Abstract
Rapid and frequent screening of cytokines as immunomodulation agents is necessary for precise interventions in severe pathophysiological conditions. In addition to high-sensitivity detection of such analytes in complex biological fluids such as blood, saliva, and cell culture medium samples, it is also crucial to work out miniaturized bioanalytical platforms with potential for high-density integration enabling screening of multiple analytes. In this work, we show a compact, point-of-care-ready bioanalytical platform for screening of cytokines such as interleukin-4 (IL-4) and interleukin-2 (IL-2) based on one-dimensional ion-sensitive field-effect transistors arrays (nanoISFETs) of silicon fabricated at wafer-scale via nanoimprint lithography. The nanoISFETs biofunctionalized with receptor proteins alpha IL-4 and alpha IL-2 were deployed for screening cytokine secretion in mouse T helper cell differentiation culture media, respectively. Our nanoISFETs showed robust sensor signals for specific molecular binding and can be readily deployed for real-time screening of cytokines. Quantitative analyses of the nanoISFET-based bioanalytical platform was carried out for IL-4 concentrations ranging from 25 fg/mL (1.92 fM) to 2.5 μg/mL (192 nM), showing a limit of detection down to 3–5 fM, which was found to be in agreement with ELISA results in determining IL-4 concentrations directly in complex cell culture media.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
Cytokines are secreted by immune cells such as lymphocytes (T cells) and epithelial cells, as low molecular weight proteins which are responsible for management of various host defense mechanisms. In response to an infection, specialized T cells change their function and provide biomolecular triggers to the immune system by initiating a pro-inflammatory cascade, i.e., excessive production of pro-inflammatory cytokines including interleukin-2 (IL-2), interleukin-6 (IL-6), interferon gamma (IFNγ), and tumor necrosis factor alpha (TNF alpha). In a subsequent anti-inflammatory phase, the immune response switches to production of anti-inflammatory cytokines, including interleukin-10 (IL-10), transforming growth factor (TGF-β), and interleukin-4 (IL-4) [1]. Secretion profiles of different cytokines are therefore important biomarkers in disease management and immunology studies [2].
The cytokine IL-4 regulates differentiation, proliferation, and apoptosis hematopoietic and non-hematopoietic cells [3, 4]. The biosignalling machinery of IL-4 is known to be very diverse. For instance, IL-4 plays a crucial role in differentiation of naïve CD4+ T cells (or Th cells) by promoting Th2 cell differentiation acting against parasitic infections and rheumatoid arthritis [1, 2]. IL-4 is, conversely, involved in promoting diseases such as allergic asthma [2]. Similarly, IL-2 have essential role in key immune response mechanisms where they directly impact the differentiation of T cells, effector T cells, and memory T cells, helping the immune system act against infection [5, 6]. Expression of IL-2 is tightly regulated for their role in maintaining transient feedback loops controlling the immune response and also playing a key role in cell-mediated immunity [4, 6]. Due to highly dynamic and heterogeneous secretion profiles in different tissues, precise monitoring of cytokine during a disease progression and treatment pose a real challenge. Generally, labelling-based methods are performed for qualitative tissue analysis while enzyme-linked immunosorbent spot (ELISpot) and enzyme-linked immunosorbent assay (ELISA) are employed for quantitative analysis [1, 7]. Though high sensitivity and multiplexing ability, ELIspot and ELISA are not suitable for rapid immune status monitoring and generally require several days before completion [1, 7, 8]. Given ultrafast and efficient screening of cell-secreted cytokines, personalized care in different inflammatory conditions, such as sepsis, cancer, lupus, and graft-versus-host disease (GVHD), is expected to be critically improved [7, 9].
Point-of-care (PoC) immunodiagnostic platforms based on miniaturized optical and electrical biosensors and smart instrumentation such as real-time readout, data transfer, and analysis tools are in the rise [10, 11]. Such smart PoCs, when integrated in high-density arrays, have the potential to revolutionize the immune profiling technology in the laboratory and clinical environments. In recent years, mechanical (quartz crystal microbalance), optical (fluorescence, colorimetric, photonic resonators, and Raman etc.), and electrical (electrochemical, impedance and field-effect based) sensor platforms have been realized for label-free screening of cytokines [7,8,9, 12,13,14,15,16,17,18,19,20,21,22]. Notably, localized surface plasmon resonance (LSPR)-based multi-arrayed biosensors were shown for label-free, real-time detection of IL-2, IL-4, IL-6, IL-10, TNF-α, and IFNγ from serum [9, 23]. Most of the optical methods, however, still require expensive instrumentation and are not suitable for PoC integration [7, 8, 24]. Among electrical biosensors, one-dimensional (1D) and two-dimensional (2D) ion-sensitive field-effect transistors (ISFETs) based on silicon (Si), graphene, and molybdenum disulfide (MoS2) have been employed for detection of cytokines such as TNF-α, IL-1β, IL-6, and IL-8 [12, 25,26,27,28,29,30,31,32,33,34,35]. A comparative analysis (in terms of sensitivity and response time) of different electrical sensors realized for detection of cytokines in last 5 years is listed in Table 1. Here, top-down fabricated electrical biosensors such as Si-based ISFETs show a real promise for high-throughout readout for detection of ultralow concentrations of cytokines due to the highly identical sensor behavior and due to their reliable multiplexed detection capabilities [46, 47].
