Ultrasonics Sonochemistry 16 (2009) 692–697
Contents lists available at ScienceDirect
Ultrasonics Sonochemistry
journal homepage: www.elsevier.com/locate/ultsonch
Cavitation occurrence around ultrasonic dental scalers q
Bernhard Felver a, David C. King b, Simon C. Lea a, Gareth J. Price b,*, A. Damien Walmsley a,*
a
b
School of Dentistry, University of Birmingham, St. Chad’s Queensway, Birmingham, B4 6NN, UK
Department of Chemistry, University of Bath, Bath BA2 7AY, UK
a r t i c l e
i n f o
Article history:
Received 22 September 2008
Received in revised form 2 November 2008
Accepted 3 November 2008
Available online 13 November 2008
Keywords:
Sonochemiluminescence
Scanning laser vibrometry
Dental instruments
Cavitation
a b s t r a c t
Ultrasonic scalers are used in dentistry to remove calculus and other contaminants from teeth. One
mechanism which may assist in the cleaning is cavitation generated in cooling water around the scaler.
The vibratory motion of three designs of scaler tip in a water bath has been characterised by laser vibrometry, and compared with the spatial distribution of cavitation around the scaler tips observed using
sonochemiluminescence from a luminol solution. The type of cavitation was confirmed by acoustic emission analysed by a ‘Cavimeter’ supplied by NPL. A node/antinode vibration pattern was observed, with the
maximum displacement of each type of tip occurring at the free end. High levels of cavitation activity
occurred in areas surrounding the vibration antinodes, although minimal levels were observed at the free
end of the tip. There was also good correlation between vibration amplitude and sonochemiluminescence
at other points along the scaler tip. ‘Cavimeter’ analysis correlated well with luminol observations, suggesting the presence of primarily transient cavitation.
Ó 2008 Elsevier B.V. All rights reserved.
1. Introduction
Ultrasonic scalers are used in dentistry to remove both subgingival (below gumline) and supragingival (above gumline) deposits
from teeth. Compared with traditional hand instrumentation techniques, ultrasonic scalers have the primary advantage of significantly reduced treatment duration [1,2]. The design of scalers
varies between manufacturers, but as shown in Fig. 1 is generally
a J-shaped metal (often titanium) tip, approximately 25 mm in
length, attached to a handpiece that is manipulated by the clinician. During use, a transducer in the handpiece induces vibration
at ultrasonic frequencies in the tip, and the side of the free end
of the tip is placed against the tooth, mechanically removing
deposits on the surface. Frictional heating of the tooth is reduced
by passing an irrigant solution through a small outlet on the underside of the tip [3]. It has been suggested that cavitation might occur
in the cooling water as it flows over the tip and that this might aid
the cleaning process [4].
Inertial cavitation occurs in liquids when the acoustic pressure
during rarefaction becomes more negative than the saturated vapour pressure of the liquid, leading to rupture of the liquid and formation of cavities [5]. A sufficient negative pressure, in pure water
q
This work was presented at the 11th meeting of the European Society of
Sonochemistry at La Grande-Motte, France, June 2008.
* Corresponding authors. Tel.: +44 1225 386504.
E-mail addresses: bxf525@bham.ac.uk (B. Felver), d.c.king@bath.ac.uk (D.C. King),
s.lea@bham.ac.uk (S.C. Lea), g.j.price@bath.ac.uk (G.J. Price), a.d.walmsley@bham.
ac.uk (A.D. Walmsley).
1350-4177/$ - see front matter Ó 2008 Elsevier B.V. All rights reserved.
doi:10.1016/j.ultsonch.2008.11.002
of the order of 100–120 MPa, can be induced by ultrasonic scalers,
such that cavitation will occur in liquid surrounding the tip [6].
Collapse of these cavities results in localised extreme conditions
of pressure and temperature, up to 5000 K and 500 bar, which
can in turn lead to pyrolysis of species which can volatilise into
the bubble from solution [7]. Sonolysis products of water, specifically hydroxyl radicals and oxygen atoms, are highly oxidising species, as are secondary products such as hydrogen peroxide. Irrigant
solutions often contain hydrogen peroxide to enhance the cleaning
process. The physical effects of cavity collapse are also thought to
aid surface cleaning, via the formation of microjets that impact
on a surface when collapse occurs within close proximity [8].
