Landsat-8 Operational Land Imager Radiometric Calibration and Stability
"> Figure 1
<p>OLI focal plane showing 14 Focal Plane Modules (FPMs) and individual band spectral filters in their butcher-block assemblies.</p> "> Figure 2
<p>(<b>a</b>) OLI lamp assembly drawing; (<b>b</b>) OLI lamp assemblies mounted on OLI aperture stop; (<b>c</b>) Calibration device locations in OLI optical path.</p> "> Figure 3
<p>Working lamp pair illumination pattern on the OLI focal plane; (bias corrected and linearized average across 30 s after 2.2 min of warm-up). The discontinuities occur at the boundaries of the FPMs as the odd and even numbered FPMs are displaced relative to each other (<a href="#remotesensing-06-12275-f001" class="html-fig">Figure 1</a>).</p> "> Figure 4
<p>Warm-up behavior of the “working” pair of lamps as observed by the OLI detectors and monitor photodiodes during a pre-launch test. The OLI detector data are bias subtracted, linearized and converted to radiance, as described in <a href="#sec2dot1-remotesensing-06-12275" class="html-sec">Section 2.1</a>.</p> "> Figure 5
<p>Components of the calibration assembly.</p> "> Figure 6
<p>“Working” Diffuser On-Orbit Data for 28 April 2014, NIR band.</p> "> Figure 7
<p>The OLI scan pattern of the moon over two orbits on even number months (gray is orbit #1; green is orbit #2). The center and edge detectors show the extent of each FPM in the outer and inner bands (bands 8 and 9).</p> "> Figure 8
<p>OLI Lunar Images, Pan band, 26 March 2013, FPM 7. (<b>a</b>) Level 1R; (<b>b</b>) Level 1G, radiometrically and geometrically corrected (enlarged ~8× relative to (a)).</p> "> Figure 9
<p>Ball Aerospace & Technologies Corp (BATC) artifacts used for OLI calibration: (<b>a</b>) Death Star Source; (<b>b</b>) Small Sphere Source; (<b>c</b>) COTS Transfer Radiometer (CXR).</p> "> Figure 10
<p>(<b>a</b>) Radiance traceability flow from NIST to BATC: (1) Calibration of the Small Sphere Source (SSS) by FASCAL at NIST; (2) Transport of SSS from NIST to BATC with checks by the Ball Standard Radiometer (BSR), University of Arizona radiometers (UAR’s) and government (NIST and NASA) radiometers before and after transport; (3) Transfer of the calibration from the SSS to the COTS Transfer Radiometer (CXR) by measurements at BATC. The Spectral Irradiance and Radiance responsivity Calibrations using Uniform Sources (SIRCUS), a NIST facility [<a href="#B21-remotesensing-06-12275" class="html-bibr">21</a>], was used to provide the CXR spectral response. The NIST Portable Radiometric Source (NPR) [<a href="#B18-remotesensing-06-12275" class="html-bibr">18</a>] provided an additional check on the stability of the radiometers; (<b>b</b>) Radiance traceability flow from BATC to on-Orbit: (4) Calibration of the Death Star Source (DSS) by the CXR through the CATS-RC chamber window; (5) Calibration of the OLI instrument in the CATS-RC chamber; (6) Transfer of calibration to the illuminated flight diffusers with the heliostat in CATS-RC; (7) Recalibration of the OLI by the solar illuminated diffuser on-orbit; (8) Observations of the Earth by the OLI.</p> "> Figure 10 Cont.