In this work, top-down fabricated nanoscale ISFET-arrays (nanoISFETs) of Si are shown as a proof-of-concept PoC-ready biosensor platform for label-free electrical detection of cell-secreted IL-4 and IL-2. Fluidics encapsulated nanoISFETs were surface modified with IL-4- and IL-2-specific receptor molecules and deployed for field-effect-based detection. The field-effect behavior of the sensor platform was thoroughly characterized upon biofunctionalization and analyte binding steps. The biosensor platform was also successfully deployed for real-time screening of cytokines. The results obtained through electrical screening of IL-4 were found to be in agreement with those obtained from standard ELISA as a reference method. Therefore, nanoISFET array-based sensor platforms such as presented here are apt for further deployment as integrated total bioanalytical systems for screening of immune balance in PoC approaches.
Materials and methods
IL-4 and IL-2
The cytokines IL-4 and IL-2 used in this work were secreted from T helper cells into their culture medium. Isolation of naïve T cell from the spleen and lymph nodes of mice and their differentiation was based on our earlier works [48, 49]. The method used for isolation and differentiation of Th cells for expression of IL-4 and IL-2, and for minimal expression of cytokines for control experiments, is discussed in the Electronic Supplementary Material (ESM, section S1). Anti-IL-4 and anti-IL-2 antibodies used as receptors towards specific binding of IL-4/IL-2 expressed in the cell culture media were obtained from eBiosciences (Thermo Fisher), Germany. A recombinant IL-4 protein specific to anti-IL-4 was also commercially procured (from eBioscience, Germany) in order to ascertain the molecular binding and sensor performance. In addition, recombinant IL-2 protein (eBioscience, Germany) non-specific to the anti-IL-4 was obtained for control experiments. Ten millimolar phosphate buffer saline (PBS) with pH 7.4 was used for all the samples with commercially procured biomolecules.
Assembly of the biosensor platform
NanoISFET arrays of silicon used for the assembly of the electrical biosensor platform were fabricated in a top-down lithography process. A combination of nanoimprint lithography, wet-chemical etching, and photolithography was used to cast out Si nanowires (NWs) in the top Si layer of prime quality silicon-on-insulator (SOI) wafers (procured from SOITec, France). The fabrication protocol for realization of nanoISFETs with near-identical biosensor characteristics was reported earlier [50]. Such nanoISFET arrays were earlier deployed as an aptamer-based sensor platform for label-free detection of prostate-specific antigens (PSA)—a biomarker for the onset of prostate cancer [51]. Assembly of nanoISFET array-based miniaturized sensor platform used in this work is illustrated in Fig. 1. Individual NWs in the arrays form a trapezoidal cross section due to anisotropic etching and measure 14 μm in length, 250 nm in width, and 60 nm in height. Each sensor chip consists of 8 sets of such NWs (a total of 32 NWs) connected to a common source but individually addressable drain electrodes. Eight-millimeter silicon dioxide (SiO2) layer serves as an dielectric on top of the NWs as well as a passivation layer in the contact line regions. As shown in Fig. 1b, sensor chips were mounted on a chip carrier and wire-bonded. Polymethylmethacrylate (PMMA)- and polydimethoxysiloxane (PDMS)-based fluidic layers were used in order to facilitate efficient handling of small sample volumes (20 μl) and to protect the electrical contacts from the liquid environment. Figure 1c shows the complete measurement setup, connected to a miniaturized readout system via a channel switch. A portable four-channel device built around a 32-bit PIC microcontroller as the readout tool [52]. The readout is equipped with a user-friendly LabView-based software interface, communicating with the hardware to record and display the electrical measurements on a computer screen. In Fig. 1d, the sensor platform is shown from the side (without the top PMMA cover) with a miniaturized, electrochemical silver/silver chloride (Ag/AgCl) reference electrode (World Precision Instruments Europe, DRI-Ref, 450 μm diameter), which was used to apply the gate voltage (Vg) to the electrolyte solution.
