Abstract
Purpose
The accuracy of 4D-CT registration is limited by inconsistent Hounsfield unit (HU) values in the 4D-CT data from one respiratory phase to another and lower image contrast for lung substructures. This paper presents an optical flow and thin-plate spline (TPS)-based 4D-CT registration method to account for these limitations.
Methods
The use of unified HU values on multiple anatomy levels (e.g., the lung contour, blood vessels, and parenchyma) accounts for registration errors by inconsistent landmark HU value. While 3D multi-resolution optical flow analysis registers each anatomical level, TPS is employed for propagating the results from one anatomical level to another ultimately leading to the 4D-CT registration. 4D-CT registration was validated using target registration error (TRE), inverse consistency error (ICE) metrics, and a statistical image comparison using Gamma criteria of 1 % intensity difference in \(2\,\hbox {mm}^{3}\) window range.
Results
Validation results showed that the proposed method was able to register CT lung datasets with TRE and ICE values \(<\)3 mm. In addition, the average number of voxel that failed the Gamma criteria was \(<\)3 %, which supports the clinical applicability of the propose registration mechanism.
Conclusion
The proposed 4D-CT registration computes the volumetric lung deformations within clinically viable accuracy.
Similar content being viewed by others
References
Keall PJ, Mageras GS, Balter JM, Emery RS, Forster KM, Jiang SB, Kapatoes JM, Low DA, Murphy MJ, Murray BR, Ramsey CR, van Herk MB, Vedam SS, Wong JW, Yorke E (2006) The management of respiratory motion in radiation oncology report of AAPM Task Group 76. Med Phys 33(10):3874–3900
Bortfeld T, Jiang SB, Rietzel E (2004) Effects of motion on the total dose distribution. Semin Radiat Oncol 14(1):41–51
Starkschall G, Britton K, McAleer MF, Jeter MD, Kaus MR, Bzdusek K, Mohan R, Cox JD (2009) Potential dosimetry benefits of four-dimensional radiation treatment planning. Med Phys 73(5):1560–1565
Vinogradskiy YY, Balter P, Followill DS, Alvarez PE, White RA, Starkschall G (2009) Comparing the accuracy of 4D photon dose calculation with 3D calculating using moving and deforming phantoms. Med Phys 36(11):5000–5006
Vinogradskiy YY, Balter P, Followill DS, Alvarez PE, White RA, Starkschall G (2009) Verification of four dimensional dose calculations. Med Phys 36(8):3438–3447
McClelland JR, Hughes S, Modat M, Qureshi A, Ahmad S, Landau D, Ourselin S, Hawkes DJ (2011) Inter-fraction variations in respiratory motion models. Phys Med Biol 56(2011):251–272
Metaxas DN (1996) Physics-based deformable models: applications to computer vision, graphics, and medical imaging, 1st edn. Kluwer Academic Publishers Norwell, MA
Santhanam AP, Willoughby T, Shah A, Meeks SL, Rolland JP, Kupelian P (2008) Real-time simulation of 4D lung tumor radiotherapy using a breathing model. Lect Notes Comput Sci 2:710–717
Klinder T, Lorenz C, von Berg J, Renisch S, Blaffert T, Ostermann J (2008) 4DCT image-based motion field extraction and analysis. In: SPIE Medical, Imaging. SPIE
Guerrero T, Sanders K, Castillo E, Zhang Y, Bidaut L, Pan T, Komaki R (2006) Dynamic ventilation imaging from four-dimensional computed tomography. Phys Med Biol 51(4):777–791
Zhang GG, Huang T-C, Guerrero T, Lin K-P, Stevens C, Starkschall G, Forster K (2008) Use of three-dimensional optical flow method in mapping 3D anatomic structure and tumor contours across four-dimensional computed tomography data. J Appl Clin Med Phys 9(1):2738
Brock KK (2007) A multi-institution deformable registration accuracy study. Int J Radiat Oncol Biol Phys 69(3):S44
Sarrut D, Boldea V, Ayadi M, Badel J-N, Gineset C, Clippe S, Carrie C (2005) Non-rigid registration method to assess reproducibility of breathing-holding with ABC in lung cancer. Int J Radiat Oncol Biol Phys 61(2):594–607
Murphy K, van Ginneken B, Pluim JPW, Klien S, Staring M (2008) Quantitative assessment of registration in thoracic CT. In: First international workshop on pulmonary image processing. pp 203–211
Betke M, Hong H, Thomas D, Chekema P, Ko JP (2003) Landmark detection in the chest and registration of lung surfaces with an application to nodule registration. Med Image Anal 7:265–281
Yin Y, Hoffman EA, Lin CL (2009) Mass preserving non-rigid registration of CT lung images using cubic b-spline. Med Phys 36(9):4213–4222
Fan L, Chen CW (1999) 3D warping and registration from lung images. In: Medical Imaging. SPIE, pp 459–470
Fan L, Chen CW, Reinhardt JM, Eric AH (2001) Evaluation and application of 3D lung warping and registration model using HRCT images. In: Medical Imaging. SPIE, pp 234–243