<p>(<b>a</b>) Radiance traceability flow from NIST to BATC: (1) Calibration of the Small Sphere Source (SSS) by FASCAL at NIST; (2) Transport of SSS from NIST to BATC with checks by the Ball Standard Radiometer (BSR), University of Arizona radiometers (UAR’s) and government (NIST and NASA) radiometers before and after transport; (3) Transfer of the calibration from the SSS to the COTS Transfer Radiometer (CXR) by measurements at BATC. The Spectral Irradiance and Radiance responsivity Calibrations using Uniform Sources (SIRCUS), a NIST facility [<a href="#B21-remotesensing-06-12275" class="html-bibr">21</a>], was used to provide the CXR spectral response. The NIST Portable Radiometric Source (NPR) [<a href="#B18-remotesensing-06-12275" class="html-bibr">18</a>] provided an additional check on the stability of the radiometers; (<b>b</b>) Radiance traceability flow from BATC to on-Orbit: (4) Calibration of the Death Star Source (DSS) by the CXR through the CATS-RC chamber window; (5) Calibration of the OLI instrument in the CATS-RC chamber; (6) Transfer of calibration to the illuminated flight diffusers with the heliostat in CATS-RC; (7) Recalibration of the OLI by the solar illuminated diffuser on-orbit; (8) Observations of the Earth by the OLI.</p> "> Figure 11
<p>(<b>a</b>,<b>b</b>) Validations of the BATC DSS radiance scale by comparison to NASA and University of Arizona radiometers. One spectrometer (ASD6172) and two of the participating radiometers (LXR, UA SWIR) are shown. Two DSS radiance levels are shown, where the band, e.g., “blue”, designates the control wavelength and the “Level” is a brightness designation, with Level 10 being the brightest.</p> "> Figure 12
<p>OLI radiance responsivity (bias corrected, linearized, 14 bit counts/(W/m<sup>2</sup> sr μm) for all OLI NIR (Band 5) detectors based on NIR sphere Level 10).</p> "> Figure 13
<p>OLI radiometric stability over a 16-day period using different sources.</p> "> Figure 14
<p>OLI radiometric response to working lamp during thermal vacuum testing. (<b>a</b>) Pan band response; (<b>b</b>) Photodiode response; (<b>c</b>) Photodiode Temperature Sensitivity; (<b>d</b>) Temperature Corrected Photodiode Response.</p> "> Figure 15
<p>OLI diffuser in holder for reflectance measurement. The five measurement locations and approximate fields-of-view of the radiometer are shown in relation to the centers of fields of view of the OLI detectors on the FPMs 1–14.</p> "> Figure 16
<p>OLI working diffuser Bidirectional Reflectance Factors: Row 1: (<b>a</b>) In-plane, 445 nm, all positions; (<b>b</b>) Varying azimuths, 445 nm, position 3; Row 2: (<b>c</b>) In-plane, 863 nm, all positions; (<b>d</b>) Varying azimuths, 863 nm, position 3; Row 3: (<b>e</b>) In-plane, 2206 nm, all positions; (<b>f</b>) Varying azimuths, 2206 nm, position 3.</p> "> Figure 17
<p>OLI pristine diffuser Bidirectional Reflectance Factors: Row 1: (<b>a</b>) In-plane 445 nm, all positions (<b>b</b>) Varying azimuths, 445 nm, position 3; Row 2: (<b>c</b>) In-plane, 863 nm, all positions; (<b>d</b>) Varying azimuths, 863 nm, position 3; Row 3: (<b>e</b>) In-plane, 2206 nm, all positions; (<b>f</b>) Varying azimuths, 2206 nm, position 3.</p> "> Figure 18
<p>BATC heliostat test facility.</p> "> Figure 19
<p>Comparison of Diffuser Reflectance to OLI Response to Diffuser.</p> "> Figure 20
<p>Atmospheric and heliostat transmissions used to correct OLI transfer to orbit data to top of atmosphere values. (<b>a</b>) VNIR; (<b>b</b>) SWIR.</p> "> Figure 21
<p>OLI and stim lamp photodiode monitor changes in response to stim lamps. Horizontal bars indicate the bandpasses. Blackbody temperature shifts matching the panchromatic band observed change (blue) and the photodiode observed change (red) are shown for reference.</p> "> Figure 22
<p>Estimated OLI responsivity changes on transfer to orbit using heliostat test (from <a href="#remotesensing-06-12275-t002" class="html-table">Table 2</a>) and stim lamps with photodiode response.</p> "> Figure 23