Biofunctionalization of interleukin-specific receptors
The strategy to render the nanoISFETs biologically specific towards binding of IL-4 and IL-2 is illustrated in Fig. 2a and b. NWs were functionalized with 1 μg/mL of anti-IL-4 and anti-IL-2 antibodies (obtained from eBioscience, Germany) in a surface modification process similar to other biofunctionalization strategies reported in previous works [50, 51, 53]. Briefly, gas-phase silanization with (3-glycidyloxypropyl)triethoxysilane (GPTES—Sigma-Aldrich, Germany) was used for surface modification of the nanoISFET surfaces and the immobilization of the IL-specific antibodies was done overnight at 37 °C. Control experiments for non-specific interactions were carried out using Th0 cultures with minimal expressions of IL-4 and IL-2. Further, control experiments evaluating non-specific adsorption of IL-4 on nanoISFET surface were also performed, which are included in the ESM (Fig. S3).
Results and discussions
The biofunctionalization of nanoISFETs with cytokine-specific receptor molecules was confirmed by atomic force microscopy (AFM) as shown in Fig. 2c. The three-dimensional AFM scan images in the left and right compare the topography of ISFETs before and after the anti-IL-4 biofunctionalization steps. The graph in Fig. 2d compares height profiles of 4 nanoISFETs. A height increase of about 15 nm was measured after the biofunctionalization step. Further, Fig. 2e compares roughness of the measured areas before and after biofunctionalization in the histogram. This histogram counts the number of events recorded for the heights in the scan areas and up to 160 nm. For the nanoISFETs before functionalization, maximum numbers of events are recorded for heights around 26 nm and for a height of 130 nm. After the biofunctionalization, this event distribution shifts to around 40 nm and thereafter for a height of 146 nm. Increase in the number of events from 30 to 40 represents deposition of biomolecules in the chip surface, while from 130 to 146 represents an increase in the height of NWs.
Sensor measurements were carried out by measuring the changes in threshold voltages (Vth) of the nanoISFETs after the biofunctionalization and upon placing small volumes of culture media with IL-4 and IL-2. Figure 3 sums up the sensor measurements carried out for the detection of IL-4 from the culture media. Figure 3a shows typical field-effect or transfer characteristics of a nanoISFET at the three stages, i.e., after surface modification with GPTES (or also mentioned as “before,” where the field-effect curve is shown in black), after the biofunctionalization (red curve), and in the cell culture medium with IL-4 cytokines (green curve). In the measurement graphs, the drain current (Id) is plotted against the gate-source voltage (Vgs) (0 to − 2 V) while applying a drain source bias (Vds) of − 100 mV. Successive field-effect measurements followed the same parameters unless specifically mentioned. Vth values for the field-effect curves were calculated from the first-order derivatives by extrapolating the linear regions of the curves. When comparing the linear regions of the field-effect curves, a shift towards a more positive Vg can be noticed after anti-IL-4 biofunctionalization. This shift towards more positive Vg is expected for anti-IL-4, because it represents the alpha chain of the IL-4 protein receptors, which is a 25-kDa transmembrane protein with an isoelectric point (pI) around 6 adding negatively charged functional groups onto the nanoISFET surfaces. Upon exposure to the culture medium with IL-4, the field-effect curves shift towards negative gate voltages. This shift is related to the specific binding of IL-4 (a 17.5-kDa protein) with pI value of around 9.1 bringing an overall increase of net positive charges at the interface [58]. Figure 3b shows the threshold values of 15 nanoISFETs associated with this molecular recognition event at the NW surface. The Vth values generally increase upon exposure to the culture media (green curve with symbols) with IL-4 as a result of IL-4 recognition at the nanoISFET surface. Further, in order to confirm molecular recognition at the nanoISFET surface, measurements were carried out in Th0 culture medium (with negligible concentrations of IL-4 among other ILs). Results from such measurements are shown in Fig. 3c, where threshold values are shown with the blue symbol. These curves represent the non-specific binding effect of ILs, which was remarkably lower than that for the specific binding. An overall comparison between the specific and non-specific binding is shown in the graph of Fig. 3d. For specific binding, an average change in threshold voltage was calculated to − 15.8 (± 6.9) mV against − 3.5 (± 3) mV for the non-specific binding. A statistical comparison between specific and non-specific binding using unpaired T test analysis suggested extremely significant signal with P value equivalent to 0.0001. In addition, we also carried out control measurements characterizing the free adsorption, where changes in Vth values were recorded for completely non-modified nanoISFETs in Th2 culture medium. While average changes in the Vth values − 3.2 (± 2) mV are shown in Fig. 3d, more detailed measurements are included in the ESM. With these measurements, we confirmed that anti-IL-4 antibody bioimmobilized nanoISFETs act as robust electrical biosensor platform, distinguishing specific molecular interactions even directly in the complex culture media. Even though it is difficult to access the exact molecular conformations and distributions of charges in IL-4 complexes (anti-IL-4 antibody-IL-4-antigen) with respect to their position within or outside the Debye length, regular changes in Vth values suggest for significant changes in surface charge density of biofunctional layer resulting in reproducible sensor signals.
