zyxwvutsrqponmlkjihg
zyxwvutsrqponm
zyxwvutsr
zyxwvutsrqponmlk
zyxwvutsrqponmlkjih
Artificicii Orgrins
20(6):618-4?0,Blackwell Science. Inc., Boston
R1 IYYh Intel-national Society for Artificial Organs
zyxw
Controller for an Axial Flow Blood Pump
"Hiroaki Konishi, *tJames F. Antaki, "lDevin V. Amin, tJ.R. Boston,
lJohn P. Kerrigan, tWilliam A. Mandarino, "Philip Litwak, "Kenji Yamazaki,
"Mahender Macha, $Kenneth C. Butler, *?Harvey S. Borovetz, and
"Robert L. Kormos
zyxwvuts
University of Pittsburgh, Artuirinl. Heart crnd tirng Program, Schools ~ f ' * M e d i c i nae n d ?Engineering, Pittsburgh,
Pennsvlwinia; and $Nirnhii.v, Inc.. Rancho Cordovu, Calijbrnia, U . S . A .
Abstract: A rotary blood pump inherently provides only
one noninvasive "observable" parameter (motor current)
and allows for only one "controllable" parameter (pump
speed). 'Yo maintain the systemic circulation properly, the
pump speed must be controlled to sustain appropriate
outlet flows and perfusion pressure while preventing pulmonary damage caused by extremes in preload. Steadystate data were collected at repeated intervals during
chronic trials of the Nimbus AxiPump (Nimbus, Inc.,
Rancho Cordova, California, U.S.A.) in sheep (n = 7)
and calves (n = 12). For each data set, the pump speed
was increased at increments of 500 rpm until left ventricular and left atrial emptying was observed by left atrial
pressure diminishing to zero. The effect of decreasing
preload was evaluated perioperatively by inferior vena
cava occlusion at a constant pump speed. Fourier analysis established a relationship between changes in the
pump preload and the power spectra of the pump current
waveform. Based on these results, a control method was
devised to avoid ventricular collapse and maintain the
preload within a physiologic range. The objective of this
controller is the minimization of the second and third harmonic of the periodic current waveform. This method is
intended to provide a noninvasive regulation of the pump
by eliminating the need for extraneous transducers. Key
Words: Axial flow blood pump-Noninvasive pump regulation.
The axial flow blood pump design has numerous
advantages for the next generation of left ventricular assist devices (LVAD): small size, high efficiency, and simplicity of implantation (I). These
and other advantages are bought at the expense of
inherently greater demand for external control. Unlike their positive displacement counterparts, these
devices intrinsically provide only one observable
parameter, pump current, and one controllable parameter, pump speed. The characteristic performance of this class of pumps dictates that flow and
differential pressure are interdependent. Therefore,
it is difficult to control pump flow without flow sensor feedback, or knowledge of afterload or differential pressure (2). To overcome this difficulty, we
are attempting to develop a control algorithm based
exclusively on the time-varying electrical power
drawn by the pump motor (3,4).
zyxwvutsrq
~
KecelVed JdnUdtry 1996.
Addres\ correyondence and reprint requests to Dr. Robert 1,
Kormoq, at Division of Cardiothoracic Surgery, University of
Pittsbuigh Medical Center, Suite C-700, Presbyterian Univer5ity
Hozpital. Lothrop Street, Pittsburgh, PA 15213, U . S . A
MATERIALS AND METHODS
An axial flow blood pump (AxiPump; Nimbus,
Inc., Rancho Cordova, California, U.S.A.) was implanted in 7 sheep weighing 60-80 kg and in 12
calves weighing 90-1 10 kg in chronic preparations
for 7 to 150 days. The inflow cannula was inserted
into the left ventricular apex, and the outflow graft
was anastomosed to the descending aorta. Pump
flow was obtained by a customized ultrasonic flow
probe (Transonic Systems, Inc., Ithaca, New York,
U.S.A.) incorporated into the urethane portion of
the outflow graft. The pump was placed at the abdominal wall under the muscle. The drive line was
connected percutaneously to the drive console,
which could vary the pump speed from 6,000 rpm to
13,000 rpm.