In order to assess the role that cavitation plays in deposit removal from teeth, it is first necessary to identify the distribution
of cavitation around each tip, so that possible differences due to
variations in tip design can be elucidated. No data on cavitation
distribution around ultrasonic scaler tips exists in the literature.
Our hypothesis was that regions of high cavitation activity would
be associated with the tip regions that have the greatest displacement during scaler operation.
Scanning laser vibrometry, SLV, has successfully been used to
analyse the vibrational motion of ultrasonic scaler tips [9,10].
Briefly, the technique works by measuring the Doppler shift of a laser beam that is reflected off the target surface, giving the velocity
of the surface with respect to the incident beam. There exist a wide
variety of methods for measuring cavitation activity [11–13].
Luminol photography was employed for this work [14], since it
gives spatially-resolved information on the occurrence of cavitation. Luminol sonochemiluminescence occurs through reaction
B. Felver et al. / Ultrasonics Sonochemistry 16 (2009) 692–697
693
Fig. 2. Example of scan points layout for laser vibrometry, in this case along the P
tip.
the PS tip. The maximum displacement of the tip at each scan point
was measured and the average of ten cycles was recorded for each
tip, at each power setting.
2.2. Luminol photography
Fig. 1. A, P and PS tips, from left to right. The free end of the tip is placed side-on to
the tooth during treatment. The threaded end of the tip is attached to the
handpiece, containing an ultrasonic transducer.
with pyrolysis products of cavitation in aqueous systems. A longexposure photograph can be used to assess regions of cavitation
activity in a solution. A ‘Cavimeter’ [13], recently developed by
the National Physical Laboratory, NPL, was used to analyse the
acoustic output from a scaler tip in solution. A measure of the
intensity of the driving frequency and the cavitation activity were
obtained in order to correlate with luminescence measurements.
A full analysis of the role that cavitation might play needs the
determination of scaler behaviour under a wide range of conditions. As a first step, we relate in this work the cavitation output
from three different styles of dental tips, measured with a ‘Cavimeter’ from NPL, and observed via luminol photography, with the
vibration characteristics measured by SLV in order to determine
the spatial distribution of cavitation.
2. Experimental
All reagents were analytical grade and obtained from Sigma–Aldrich Co., UK, unless otherwise stated, and were used as received.
The ultrasonic scaler was a Piezon miniMaster provided by Electro
Medical Systems, Nyon, Switzerland, and was supplied with A, P
and PS tips, as shown in Fig. 1. The scaler operates at a nominal frequency of 30 kHz, and can be set to any of ten incremental power
settings from a control panel. For this work, power settings of 1/10,
5/10 and 10/10 were used.
2.1. Assessment of tip vibration characteristics
The vibrational motion of each tip was characterised by SLV as
described by Lea et al. [10]. A tip was clamped into position so that
it was fully-immersed in a water bath. Care was taken to ensure
that the position was reproducible between different experiments.
The SLV system was a PSV 300-F/S from Polytec GmbH, PolytecPlatz, Waldbronn, Germany and used a class II, He–Ne laser operating
at 632.8 nm. The motion in the plane of the tip was measured; any
lateral motion is not reported here. It has been measured and is
significantly smaller than the longitudinal vibration, which is commonly used as a measure of the performance or output of the scaler
tip. A number of equally-spaced scan points were chosen along the
length of each tip, from the free end to as close to the handpiece as
could be measured, as in Fig. 2. Variations in the length and curvature of each tip led to differences in the distance between scan
points such that the inter-point distance was 0.54 mm for the A
tip, 0.44 mm for the P tip and 0.55 mm for the PS tip, and the number of scan points was 15 for the A tip, 23 for the P tip and 19 for
Observation of cavitation activity was from the luminescence of
a luminol solution consisting of 1 10 3 mol dm 3 luminol (5amino-2,3-dihydro-1,4-phthalazinedione),
1 10 4 mol dm 3
hydrogen peroxide and 1 10 4 mol dm 3 EDTA (ethylenediaminetetraaceticacid), prepared in a deionised water solution and
made up to pH 12 with the addition of sodium hydroxide [14].