Deploying nanoISFETs as an immunosensor platform as established from label-free detection of IL-4s in culture medium, we further modified the nanoISFETs with anti-IL-2 receptors. Representative field-effect measurements from a nanoISFET are shown in Fig. 4a. Field-effect curves in black, red, and green curves represent the successive steps of surface modification, biofunctionalization, and exposure to IL-2 containing culture medium, respectively (iTregs—which was differentiated using 2.5 ng/mL TGF-β and 5 ng/mL IL-2) [54]. With the biofunctionalization, the field-effect curve shifts towards positive gate voltages. After the exposure to culture medium with IL-2, the curve further shifts towards positive gate voltages. The Vth values associated with the shift of field-effect curves representing specific binding between alpha IL-2 and IL-2 are shown in Fig. 4b. A total of 20 nanoISFETs were measured which showed a Vth shift towards positive gate voltages. In addition, control measurements carried out with Th0 culture medium with negligible concentration of IL-2 did not show significant changes in Vth (Fig. 4c). A comparison for specific binding and control experiment is made by plotting the average changes in Vth values (Fig. 4d) for 8 nanoISFETs. For IL-2 detection, the average change in Vth was at − 41.4 (± 9) mV, while it was only − 18.2 (± 7) mV for the control experiments. Statistical comparison between specific and non-specific binding using unpaired T test analysis suggested extremely significant signal with P value equivalent to 0.0001. Normalized threshold voltage changes as sensor response upon specific and non-specific binding of IL-4 and IL-2 are included in ESM (Fig. S4).
From the measurements shown in Figs. 3 and 4, we conclude that nanoISFETs can be used as a reliable sensor platform for the detection of cytokines in complex culture media. The sensor responses are compared for PoC applications; such nanoISFETs are expected to be an ideal candidate delivering molecular quantification in very short times. In order to evaluate the performance of our nanoISFET immunosensor platform for such applications, we characterized the sensor dynamic ranges and biosensor behavior in real time using our portable sensor readout setup. For this, anti-IL-4-modified nanoISFETs were deployed for time-dependent and quantitative measurements of IL-4. For sensor calibration and in order to avoid expressing precise concentrations of IL-4 in culture media with identical matrix compositions, we decided to use commercially procured IL-4s and prepared different calibration solutions by spiking into PBS. The setup used for the real-time quantitative measurements required no changes from what is shown in Fig. 1c, d. In addition to field-effect measurements, the DC readout tool developed in house is also able to measure an output current at selected working set points as a function of time. IL-4 concentrations of 1.92 fM (equivalent to 2.5 pg/mL), 19.2 fM, 19.2 pM, 1.92 nM, and 192 nM (equivalent to 2.5 μg/mL) in buffer were used for the measurements. The real-time detection of different concentrations of IL-4 is shown in Fig. 5. Figure 5a shows a typical measurement, where the current output from the nanoISFET on y-axis is plotted against time while injecting different concentrations of IL-4. All the measurements carried out were performed while applying a fixed Vg to the Ag/AgCl reference electrode at − 1.5 V and a drain source bias of − 1 V. The black curve in the graph represents the drain output of a nanoISFET in PBS for up to 15 min showing a stable electrical response. The rest of the measurements were carried out in continuum, where current output of same biofunctionalized nanoISFET was firstly measured in PBS (red curve) and thereafter stepwise adding different concentrations of IL-4 in PBS. It can be seen that after the biofunctionalization step, the output current decreased significantly, which also reflects the situation shown in Fig. 3, where the field-effect curve shifted towards positive gate voltages. At a constant working point, a shift towards positive gate voltages means a decreased output current. Then after, the current output from the nanoISFET started increasing upon stepwise injecting the IL-4 concentrations. The stepwise increase in output current is also in agreement with the endpoint measurements shown in Fig. 3, where the field-effect curve shifts towards negative gate voltages due to specific binding of IL-4 on the nanoISFET surface. At a constant working point, a shift towards more negative gate voltages would amount to an increased output current. For real-time measurements, each concentration of IL-4 was allowed in the fluidic chamber for 15 to 20 min. Real-time measurements were carried out similarly for other sensor assemblies (see ESM). For determination of sensor behavior in terms of dynamic range and quantification, a dose response curve was worked out after taking an average of current output values for each concentration of IL-4 (Fig. 5b). For this, equivalent Vth values were calculated from average drain outputs by using the equation Vgs = Ids/gm, (where gm represents the transconductance of nanoISFET as calculated from the field-effect characterization of the device at Vgs = − 1.5 V, Vds = − 1 V). Figure 5c shows such equivalent Vth values converted from the output drain current values of the ISFETs against different IL-4 concentrations. The dose response curve is in agreement with the IL-4 measurements carried out in culture media and shows a sigmoidal fitting (red curve) typical of a molecular binding isotherm.
