Abstract
This paper describes a precision-guided surgical navigation system for minimally invasive surgery using fusion of laser guidance technique and three-dimensional (3-D) autostereoscopic image overlay technique. The images superimposed onto the patient are created by employing an animated autostereoscopic image called integral videography (IV), which display geometrically accurate 3-D autostereoscopic images and reproduce motion parallax without the need for special viewing or tracking devices. To improve the insertion accuracy of surgical instrument, we integrated the image overlay system with laser guidance for visualization of insertion point and orientation of the surgical instrument. We designed and manufactured a laser guidance device and mounted it to the IV image overlay device. Accuracy evaluations showed that the system could guide a linear surgical instrument toward a target with an average error of 2.48 mm and standard deviation of 1.76 mm. Improvement in the design of the laser guidance device and the patient-image registration of the IV image overlay will make this system practical and its use will increase surgical accuracy and reduce invasiveness.
Preview
Unable to display preview. Download preview PDF.
Similar content being viewed by others
References
Rosen, M., Ponsky, J.: Minimally Invasive Surgery. Endoscopy 33, 358–366 (2001)
Breedveld, P., Wentink, M.: Eye-hand coordination in laparoscopy - an overview of experiments and supporting aids. Minim. Invasiv. Ther. 10(3), 155–162 (2001)
Breedveld, P., Stassen, H.G., Meijer, D.W., Stassen, L.P.S.: Theoretical background and conceptual solution for depth perception and eye-hand coordination problems in laparoscopic surgery. Minim. Invasiv. Ther. 8, 227–234 (1999)
Stetten, G.D., Clib, V.S.: Overlaying Ultrasonographic Images on Direct Vision. J. Ultrasound Med. 20, 235–240 (2001)
Rosenthal, M., State, A., Lee, J., Hirota, G., Ackerman, J., Keller, K., Pisano, E.D., Jiroutek, M., Muller, K., Fuchs, H.: Augmented reality guidance for needle biopsies: An initial randomized, controlled trial in phantoms. Medical Image Analysis 6, 313–320 (2002)
Fichtinger, G., Deguet, A., Masamune, K., Balogh, E., Fischer, G.S., Mathieu, H., Taylor, R.H., Zinreich, S.J., Fayad, L.M.: Image Overlay Guidance for Needle Insertion in CT Scanner. IEEE Trans. Biomedical Engineering 52(8), 1415–1424 (2005)
Edwards, P.J., King, A.P., Maurer Jr., C.R., Cunha, D.A., Hawkes, D.J., Hill, D.L.G., Gaston, R.P., Fenlon, M.R., Jusczyzck, A., Strong, A.J., Chandler, C.L., Gleeson, M.J.: Design and evaluation of a system for microscope-assisted guided interventions (MAGI). IEEE Trans. Med. Imag. 19, 1082–1093 (2000)
Shahidi, R., Bax, M.R., Maurer, C.R., Johnson, J.A., Wilkinson, E.P., Wang, B., West, J.B., Citardi, M.J., Manwaring, K.H., Khadem, R.: Implementation, calibration and accuracy testing of an image-enhanced endoscopy system. IEEE Trans. Med. Imag. 21, 1524–1535 (2002)
Sasama, T., Sugano, N., Sato, Y., Momoi, Y., Nakajima, Y., Koyama, T., Nakajima, Y., Sakuma, I., Fujie, M., Yonenobu, K., Ochi, T., Tamura, S.: A Novel Laser Guidance System for Alignment of Linear Surgical Tool: Its Principles and Performance Evaluation as a Man-Machine System. In: Dohi, T., Kikinis, R. (eds.) MICCAI 2002. LNCS, vol. 2489, pp. 125–132. Springer, Heidelberg (2002)
Nakajima, Y., Yamamoto, H., Sato, Y., Sugano, N., Momoi, Y., Sasama, T., Koyama, T., Tamura, T., Yonenobu, K., Sakuma, I., Yoshikawa, H., Ochi, T., Tamura, S.: Available range analysis of laser guidance system and its application to monolithic integration with optical tracker. In: CARS 2004, pp. 449–454 (2004)
Liao, H., Hata, N., Nakajima, S., Iwahara, M., Sakuma, I., Dohi, T.: Surgical Navigation by Autostereoscopic Image Overlay of Integral Videography. IEEE Transactions on Information Technology in Biomedicine 8(2), 114–121 (2004)
Liao, H., Iwahara, M., Hata, N., Dohi, T.: High-quality integral videography using a multi-projector. Optics Express 12(6), 1067–1076 (2004)
Lippmann, M.G.: Epreuves reversibles donnant la sensation du relief. J. de Phys. 7(4th series), 821–825 (1908)
Liao, H., Matsui, K., Dohi, T.: Design and Evaluation of Surgical Navigation for Anterior Cruciate Ligament Reconstruction using Autostereoscopic Image Overlay of Integral Videography. In: 27th Annual International Conference of the IEEE Engineering In Medicine and Biology Society (EMBS) 2005. CD-ROM (September 2005)
Liao, H., Inomata, T., Sakuma, I., Dohi, T.: Surgical Navigation of Integral Videography Image Overlay for Open MRI-Guided Glioma Surgery. In: Yang, G.-Z., Jiang, T., Shen, D., Gu, L., Yang, J. (eds.) MIAR 2006. LNCS, vol. 4091, pp. 187–194. Springer, Heidelberg (2006)
Author information
Authors and Affiliations
Editor information
Rights and permissions
Copyright information
© 2008 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Liao, H., Ishihara, H., Tran, H.H., Masamune, K., Sakuma, I., Dohi, T. (2008). Fusion of Laser Guidance and 3-D Autostereoscopic Image Overlay for Precision-Guided Surgery. In: Dohi, T., Sakuma, I., Liao, H. (eds) Medical Imaging and Augmented Reality. MIAR 2008. Lecture Notes in Computer Science, vol 5128. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-540-79982-5_40
Download citation
DOI: https://doi.org/10.1007/978-3-540-79982-5_40
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-540-79981-8
Online ISBN: 978-3-540-79982-5
eBook Packages: Computer ScienceComputer Science (R0)