A scaler tip was immersed in 50 cm3 of luminol solution in a
flat-sided glass container, and clamped into position in sight of a
Canon EOS 30D digital camera, set to ISO 3200 sensitivity. The Cavimeter sensor and the tip handpiece were clamped onto the same
stand and kept in the same position for each experiment. A Canon
EF 100 mm f/2.8 USM Macro lens was used for imaging at the
close-focus distance, corresponding to 1:1 reproduction. The apparatus was sealed in a light-proof box, and light emission recorded
for 30 s of scaler operation. The procedure was repeated for each
tip design, at low, medium and high power settings, with
freshly-made luminol solution used in each case. Daylight photographs were also taken. The total intensity of the luminol emission
was calculated after subtraction of background levels using ImageJ
software [15].
2.3. ‘Cavimeter’
The ‘Cavimeter’ was developed by NPL [16] to provide a quantitative estimate of cavitation activity in a wide variety of circumstances. It consists of a strip of piezoelectric material embedded
in a sound absorbing plastic cylinder. The design of the sensor is
such that the response arises solely from activity inside this cylinder. The signal produced is integrated over a 2 s period and analysed in three ways; the ‘direct field’ value correlates the
response at the fundamental driving frequency (in this case
30 kHz) arising from the source as well as linearly oscillating
bubbles. The ‘subharmonic’ measurement is recorded at frequencies of one-half and one-quarter of the fundamental and is indicative of the onset of transient cavitation. This type of cavitation is
also quantified by measuring the ‘cavitation’ signal from emission
at high frequencies between 1.5 and 5 MHz which arise from shock
waves emitted by the collapse of transient bubbles [17,18]
although stable cavitation can also give signals in this region.
The scaler tip under investigation was surrounded by the cylindrical ‘Cavimeter’ sensor, placed into a glass beaker and immersed
in 50 cm3 of freshly-drawn deionised but not degassed water. The
‘Cavimeter’ monitor was attached to the sensor, and the zero levels
set. Care was taken to ensure that the scaler tip was completely immersed in water and positioned in the same place relative to the
cavimeter for each experiement. The handpiece was aligned parallel to the cylindrical axis of the sensor and inserted so that the top
and bottom of the exposed tip were equidistant from the top and
694
B. Felver et al. / Ultrasonics Sonochemistry 16 (2009) 692–697
bottom of the sensor. The ultrasonic scaler was switched on for
approximately 10 s, until the ‘Cavimeter’ readings were stable.
An average of three successive measurements of the ‘direct field’
and ‘cavitation’ (i.e high frequency output) measurements were recorded while changing the power. No hysteresis effects were noted
on increasing and decreasing the power. The procedure was repeated for each tip design, at all power settings and with freshlydrawn water in each case.
3. Results and discussion
3.1. Vibration characteristics of ultrasonic scaler tips
The measured maximum displacement amplitudes of the ultrasonic scaler tips are shown in Fig. 3. Only the magnitude of the displacement is presented, the sign is omitted for clarity. The
direction of motion of the end of the tip is opposite that of the middle section. The exaggerated displacement of a P tip at extremes of
displacement can be seen in the computer reconstruction of the
SLV data in Fig. 4. The vibrational antinode with the greatest displacement amplitude was found to always occur at the free end
of each tip. Maximum displacement amplitudes at power 10/10
were measured to be 35.7 ± 6.3 lm for the A tip, 21.9 ± 3.9 lm
for the P tip and 28.6 ± 7.4 lm for the PS tip. A further antinode
Fig. 4. Computer generated, exaggerated images of a P tip at maximum
displacement.
was found to occur for each tip, 7.6 mm, 9.7 mm and 6.1 mm from
the free end for A, P and PS tips, respectively.
3.2. Luminol photography of ultrasonic scaler tips
The effect of the vibration pattern measured by SLV can be seen
in the optical photographs illustrated in Fig. 5, taken during tip
operation. The cloud of cavitation bubbles is localised at points
along the tips. Scaler operation also caused significant degassing
with large air bubbles appearing after only 10–15 s of operation
at the highest power. The position of the cavitation bubble cloud
correlates well with the maximum displacements shown in
Fig. 3, except that there appears to be little cavitation generated
at the free ends of the scaler tips.
To confirm that these are cavitation bubbles, emission from
luminol was recorded under the same conditions and the results
are shown in Fig. 6. Although all power settings were used, only
photographs from the highest power setting experiments are
shown. Data from other experiments are presented in Table 1.