It is to be noted that when measuring IL-4 concentrations in the culture media as shown in Fig. 3, we measured an average change in Vth at − 15.8 mV. This average change in Vth, representative of IL-4 concentrations in Th2 culture, is shown in Fig. 5c by the horizontal dashed line in blue color. The intersection point of this line on the calibration plot relates to IL-4 concentration of 10.1 pM as shown by the vertical bar in blue color. The vertical bar in red with the dash-dot line represents the ELISA reference measurements of IL-4 concentrations in culture media that was used for the measurements as shown in Fig. 3. Figure 5d shows the results from the ELISA measurements used for the determination of IL-4 in Th0 (0.6 pM) and Th2 (3.2 pM) culture media. Therefore, the nanoISFET measurements overestimate the IL-4 concentration as compared with ELISA by about 7 pM, which falls within the error margins of the nanoISFET calibration plot. Even though the electrical measurements come quite close to the results obtained with ELISA, higher accuracy would be needed for practical applications. One cannot directly compare the two methods, for the different surface properties and modification requirements are expected to have an influence. In future, it would be interesting to employ advanced techniques, where ELISA or other optical measurements such as surface plasmon resonance measurements can be carried out simultaneously employing the same surface, i.e., Si/SiO2, representing identical biofunctional interfaces and molecular interactions. Here, we assume that the higher ionic strengths and complex compositions of culture media also cause the overestimation of the IL-4 concentrations, since the calibration plot was only worked out in PBS solution without complex biomolecule background.
It is worth mentioning that the development of advanced nanoelectronic biosensor platforms such as presented in this work is mainly driven by the requirements from biomedical research. Active collaborations are fruitful and demanding for us towards the optimization of this novel PoC method and for future protocols towards real biomedical applications. In this respect, we look forward to further develop our nanoISFET platform for molecular screening in clinical settings overcoming concurrent challenges in the field. Use of novel receptors molecules such as aptamers in advanced anti-biofouling matrices on top of sensor surfaces is especially interesting for enhanced charge-based detection for in vivo and in vitro immunosensor applications [7, 55,56,57]. We have reported earlier on the correlation of interleukin expressions with mRNAs (such as for IL-4 and IL-2 used in this work with the IL-4 and IL-2 for positive and negative controls) in the differentiated T helper cells [48, 54]. Such mRNA correlations are expected to further strengthen on-chip immune strategies such as presented in this work for different pathophysiological conditions and biological development.
Conclusion
In summary, we present an elegant electrical screening platform that comes with advantages of top-down fabricated nanoISFETs with reproducible sensor behavior and is system-integrated for deployment as PoC biosensors. The nanoISFET-based PoC biosensor platform not only showed an interleukin detection range down to sub-picomolar concentrations in PBS but also demonstrated robust biosensor behavior directly inside complex culture media of mouse T cells. Such miniature electrical detection platforms are expected to enhance bioanalytical screening capabilities in biochemistry and biotechnology domains. In particular, for immunology applications, expression of interleukins panels is associated with complex immune response pathways and as shown in this work, PoC nanoISFETs will be a viable alternative for their capabilities to reduce the cost of screening. Future versions of nanoISFET-based biosensor platforms are expected to take advantage of high-density integration, localized surface modification methods, and developments in readout capabilities, in putting together more compact versions of multiplexed immunosensing devices.
References
Stenken JA, Poschenrieder AJ. Bioanalytical chemistry of cytokines--a review. Anal Chim Acta. 2015;853:95–115.
Kelso A. Cytokines: principles and prospects. Immunol Cell Biol. 1998;76(4):300–17.
Banchereau J, Pascual V, O’Garra A. From IL-2 to IL-37: the expanding spectrum of anti-inflammatory cytokines. Nat Immunol. 2012;13(10):925–31.
Zamorano J, Rivas MD, Pérez-G M. Interleukin-4: a multifunctional cytokine. Inmunología. 2003;22(2):215–24.
Arenas-Ramirez N, Woytschak J, Boyman O. Interleukin-2: biology, design and application. Trends Immunol. 2015;36(12):763–77.
Liao W, Lin J-X, Leonard WJ. IL-2 family cytokines: new insights into the complex roles of IL-2 as a broad regulator of T helper cell differentiation. Curr Opin Immunol. 2011;23(5):598–604.