Aortic blood pressure (AoP), left atrial pressure
zy
zyxwvut
zyxw
zyxwvutsrq
zyxwvutsrq
619
CONTROLLER FOR A N AXIAL FLOW BLOOD P U M P
(LAP), central venous pressure (CVP), pump
speed, and pump motor current were collected at
150 H z on a computer workstation (Apollo,
DN3550; Hewlett Packard, Inc. Chelmsford, Masachusetts, U.S.A.). During each data collection, the
pump speed was increased at 500 rpm increments
from 8,000 rpm to 13,000 rpm. In 3 of these cases,
the inferior vena cava (IVC) was occluded gradually
before chest closure to decrease the preload of the
pump for 15 s.
5
1
08
4 8
$06
3 E
=04
02
2:
0
0
14
9
10
11
13
12
Pump Speed ( xl000 rpm)
RESULTS
Figure 1 shows typical hemodynamic changes observed while the pump speed was increased. Mean
AoP increased slightly, and the increase in diastolic
pressure was higher than the increase in systolic
pressure. Accordingly, the pulse pressure decreased from 40 mm Hg to less than 15 mm Hg.
When the pump speed was maximized, collapse of
the left ventricle occurred, and the AoP was observed to fluctuate aperiodically. Concurrently,
LAP decreased and evidenced an abrupt onset of
oscillation as the left atrium was observed to collapse. Continuous or prolonged operation under
these conditions, leading to negative LAP and damage to the pulmonary circulation, must be avoided.
At the lower extreme, when pump speed was reduced excessively, the flow trace displayed regurgitation from the aorta to the left ventricle during
ventricular diastole. Within the preferred operating
speed range, the "optimal point", there is no evidence of negative flow.
The shape of the pump motor current waveform
was seen to display characteristic changes that were
dependent on the pump speed and preload (Fig. 2).
FIG. 2. The pump motor current waveform and the relationship between NSH and LAP are shown. NSH, normalized second harmonic.
A relationship was determined to exist between the
optimal pump speed and the normalized second harmonic, defined as the ratio of the second harmonic
to the first harmonic (normalized second harmonic
[NSH]). The relationship between the NSH and
LAP typically revealed a unimodal behavior with a
single minimum coinciding with the minimum LAP.
This result would imply that if the pump speed is
maintained at the point at which the NSH value is
minimal, the ventricle and atrium do not collapse
and LAP can be maintained within the lower side of
its physiologic range.
Figure 3 shows the results during IVC occlusion.
During the occluding phase, the LAP and AoP decreased rapidly, and the current waveform became
irradic, changing beat by beat. However, during the
recovery phase, the preload of the pump returned
slowly to physiologic range, thereby displaying the
damped response of the current waveform to the
changing preload.
DISCUSSION
In developing an axial flow blood pump as an
implantable LVAD, long-term survival has been ob-
+IVC occlusion
+
20
15
1
I" 10
-
EE5
0
-5
'
lo
3
2 51
Normallzed second harmonic
zyxwvut
zyxwvutsrqponmlk
zyxwvutsrqponmlkjihgf
zyxwvutsrqponmlkjihgfedc
2
I
7
10
40
80
30
Irnb
70
B"
90
loo
1!0
120
Iler*ms/
FIG. 1. Hemodynamic changes while the pump speed was
increased. AOP, aortic blood pressure; LAP, left atrial pressure. The arrow indicates ventricular collapse.
15
1
05
0
FIG. 3. Results of the inferior vena cava occlusion test are
shown. IVC, inferior vena cava.
