative pressure values, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Adopt absolute pressure transducer to measure to the relative pressure relay, obtain absolute pressure value, utilize the P that is stored in advance in the instrument then
Local air pressureRevise, promptly pass through relational expression: P
The relative pressure of test=P
The absolute pressure of test-P
Local air pressure, obtain the relative pressure value, temperature value and SF during again according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure values, makes test result accurate, finishes this SF
6The performance verification of gas density relay.
To the absolute pressure value that the relative pressure relay adopts absolute pressure transducer to measure, can also handle according to the following steps: temperature value and SF during promptly earlier according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value P with the absolute pressure value that measures
The absolute pressure of 20 tests, utilize the P that is stored in advance in the instrument again
Local air pressureRevise, obtain corresponding 20 ℃ of relative pressure value: P
The relative pressure of 20 tests=P
The absolute pressure of 20 tests-P
Local air pressureFinish this SF
6The performance verification of gas density relay.
This relative pressure relay is if represent that with the relative pressure value its test result also is corresponding 20 ℃ of relative pressure values; If represent that with absolute pressure value its test result also can be converted into corresponding 20 ℃ of absolute pressure value, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Embodiment 8
Apply same pressure with relative pressure sensor and external absolute pressure gauge meter, measure local atmospheric pressure: P
Local air pressure=P
The absolute pressure gauge pressure-P
Relative cell pressureBy man-machine conversation, import local atmospheric pressure: P then
Local air pressureAnd this local atmospheric pressure P
Local air pressureBe stored in the instrument; Perhaps directly import P by man-machine conversation
The absolute pressure gauge pressure, calculate local atmospheric pressure: P automatically by instrument
Local air pressure=P
The absolute pressure gauge pressure-P
Relatively Cell pressure, and this local atmospheric pressure P
Local air pressureBe stored in the instrument.
Can be 0 also, measure local atmospheric pressure: P with the absolute pressure gauge instrumentation by applied pressure
Work as ground vapour Press=P
The absolute pressure gauge pressureAnd this local atmospheric pressure P
Local air pressureBe stored in the instrument.
At the SF that is tested
6Density monitor is absolute pressure relay or relative pressure relay, utilize man-machine communication, (man-machine conversation or artificial selection form can be versatile and flexible, and be unrestricted, for example can adopt external selector button, keyboard or a switch promptly to utilize man-machine conversation or artificial selection; Also can adopt the keyboard on the liquid crystal, also can adopt and put question to the mode of answering, also can adopt the touch-screen on the touch LCD to operate or the like), confirm to instrument.
All utilize the sensor of measuring relative pressure to test to two types density monitors.
Adopt relative pressure sensor to measure to the relative pressure relay, obtain the relative pressure value, temperature value and SF during then according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure values, finishes this SF
6The performance verification of gas density relay.
This relative pressure relay is if represent that with the relative pressure value its test result also is corresponding 20 ℃ of relative pressure values; If represent that with absolute pressure value its test result also can be converted into corresponding 20 ℃ of absolute pressure value, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Adopt relative pressure sensor to measure to the absolute pressure relay, obtain the relative pressure value, then with the P that is stored in advance in the instrument
Local air pressureRevise, promptly pass through relational expression: P
The absolute pressure of test=P
The relative pressure of test+ P
Local air pressure, obtain absolute pressure value, temperature value and SF during again according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value, makes test result accurate, finishes this SF
6The performance verification of gas density relay.
To the relative pressure value that the absolute pressure relay adopts relative pressure sensor to measure, can also handle according to the following steps: temperature value and SF during earlier according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure value: P
The relative pressure of 20 tests, utilize the P that is stored in advance in the instrument again
Local air pressureRevise, obtain corresponding 20 ℃ of absolute pressure value: P
The absolute pressure of 20 tests=P
The relative pressure of 20 tests+ P
Local air pressureThereby finish this SF
6The performance verification of gas density relay.
This absolute pressure relay is if represent that with absolute pressure value its test result also is corresponding 20 ℃ of absolute pressure value; And if represent that with the relative pressure value its test result also can be converted into corresponding 20 ℃ of relative pressure values, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Press relatively Power+ P
Standard atmospheric pressure
Embodiment 9
By man-machine conversation, import local atmospheric pressure: P earlier
Local air pressureAnd this local atmospheric pressure P
Local Air pressureBe stored in the instrument.