Chen P, Huang NT, Chung MT, Cornell TT, Kurabayashi K. Label-free cytokine micro- and nano-biosensing towards personalized medicine of systemic inflammatory disorders. Adv Drug Deliv Rev. 2015;95:90–103.
Pui TS, Kongsuphol P, Arya SK, Bansal T. Detection of tumor necrosis factor (TNF-α) in cell culture medium with label free electrochemical impedance spectroscopy. Sensors Actuators B Chem. 2013;181:494–500.
Chen P, Chung MT, McHugh W, Nidetz R, Li Y, Fu J, et al. Multiplex serum cytokine immunoassay using nanoplasmonic biosensor microarrays. ACS Nano. 2015;9(4):4173–81.
Thaler M, Luppa PB. Highly sensitive immunodiagnostics at the point of care employing alternative recognition elements and smartphones: hype, trend, or revolution? Anal Bioanal Chem. 2019;411(29):7623–35.
Poschenrieder A, Thaler M, Junker R, Luppa PB. Recent advances in immunodiagnostics based on biosensor technologies-from central laboratory to the point of care. Anal Bioanal Chem. 2019;411(29):7607–21.
Wu JY, Lin CH, Feng MH, Chen CH, Su PC, Yang PW, et al. Preparation of Silicon Nanowire Field-effect Transistor for Chemical and Biosensing Applications. J Vis Exp. 2016;110:e53660.
Baraket A, Lee M, Zine N, Sigaud M, Bausells J, Errachid A. A fully integrated electrochemical biosensor platform fabrication process for cytokines detection. Biosens Bioelectron. 2017;93:170–5.
Song Y, Chen P, Chung MT, Nidetz R, Park Y, Liu Z, et al. AC electroosmosis-enhanced nanoplasmofluidic detection of ultralow-concentration cytokine. Nano Lett. 2017;17(4):2374–80.
Park Y, Ryu B, Oh BR, Song Y, Liang X, Kurabayashi K. Biotunable nanoplasmonic filter on few-layer MoS2 for rapid and highly sensitive cytokine optoelectronic immunosensing. ACS Nano. 2017;11(6):5697–705.
Liu G, Zhang K, Nadort A, Hutchinson MR, Goldys EM. Sensitive cytokine assay based on optical fiber allowing localized and spatially resolved detection of interleukin-6. ACS Sens. 2017;2(2):218–26.
Kim D, Kwon HJ, Shin K, Kim J, Yoo RE, Choi SH, et al. Multiplexible wash-free immunoassay using colloidal assemblies of magnetic and photoluminescent nanoparticles. ACS Nano. 2017;11(8):8448–55.
Diacci C, Berto M, Di Lauro M, Bianchini E, Pinti M, Simon DT, et al. Label-free detection of interleukin-6 using electrolyte gated organic field effect transistors. Biointerphases. 2017;12(5):05F401.
Usuba R, Yokokawa M, Ackermann TN, Llobera A, Fukunaga K, Murata S, et al. Photonic lab-on-a-chip for rapid cytokine detection. Acs Sensors. 2016;1(8):979–86.
Sharma R, Deacon SE, Nowak D, George SE, Szymonik MP, Tang AAS, et al. Label-free electrochemical impedance biosensor to detect human interleukin-8 in serum with sub-pg/ml sensitivity. Biosens Bioelectron. 2016;80:607–13.
Lau UY, Saxer SS, Lee J, Bat E, Maynard HD. Direct write protein patterns for multiplexed cytokine detection from live cells using electron beam lithography. ACS Nano. 2016;10(1):723–9.
Rahimian A, Siltanen C, Feyzizarnagh H, Escalante P, Revzin A. Microencapsulated immunoassays for detection of cytokines in human blood. ACS Sens. 2019;4(3):578–85.
Oh BR, Chen P, Nidetz R, McHugh W, Fu J, Shanley TP, et al. Multiplexed nanoplasmonic temporal profiling of T-cell response under immunomodulatory agent exposure. ACS Sens. 2016;1(7):941–8.
Singh M, Truong J, Reeves WB, Hahm JI. Emerging cytokine biosensors with optical detection modalities and nanomaterial-enabled signal enhancement. Sensors (Basel). 2017;17(2):428–73.
Chen H, Choo TK, Huang JF, Wang Y, Liu YJ, Platt M, et al. Label-free electronic detection of interleukin-6 using horizontally aligned carbon nanotubes. Mater Design. 2016;90:852–7.
Pui TS, Agarwal A, Ye F, Huang Y, Chen P. Nanoelectronic detection of triggered secretion of pro-inflammatory cytokines using CMOS compatible silicon nanowires. Biosens Bioelectron. 2011;26(5):2746–50.
