Despite the rapid development of consumer-grade electronic radon monitors, their capabilities to assess long-term average radon concentrations are not systematically investigated. We present the results of a year-long measurement campaign in 32 dwellings and workplaces in two areas with typical and high radon concentrations in Bulgaria. A systematic comparison was made between electronic monitors and the gold standard – solid-state nuclear track detectors (SSNTDs). RadonEye Plus2 (RE) monitors were used, calibrated and metrologically tested at the three participating laboratories. Two SSNTD-based detectors were utilized: passive radon detectors (PRDs) developed by UKHSA, and CDs/DVDs, developed at SU. Two PRDs for quarterly radon measurements, a RE, a CD and a DVD were placed at each location. Additional old CDs/DVDs were collected for retrospective radon estimates. The results for the annual average radon concentrations, estimated by the four methods are presented. The average difference between the REs and PRDs estimate was 8%
(median 4%), between the REs and CDs was 5% (median 0.4%) and between REs and DVDs was -10% (median-12%). The Z-score, representing the difference normalized to its uncertainty, in most cases fell within -1 and 1, indicating excellent agreement. The retrospective estimates also demonstrated excellent agreement with the other data. However, in some cases, small but statistically significant biases were
identified between REs and SSNTDs or between different SSNTDs, requiring an inclusion of a reference instrument in future studies. Overall, the results imply that with sound metrological assurance, RadonEye Plus2 monitors can assess long-term average radon concentrations with sufficient accuracy.
Mots clés
Indoor radon measurements
RadonEye plus2 electronic detector
Continuous radon monitors
Radon solid-state nuclear track detector
radon
Diffusion chambers with PADC
Long-term radon average
radioactivity
ionizing radiation
instrumentation
nuclear instrumentation
metrology
detector
calibration
passive radon detector
accuracy
Origine | Fichiers produits par l'(les) auteur(s) |
---|
Contributeur MAP CEA : Connectez-vous pour contacter le contributeur
https://cea.hal.science/cea-04718640
Soumis le : mercredi 2 octobre 2024-16:15:13
Dernière modification le : jeudi 31 octobre 2024-16:12:02
Dates et versions
- HAL Id : cea-04718640 , version 1
- DOI : 10.1016/j.radphyschem.2024.112212
Citer
Ivelina Dimitrova, J.M. Wasikiewiecz, Vladislav Todorov, Strahil Georgiev, Z. Daraktchieva, et al.. Coherent long-term average indoor radon concentration estimates obtained by electronic and solid state nuclear track detectors. Radiation Physics and Chemistry, 2025, 226, pp.112212. ⟨10.1016/j.radphyschem.2024.112212⟩. ⟨cea-04718640⟩
Collections
24
Consultations
10
Téléchargements