Christensen GE, Song JH, Lu W, Naga IE, Daniel AL (2007) Tracking lung tissue motion and expansion/compression with inverse consistent image registration and spirometry. Med Phys 34(6):2155–2167
Christensen GE, Johnson H (2001) Consistent image registration. IEEE Trans Med Imaging 20(7):568–572
Christensen GE (2005) Inverse consistent image registration. In: Handbook of biomedical image analysis vol III. pp 219–250
Ding K, Cao K, Christensen GE, Raghavan ML, Hoffman EA, Reinhardt JM (2008) Registration based lung tissue mechanics assessment during tidal breathing. In: Medical Image Analysis. Lecture notes on computer science, pp 63–72
Pan Y, Kumar D, Hoffman EA, Christensen GE, Mclennan G, Song JH, Ross A, Simon BA, Reinhardt JM (2005) Estimation of regional lung expansion via 3D image registration. In: Medical, Imaging. SPIE
Horn BKP, Schunck BG (1981) Determining optical flow. Artif Intell 17:185–204
Lorenzen P, Davis B, Shen J (2004) Model-based symmetric information theoretic large deformation multi-modal image registration. In: International symposium of Biomedical Imaging. pp 72–723
Wu G, Yap P, Kim M, Shen D (2010) TPS-HAMMER: Improving HAMMER registration algorithm by soft correspondence matching and thin plate splines based deformation interpolation. NeuroImage 49:2225–2233
Shen D, Davatzikos C (2002) HAMMER: Hierarchical attribute matching mechanism for elastic registration. IEEE Trans. Med. Imaging 21(11):1421–1439
Wu G, Wang Q, Lian J, Shen D (2011) Re-construction of 4D-CT from a single free-breathing 3D-CT by spatial temporal image registration. In: Information Processing in Medical Imaging. pp 686–698
Vik T, Kabus S, von Berg J, Ens K, Dries S, Klinder T, Lorenz C (2008) Validation and comparison of registration methods for free-breathing 4D lung-CT. In: Medical, Imaging. SPIE
Pflicke H, Sixt M (2009) Preformed portals facilitate dendritic cell entry into afferent lymphatic vessels. J Exp Med 206(13):2925
Yang D, Brame S, El Naqa I, Aditya A, Wu Y, Goddu SM, Mutic S, Deasy JO, Low DA (2010) DIRART-a software suite for deformable image registration and adaptive radiotherapy research. Med Phys 30(1):67–77
Low D (2010) Gamma dose distribution evaluation tool. Journal of Physics conference series (250)
Fraass B, Doppke K, Hunt M, Kutcher G, Starkschall G, Stern R, Van Dyke J (1998) American association of physicists in medicine radiation therapy committee task group 53: quality assurance for clinical radiotherapy treatment planning. Med Phys 25(10):1773–1829
Latifi K, Zhang G, Stawicki M, van Elmpt W, Dekker A, Forster K (2013) Validation of three deformable image registration algorithms for the thorax. J Appl Clin Med Phys 14(1):3834
Vandemeulebrouche J, Sarrut D, Clarysse P (2007) The POPI-model, a point-validated pixel-based breathing thorax model. International Conference on the Use of Computers in Radiation Therapy
Murphy KJ, van Ginneken B, Reinhardt JM, Kabus S, Ding K, Deng X, Cao K, Du K, Christensen GE, Garcia V, Vercauteren T, Ayache N, Commowick O, Malandain G, Glocker B, Paragios N, Navab N, Gorbunova V, Sporring J, de Bruijne M, Han X, Heinrick MP, Schnabel JA, Jenkinson M, Lorenz C, Modat M, McClelland JR, Ourselin S, Muenzing SEA, Viergever MA, Nigris DD, Collins DL, Arbel T, Peroni M, Li R, Sharp GC, Schmidt-Richberg A, Ehrhardt J, Werner R, Smeets D, Loeckx D, Song G, Tustison N, Avants B, Gee JC, Staring M, Klien S, Stoel BC, Urschler M, Werlberger M, Vandemeulebroucke J, Rit S, Sarrut D, Pluim PW (2011) Evaluation of registration methods on thoracic CT: the Empire 10 challenge. IEEE Trans Med Imaging 30(11):1901–1920
Min Y, Santhanam AP, Neelakkantan H, Ruddy BH, Meeks SL, Kupelian P (2010) A GPU-based framework for modeling real-time 3D lung tumor conformal dosimetry with subject specific lung tumor motion. Phys Med Biol 55(17):5137–5150
Santhanam AP, Min Y, Mudur SP, Rastogi A, Ruddy BH, Shah A, Divo E, Kassab A, Rolland JP, Kupelian P (2010) An inverse hyper-spherical harmonics based formulation for reconstructing 3D volumetric lung deformations. CR MECANIQUE 338(7–8):461–473
Acknowledgments
This project is supported in part by James and Esther King Grant Florida, National Science Foundation (1200579), and the University of California, Los Angeles.
Conflict of interest
Yugang Min, John Neylon, Amish Shah, Sanford Meeks, Percy Lee, Patrick Kupelian and Anand P. Santhanam declare no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Min, Y., Neylon, J., Shah, A. et al. 4D-CT Lung registration using anatomy-based multi-level multi-resolution optical flow analysis and thin-plate splines. Int J CARS 9, 875–889 (2014). https://doi.org/10.1007/s11548-013-0975-7
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11548-013-0975-7