Hakim MM, Lombardini M, Sun K, Giustiniano F, Roach PL, Davies DE, et al. Thin film polycrystalline silicon nanowire biosensors. Nano Lett. 2012;12(4):1868–72.
Zhang Y, Chen R, Xu L, Ning Y, Xie S, Zhang GJ. Silicon nanowire biosensor for highly sensitive and multiplexed detection of oral squamous cell carcinoma biomarkers in saliva. Anal Sci. 2015;31(2):73–8.
Mao Y, Shin KS, Wang X, Ji Z, Meng H, Chui CO. Semiconductor electronic label-free assay for predictive toxicology. Sci Rep. 2016;6:24982.
Ryu B, Nam H, Oh BR, Song Y, Chen P, Park Y, et al. Cyclewise operation of printed MoS2 transistor biosensors for rapid biomolecule quantification at femtomolar levels. ACS Sens. 2017;2(2):274–81.
Mu L, Droujinine IA, Lee J, Wipf M, Davis P, Adams C, et al. Nanoelectronic platform for ultrasensitive detection of protein biomarkers in serum using DNA amplification. Anal Chem. 2017;89(21):11325–31.
Liu Y, Rahimian A, Krylyuk S, Vu T, Crulhas B, Stybayeva G, et al. Nanowire aptasensors for electrochemical detection of cell-secreted cytokines. ACS Sens. 2017;2(11):1644–52.
Kim TH. A simple and real-time sensing of human serum albumin using antibody-modified CNT-FET. Biochip J. 2017;11(2):116–20.
Pham Ba VA, Han YM, Cho Y, Kim T, Lee BY, Kim JS, et al. Modified floating electrode-based sensors for the quantitative monitoring of drug effects on cytokine levels related with inflammatory bowel diseases. ACS Appl Mater Interfaces. 2018;10(20):17100–6.
Nam H, Oh BR, Chen P, Chen M, Wi S, Wan W, et al. Multiple MoS2 transistors for sensing molecule interaction kinetics. Sci Rep. 2015;5:10546.
Parate K, Rangnekar SV, Jing D, Mendivelso-Perez DL, Ding S, Secor EB, et al. Aerosol-jet-printed graphene immunosensor for label-free cytokine monitoring in serum. ACS Appl Mater Interfaces. 2020;12(7):8592–603.
Cruz A, Queiros R, Abreu CM, Barata C, Fernandes R, Silva R, et al. Electrochemical immunosensor for TNFalpha-mediated inflammatory disease screening. ACS Chem Neurosci. 2019;10(6):2676–82.
Tertiş M, Ciui B, Suciu M, Săndulescu R, Cristea C. Label-free electrochemical aptasensor based on gold and polypyrrole nanoparticles for interleukin 6 detection. Electrochim Acta. 2017;258:1208–18.
Arya SK, Kongsuphol P, Park MK. On-chip electrochemical immunoassay platform for specific protein biomarker estimation in undiluted serum using off-surface membrane matrix. Biosens Bioelectron. 2017;91:721–7.
Ni SN, Qiao LC, Shen ZP, Gao YF, Liu GZ. Physical absorption vs covalent binding of graphene oxide on glassy carbon electrode towards a robust aptasensor for ratiometric electrochemical detection of vascular endothelial growth factor (VEGF) in serum. Electrochim Acta. 2020;331.
Gentili D, D’Angelo P, Militano F, Mazzei R, Poerio T, Brucale M, et al. Integration of organic electrochemical transistors and immuno-affinity membranes for label-free detection of interleukin-6 in the physiological concentration range through antibody-antigen recognition. J Mater Chem B. 2018;6(33):5400–6.
Barhoumi L, Baraket A, Bellagambi FG, Karanasiou GS, Ali MB, Fotiadis DI, et al. A novel chronoamperometric immunosensor for rapid detection of TNF-α in human saliva. Sensors Actuators B Chem. 2018;266:477–84.
Berto M, Diacci C, D’Agata R, Pinti M, Bianchini E, Di Lauro M, et al. EGOFET peptide aptasensor for label-free detection of inflammatory cytokines in complex fluids. Adv Biosyst. 2018;2(2).
Wang ZR, Hao Z, Yu SF, De Moraes CG, Suh LH, Zhao XZ, et al. An ultraflexible and stretchable aptameric graphene nanosensor for biomarker detection and monitoring. Adv Funct Mater. 2019;29(44).
Hao Z, Wang Z, Li Y, Zhu Y, Wang X, De Moraes CG, et al. Measurement of cytokine biomarkers using an aptamer-based affinity graphene nanosensor on a flexible substrate toward wearable applications. Nanoscale. 2018;10(46):21681–8.
Mu L, Chang Y, Sawtelle SD, Wipf M, Duan XX, Reed MA. Silicon nanowire field-effect transistors-a versatile class of potentiometric nanobiosensors. Ieee Access. 2015;3:287–302.