The data shows intense regions of activity surrounding the bend
for each tip, while little to no activity was observed at the free
end of the tip. The intensity of all pixels in each image was
summed and the results are presented in Table 1. As expected,
higher power settings produced the greatest intensity, and therefore the greatest amount of cavitation, for each tip. Comparison
of Figs. 4 and 5 shows that the spatial distribution of luminol emission is similar to the bubble clouds observed during scaler
operation.
3.3. ‘Cavimeter’ analysis
Fig. 3. Maximum displacement amplitude for each scan point of each tip. (a) A tip,
(b) P tip, (c) PS tip. The origin on the displacement axis corresponds to the ‘angle’ of
the scaler tip; the largest distance corresponds to the free end.
Monitoring luminol emission cannot differentiate between ‘stable’ and ‘transient’ (or inertial) cavitation. The latter would be
expected to be dominant in the type of sound field produced by
a scaler. There was a good correlation between the ‘direct field’
695
B. Felver et al. / Ultrasonics Sonochemistry 16 (2009) 692–697
Fig. 5. Photographs of scaler tips in operation at power 10/10. (a) A tip, (b) P tip, (c)
PS tip.
measurements from the ‘Cavimeter’ and the luminol emission
showing that higher power settings produced larger amounts of
cavitation. ‘Cavimeter’ readings also produced some evidence of
sub-harmonic signals although there was considerable variation
in the values obtained. More convincing evidence of the presence
of transient cavitation is presented in Fig. 7 which shows a comparison of ‘Cavimeter’ measurements of transient cavitation (i.e. high
frequency broadband signals) and luminol emission measurements
for the three tips. The absolute measurements have been scaled
relative to the maximum signal.
While the agreement is not perfect, both measurements show
the same trend in increase, strongly suggesting that the acoustic
emission and luminol chemiluminescence have the same origin,
Fig. 6. Luminol photography of A, P and PS tips at power 10/10. Light regions
indicate areas of high cavitation activity, with dark regions indicating little or no
activity.
Table 1
Relative intensity of luminol emission.
Relative intensity of luminol emission
Tip
A
P
PS
Power 1/10
3.31
1.00
1.00
Power 5/10
8.64
2.05
1.15
Power 10/10
11.54
12.85
5.92
i.e. transient cavitation bubbles generated around the vibrating
scaler. The agreement for the PS tips is less satisfactory although
it should be noted that this is not unexpected since the absolute
values of both luminol and acoustic emissions were by far the lowest for this tip and the variation in signals much smaller.
696
B. Felver et al. / Ultrasonics Sonochemistry 16 (2009) 692–697
compare the paddle of an oar to the shaft and the difference in effect is plain. Therefore, the wider cross sections of the A and P tips
would be expected to displace larger volumes of water than the
cylindrical PS tip, leading to the observed difference in cavitation
activity between the tips. The highest levels of cavitation activity
were observed around vibration antinodes close to the bend in
each tip. Surprisingly, while the displacement amplitude was
greatest at the free end of the tip, little to no cavitation was observed at the free end using luminol photography.
A similar explanation may apply here. Each tip tapers towards
the free end to a narrow end. Thus, even though the end shows
large motion, its shape may mean that water flows easily around
it so that the large negative pressures required for cavitation are
not generated in its wake.
In typical clinical use, the tip of the scaler is contacted against a
tooth and so applies a load at the free end of the tip orthogonal to
the direction of vibration. Further work is underway to ascertain
the effects of clinically-relevant loading of the scaler. While the
work reported here concerns unloaded tips, it is relevant where
scalers may be operated in pools of irrigant or liquid between teeth
and gums. It is also possible that physical effects, such as streaming, may be more relevant to cleaning efficiency than cavitation
activity and measurements of these factors will be reported in
due course. With the current drive towards slimmer scaler tips
[20], an understanding of how the shape and design affects the
generation of cavitation is important and the results reported here
provide a base for comparison of measurements under other
conditions.
4. Conclusion
A good correlation has been found between vibrational antinodes occurring during operation of ultrasonic scalers with cavitation activity around the scaler tip. Significant differences in the
amount of cavitation generated have been found between different
tip designs which tentatively suggest an influence of tip profile on
the resultant cavitation. Despite showing the largest motion, little
cavitation was detected at the free end of the tips and this again
may be due to the shape of the tip at its end. Good agreement
was found between optical, acoustic emission and luminol emission methods for detecting and measuring cavitation.