Artf O r f a n s , Vol. 20, N o . 6 , 1996
620
zyxwvutsrqp
zyxwvutsrq
H . KONISHI ET AL.
tained in animals ( 5 ) . However, there still is no systematic algorithm to control these types of pumps in
the chronic setting. Dynamically characterizing the
relationship among pump flow, pressure increase,
and speed would provide an improved opportunity
to control speed for optimal hemodynamics. Although the quasi-static H-Q response of these
pumps usually is well defined, the effect of pulsatile
hemodynamics introduces significant hysteresis
and nonuniqueness ( 6 ) (Fig. 4). Accordingly, the
ability to identify flow or pressure from speed or
motor current becomes much more difficult, if not
impossible. In the next-generation controller, to
maintain normal hemodynamics under the wide
range of conditions expected, pump flow must be
adjusted rapidly to accommodate physiologic demand. Because the pump speed is not inherently
sensitive to changes in the preload or afterload, the
ventricle and atrium will collapse if the pump speed
is too high. If the pump speed is too low, regurgitation occurs from the aorta into the left ventricle
during ventricular diastole (Fig. 5). Detection of the
minimum NSH is one of the noninvasive methods
used to avoid these situations. To improve the reliability of the detection of the minimum NSH, it is
necessary to obtain a stable motor current signal. In
the development of the next stage of this controller,
the algorithm used to seek the minimum NSH continuously must be developed. In the secondgeneration controller, the flow estimation will be
derived from the motor current. This work includes
**
Increasing pump speed
I'I
Time
FIG. 5. Pump flow while the pump speed was increased is
shown, *, regurgitation from the aorta into the left ventricle;
**, collapse of the left ventricle or atrium.
the application of matched filters to analyze the
overall waveform characteristics. Ongoing work to
make the control algorithm more robust and enable
the control structure to be tuned or adapted to individual patients is another part of this ongoing effort.
REFERENCES
zyxwvutsrq
I . Antaki JF, Butler KC, Kormos RL, Kawai A, Konishi H ,
Kcrrigan JP, Borovetz HS, Maher TR, Kameneva MV, Griffith BP. In vivo evaluation of the Nimbus axial flow ventricular assist system: criteria and methods. A m Soc Artiflntern
Orguns 1993;39:M231-M236.
2. Schimd H. Trubel W, Moritz A, Wieselthaler G , Stohr HG,
Thoma H, Losert U, Wolner E. Noninvasive monitoring of
rotary blood pumps: necessity, possibilities. and limitations.
Am Soc ArtifIntern Organs 1992;16(2):195-202.
3 . Konishi H, Antaki JF, Boston JK, Borovetz HS, Kerrigdn
JP. Mandarin0 WA, Butler KC, Kormos RL. Controller for
an axial flow blood pump LVAD. Am Soc Artiffnterfi Orguns 1994;23:35.
4. Trinkl J. Havlik P, Mesana T, Mitsui N , Morita S , Demunck
J-L, Tourres J-L, Monties J-R. Control of a rotary pulsatile
cardiac assist pump driven by an electric motor without a
pressure sensor to avoid collapse of the pump inlet. Am Soc
Artiffntern Organs 1993;39:M237-M241.
5. Macris MP, Myers TJ, Jarvik R, Robinson JL, Fuqua JM,
Parnis SM, Frazier OH. In vivo evaluation of an intraventricular electric axial flow pump for left ventricular assistance. Am Soc Artif Intern Organs 1994;40:M719-M722.
6 . Yamazaki K , Umezu M, Koyanagi H, Outa E, Ogino S ,
Ordke Y.Development of a miniature intraventricdar axial
flow blood pump. Am Soc Artif Intern Organs 1993;39:
M22CM230.
zyxwvutsrqp
zyxwvutsrqpon
zyxwvuts
zyxwvutsrqponm
zyxwvutsrqpon
h
ul
g
(L / min)
'O1
-2
E
2
zyxwvut
140-
Quasi - steady
120100-
80-
u)
-m
'g
60-
40-
Pulsatile
b)
Pump flow (Lhin)
FIG. 4. Pump performance (H-Qcurve) (Pump speed
Krpm) is shown.
Artif Orguns, Vol. 20. N o , 6, 1996
=
10