<p>OLI on-orbit radiometric calibration results across the current mission lifetime for the pan band.</p> "> Figure 24
<p>OLI on-orbit radiometric calibration results across the current mission lifetime for each band.</p> "> Figure 25
<p>OLI Telemetry Mission Trends.</p> "> Figure 26
<p>OLI stability relative to desert calibration sites.</p> "> Figure 27
<p>Comparisons of OLI operational radiometric calibrations to other measurements [<a href="#B2-remotesensing-06-12275" class="html-bibr">2</a>,<a href="#B3-remotesensing-06-12275" class="html-bibr">3</a>]: (<b>a</b>) Reflectance calibration; (<b>b</b>) Radiance calibration.</p> "> Figure 28
<p>OLI inferred solar irradiance as compared to ChKur solar irradiance. Estimated uncertainty in OLI derived solar irradiance is shown.</p> ">
Abstract
:1. Introduction
Band (#) | Band Name | Center Wavelength (nm) | Bandwidth (nm) | IFOV (m) | Active Detectors (#) |
---|---|---|---|---|---|
1 | Coastal-Aerosol (CA) | 443 | 16 | 30 | 6916 (14 × 494) |
2 | Blue | 482 | 60 | 30 | 6916 |
3 | Green | 561 | 57 | 30 | 6916 |
4 | Red | 655 | 38 | 30 | 6916 |
5 | Near-IR (NIR) | 865 | 28 | 30 | 6916 |
6 | Short Wave IR-1 (SWIR 1) | 1609 | 85 | 30 | 6916 |
7 | SWIR 2 | 2201 | 187 | 30 | 6916 |
8 | Panchromatic (Pan) | 590 | 172 | 15 | 13,832 (14 × 988) |
9 | Cirrus | 1373 | 20 | 30 | 6916 |
2. OLI In-Flight Calibration Capabilities
2.1. OLI Data and Data Processing
2.2. On-Board Lamps
2.3. Solar Diffusers
2.4. Lunar Calibration
3. OLI Pre-Launch Radiance Calibration and Stability Monitoring
3.1. Absolute Radiance Calibration
3.2. Radiometric Stability
4. OLI Pre-Launch Reflectance Calibrator Characterization
5. OLI Transfer to Orbit Experiment
- (1)
- The OLI diffusers would be the most stable artifact through the launch and early-on-orbit period (as opposed to the lamps or the OLI itself)
- (2)
- The OLI response to the diffusers could be measured both prior to launch and on-orbit using the same illumination source (namely the sun).
- (3)
- The transmission losses in the atmosphere and heliostat could be sufficiently quantified for all but the cirrus band to allow the pre-launch test to be corrected to post launch conditions with an uncertainty of a few percent or less.
- Lband, helio is the OLI measured radiance of the diffuser illuminated via the heliostat using OLI’s
- DSS-based calibration
- Tband, helio is the measured transmittance of heliostat
- Tband, atm is the transmittance of the atmosphere at the time of the heliostat measurement.
- DE − S, helio is the Earth-Sun distance at the time of the heliostat observation
- DE − S, orbit is the Earth-Sun distance at the time of the on-orbit observation.
- Tband, atm was calculated using MODTRAN [25] based on the atmospheric pressure (Trayleigh), the solar radiometer measurements (Taerosol) and atmospheric gaseous abundances (Tgases).
CA | Blue | Green | Red | NIR | SWIR 1 | SWIR 2 | Pan | |
---|---|---|---|---|---|---|---|---|
Mean | 1.033 | 1.040 | 1.029 | 1.007 | 1.038 | 1.045 | 1.051 | 1.041 |
StDev | 0.009 | 0.007 | 0.005 | 0.006 | 0.005 | 0.004 | 0.005 | 0.005 |
6. OLI On-Orbit Reflectance Calibration
- Gr[d] is the reflectance Gain for detector d in DN/unit reflectance
- BRF[d] is the Bidirectional Reflectance Factor of the solar diffuser for the illumination and view angles for detector d
- θ s,SD is the solar incidence angle for this diffuser observation (always 45°)
- QSDnetlin[d] is the bias and non-linearity corrected response for detector d to the diffuser for this observation in DN
- DES,SD is the Earth-Sun distance for this diffuser observation in Astronomical Units
7. Calibration Updates
- (1)
- The cirrus band reflectance calibration was corrected for an error found in the calculation; this increased the reflectance in this band by ~7%, without changing the radiance calibration.