Acknowledgements
Fig. 7. A comparison of the results obtained via the ‘Cavimeter’ device and by
luminol photography.
We gratefully acknowledge the Engineering and Physical Sciences Research Council for funding via grants EP/C536894/1 and
EP/C536908/1. We would also like to acknowledge the loan of a
Piezon miniMaster ultrasonic scaler and set of tips by Electro
Medical Systems. We are also grateful to Dr. Mark Hodnett at the
National Physical Laboratory, UK for the loan of a ‘Cavimeter’.
3.4. Further discussion
It has been suggested that cavitation can be generated in the irrigant solution that is pumped over the surface of the tip [19]. This
work has shown that use of higher generator power settings increases cavitation activity over that of low power settings, in line
with previous observations [6], and with both luminol and
‘Cavimeter’ techniques. However, the amount of cavitation
produced depended on the tip employed. Higher levels of cavitation activity were observed around the A and P tips, with only half
as much observed around the PS tip. The almost-cylindrical cross
section of the PS tip may be a factor in this difference. A cylinder
is a remarkably ineffective shape for displacement of water (and
hence generation of negative pressures), whereas a thin, rectangular cross section is significantly more effective. As a simple analogy,
References
[1] A.D. Walmsley, S.C. Lea, G. Landini, A.J. Moses, J. Clin. Periodontol. 35 (S8)
(2008) 22–28.
[2] J. Tunkel, A. Heinecke, T.F. Flemmig, J. Clin. Periodontol. 29 (2002) 72–81.
[3] S.C. Lea, G. Landini, A.D. Walmsley, J. Clin. Periodontol. 31 (2004) 370–375.
[4] A.D. Walmsley, W.R.E. Laird, A.R. Williams, J. Dent. Res. 63 (1984) 1162–1165.
[5] T.G. Leighton, The Acoustic Bubble, Academic Press, London, England, 1994.
[6] S.C. Lea, G.J. Price, A.D. Walmsley, Ultrason. Sonochem. 12 (2005) 233–236.
[7] D. Krefting, R. Mettin, W. Lauterborn, Ultrason. Sonochem. 11 (2004) 119–123.
[8] A. Henglein, Ultrason. Sonochem. 2 (1995) S115–S121.
[9] S.C. Lea, G. Landini, A.D. Walmsley, J. Clin. Periodontol. 30 (2003) 505–510.
[10] S.C. Lea, G. Landini, A.D. Walmsley, J. Sound Vib. 271 (2004) 1113–1120.
[11] T.G. Leighton, P.R. Birkin, M. Hodnett, B. Zeqiri, J.F. Power, G.J. Price, T. Mason,
M. Plattes, N. Dezhkunov, A.J. Coleman, in: A.A. Doinikov (Ed.), Bubble and
Particle Dynamics in Acoustic Fields: Modern Trends and Applications,
Research Signpost, Trivandrum, India 2005, pp. 37–94.
B. Felver et al. / Ultrasonics Sonochemistry 16 (2009) 692–697
[12]
[13]
[14]
[15]
X. Fang, G. Mark, C. von Sonntag, Ultrason. Sonochem. 3 (1996) 57–63.
M. Hodnett, R. Chow, B. Zeqiri, Ultrason. Sonochem. 11 (2004) 441–454.
H.N. McMurray, B.P. Wilson, J. Phys. Chem. A 103 (1999) 3955–3962.
M.D. Abramoff, P.J. Magelhaes, S.J. Ram, Biophotonics Int. 11 (2004)
36–42.
[16] B. Zeqiri, P.N. Gelat, M. Hodnett, N.D. Lee, IEEE Trans. Ferro. Ultrasonics Freq.
Control 50 (2003) 1342.
697
[17] C.-D. Ohl, T. Kurz, R. Geisler, O. Lindau, W. Lauterborn, Phil. Trans. R. Soc. Lond.
A 357 (1999) 269–294.
[18] M. Ashokkumar, M. Hodnett, B. Zeqiri, F. Grieser, G.J. Price, J. Amer. Chem. Soc.
129 (2007) 2250–2258.
[19] A.D. Walmsley, W.R.E. Laird, A.R. Williams, J. Clin. Periodontol. 15 (1988) 539–
543.
[20] M. Sanz, W. Teughels, J. Clin. Periodontol. 35 (S8) (2008) 3–7.