Salhi B, Hossain MK, Mukhaimer AW, Al-Sulaiman FA. Nanowires: a new pathway to nanotechnology-based applications. J Electroceram. 2016;37(1):34–49.
Singh Y, Garden OA, Lang F, Cobb BS. Retroviral transduction of helper T cells as a genetic approach to study mechanisms controlling their differentiation and function. J Vis Exp. 2016;117:e54698.
Singh Y, Garden OA, Lang F, Cobb BS. MicroRNAs regulate T-cell production of interleukin-9 and identify hypoxia-inducible factor-2alpha as an important regulator of T helper 9 and regulatory T-cell differentiation. Immunology. 2016;149(1):74–86.
Rani D, Pachauri V, Mueller A, Vu XT, Nguyen TC, Ingebrandt S. On the use of scalable nanoISFET arrays of silicon with highly reproducible sensor performance for biosensor applications. ACS Omega. 2016;1(1):84–92.
Rani D, Pachauri V, Madaboosi N, Jolly P, Vu XT, Estrela P, et al. Top-down fabricated silicon nanowire arrays for field-effect detection of prostate-specific antigen. ACS Omega. 2018;3(8):8471–82.
Nguyen TC, Schwartz M, Vu XT, Blinn J, Ingebrandt S. Handheld readout system for field-effect transistor biosensor arrays for label-free detection of biomolecules. Phys Status Solidi A. 2015;212(6):1313–9.
Munief WM, Heib F, Hempel F, Lu X, Schwartz M, Pachauri V, et al. Silane deposition via gas-phase evaporation and high-resolution surface characterization of the ultrathin siloxane coatings. Langmuir. 2018;34(35):10217–29.
Singh Y, Garden OA, Lang F, Cobb BS. MicroRNA-15b/16 enhances the induction of regulatory T cells by regulating the expression of Rictor and mTOR. J Immunol. 2015;195(12):5667–77.
Vu CA, Hu WP, Yang YS, Chan HW, Chen WY. Signal enhancement of silicon nanowire field-effect transistor immunosensors by RNA aptamer. ACS Omega. 2019;4(12):14765–71.
Poghossian A, Schoning MJ. Label-free sensing of biomolecules with field-effect devices for clinical applications. Electroanalysis. 2014;26(6):1197–213.
Sabate Del Rio J, Henry OYF, Jolly P, Ingber DE. An antifouling coating that enables affinity-based electrochemical biosensing in complex biological fluids. Nat Nanotechnol. 2019;14(12):1143–9.
Gasteiger E, Gattiker A, Hoogland C, Inanyi I, Appel RD, Bairoch A. Nucleic Acids Res. 2003; 31(13): 3784-88
Acknowledgements
We thank Detlev Cassel, Achim Müller, and Rainer Lilischkis (all University of Applied Sciences Kaiserslautern, Germany) for their help in the clean-room processes and with surface characterizations.
Funding
Open Access funding is provided by Projekt DEAL. The work was supported by the Marie Curie Initial Training Network PROSENSE (grant no. 317420, 2012−2016, www.prosense-itn.eu). V.P. received industrial support for the funding of his position from the EUROIMMUN Medizinische Labordiagnostika AG in Lübeck, Germany, and from the Stiftung Rheinland-Pfalz für Innovation (grant no. 1082). M.S.S. received the Margarete von Wrangell (MvW 31-7635.41/118/3) habilitation scholarship co-funded by the Ministry of Science, Research and the Arts (MWK) of the state of Baden-Württemberg and by the European Social Funds.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflicts of interest.
Ethical approvals
Naïve CD4+ T cells were isolated from C57BL/6 mice (male and female) between 8 and 16 weeks of age. All the animals were kept in standard housing conditions with 12± dark/light cycle and fed on a standard chow diet and had ad lib access to drinking water. All experiments were performed according to the EU Animals Scientific Procedures Act and the German law for the welfare of animals. All procedures were approved by local government authorities (Regierungspräsidium Tübingen - review committee according to §4 Animal Welfare Act on 29/05/2015 and 20/02/2017) of the state of Baden-Württemberg, Germany. No drug treatment or in vivo experiments were performed on live animals.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Published in the topical collection ABC Highlights: authored by Rising Stars and Top Experts
Electronic supplementary material
ESM 1
(PDF 898 kb)
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Rani, D., Singh, Y., Salker, M. et al. Point-of-care-ready nanoscale ISFET arrays for sub-picomolar detection of cytokines in cell cultures. Anal Bioanal Chem 412, 6777–6788 (2020). https://doi.org/10.1007/s00216-020-02820-4
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-020-02820-4