- (2)
- The SWIR-2 band gain was adjusted for the difference observed prior to launch between the BATC and validation results; this increased the radiance by about 3%, without changing reflectance calibration.
- (3)
- All bands gains were adjusted to correct for an outdated radiance file that was mistakenly used in the pre-launch calibration data reduction for all bands; this decreased the radiance in all bands by ~2%, without changing the reflectance calibration.
- (4)
- The precision of the radiance to reflectance conversion was improved for all bands; this was a less than 1% effect in reflectance.
8. Stability over Mission
- (1)
- Two of the early lunar acquisitions were at ~−7° phase angle versus the ~+7° phase angle for the remainder of the collects. These two negative phase angle points are outliers.
- (2)
- A safe hold event at about 0.6 years into the mission power cycled the OLI instrument. This resulted in small changes in the responsivity of bands 1–4 and 8 of 0.1 to 0.2%. This change in response recovered about 0.1 years later for reasons currently unknown.
- (1)
- The OLI average response is stable to within <1% over the first 1.6 years.
- The CA band has the largest change, about 1.0%, and the change is consistent across all of the calibration techniques.
- (2)
- Different calibration sources generally agree to within a few tenths of a percent in the trends.
- Several bands show a 0.1% to 0.3% increase in response apparent in the lamp and solar data, but less so in the lunar data.
- The origin of lower precision of the lunar results, particularly in the SWIR bands, is currently under study and may be related to the integration technique.
- (3)
- The divergence in the lamp and working diffuser results may indicate that the working diffuser is brightening, particularly in the NIR and SWIR.
- (1)
- The telemetry indicates that constant current is maintained through the stim lamps at the better than ±0.01% level.
- (2)
- The OLI focal plane has been maintained to better than ±0.1 °C across the mission; some seasonal variation is evident.
- (3)
- The lamp voltage varies by about ±0.1% across the mission according to the telemetry and shows a correlation to the instrument temperature.
- (4)
- The photodiode monitors show an apparent decrease in the stim lamp radiance of ~0.5% over the mission lifetime and there is apparent temperature sensitivity in the trend. Adjusting for the temperature sensitivity (Figure 14d) removes most of the structure, but leaves a nearly linear degradation with time. The spectrally comparable OLI panchromatic band (band 8) shows a flat to slightly increasing response to the working stim lamps, which closely matches the response to the diffusers, suggesting the change is in the monitoring circuit.
9. Comparison to Other Measurements
10. Discussion
11. Conclusions
Acknowledgments
Author Contributions
Conflicts of Interest
References
- Knight, E.J.; Kvaran, G. Landsat-8 Operational Land Imager design, characterization, and performance. Remote Sens. 2014, 6, 10286–10305. [Google Scholar] [CrossRef]
- Czapla-Myers, J.; McCorkel, J.T.; Anderson, N.; Thome, K.J.; Bigger, S.; Helder, D.L.; Aaron, D.; Leigh, L.; Mishra, N. The ground-based absolute radiometric calibration of Landsat 8 OLI. Remote Sens. 2014, in press. [Google Scholar]
- Mishra, N.; Haque, M.O.; Leigh, L.; Aaron, D.; Helder, D.L.; Markham, B.L. Radiometric cross calibration of Landsat 8 Operational Land Imager (OLI) & Landsat 7 Enhanced Thematic Mapper Plus (ETM+). Remote Sens. 2014, in press. [Google Scholar]
- Kvaran, G.; Markham, B.; Zalewski, E. Overview of the radiometric calibration of the Operational Land Imager (OLI). In Proceedings of the 19th Annual Conference on Characterization and Radiometric Calibration for Remote Sensing (CalCon), Logan, UT, USA, 23–26 August 2010.
- Tsai, B.; Allen, D.; Cooksey, C.; Yoon, H.; Fulton, L.; Kvaran, G.; Biggar, S.; Markham, B. Calibration of Landsat/OLI panels using the NIST BRDF scale from 1100 to 2400 nm. In Proceedings of the 19th Annual Conference on Characterization and Radiometric Calibration for Remote Sensing (CalCon), Logan, UT, USA, 23–26 August 2010.