At the SF that is tested
6Density monitor is absolute pressure relay or relative pressure relay, utilize man-machine communication, (man-machine conversation or artificial selection form can be versatile and flexible, and be unrestricted, for example can adopt external selector button, keyboard or a switch promptly to utilize man-machine conversation or artificial selection; Also can adopt the keyboard on the liquid crystal, also can adopt and put question to the mode of answering, also can adopt the touch-screen on the touch LCD to operate or the like), confirm to instrument.
All utilize the sensor of measuring relative pressure to test to two types density monitors.
Adopt relative pressure sensor to measure to the relative pressure relay, obtain the relative pressure value, temperature value and SF during then according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure values, finishes this SF
6The performance verification of gas density relay.
This relative pressure relay is if represent that with the relative pressure value its test result also is corresponding 20 ℃ of relative pressure values; If represent that with absolute pressure value its test result also can be converted into corresponding 20 ℃ of absolute pressure value, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Adopt relative pressure sensor to measure to the absolute pressure relay, obtain the relative pressure value, then with the P that is stored in advance in the instrument
Local air pressureRevise, promptly pass through relational expression: P
The absolute pressure of test=P
The relative pressure of test+ P
Local air pressure, obtain absolute pressure value, temperature value and SF during again according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value, makes test result accurate, finishes this SF
6The performance verification of gas density relay.
To the relative pressure value that the absolute pressure relay adopts relative pressure sensor to measure, can also handle according to the following steps: temperature value and SF during earlier according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure value: P
The relative pressure of 20 tests, utilize the P that is stored in advance in the instrument again
Local air pressureRevise, obtain corresponding 20 ℃ of absolute pressure value: P
The absolute pressure of 20 tests=P
The relative pressure of 20 tests+ P
Local air pressureThereby finish this SF
6The performance verification of gas density relay.
This absolute pressure relay is if represent that with absolute pressure value its test result also is corresponding 20 ℃ of absolute pressure value; And if represent that with the relative pressure value its test result also can be converted into corresponding 20 ℃ of relative pressure values, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Press relatively Power+ P
Standard atmospheric pressure
Embodiment 10
According to different areas, instrument is in advance the atmospheric pressure of locality: P
Local air pressureBe stored in the instrument.
At the SF that is tested
6Density monitor is absolute pressure relay or relative pressure relay, utilize man-machine communication, (man-machine conversation or artificial selection form can be versatile and flexible, and be unrestricted, for example can adopt external selector button, keyboard or a switch promptly to utilize man-machine conversation or artificial selection; Also can adopt the keyboard on the liquid crystal, also can adopt and put question to the mode of answering, also can adopt the touch-screen on the touch LCD to operate or the like), confirm to instrument.
All utilize the sensor of measuring relative pressure to test to two types density monitors.
Adopt relative pressure sensor to measure to the relative pressure relay, obtain the relative pressure value, temperature value and SF during then according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure values, finishes this SF
6The performance verification of gas density relay.
This relative pressure relay is if represent that with the relative pressure value its test result also is corresponding 20 ℃ of relative pressure values; If represent that with absolute pressure value its test result also can be converted into corresponding 20 ℃ of absolute pressure value, the conversion relation between absolute pressure and relative pressure is: P
Absolute pressure=P
Relative pressure+ P
Standard atmospheric pressure
Adopt relative pressure sensor to measure to the absolute pressure relay, obtain the relative pressure value, then with the P that is stored in advance in the instrument
Local air pressureRevise, promptly pass through relational expression: P
The absolute pressure of test=P
The relative pressure of test+ P
Local air pressure, obtain absolute pressure value, temperature value and SF during again according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of absolute pressure value, makes test result accurate, finishes this SF
6The performance verification of gas density relay.
To the relative pressure value that the absolute pressure relay adopts relative pressure sensor to measure, can also handle according to the following steps: temperature value and SF during earlier according to test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of relative pressure value: P
The relative pressure of 20 tests, utilize the P that is stored in advance in the instrument again
Local air pressureRevise, obtain corresponding 20 ℃ of absolute pressure value: P
The absolute pressure of 20 tests=P
The relative pressure of 20 tests+ P
Local air pressureThereby finish this SF
6The performance verification of gas density relay.