- Cutlip, H.; Chase, H.; Fulton, L.; Schneider, L.; Pierce, A.; Gleeson, D.; Kvaran, G.; Canova, B.; Johnson, C.; Saunders, R.; et al. OLI radiometric calibration: Ground support equipment characterization and calibration. In Proceedings of the 19th Annual Conference on Characterization and Radiometric Calibration for Remote Sensing (CalCon), Logan, UT, USA, 23–26 August 2010.
- Collins, S.; Good, W.; Cutlip, H.; Seerveld, J.; Kaptchen, P.; Kvaran, G. Operational Land Imager: Using the Sun as the source for ground based radiometric calibration. In Proceedings of the 19th Annual Conference on Characterization and Radiometric Calibration for Remote Sensing (CalCon), Logan, UT, USA, 23–26 August 2010.
- Kvaran, G.; Markham, B.; Leisso, N.; Czapla-Myers, J.; Biggar, S. OLI radiometric calibration. In Proceedings of the 20th Conference on Characterization and Radiometric Calibration for Remote Sensing (CalCon), Logan, UT, USA, 29 August–1 September 2011.
- Biggar, S.; Anderson, N.; Fulton, L.; Kvaran, G.; Kortmeyer, H. Reflectance factor measurements of the OLI flight diffusers. In Proceedings of the 20th Conference on Characterization and Radiometric Calibration for Remote Sensing (CalCon), Logan, UT, USA, 29 August–1 September 2011.
- Cutlip, H.; Cole, J.; Johnson, B.C.; Maxwell, S.; Markham, B.; Ong, L.; Hom, M.; Biggar, S. The OLI radiometric scale realization round robin measurement campaign. In Proceedings of the 20th Conference on Characterization and Radiometric Calibration for Remote Sensing (CalCon), Logan, UT, USA, 29 August–1 September 2011.
- Kvaran, G.; Knight, E.; Ansari, K.; Lee, K.; Canova, B.; Donley, B.; Kieffer, H. Landsat-8 Operational Land Imager (OLI) initial on-orbit performance. In Proceedings of the 22th Conference on Characterization and Radiometric Calibration for Remote Sensing (CalCon), Logan, UT, USA, 19–22 August 2013.
- Vanderwerff, K.; Morfitt, R. Bias estimation for the Landsat 8 Operational Land Imager. Prof. SPIE 2011, 8153. [Google Scholar] [CrossRef]
- Morfitt, R.; Markham, B.L.; Micijevic, E.; Scaramuzza, P.; Barsi, J.A.; Levy, R.; Ong, L.; Vanderwerff, K. OLI radiometric performance on-orbit. Remote Sens. 2014, in press. [Google Scholar]
- Eplee, R.; Barnes, R.; Patt, F.; Meister, G.; McClain, C. SeaWIFS lunar calibration methodology after six years on orbit. Proc. SPIE 2004. [Google Scholar] [CrossRef]
- Sun, J.; Xiong, X.; Barnes, W.; Guenther, B. MODIS reflective band on-orbit lunar calibration. IEEE Trans. Geosci. Remote Sens. 2007, 7, 2383–2393. [Google Scholar] [CrossRef]
- Mendenhall, J.; Lencioni, D.; Evans, J.B. Spectral and radiometric calibration of the advanced land imager. Linc. Lab. J. 2005, 15, 207–224. [Google Scholar]
- Lacherade, S.; Aznay, O.; Fougnie, B. POLO Pleiades Orbital Lunar Observations: Intensive study of the Moon and comparison to ROLO model. In Proceedings of the 22th Conference on Characterization and Radiometric Calibration for Remote Sensing (CalCon), Logan, UT, USA, 19–22 August 2013.
- Kieffer, H.H.; Stone, T.C. The spectral irradiance of the Moon. Astron. J. 2005, 129, 2887–2901. [Google Scholar] [CrossRef]
- University of Arizona, College of Optical Sciences, Remote Sensing Group, Transfer Radiometers. Available online: http://fp.optics.arizona.edu/RSG/resources.php?content=transfer (accessed on 15 May 2014).
- Butler, J.; Barnes, R. The use of transfer radiometers in validating the visible to shortwave infrared calibrations of radiance sources used by instruments in NASA’s Earth Observing System. Metrologia 2003, 40, S70–S77. [Google Scholar] [CrossRef]
- Brown, S.W.; Eppledauer, G.P.; Lykke, K.R. Facility for Spectral Irradiance and Radiance Responsivity Calibrations using Uniform Sources (SIRCUS). Appl. Opt. 2006, 45, 8218–8237. [Google Scholar] [CrossRef] [PubMed]
- Landsat-8 Calibration Parameter Files (CPF) and Bias Parameter Files (BPF). Available online: http://landsat.usgs.gov/cpfbpf.php (accessed 1 June 2014).
- Kuester, M.A.; Czapla-Myers, J.; Kaptchen, P.; Good, W.; Lin, T.; Biggar, S.; Thome, K. Development of a heliostat facility for solar-radiation-based calibration of Earth observing sensors. Proc. SPIE 2008. [Google Scholar] [CrossRef]
- Czapla-Myers, J.; Thome, K.; Anderson, N.; McCorkel, J.; Leisso, N.; Good, W.; Collins, S. Transmittance measurement of a heliostat facility used in the preflight radiometric calibration of Earth-observing sensors. Proc. SPIE 2009. [Google Scholar] [CrossRef]
- Berk, A.; Anderson, G.; Acharya, P.; Shettle, E. MODTRAN 5.2.0.0. User’s Manual; Air Force Geophysical Laboratory: Hanscom AFB, MA, USA, 2008; p. 100. [Google Scholar]
- Space Based Measurements of Ozone and Air Quality in the Ultraviolet and Visible. Available online: http://ozoneaq.gsfc.nasa.gov/ozone_overhead_all_v8.md (accessed on 1 May 2014).
- Butler, J.J.; Johnson, B.C.; Barnes, R.A. The calibration and characterization of Earth remote sensing and environmental monitoring instruments. In Optical Radiometry; Parr, A.C., Datla, R.U., Gardner, J.L., Eds.; Elsevier Academic Press: New York, NY, USA, 2005; pp. 488–494. [Google Scholar]
- Cosnefroy, H.; Leroy, M.; Briottet, X. Selection and characterization of Saharan and Arabian desert sites for the calibration of optical satellite sensors. Remote Sens. Environ. 1996, 58, 101–114. [Google Scholar] [CrossRef]
- Markham, B.L.; Thome, K.J.; Barsi, J.A.; Kaita, E.; Helder, D.L.; Barker, J.L.; Scaramuzza, P.L. Landsat-7 ETM+ on-orbit reflective-band radiometric stability and absolute calibration. IEEE Trans. Geosci. Remote Sens. 2004, 42, 2810–2820. [Google Scholar] [CrossRef]
© 2014 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/4.0/).
Share and Cite
Markham, B.; Barsi, J.; Kvaran, G.; Ong, L.; Kaita, E.; Biggar, S.; Czapla-Myers, J.; Mishra, N.; Helder, D. Landsat-8 Operational Land Imager Radiometric Calibration and Stability. Remote Sens. 2014, 6, 12275-12308. https://doi.org/10.3390/rs61212275
Markham B, Barsi J, Kvaran G, Ong L, Kaita E, Biggar S, Czapla-Myers J, Mishra N, Helder D. Landsat-8 Operational Land Imager Radiometric Calibration and Stability. Remote Sensing. 2014; 6(12):12275-12308. https://doi.org/10.3390/rs61212275
Chicago/Turabian StyleMarkham, Brian, Julia Barsi, Geir Kvaran, Lawrence Ong, Edward Kaita, Stuart Biggar, Jeffrey Czapla-Myers, Nischal Mishra, and Dennis Helder. 2014. "Landsat-8 Operational Land Imager Radiometric Calibration and Stability" Remote Sensing 6, no. 12: 12275-12308. https://doi.org/10.3390/rs61212275
APA StyleMarkham, B., Barsi, J., Kvaran, G., Ong, L., Kaita, E., Biggar, S., Czapla-Myers, J., Mishra, N., & Helder, D. (2014). Landsat-8 Operational Land Imager Radiometric Calibration and Stability. Remote Sensing, 6(12), 12275-12308. https://doi.org/10.3390/rs61212275