This absolute pressure relay is if represent that with absolute pressure value its test result also is corresponding 20 ℃ of absolute pressure value; And if represent that with the relative pressure value its test result also can be converted into corresponding 20 ℃ of relative pressure values, the conversion relation between absolute pressure and relative pressure is: P absolute pressure=P relative pressure+P standard atmospheric pressure.
The principal feature of the method for calibration of SF6 gas density relay of the present invention is:
1, by instrument and tester's communication, can be during test to the SF that is tested
6Density monitor carries out work (absolute pressure relay) with measuring the absolute pressure value method, and still (the relative pressure relay) that carries out work with measurement relative pressure value method selected or confirmed, can tell the SF that instrument is tested
6The type of density monitor.
2, by the test of instrument own directly or indirectly obtain the local atmospheric pressure of local atmospheric pressure or input or in advance the local atmosphere of storage be pressed onto on the instrument, and make and store local atmospheric pressure on the instrument.
3, can be according to the SF that is tested
6The type of gas density relay, according to the force value method of testing (absolute pressure transducer or relative pressure sensor) of the sensor that adopted, and local atmospheric pressure, revise accordingly, make the force value test result accurate, can not influence measuring accuracy because of atmospheric influence.
4, can be according to the temperature value and the SF in when test
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of force value automatically, finishes SF
6The accurate test of gas density relay performance; Perhaps can be according to the SF that is tested
6The force value method of testing of gas density relay, and the temperature value can be according to test the time are then according to SF
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of force value automatically, again according to the force value method of testing of the sensor that adopted, and local atmospheric pressure, revise accordingly again and obtain 20 ℃ of force value accurately, make test result accurate, can be because of atmospheric influence does not influence measuring accuracy, thus finish SF
6The accurate test of gas density relay performance.
5, by instrument and tester's communication, can be during test to the SF that is tested
6Density monitor carries out work (absolute pressure relay) with measuring the absolute pressure value method, and still (the relative pressure relay) that carries out work with measurement relative pressure value method selected or confirmed, can tell the SF that instrument is tested
6The method of testing of density monitor is promptly with measuring that the absolute pressure value method is carried out work or carrying out work with measurement relative pressure value method; Can be according to the SF that is tested
6The method of testing of gas density relay selects corresponding sensor to measure, and makes the force value test result accurate, can not influence measuring accuracy because of atmospheric influence.And the temperature value can be according to test the time, then according to SF
6The pressure-temperature characteristic relation of gas is converted to corresponding 20 ℃ of force value automatically, finishes SF
6The accurate test of gas density relay performance.
6, instrument and tester's communication way can be versatile and flexible, unrestricted, for example can adopt an external selector button, selector switch, keyboard, also can adopt keyboard on the liquid crystal, also can adopt the touch-screen on the touch LCD to operate, also can utilize man-machine conversation or select and link up or the like.
7, to the absolute pressure relay if represent that with absolute pressure value its test result also is corresponding 20 ℃ of absolute pressure value; If represent that with the relative pressure value its test result also can be corresponding 20 ℃ of relative pressure values.
8, to the relative pressure relay if represent that with absolute pressure value its test result also can be corresponding 20 ℃ of absolute pressure value; If represent that with the relative pressure value its test result also is corresponding 20 ℃ of relative pressure values.
9, instrument can adopt the pressure transducer (absolute pressure transducer) of measuring absolute pressure value and the pressure transducer (relative pressure sensor) of measuring the relative pressure value to measure respectively.
10, instrument can all adopt the pressure transducer (absolute pressure transducer) of measuring absolute pressure value to measure.
11, instrument can all adopt the pressure transducer (relative pressure sensor) of measuring the relative pressure value to measure.
Those of ordinary skill in the art will be appreciated that, above embodiment is used for illustrating the present invention, and be not to be used as limitation of the invention, as long as in connotation scope of the present invention, and all will drop in claims scope of the present invention variation, the modification of the above embodiment.